Uncertainty Analysis
Error propagation
Design-Stage Uncertainty
Zero-Order Uncertainty
Instrument Uncertainty
Root-sum-square method
Design-Stage Uncertainty
calibration,
concomitant methodology,
interlaboratory comparisons, or
judgment/experience.
Random Error
When repeated measurements are made under fixed operating conditions,
random errors manifest themselves as scatter of the measured data.
Random error4 is introduced through the
repeatability and resolution of the measurement system components,
calibration,
and measurement procedure and technique;
by the measured variables own temporal and spatial variations; and by the variations in the process
operating and environmental conditions from which measurements are taken.
Error propagation
Example 5.4:
As an exercise, obtain and examine a dial oven thermometer. How would you
assess the zero- and first-order uncertainty in the measurement of the
temperature of a kitchen oven using the device?
Example 5.5:
A stopwatch is to be used to estimate the time between the start and end of
an event. Event duration might range from several seconds to 10 minutes.
Estimate the probable uncertainty in a time estimate using a hand-operated
stopwatch that claims an accuracy of 1 min/month (95%) and a resolution of
0.01 s.
Example 5.6:
A flow meter can be calibrated by providing a known flow rate through the meter
and measuring the meter output. One method of calibration with liquid systems
is the use of a catch and time technique whereby a volume of liquid, after
passing through the meter, is diverted to a tank for a measured period of time
from which the flow rate volume/time, is computed. There are two procedures
that can be used to determine the known flow rate Q in, say, ft3/min:
The volume of liquid, V, caught in known time t can be measured.
Suppose we arbitrarily set t=6 s and assume that our available facilities can determine volume (at Nth order)
to 0.001 ft . Note: The chosen time value depends on how much liquid we can accommodate in the tank.
3
Suppose we determine an Nth-order uncertainty in time of 0.15 s. In either case the same instruments are
used. Determine which method is better to minimize uncertainty over a range of flow rates based on these
preliminary estimates
Advanced-Stage Uncertainty
Analysis
In designing a measurement system, a pertinent question is, How would it affect the result if
this particular aspect of the technique or equipment were changed?
In design-stage uncertainty analysis, we only considered the errors due to a measurement systems
resolution and estimated instrument 176 Chapter 5 Uncertainty Analysis calibration errors.
Advanced-Stage Uncertainty
Analysis (cont.)
Zero-Order Uncertainty
At zero-order uncertainty, all variables and parameters that affect the outcome of the
measurement, including time, are assumed to be fixed except for the physical act of
observation itself.
Higher-Order Uncertainty
Higher-order uncertainty estimates consider the controllability of the test operating
conditions and the variability of all measured variables.
For example, at the first-order level, the effect of time as an extraneous variable in the
measurement might be considered.
The uncertainty at u includes the effects of resolution, u . So only when u u is time
not a factorin the test.
In itself, the first-order uncertainty is inadequate for reporting of test results.
1
Advanced-Stage Uncertainty
Analysis (cont.)
Nth-Order Uncertainty
At the Nth-order estimate, instrument calibration characteristics are entered into
the scheme through the instrument uncertainty uc.
Note that as a minimum, design-stage analysis includes only the effects found as a
result of u0 and uc. It is those in-between levels that allow measurement procedure
and control effects to be considered in the uncertainty analysis scheme. The Nthorder uncertainty estimate provides the uncertainty value sought in advanced
design stage or in single-measurement analyses. It is an appropriate value to be
used to report the results from single-measurement tests.