Professional Documents
Culture Documents
VetOphthalmolPerkinsdogscatsv 12n 5p 277a2842009
VetOphthalmolPerkinsdogscatsv 12n 5p 277a2842009
net/publication/26809514
CITATIONS READS
20 4,002
7 authors, including:
All content following this page was uploaded by Silvia Franco Andrade on 29 May 2018.
sphere is the same as the pressure inside that sphere (Law of authors compared the accuracy of IOP measurements using
Imbert-Fick).2 In human ophthalmology, applanation tonometry the Perkins, Tono-Pen XL, and TonoVet tonometers,
for the Goldmann method is most commonly used and and again, the Perkins tonometer showed the best accuracy
considered as the standard method for IOP measurement, in IOP measurement in rabbits.
as it is accurate and practical. The Goldmann method is a The Perkins tonometer costs, on average, three to five
variable force tonometer consisting of a double prism, with times less than the Tono-Pen, and could represent an
a 3-mm formation of semicircles of fluorescein, that is interesting alternative for the IOP exam in dogs and cats.
adjusted during the tonometry, and is used often in human However, at present, there are no reports of the use of this
ambulatory practice.2,6 tonometer in dogs and there has only been one report of
The Perkins handheld applanation tonometer utilizes the use of the Perkins tonometer in cats, an old report from
the Goldmann method,2 and has a small source of blue light 198410 that used the Alcon pneumatonograph (PTG) and
for the formation of semicircles of fluorescein (Fig. 1). This Perkins tonometer on rabbits and cats. This report empha-
instrument requires some practice in order to achieve perfect sizes use of the PTG on rabbits more than the Perkins
readings.6 The scale of the tonometer ranges from 0 to 5, tonometer on cats; therefore, more research is needed
with smaller divisions of 0.2, and the examination should be on this.
started with the scale at number 1 (Fig. 1e). In humans, readings The objective of this study was to evaluate and validate the
measured on this scale are multiplied by 10.2,6 The accuracy accuracy of the Perkins handheld applanation tonometer
of this type of applanation tonometer depends on several in the measurement of IOP in dogs and cats. To do so, IOP
factors, such as the excessive use of fluorescein, excessive readings from this tonometer, in mmHg, were converted
pressure of the prism on the cornea, incorrect calibration, using the calibration curve generated using values measured
anatomical and biomechanical characteristics of the cornea via manometry with a mercury column manometer (postmortem
(thickness, diameter, hydration, elasticity, viscosity, etc.), study). The tonometry was then validated by comparing the
pathologies (such as edema and cornea ulceration, which values of IOP in mmHg from this tonometer with the mano-
can generate incorrect readings), evaluation technique, metric values (in vivo study) in this species. Also, the IOP
apnea, muscular contraction, and contact with the adnexal measurement was evaluated in conscious dogs and cats in a
structures, etc.6,16–18 clinical set.
The few studies using the Perkins tonometer in domestic
animals have been performed in cattle,8 rabbits and cats,10
MATERIALS AND METHODS
and rabbits.11,12 One study11 used rabbits to compare the
effectiveness and accuracy of three types of tonometers, This experiment was approved by the Ethical Committee of
including the Tono-Pen XL, Perkins, and Ocular Blood UNOESTE (Protocol no. 022/07 and no. 023/07), and of
Flow (OBF) pneumatonometer. In this study, the Perkins UNIFESP (Protocol no. 1417/07). For the postmortem and
tonometer demonstrated the best accuracy in IOP measure- in vivo study the animals were obtained from the kennel
ment. In another recent study,12 also done in rabbits, the and cat pound of the UNOESTE. The conscious animals
were obtained from the clinic of Veterinary Teaching Hospital animals were given one drop of antibiotic eye drops
of UNOESTE, Presidente Prudente, SP, Brazil. Only (chloramphenicol-Visalmin, Bunker, São Paulo, SP) and
normal eyes were used in this study after being submitted to anti-inflammatory eye drops (diclofenac sodium – Still,
ophthalmic exams. Allergan), twice daily, for 1 week, and were evaluated through
For the postmortem study, both eyes from 10 mixed-breed daily basic ophthalmic exams (ophthalmoscopy, pupillary
adult dogs (four males and six females, 2–8 years of age, light reflex, Schirmer tear test and fluorescein test) until the
mean weight 9.2 ± 5.8 kg) and 10 mixed-breed adult cats induced corneal lesion healed.
(five males and five females, years of age, mean weight For the clinical set study, the cornea of the animals was
3.3 ± 0.8 kg), were obtained immediately after sacrifice from topically anesthetized with two drops of proxymetacaine
several cases authorized for necropsy. For the in vivo study, 0.5% (Anestalcon, Allergan) (Fig. 2a) and were instilled
both eyes from 10 clinically normal and anesthetized mixed- one drop of 1% fluorescein eye drops (Allergan) (Fig. 2b) for
breed adult dogs (five males and five females, 2–5 years of the formation of the fluorescein semicircles, and then three
age, mean weight 11.9 ± 4.6 kg) and 10 clinically normal and IOP readings were measured with the Perkins tonometer
anesthetized mixed-breed adult cats (five males and five and the mean was calculated (Fig. 2c,d).
females, 1–5 years of age, mean weight of 3.6 ± 0.7 kg) were To avoid transmission of infectious ocular diseases, after
used. For the IOP measurement in conscious animals, 20 each use of the Perkins tonometer, the prism was removed
dogs (four Miniature Poodle, three Dachshund, two Border and washed in a 0.9% physiologic saline solution. It was then
Collie, one Basset Hound, one Cocker Spaniel and nine mixed- submerged for 10 min in a solution of 3% hydrogen peroxide
breed; 11 males and nine females; 1–12 years of age; mean and washed again in 0.9% saline solution and dried with
weight 17.3 ± 11.6 kg) and 20 cats (five Siamese, four sterile gauze.21
Persian and 11 mixed-breed; eight males and 12 females; The mean values of IOP measurement obtained using ocular
1–11 years of age; mean weight of 3.2 ± 0.8 kg) were used. manometry and tonometry in the postmortem and in vivo
For the postmortem study, to manipulate and measure study, and the Perkins tonometer reading and the Perkins
IOP manometrically, the eyelids were separated with eyelid tonometer reading correct by the calibration curve in the
speculum, and the eye was cannulated with a 23-gauge conscious animals, were compared and analyzed by Student
needle (Embramac, Itapira, SP, Brazil), approximately 2 mm t-tests. A significance level of P < 0.05 was adopted. Linear
lateral to the limbal area, at the 10 o’clock position in the OD regression analysis was performed to analyze the relationship
and the 2 o’clock position in the OS. Cyanoacrylate between manometry vs. tonometry IOP measurements. A
glue (Superbond, Loctite, Itapevi, SP, Brazil) was applied correlation coefficient (r2) was calculated.22 The Bland–Altman
around the needle to prevent leakage of the aqueous humor. assessment for agreement was used to compare the two
The needle was connected via a polyethylene tube to a three- methods of IOP measurement. A range of agreement was
way stopcock, which was also connected to a reservoir of defined as mean bias ±2 standard-deviation.23
10 mL physiological saline solution 0.9% (Fresenius Kabi,
Campinas, SP, Brazil) and a mercury column manometer
RESULTS
(Unitec, Rio de Janeiro, RJ, Brazil) that was in the zero
position in relation to the center of the eye. The calibration The correlation coefficient (r2) between the manometry measure-
curve for manometry vs. tonometry was determined by ments and the Perkins tonometer measurements were 0.982
raising the IOP artificially in 5 mmHg increments to reach (dogs) and 0.988 (cats), and the corresponding linear regression
50 mmHg (5–50 mmHg) in open stopcock mode. Three equation were y ¼ 0.0893x + 0.1105 (dogs) and y ¼ 0.0899x +
IOP readings were measured with the Perkins tonometer 0.1145 (cats) in the postmortem study (Figs. 3 and 4).
(Clement Clarke, Harlow, UK), and the mean was calculated. The results of IOP measurements taken via manometry
Prior to the tonometer IOP reading, one drop of 1% fluorescein and the applanation tonometer in the in vivo study are shown in
eye drops were instilled (Allergan, Guarulhos, SP, Brazil) for Table 1 and Table 2. In the in vivo study the mean IOP readings
the formation of the fluorescein semicircles. in dogs were 14.5 ± 2.8 mmHg (range 9.0–18.0 mmHg) with
For the in vivo study, the dogs and cats were anesthetized the manometer and were 15.3 ± 2.9 mmHg (range 10.0–
with diazepam (Compaz, Sigma Pharma, Hortolândia, SP, 17.8) with the Perkins tonometer after calibration curve
Brazil), 0.5 mg/kg i.v., and thiopental (Thiopentax, Itapira, SP, correction, and in cats were 15.0 ± 1.7 mmHg (range 13.0–
Brazil), 12.5 mg/kg i.v. The cornea was topically anesthetized 18.0 mmHg) with the manometer and were 16.4 ± 2.6 mmHg
with two drops of proxymetacaine 0.5% (Anestalcon, Allergan). (13.1–20.9 mmHg) with the Perkins tonometer after calibra-
Manometry and tonometry measurements (Fig. 3a,b) were tion curve correction. In the in vivo study, in both species,
taken as described for the postmortem study. there was no significant difference (P > 0.05) between the IOP
After the IOP readings were made, the fine needle was measurements obtained via manometry and those obtained
removed from the anterior chamber and soon afterward, by the Perkins applanation handheld tonometer and
cyanoacrylate glue was instilled with a 25 · 7 mm needle (Injex, corrected by the calibration curve (Tables 1 and 2).
Ourinhos, SP, Brazil) at the site of the corneal puncture, The mean IOP readings with the Perkins tonometer
to protect against perforation.19,20 After the procedure, the after calibration curve correction in conscious dogs and cats
Figure 2. Use of the Perkins tonometer in normal eyes of conscious dog (female Basset Hound) and conscious cat (male mixed-breed) in the ambulatory
of the Veterinary Teaching Hospital of UNOESTE. (a) Instillation of two drops of proxymetacaine 0.5% eye drops; (b) Instillation of one drop of 1%
fluorescein eye drops; (c) and (d) Use of the Perkins tonometer after the instillation of the two eye drops.
Table 1. Values of in vivo IOP in mmHg by ocular manometry and the Perkins tonometer* in dogs (n ¼ 10)
Table 2. Values of in vivo IOP in mmHg by ocular manometry and the Perkins tonometer* in cats (n ¼ 10)
Figure 3. Statistical values for regression lines of IOP readings using the
Perkins applanation tonometer vs. direct ocular manometry in 10 dogs
(a) and in 10 cats (b).
Table 3. Values of IOP in mmHg by the Perkins tonometer* in healthy eyes of conscious dogs (n ¼ 20) assisted at the ambulatory of the Veterinary
Teaching Hospital of the University of Oeste Paulista, Presidente Prudente, SP, Brazil
Mean Perkins Mean Perkins Mean Perkins Mean Perkins Mean Perkins Mean Perkins
Animal tonometer tonometer tonometer tonometer reading tonometer reading tonometer reading
number reading OD reading OS reading correct† OD correct† OS correct†
1 1.4 1.4 1.4 14.4 14.4 14.4
2 1.6 1.6 1.6 16.6 16.6 16.6
3 1.5 1.6 1.6 15.5 16.6 16.0
4 1.5 1.5 1.5 15.5 15.5 15.5
5 1.6 1.5 1.6 16.6 15.5 16.0
6 1.6 1.6 1.6 16.6 16.6 16.6
7 1.4 1.4 1.4 14.4 14.4 14.4
8 1.7 1.6 1.7 17.8 16.6 17.2
9 1.4 1.4 1.4 14.4 14.4 14.4
10 1.4 1.4 1.4 14.4 14.4 14.4
11 1.2 1.2 1.2 12.2 12.2 12.2
12 1.6 1.5 1.6 16.6 15.5 16.0
13 1.2 1.2 1.2 12.2 12.2 12.2
14 1.4 1.4 1.4 14.4 14.4 14.4
15 1.2 1.2 1.2 12.2 12.2 12.2
16 1.4 1.5 1.5 14.4 15.5 15.0
17 1.6 1.6 1.6 16.6 16.6 16.6
18 1.3 1.3 1.3 13.3 13.3 13.3
19 1.6 1.6 1.6 16.6 16.6 16.6
20 1.4 1.4 1.4 14.4 14.4 14.4
Mean±SD 1.5 ± 0.2a 1.4 ± 0.1a 1.5 ± 0.2a 15.0 ± 1.7b 14.9 ± 1.5b 14.9 ± 1.6b
treatment. Following the application of antibiotics and handheld Perkins tonometer is often used in clinics and for
nonsteroid anti-inflammatory eye drops, the lesions resolved bedridden patients and those patients unable to complete a
well within 24 h after the experiment. fixed exam.2,6 This method is attractive because it carries a
It was relatively easy the IOP measurement with the Perkins much more accessible cost, around three to five times less
tonometer in conscious animals in the ambulatory routine. than that of the Tono-Pen and TonoVet. Until now, the
Only three dogs and two cats needed a more firm manual Goldmann methodology has not been examined extensively
restrain for the IOP measurement. in the veterinary setting, with only a few studies reporting work
in cats, cattle and rabbits. However, these few studies, per-
formed mainly in rabbits, have shown an excellent correlation
DISCUSSION
in IOP measurements.8,10–12
In veterinary medicine, IOP measurement is most often Anecdotally, some veterinary ophthalmologists have
performed via the applanation method, using handheld suggested that, in animals, the Goldmann methodology is
tonometers such as the Tonopen, which facilitate the execution impossible, because the animal cannot be restrained ade-
of this exam in animals.1,3–5 Most of these tonometers were quately to permit the formation of the semicircles. However,
originally manufactured for use in humans. Therefore, before if this were actually so, this exam would also be very difficult
they are used in animals, they need to be validated based on in children. Yet this exam is frequently utilized in children,2,6
manometry measurements, especially given the differences and, because the exam lasts only seconds, it is, in fact, easy to
found in animal corneas, such as differences in corneal diameter, execute. In our experiment, we could notice that the ability
thickness, curvature and viscoelastics properties, as well in to use the Perkins tonometer was acquired in a few days,
other factors, such as surface tension and lacrimal film charac- inside of the minimum patterns of appropriate quality of the
teristics, when compared to human corneas.4,8–10 exam, what turns the same reasonably feasible. Besides, this
A significant barrier to the popularization of this exam in tonometer can be used in any position, is extremely resistant,
the veterinary setting has been the high cost of tonometers easy to calibrate and disinfect.
like the Tonopen and Tonovet, especially in poor countries. The correlation coefficient for the use of the Perkins
In human medicine, the Goldmann methodology using the tonometer in dogs (r2 ¼ 0.982) and cats (r2 ¼ 0.988) was
Table 4. Values of IOP in mmHg by the Perkins tonometer* in healthy eyes of conscious cats (n ¼ 20) assisted at the ambulatory of the Veterinary
Teaching Hospital of the University of Oeste Paulista, Presidente Prudente, SP, Brazil
Mean Perkins Mean Perkins Mean Perkins Perkins Perkins Mean Perkins
Animal tonometer tonometer tonometer tonometer reading tonometer reading tonometer reading
number reading OD reading OS reading correct* OD correct* OS correct*
1 1.7 1.7 1.7 17.6 17.6 17.6
2 1.4 1.6 1.5 14.3 16.5 15.4
3 1.4 1.5 1.5 14.3 15.4 14.9
4 1.4 1.4 1.4 14.3 14.3 14.3
5 1.4 1.4 1.4 14.3 14.3 14.3
6 1.6 1.6 1.6 16.5 16.5 16.5
7 1.4 1.5 1.5 14.3 15.4 14.9
8 1.8 1.8 1.8 18.7 18.7 18.7
9 1.5 1.4 1.5 15.4 14.3 14.9
10 1.4 1.4 1.4 14.3 14.3 14.3
11 1.6 1.8 1.6 16.5 18.7 17.6
12 1.2 1.2 1.2 12.1 12.1 12.1
13 1.6 1.6 1.6 16.5 16.5 16.5
14 1.4 1.4 1.4 14.3 14.3 14.3
15 1.5 1.5 1.5 15.4 15.4 15.4
16 1.5 1.6 1.6 15.4 16.5 15.6
17 1.4 1.4 1.4 14.3 14.3 14.3
18 1.5 1.5 1.5 15.4 15.4 15.4
19 1.3 1.3 1.3 13.2 13.2 13.2
20 1.2 1.2 1.2 12.1 12.1 12.1
Mean±SD 1.5 ± 0.2a 1.5 ± 0.2a 1.5 ± 0.2a 15.0 ± 1.7b 15.3 ± 1.9b 15.1 ± 1.7b
considered excellent (Fig. 3). The Bland-Altman graph (conscious cats), are very close to the values report by
also reveals a good correlation between IOP readings of Nagy et al.26 with IOPcc (cornea-compensated IOP) of
Perkins tonometer and manometer (Fig. 4). In humans,2,6 16.0 ± 5.4 mmHg.
tonometer IOP readings are multiplied by 10. We have found In the present study, the use of the mercury column
that, based on our calibration curve for dogs (y ¼ 0.0893x + manometer, instead of an electronic monitor, was shown to
0.1105) and cats (y ¼ 0.0899x + 0.1145), we can approximate be an accurate, practical and low cost option for experimental
the equation and multiply IOP measurements by a factor IOP measurement, and was in agreement with other
of 10 for dogs and cats as well. This would facilitate the use reports.13 The advantages observed with the use of the
of the Perkins tonometer for obtaining IOP measurements Perkins tonometer in the measurement of IOP in dogs and
in ambulatory veterinary practice. cats were accuracy, reproducibility, maintenance, calibration,
The mean IOP values in dogs measured with the Perkins and disinfection of the prism tonometer. The disadvan-
tonometer in the in vivo study (15.2 ± 3.0, range 10.0– tages were the limit of IOP reading of the Perkins tonome-
17.8 mmHg) and in conscious dogs (14.9 ± 1.6, range ter (upper limit of 50 mmHg) and a larger time training
12.2–17.2 mmHg) are smaller of those reported by necessary for the correctly use of the tonometer.
Gellat and MacKay,24 measured with the Tono-Pen XL The results of our experiment demonstrate that the Per-
(19.2 ± 5.9 mmHg) and the Mackay-Marg (18.8 ± 5.5 mmHg). kins handheld applanation tonometer could be in the
However, the values measured by the Perkins tonometer future an excellent tool for glaucoma detection or
are closer of the real IOP values measured by manometry experimental work in dogs and cats, and could possibly
(14.4 ± 2.8, range 9.0–18.0 mmHg). In this study, we didn’t increase the popularity of this ophthalmic exam in small
compare the Perkins tonometer with others tonometers in animal clinics because of its more accessible cost as com-
conscious animals; therefore, more research should be made pared to other tonometers currently used in veterinary
in this sense to prove the accuracy of the Perkins tonometer clinics. More researches should be made comparing this
in a clinical set.25 tonometer with others tonometers, such as the Tonopen
In cats, the mean value of IOP measurements using or the Tonovet, in conscious dogs and cats, to prove
the tonometer, corrected by the calibration curve to your accuracy in the IOP measurement in the clinical
16.4 ± 2.6 mmHg (in vivo study) and 15.1 ± 1.7 mmHg setting.
ACKNOWLEDGMENT 13. Passaglia CL, Guo X, Troy JB. Tono-Pen XL calibration curves for
cats, cows and sheep. Veterinary Ophthalmology 2004; 7: 261–264.
The authors thank the post-graduate program of UNOESTE 14. Knollinger AM, La Croix NC, Barrett PM et al. Evaluation of a
and UNIFESP for financial support and orientation, respectively, rebound tonometer for measuring intraocular pressure in dogs and
and Dr Rogério Giuffrida for statistical analysis. horses. Journal of American Veterinary Medical Association 2005; 227:
244–248.
15. Görig C, Roel TI, Stades FC et al. Comparison of the use of new
REFERENCES handheld tonometers and established applanation tonometers in
dogs. American Journal of Veterinary Research 2006; 67: 134–144.
1. Gelatt KN. Veterinary Ophthalmology, 3rd edn. Lippincott Williams
16. Kniestedt C, Lin S, Choe J et al. Clinical comparison of contour
& Wilkins, Philadelphia, 1999.
and applanation tonometry and their relationship to pachymetry.
2. Kanski JK. Clinical Ophthalmology – A Systematic Approach, 5th edn.
Archive of Ophthalmology 2005; 123: 1532–1537.
Butterworth-Heinemann, Edinburgh, 2003.
17. Ceruti P, Morbio R, Marraffa M et al. Comparison of dynamic
3. Slatter DH. Fundamentals of Veterinary Ophthalmology, 3rd edn.
contour tonometry and Goldmann applanation tonometry in
W.B. Saunders, Philadelphia, 2001.
deep lamellar and penetrating keratoplasties. American Journal of
4. Gelatt KN, Peiffer RL, Gum GG et al. Evaluation of applanation
Ophthalmology 2008; 145: 215–221.
tonometers for the dog eye. Investigative Ophthalmology and Visual
18. Medeiros FA, Sample PA, Weinreb RN. Comparison of dynamic
Science 1977; 16: 963–968.
contour tonometry and Goldmann applanation tonometry in
5. Peruccio C, Pizzirani S, Peiffer RL. Diagnostics. In: Small Animal
African American subjects. Ophthalmology 2007; 114: 658–665.
Ophthalmology – A Problem-Oriented Approach, 2nd edn. Saunders,
19. Ueda EL, Ottaiano JAA. Comparação dos gastos com material de
London, 1997.
reparação de perfurações de córnea com cola de cianoacrilato e sutura
6. Chang DF. Ophthalmic exam. In: General Ophthalmology, 14th edn.
de córnea. Arquivos Brasileiros de Oftalmologia 2004; 67: 97–101.
(eds Vaughan D, Asbury T, Riordan-Eva P) Appleton & Lange,
20. Felberg S, Lake JC, Lima FA et al. Adesivo de cianoacrilato no trata-
Stamford, 1999.
mento de afinamentos e perfurações corneais: técnica e resultados.
7. Takatsuji K, Sato Y, Izuka S et al. Animal model of closed angle
Arquivos Brasileiros de Oftalmologia 2003; 66: 345–349.
glaucoma in albino mutant quails. Investigative Ophthalmology and
21. Lingel NJ, Coffey B. Effects of disinfecting solutions recommended
Visual Science 1986; 27: 396–400.
by the centers for disease control on goldmann tonometer biprisms.
8. Gerometta R, Podos SM, Candia MD et al. Steroid-induced ocular
Journal of American Optometrists Association 1992; 63: 43.
hypertension in normal cattle. Archives of Ophthalmology 2004; 122:
22. Steel RGD, Torrie JH. Principles and Procedures of Statistics: A
1492–1497.
Biometrical Approach, 2nd edn. Mc-Graw Hill, New York, 1981.
9. Dziezyc J, Milichamp J, Smith WB. Comparison of applanation
23. Bland JM, Altman DG. Statistical methods for assessing agreement
tonometers in dogs and horses. Journal of American Veterinary
between two methods of clinical measurement. Lancet 1986; 1: 307–
Medical Association 1992; 201: 430–433.
310.
10. Hammond BR, Bhattacherjee P. Calibration of the Alcon applana-
24. Gelatt KN, MacKay EO. Distribution on intraocular pressure in
tion pneumatograph and Perkins tonometer for use in rabbits and
dogs. Veterinary Ophthalmology 1998; 1: 109–114.
cats. Current Eye Research 1984; 3: 1155–1158.
25. Leiva M, Naranjo C, Pena MT. Comparison of the rebound
11. Lim KS, Wickremasinghe SS, Cordeiro MF et al. Accuracy of
tonometer (ICare) to the applanation tonometer (Tonopen XL)
intraocular pressure measurements in New Zealand white rabbits.
in normotensive dogs. Veterinary Ophthalmology 2006; 9: 17–21.
Investigative Ophthalmology and Visual Science 2005; 46: 2419–2423.
26. Nagy LJ, MacRae S, Yoon G et al. Photorefractive keratectomy in
12. Kalesnykas G, Uusitalo H. Comparison of simultaneous readings of
the cat eye: biological and optical out comes. Journal of Cataract
intraocular pressure in rabbits using Perkins handheld, Tono-Pen
Refractive Surgery 2007; 33: 1051–1064.
XL, and TonoVet tonometers. Graefe’s Archives of Clinical Experi-
mental Ophthalmology 2007; 245: 761–762.