Professional Documents
Culture Documents
MODEL 19032
Key Features :
Twin port patented unique feature allows Hipot and Ground Bond be tested at the same time. TUV approved CE certified Programmable output voltage up to 5kV AC and 6kV DC Ground Bond Test up to 30A AC with adjustable limit Line/Earth leakage current measurement (UL2601-1/UL1950) with 5 human body models and 4 measurement RC networks Insulation resistance measurements from 1M to 50G Ground fault Interruption shutdown the instrument when imbalance current > 0.5mA, provide the highest protection capability Arc detection with programmable limit and 4 pluse width
Simulated leakage current measurement (UL 1950) Built-in 8 channels scanner option Large LCD display (320 x 240dot matrix) with great view and test result can read easily Programmable ramp and test times Continuous leakage current monitoring Trip current programmable up to 40mA AC and 12mA DC Front panel lockout via password Storage of 50 tests setups with 100 steps per setup Standard remote control interfaces for automatic control Optional IEEE,RS-232 and Printer interface Optional 40A Ground Bond
GPIB
RS-232C
PRINTER
Leakage Current
The steady-state current that flow through the insulation
production test.
Conventional SAfety Analyzer
Line S2 120V
Hipot
Neutral S1
S4
Table
S3
19032
2 Sec.
Probe 2 S5
Typical Test Value for Information equipment & Experiment Leakage Current Test, applying for IEC 60065 & IEC 60950 standards
Standard
Level
IEC 60065 34V peak or 100V DC 0.7mA peak or 2mA DC with 50k 34V peak or 100V DC 0.7mA peak or 2mA DC with 50k 34V peak or 100V DC 0.7mA peak or 2mA DC with 50k 34V peak or 100V DC 0.7mA peak or 2mA DC with 50k 34V peak or 100V DC 0.7mA peak or 2mA DC with 2k
IEC 60950 0.25mA, LCC requirements 0.75mA, LCC requirements 3.5mA, LCC requirements 3.5mA or 5% of input current if permanently LCC requirements Subjected to IEC 60065
Class I plug able type A and moveable Class I plug able type B
Isolating Transformer Leakage Current Test Applying for IEC 950 standards of Information equipment & Experiment Leakage Current Test
2 Selector Switch
Leakage Current Meter -Power actual working frequency up to several ten thousand Hz, safety requirement using simulation human body impedance to proceed Earth Leakage Current Measurement
10k 02F
Operation Caution
S.M.P.S. Test - No Floating Ground (standard) Test items: Ground Bond, P (L+N)-Case - With Grounded output and Floating Ground output simultaneously Test items: Ground Bond, P-Case+S (grounded) P-S (Floating ground), S (floating)-C
1500
0.022F 500
Voltmeter
No Ground Lead Test item: P-Case+S (ground) Plastic sealed and two line input Test item: P-S(ground) Plastic sealed and three line input (part design will use a connection line with chock between the secondary ground and the primary ground) Test items: Ground Bond, P-Case+S(ground) Transformer Test - Two groups coil (ARCing level 5 mA)
.......
* Measuring circuits for D.C. and A.C. with sinusoidal frequencies up to 1MHz
Arcing
Arc is electrical character occurred by voltage or current quickly changing. At the same time, it may happen short or zapping sound. No arcover or "sparking" should occur in an insulation stress test. When Arcs is lower 10 sec, it is no need to concern its effect degree, can regard as no harm spike discharge. you can choose one of 4 minimum Pulse Width (40/20/10/4sec) to test Arc for designing and product's reliable test. Arcing Test can help you to solve product's quality of screw loosen, bad material insulation, partial components over approaching.
Test items: P (coil 1) - S (coil 2), P (coil 1) - Core - Multi coils (ARCing level 5 mA) Test items: P (coil 1)-S1 (coil 2) + S2 (coil 3) ... S1 (coil 2) - S2 (coil 3), P (coil 1) -Core
..... .................
10 sec
Panel Description
11 29 6 7 8 30 31 32 1. LINE Switch 2. Window Display 3. Stop Key 4. Start Key 5. Cursor Keys 6. Menu Key 7. Print Key 8. Local Key 9. Cal-Enable 10. DATA Entry Key 11. Function Key 12. HV Output 13. Drive (+) 14. Sense (+) 15. Sense (-) 16. Drive (-) 27 18 24 20 17. Update Enable (Firmware is upgraded by PC directly.) 18. Remote I/O 19. LINE Voltage Selector 20. Power Cord Receptacle 21. GND Terminal 22. GPIB/Printer Interface (option) 23. Scan Interface 24. Fan 25. Optinal Rear Panel Output 26. Remote Interface 27. Scan Box (option) 28. RS-232 Interface (standard) 29. Data Entry Keys/ Program Keys 30~32. PASS/FAIL/TEST Indicator
3 4 1 2 25 22 23 19 9 17 26 10 13 14 1516 21 12
Specification
Model 19032 Mode DCV/ACV/IR/ GB/LC Withstanding Voltage Test Output Voltage DC : 0.05 ~ 6kV , AC : 0.05 ~ 5kV Load Regulation 1% + 5V Voltage Resolution 2V Voltage Accuracy 1% + 5 counts Cutoff Current DC : 12mA , AC : 40mA Current Resolution 0.1 A DC, 1 A AC Current Accuracy 1% + 5 counts Output Frequency 50/60Hz Test Time 0.3 ~ 999 sec., continue Ramp up Time 0.3 ~ 99.9 sec, Off Waveform Sine wave Insulation Resistance Test Output Voltage DC : 0.05 ~ 1kV Voltage Resolution 2V Voltage Accuracy 2% + 5V IR Range 1 M ~ 50G Resistance Resolution 0.1M Resistance Accuracy 5% typical Ground Resistance Test Output Current AC : 1 ~ 30A Current Resolution 10mA Current Accuracy 1% + 5 counts GR Range 10 ~ 510m Resistance Resolution 0.2m Resistance Accuracy 1% + 5 counts Test Method 4 wire measure Offset Function 0 ~ 100m Arc Detection Setting Mode Programmable setting Detection Current AC : 1~ 40mA, DC : 1~ 12mA Minimum Pulse Width 40/20/10/4 s selective Secure Protection Function Fast Output Cut-Off 0.4ms after NG happen Fast DC Charge < 0.2 sec. Panel Operation Lock Present password Power on Data Rally Yes Reset Protection Key Yes GO/NG Judgment GO : short sound Indication, Alarm NG : HI, LOW, ARC NG with long sound Data Hold Last tests data memories Step Hold Step signal trigger On/Off Memory Storage 50 steps with 100 groups recall Model 19032 Remote Connector Rear Panel 9 pin Input : Start, Stop, Interlock(at 11 pin terminal block only) D-type Connector Output : Under test, Pass, Fail Dynamic HV Leakage Current Scanner Option Mode LC/ VA Leakage Current Input Characteristic DC-1MHz Input Impedance 1M/20pF DUT Input Power Capacity AC : 300V/10 A Max. Short Protection 10A/250V Fuse Current Range 0.1 ~ 590A/ 500 ~ 6000 A Current Resolution 0.2 A/ 2 A Current Accuracy 2 % + 5 counts (DC - 5 kHz) Measurement Mode Normal, reverse, single fault normal, single fault reverse Measurement Devices UL 544 NP Medical appliances & equipment UL 544 P Medical appliances & equipment UL 1563 Electrical equipment & component UL 2601,IEC 601-1, Medical appliances & equipment EN 60601-1 UL 3101, UL 1950, Experiment & information UL 950, IEC 1010, equipment IEC 950, EN 60950 5 different kinds human body RC network Voltage Meter Voltage Range 0 ~ 300V AC Voltage Accuracy 1.5 % + 5 counts Current Meter Current Range 0.1 ~ 10 A AC Current Accuracy 1.5 % + 5 counts HI-LO Limit LC HI-LO Limit 0 ~ 9.99mA, 1A resolution Current HI-LO Limit 0 ~ 9.999A VA HI-LO Limit 0 ~ 2200VA VA Resolution 0.1VA General Temperature : 0 ~ 40 Operation Environment Humidity : 80% RH Power Consumption No load : < 100W With rated load : 650W Power Requirements 100V/120V/ 220V (AC 10%)/ 240V (AC +5%~ -10%),
50/60Hz
Weight Dimension(W X H X D)
Order Information
19032 A190301 A190302 A190303 A190304 A190305 A190306 A190307 A190313 A190314 A190316 A190317 A190321 A190324 A190327 A190334 Electrical Safety Analyzer 8HV Scanning Box 5HV/3GC Scanner 3HV/5GC Scanner 8HV Scanner Line Leakage Current Scanner (generally) Hipot/ Line leakage/ Probe Scanner Line leakage Current Scanner for EMI Filter 500VA Isolating Transformer 1000VA Isolating Transformer Dummy Load Bar Code Scanner GPIB Interface PRINTER Interface L-N Scanner Leakage Current Scanner 40A Ground Bond
Application
Production testing of medical electronic products and instruments, appliances, information technology equipment products such as blenders, fan and other Small Appliances, Component, in accordance with UL, CSA, and IEC TUV and standards such as UL 950, IEC 601-1, UL 2601-1, UL 1563, UL 544, EN 60335, EN 60950, EN 61010, CSA C22.2 No. 1010.1, UL 3111 and UL 1950 Electrical safety compliance testing on a variety of products and devices Transformer electrical safety testing Electric motor safety
Distributed by:
China 8F, No.4, Nanyou Tian An Industrial Estate Shenzhen, China Post Code: 518052 Tel: +86-755-2664-4598 Fax: +86-755-2641-9620 U. S. A. 7 Chrysler Irvine CA 92618 Tel: +1-949-421-0355 Fax: +1-949-421-0353 Toll Free: +1-800-478 2026 Europe Max Planckstraat 4 6716 BE Ede The Netherlands Tel: +31-318-648282 Fax: +31-318-648288