Service Manual

Pharmaceutical Refrigerator

PHC Corporation
Biomedical Div.

©2019 PHC Corporation


All rights reserved. Unauthorized copying
and distribution is a violation of law
Refrigeration circuit
Wiring diagram
Control specifications

1. Keys and switches


BUZZER (1) Control of buzzer and remote alarm
In alarming condition, press this key to stop buzzer sounding and to turn
remote alarm output off, however ALARM lamp keeps blinking.
Press this key again to sound buzzer and to turn remote alarm output on.
(2) Battery setting check (with MPR-48B1)
During power failure, press this key to show the chamber temperature for 5
seconds if you set ‘001’ in F42 with battery installed.
Press this key to show E09 without battery.
E09 is also shown when the battery switch is off position with battery
installed, or when you set ‘001’ in F42 with battery installed.

SET Press this key to be setting mode available with the 1st digit blinks. Press
this key again to memorize the value which you changed.

(Scroll key) In Setting Mode, moves between the first and second places before the
decimal point every time you press this key.
Key Lock mode is available (“L 0” display) if you press this key for 5
seconds.

(Up arrow key) In Setting Mode, increase by 1 every time you press this key.
Function modes are available (“F00” display) if you press this key for 5
seconds.

2. Temperature setting
Setting range +2℃ ~ +23℃
Display range -50 ~ +50
How to set Press the SET key and change to your desired value with up arrow key and
scroll key. Press SET key to memorize the value and return to chamber
temperature display.
Unacceptable If you input the value out of setting range and press SET key, buzzer tone
setting is emitted to inform the value unacceptable.

3. Key lock
Setting range 0 (Unlock) or 1 (Lock)
How to set In chamber temperature display, press scroll key for 5 seconds to step into
Key Lock mode with ‘L _ X’ displayed. (default: L_0)
Change blinking ‘X’ to your desired value with up arrow key.
Press SET key to memorize the value in non-volatile memory.

Function mode is available if Key Lock is effective.

4. Function mode
How to set In chamber temperature display, press up arrow key for 5 seconds to step
into Function mode with ‘F00’ displayed.
Change blinking 1st digit to your desired value with up arrow key and scroll
key. Press SET key to be Function mode available.
Unacceptable If you input unused function code or F51~F59 and press SET key, buzzer
setting tone emitted to inform the value unacceptable.
5. Error codes
Error codes (Self-diagnostic functions)
E01: Temp. sensor is disconnected
E02: Temp. sensor is short circuited
E03: DEF sensor is disconnected
E04: DEF sensor is short circuited
E05: Comp. sensor is disconnected
E06: Comp. sensor is short circuited
E07: Ambient temp. sensor is disconnected
E08: Ambient temp. sensor is short circuited
E09: Battery setting check
E10: Compressor temperature is abnormal
E13: Chamber fun alarm sensor is disconnected
E14: Chamber fun alarm sensor is short circuited
F-1: Battery replacement time
F-2: Condensing fan replacement time
E2P: Memorization of non-volatile memory is failed

Error codes descriptions


E01: Temp. sensor is disconnected.
When temp. sensor reads -60℃ or lower, E01 and ‘-50’ are alternately
displayed.
Buzzer sounds intermittently and remote alarm turns on.
You can stop the buzzer sounding and remote alarm operation by pressing
BUZZER key.

E02: Temp. sensor is short circuited.


When temp. sensor reads 60℃ or higher, E02 and ‘50’ are alternately
displayed.
Buzzer sounds intermittently and remote alarm turns on.
You can stop the buzzer sounding and remote alarm operation by pressing
BUZZER key.

Chamber temperature is set in 5℃ which is controlled by the temperature


in DEF sensor.

E03: DEF sensor is disconnected.


When DEF sensor reads -60℃ or lower, E03 and chamber temperature
are alternately displayed.
Buzzer sounds intermittently and remote alarm turns on.
You can stop the buzzer sounding and remote alarm operation by pressing
BUZZER key.

E04: DEF sensor is short circuited.


When DEF sensor reads 60℃ or higher, E04 and chamber temperature
are alternately displayed.
Buzzer sounds intermittently and remote alarm turns on.
You can stop the buzzer sounding and remote alarm operation by pressing
BUZZER key.
Defrosting is not performed correctly during E03 and E04 emitted.

E05: Comp. sensor is disconnected.


When comp. sensor reads -60℃ or lower, E05 and chamber temperature
are alternately displayed.
Buzzer sounds intermittently and remote alarm turns on.
You can stop the buzzer sounding and remote alarm operation by pressing
BUZZER key.
Specifications

Structural specifications
MPR-722 MPR-722R
External dimensions W770 x D830 x H1950 (mm)
Internal dimensions W770 x D710 x H1500 (mm)
Effective capacity 684 L 671 L
Exterior Painted steel
Interior Painted steel
Door Double layer pair glass/steel plate, automatic closing mechanism
Insulation Rigid polyurethane foamed-in place
Shelves Hard steel wire on polyester coating ------------
4 pcs.
Drawers ---------- Painted steel
5 pcs.
Access port φ30 mm in 3 locations (Left, right, top)
Cooling method Forced cool air circulation
Compressor Hermetic type Output; 200W
Condenser Wire and tube type
Evaporator Fin and tube type
Refrigerant R-513A
Defrosting Forced type, fully automatic
Internal heater 100W
Fluorescent lamp White light 40W x 1 (Glow starter; FG-4P)
Weight 174 kg 193 kg
Accessories 1 set of key,2 nylon clip, 2 binders
Optional Automatic temperature recorder (MTR-0621LH),
components Recorder fixture (MPR-S30),
Automatic temperature recorder (MTR-G04A)
Recorder fixture (MPR-S7)
Light shielding plate (MPR-72BP)
Battery box for power failure alarm (MPR-48B1)
Interface board (MTR-480、MTR-L03)
6. Alarms
High temp. alarm
When the chamber temperature is equal or higher than set temperature +
high temp. alarm set temperature + 0.09, ALARM lamp and temperature
display blink, buzzer sounds intermittently and remote alarm turns on after
delay time elapses.
High temp. alarm set temperature: +5℃ (default)
Delay time: 15 minutes (default)
When the chamber temperature is equal or lower than set temperature,
ALARM lamp goes off, temperature display stops blinking, buzzer stops
sounding and remote alarm turns off.
You can cancel the buzzer and remote alarm operation by pressing
BUZZER key during high temp. alarm.

Low temp. alarm When the chamber temperature is equal or lower than set temperature -
low temp. alarm set temperature - 0.09, ALARM lamp and temperature
display blink, buzzer sounds intermittently and remote alarm turns on after
delay time elapses.
High temp. alarm set temperature: -5℃ (default)
Delay time: 15 minutes (default)
When the chamber temperature is equal or higher than set temperature,
ALARM lamp goes off, temperature display stops blinking, buzzer stops
sounding and remote alarm turns off.
You can cancel the buzzer and remote alarm operation by pressing
BUZZER key during low temp. alarm.

0℃ alarm When the chamber temperature is lower than 0.0, ALARM lamp and
temperature display blink, buzzer sounds intermittently, remote alarm turns
on without delay.
When the chamber temperature is equal or higher than 0.0, ALARM lamp
goes off, temperature display stops blinking, buzzer stops sounding and
remote alarm turns off.
You can cancel the buzzer and remote alarm operation by pressing
BUZZER key during 0℃ alarm.

0℃ alarm has the higher priority than low temp. alarm.

Chamber fan alarm When the chamber alarm temperature is higher than the sum of setting
(default 5℃) temperature, the chamber fan alarm setting value and 0.09, ALARM lamp
and numerical value LED blink. Then buzzer sounds intermittently and
remote alarm is turn on after passing delay time 8default 15 min) .

When the chamber fan alarm temperature is or lower than the setting
temperature, ALARM lamp turns off, the display LED blinks and, buzzer
sound and remote alarm is turn off.
You can cancel the buzzer and remote alarm operation by pressing
BUZZER key during the information,
The chamber fan alarm setting value is same value as the high
temperature alarm se value on default. (default is set by F01, then the
lower limit is 5℃)
When the chamber alarm setting value is changed the independent setting
to valid “001” by F41, you can set the alarm setting value independently by
F39. However, during the door open, alarm has not sound.
During pull-down, buzzer has sounded and remote alarm has turn off.
However, Alarm lamp and the numerical value LED blink.
9. Compressor control

Compressor ON/OFF temperature (Differential point)


Compressor ON* When chamber temperature is equal to set temperature + 0.3℃
Compressor OFF When chamber temperature is equal to set temperature – 0.3℃
* Unit has 3 minutes delay time since the compressor is turned off, so the compressor is not
turned on even though the unit meets the above condition.

Compressor operation when temp. sensor is malfunction (E01, E02)


When temp. sensor is disconnected (E01) or short circuited (E02), compressor is controlled by
DEF sensor so that unit could keep chamber temperature in 5℃.
DEF sensor temp. to turn compressor ON DEF sensor temp. to turn compressor OFF
3.5℃ -3.5℃

Compressor protection
When comp. sensor reads 52℃ or higher, unit diagnoses compressor should stop operation.
When comp. sensor reads ambient temp. sensor value plus 10℃ or lower, unit diagnoses
compressor should start operation.

10. Defrosting
Mechanism
DEF sensor detects the temperature in evaporator pipe and unit diagnoses current should be
conducted to heater to defrost an evaporator.
During defrosting, compressor should stop operation, chamber temperature and “dF” are
alternately displayed.

By evaporator temperature
When DEF sensor temperature in evaporator reads equal or lower than -12℃, unit performs
defrosting.
When DEF sensor temperature in evaporator reads equal or higher than 5℃, unit stops
defrosting.

By other conditions
Unit performs defrosting if it meets the following conditions both (1) and (2).

(1) When compressor keeps running for 3 hours


(2) When DEF sensor temperature in evaporator reads lower than 0℃

The unit stops defrosting when DEF sensor temperature in evaporator reads equal or higher
than 5℃.
When the DEF sensor temperature in evaporator reads higher than 0℃ and the compressor
keeps running for 4 hours without defrosting, counting of compressor ON time is reset.

Condition to stop defrosting forcibly


When the chamber temperature is higher than set temperature +7.0℃ and DEF sensor
temperature is higher than the temperature to start defrosting, unit forcibly stops defrosting.
When this sequence repeats twice, unit never stops defrosting forcibly.

Frequency of defrosting
Unit can memorize frequency of forcible defrosting since the power is supplied.
You can check the frequency in the function mode, F36.
Specification of sensor

Temperature and resistance of thermistor sensor (Type: 502AT)


Resistance Resistance Resistance
Temp.(℃) Temp.(℃) Temp.(℃)
(kΩ) (kΩ) (kΩ)
-50 154.50 -7 17.92 12 8.17
-45 116.50 -6 17.16 13 7.85
-40 88.85 -5 16.43 14 7.55
-35 68.15 -4 15.74 15 7.27
-30 52.84 -3 15.08 16 6.99
-25 41.19 -2 14.45 17 6.73
-20 32.43 -1 13.86 18 6.48
-19 30.92 0 13.29 19 6.24
-18 29.50 1 12.74 20 6.01
-17 28.14 2 12.22 25 5.00
-16 26.87 3 11.72 30 4.18
-15 25.65 4 11.25 35 3.51
-14 24.51 5 10.80 40 2.96
-13 23.42 6 10.37 45 2.51
-12 22.39 7 9.96 50 2.14
-11 21.41 8 9.57 55 1.83
-10 20.48 9 9.20 60 1.57
-9 19.58 10 8.84
-8 18.73 11 8.49
Test Data

MPR-722 Pulldown speed(AT35 ℃)


45
230 V 50 Hz
115 V 60 Hz
35 Pullup

25
Internal temperature(℃)

15

-5

-15
0 1 2 2 3 4 5 6 7 7 8
Time scale(hour)

MPR-722 Pulldown pressure(AT35 ℃)


2.0

1.8 Peak(115 V 60 Hz) 1.778 MPa

1.6

1.4 Stabilized(230 V 50 Hz) 1.279 MPa

1.2 Peak(230 V 50 Hz) 1.765 MPa


Pd,Ps(Mpa)

Stabilized(115 V 60 Hz) 1.277 MPa

1.0 230 V 50 Hz Pd
230 V 50 Hz Ps
0.8
115 V 60 Hz Pd
0.6 Balance pressure 0.44 MPa 115 V 60 HzPs

0.4 Peak(230 V 50 Hz) 0.278 MPa


Stabilized(230 V 50 Hz) 1.168 MPa
0.2
Peak(115 V 60 Hz) 0.315 MPa Stabilized(115 V 60 Hz) 0.157 MPa
0.0
0 1 2 3 4 5 6 7 8
Time scale(hour)
Installation procedure of temperature recorder
Installation for MTR-0621LH
Recorder
Note panel
The temperature recording range of the automatic
temperature recorder (MTR-0621LH) is -60 to + 20 .
Temperature of + 20 or more can not be recorded. Shaft

CAUTION
When installing a automatic temperature recorder,
unplug the power plug. It may cause an electric shock. Fig 1

1. Remove the fixing screw on the recorder panel at the Automatic


temperature recorder mounting position and open the Recorder temperature
recorder panel toward you. (Fig. 1) Then, extend the fixture recorder
shaft shown in the figure to the left and right and remove
the recorder panel

2. Attach the automatic temperature recorder to the


recorder mounting bracket according to the installation
Fig2
procedure enclosed with the recorder mounting bracket.
(Fig 2)

3. Attach the recorder mounting bracket with the


automatic temperature recorder sensor to the self-
recording temperature mounting position of the front
cover. (Fig 3)

4. There is the access port on the right side of the top


surface of the chamber. Remove the 2 rubber caps
(inside and outside) of the measuring hole and the
insulation.

5. From the outside of the access port, pass through the Fig3
temperature-sensitive section of the self-recording
temperature sensor into the cabinet. (Fig 4)

Recorder
sensor

Fig 4
6. As shown in Fig. 5, use the supplied binder (2 pcs)
on the fan cover on the top surface to fix the
temperature-sensitive sensor of the automatic
temperature recorder sensor.

7. Fix the capillary of the temperature sensor in the


thermograph as shown in the figure using the supplied
nylon clip. (Fig 6)
Binder Fig5
8. Insert the cut into the rubber cap removed in step 4
and attach it with the insulation in the original position.
(Fig 7)

Nylon clip

Capillary

Fig 6

Capillary
Cap

Fig7
Installation for MTR-G04

1. Remove the fixing screw on the recorder panel at the


recorder mounting position and open the recorder panel Recorder
panel
towards you. (Fig. 1) Then, extend the shaft shown in the
figure to the left and right and remove the recorder panel.
Shaft
2. Attach the recorder mounting bracket to the self-
recording temperature recorder (MTR-G04) referring to
the installation manual enclosed with the recorder
mounting bracket (MPR-S7). (Fig 2) Fig1

Automatic
temperature Recorder
recorder fixture

Fig 2

3. There is the access port of the self-recording


temperature recorder on the right side of the top surface
of the chamber. Remove the 2 rubber caps (inside and
outside) of the measuring hole and the insulation.

Recorder
4. From the outside of the access port, pass through the sensor
temperature sensor of the self-recording temperature
sensor into the storage compartment. (Fig 3)

5. Fix the temperature sensitive part of the self- Fig 3


recording temperature sensor with the supplied nylon
clip (2 pcs) on the fan cover on the top as shown in Fig.4.

6. Fix the lead wire of the temperature sensor of the


self-recording temperature sensor as shown in Fig 5
using the supplied nylon clip.

7. Insert the cut into the rubber cap removed in step 3


and attach it with the insulation in the original position.
(Fig 6)
Nylon clip

Fig 4
8. Connect the connector on the temperature recorder
side to the power supply connector at the upper right
corner of the mounting position of the temperature
recorder. (Fig 7) Nylon clip

9. Align the groove of the mounting bracket with the shaft, Capillary
attach the self-recording temperature recorder to the
self-recording temperature mounting position of the
product, and fix it with one screw. (Fig 8)
Fig 5

Cap Wire

Fig 6

Power plug

Screw
Shaft

Fig 8
Automatic temperature recorder Fig 7 Slit
After installation of the optional battery box for power failure, be sure to make the following settings.
If the battery switch is turned on without the following settings, E09” and the chamber temperature is
displayed alternately in the temperature display .

● If “E09” is displayed, turn off the battery switch, then make the following settings and turn on the battery
switch again

operation key display

1 ----- Current chamber temp. is displayed

Press the up arrow key for 5 sec or


2 The first digit blinks
more
Set the first digit to 2 with the up Press the up arrow key and
3
arrow key. increase the number of settable digit
Press the scroll key and move
4
Set to F42 with scroll key and up the settable digit
arrow key Press the up arrow key and increase
5
the number of settable digit.

6 Press Set key. SET The first digit blinks.

Set the first digit to 1 with the up Press the up arrow key and increase
7
arrow key. the number of settable digit
Memorize the alarm temperature
8 Press set key SET setting value and current chamber
value