Pharmaceutical Refrigerator
PHC Corporation
Biomedical Div.
SET Press this key to be setting mode available with the 1st digit blinks. Press
this key again to memorize the value which you changed.
(Scroll key) In Setting Mode, moves between the first and second places before the
decimal point every time you press this key.
Key Lock mode is available (“L 0” display) if you press this key for 5
seconds.
(Up arrow key) In Setting Mode, increase by 1 every time you press this key.
Function modes are available (“F00” display) if you press this key for 5
seconds.
2. Temperature setting
Setting range +2℃ ~ +23℃
Display range -50 ~ +50
How to set Press the SET key and change to your desired value with up arrow key and
scroll key. Press SET key to memorize the value and return to chamber
temperature display.
Unacceptable If you input the value out of setting range and press SET key, buzzer tone
setting is emitted to inform the value unacceptable.
3. Key lock
Setting range 0 (Unlock) or 1 (Lock)
How to set In chamber temperature display, press scroll key for 5 seconds to step into
Key Lock mode with ‘L _ X’ displayed. (default: L_0)
Change blinking ‘X’ to your desired value with up arrow key.
Press SET key to memorize the value in non-volatile memory.
4. Function mode
How to set In chamber temperature display, press up arrow key for 5 seconds to step
into Function mode with ‘F00’ displayed.
Change blinking 1st digit to your desired value with up arrow key and scroll
key. Press SET key to be Function mode available.
Unacceptable If you input unused function code or F51~F59 and press SET key, buzzer
setting tone emitted to inform the value unacceptable.
5. Error codes
Error codes (Self-diagnostic functions)
E01: Temp. sensor is disconnected
E02: Temp. sensor is short circuited
E03: DEF sensor is disconnected
E04: DEF sensor is short circuited
E05: Comp. sensor is disconnected
E06: Comp. sensor is short circuited
E07: Ambient temp. sensor is disconnected
E08: Ambient temp. sensor is short circuited
E09: Battery setting check
E10: Compressor temperature is abnormal
E13: Chamber fun alarm sensor is disconnected
E14: Chamber fun alarm sensor is short circuited
F-1: Battery replacement time
F-2: Condensing fan replacement time
E2P: Memorization of non-volatile memory is failed
Structural specifications
MPR-722 MPR-722R
External dimensions W770 x D830 x H1950 (mm)
Internal dimensions W770 x D710 x H1500 (mm)
Effective capacity 684 L 671 L
Exterior Painted steel
Interior Painted steel
Door Double layer pair glass/steel plate, automatic closing mechanism
Insulation Rigid polyurethane foamed-in place
Shelves Hard steel wire on polyester coating ------------
4 pcs.
Drawers ---------- Painted steel
5 pcs.
Access port φ30 mm in 3 locations (Left, right, top)
Cooling method Forced cool air circulation
Compressor Hermetic type Output; 200W
Condenser Wire and tube type
Evaporator Fin and tube type
Refrigerant R-513A
Defrosting Forced type, fully automatic
Internal heater 100W
Fluorescent lamp White light 40W x 1 (Glow starter; FG-4P)
Weight 174 kg 193 kg
Accessories 1 set of key,2 nylon clip, 2 binders
Optional Automatic temperature recorder (MTR-0621LH),
components Recorder fixture (MPR-S30),
Automatic temperature recorder (MTR-G04A)
Recorder fixture (MPR-S7)
Light shielding plate (MPR-72BP)
Battery box for power failure alarm (MPR-48B1)
Interface board (MTR-480、MTR-L03)
6. Alarms
High temp. alarm
When the chamber temperature is equal or higher than set temperature +
high temp. alarm set temperature + 0.09, ALARM lamp and temperature
display blink, buzzer sounds intermittently and remote alarm turns on after
delay time elapses.
High temp. alarm set temperature: +5℃ (default)
Delay time: 15 minutes (default)
When the chamber temperature is equal or lower than set temperature,
ALARM lamp goes off, temperature display stops blinking, buzzer stops
sounding and remote alarm turns off.
You can cancel the buzzer and remote alarm operation by pressing
BUZZER key during high temp. alarm.
Low temp. alarm When the chamber temperature is equal or lower than set temperature -
low temp. alarm set temperature - 0.09, ALARM lamp and temperature
display blink, buzzer sounds intermittently and remote alarm turns on after
delay time elapses.
High temp. alarm set temperature: -5℃ (default)
Delay time: 15 minutes (default)
When the chamber temperature is equal or higher than set temperature,
ALARM lamp goes off, temperature display stops blinking, buzzer stops
sounding and remote alarm turns off.
You can cancel the buzzer and remote alarm operation by pressing
BUZZER key during low temp. alarm.
0℃ alarm When the chamber temperature is lower than 0.0, ALARM lamp and
temperature display blink, buzzer sounds intermittently, remote alarm turns
on without delay.
When the chamber temperature is equal or higher than 0.0, ALARM lamp
goes off, temperature display stops blinking, buzzer stops sounding and
remote alarm turns off.
You can cancel the buzzer and remote alarm operation by pressing
BUZZER key during 0℃ alarm.
Chamber fan alarm When the chamber alarm temperature is higher than the sum of setting
(default 5℃) temperature, the chamber fan alarm setting value and 0.09, ALARM lamp
and numerical value LED blink. Then buzzer sounds intermittently and
remote alarm is turn on after passing delay time 8default 15 min) .
When the chamber fan alarm temperature is or lower than the setting
temperature, ALARM lamp turns off, the display LED blinks and, buzzer
sound and remote alarm is turn off.
You can cancel the buzzer and remote alarm operation by pressing
BUZZER key during the information,
The chamber fan alarm setting value is same value as the high
temperature alarm se value on default. (default is set by F01, then the
lower limit is 5℃)
When the chamber alarm setting value is changed the independent setting
to valid “001” by F41, you can set the alarm setting value independently by
F39. However, during the door open, alarm has not sound.
During pull-down, buzzer has sounded and remote alarm has turn off.
However, Alarm lamp and the numerical value LED blink.
9. Compressor control
Compressor protection
When comp. sensor reads 52℃ or higher, unit diagnoses compressor should stop operation.
When comp. sensor reads ambient temp. sensor value plus 10℃ or lower, unit diagnoses
compressor should start operation.
10. Defrosting
Mechanism
DEF sensor detects the temperature in evaporator pipe and unit diagnoses current should be
conducted to heater to defrost an evaporator.
During defrosting, compressor should stop operation, chamber temperature and “dF” are
alternately displayed.
By evaporator temperature
When DEF sensor temperature in evaporator reads equal or lower than -12℃, unit performs
defrosting.
When DEF sensor temperature in evaporator reads equal or higher than 5℃, unit stops
defrosting.
By other conditions
Unit performs defrosting if it meets the following conditions both (1) and (2).
The unit stops defrosting when DEF sensor temperature in evaporator reads equal or higher
than 5℃.
When the DEF sensor temperature in evaporator reads higher than 0℃ and the compressor
keeps running for 4 hours without defrosting, counting of compressor ON time is reset.
Frequency of defrosting
Unit can memorize frequency of forcible defrosting since the power is supplied.
You can check the frequency in the function mode, F36.
Specification of sensor
25
Internal temperature(℃)
15
-5
-15
0 1 2 2 3 4 5 6 7 7 8
Time scale(hour)
1.6
1.0 230 V 50 Hz Pd
230 V 50 Hz Ps
0.8
115 V 60 Hz Pd
0.6 Balance pressure 0.44 MPa 115 V 60 HzPs
CAUTION
When installing a automatic temperature recorder,
unplug the power plug. It may cause an electric shock. Fig 1
5. From the outside of the access port, pass through the Fig3
temperature-sensitive section of the self-recording
temperature sensor into the cabinet. (Fig 4)
Recorder
sensor
Fig 4
6. As shown in Fig. 5, use the supplied binder (2 pcs)
on the fan cover on the top surface to fix the
temperature-sensitive sensor of the automatic
temperature recorder sensor.
Nylon clip
Capillary
Fig 6
Capillary
Cap
Fig7
Installation for MTR-G04
Automatic
temperature Recorder
recorder fixture
Fig 2
Recorder
4. From the outside of the access port, pass through the sensor
temperature sensor of the self-recording temperature
sensor into the storage compartment. (Fig 3)
Fig 4
8. Connect the connector on the temperature recorder
side to the power supply connector at the upper right
corner of the mounting position of the temperature
recorder. (Fig 7) Nylon clip
9. Align the groove of the mounting bracket with the shaft, Capillary
attach the self-recording temperature recorder to the
self-recording temperature mounting position of the
product, and fix it with one screw. (Fig 8)
Fig 5
Cap Wire
Fig 6
Power plug
Screw
Shaft
Fig 8
Automatic temperature recorder Fig 7 Slit
After installation of the optional battery box for power failure, be sure to make the following settings.
If the battery switch is turned on without the following settings, E09” and the chamber temperature is
displayed alternately in the temperature display .
● If “E09” is displayed, turn off the battery switch, then make the following settings and turn on the battery
switch again
Set the first digit to 1 with the up Press the up arrow key and increase
7
arrow key. the number of settable digit
Memorize the alarm temperature
8 Press set key SET setting value and current chamber
value