Professional Documents
Culture Documents
Analytical Considerations: Alfredo García - Arieta
Analytical Considerations: Alfredo García - Arieta
Existing guidelines
ICH Guidance for industry
Validation of analytical methods: definitions and terminology,
June 1995
Validation of analytical procedures: methodology, November
1996
2|
GCP/GLP compliance
Clinical studies have to be performed under conditions
complying with the principles of Good Clinical Practice,
and for analytical methods and sample data handling
conditions complying with the principles of Good
Laboratory Practice are required.
For older studies without statement of compliance with
the above mentioned principles, the assessor should rely
on the quality of the submitted report
3|
GCP/GLP
GCP/GLP are required to assure:
Quality and Credibility
4|
Reference standard
Analysis of drugs and their metabolites in a biological
matrix is carried out using samples spiked with calibration
(reference) standards and using quality control (QC)
samples
The purity of the reference standard used to prepare
spiked samples can affect study data
For this reason, an authenticated analytical reference
standard of known identity and purity should be used to
prepare solutions of known concentrations
5|
Reference standard
The source and lot number, expiration date, certificates of
analyses when available, and/or internally or externally
generated evidence of identity and purity should be
furnished for each reference standard
If possible, the reference standard should be identical to
the analyte
When this is not possible, an established chemical form
(free base or acid, salt or ester) of known purity can be
used
6|
Reference standard
Three types of reference standards are usually used:
(1) certified reference standards (e.g., USP compendial
standards);
(2) commercially supplied reference standards obtained from a
reputable commercial source;
(3) other materials of documented purity custom-synthesized by
an analytical laboratory or other noncommercial establishment
7|
METHOD DEVELOPMENT
A specific, detailed description of the bioanalytical method
should be written
This can be in the form of a protocol, study plan, report,
and/or SOP
Any method is valid as long as it is validated
8|
METHOD VALIDATION
Procedures that demonstrate that a method is reliable
and reproducible for the intended use.
Types:
Full validation: first time, new drug, or addition of metabolites
Partial validation: modifications of a validated method
Cross-validation: comparison between methods
9|
PRE-STUDY VALIDATION
Fundamental parameters for validation
Method used for the determination of drugs and/or
metabolites should be:
Sensitive
Selective
Accurate
Precise
Stable
10 |
Selectivity
Selectivity is the ability of an analytical method to
differentiate and quantify the analyte in the presence of
other components in the sample
For selectivity, analyses of blank samples of the
appropriate biological matrix (plasma, urine, or other
matrix) should be obtained from at least six sources
Each blank sample should be tested for interference, and
selectivity should be ensured at the lower limit of
quantification (LLOQ)
11 |
Selectivity
Potential interfering substances in a biological matrix
include endogenous matrix components, metabolites,
decomposition products, and in the actual study,
concomitant medication and other exogenous xenobiotics
If the method is intended to quantify more than one
analyte, each analyte should be tested to ensure that
there is no interference
12 |
Outliers
Reported method validation data and the determination of
accuracy and precision should include all outliers
However, calculations of accuracy and precision
excluding values that are statistically determined as
outliers can also be reported
13 |
Accuracy
The accuracy of an analytical method describes the
closeness of mean test results obtained by the method to
the true value (concentration) of the analyte
Accuracy is determined by replicate analysis of samples
containing known amounts of the analyte (QC samples)
Accuracy should be measured using a minimum of five
determinations per concentration
14 |
Accuracy
A minimum of three concentrations in the range of
expected concentrations is recommended
The mean value should be within 15% of the actual value
except at LLOQ, where it should not deviate by more than
20%
The deviation of the mean from the true value serves as
the measure of accuracy
15 |
Precision
The precision of an analytical method describes the
closeness of individual measures of an analyte when the
procedure is applied repeatedly to multiple aliquots of a
single homogeneous volume of biological matrix
Precision should be measured using a minimum of five
determinations per concentration
A minimum of three concentrations in the range of
expected concentrations is recommended
16 |
Precision
The precision determined at each concentration level
should not exceed 15% of the coefficient of variation (CV)
except for the LLOQ, where it should not exceed 20% of
the CV
Precision is further subdivided into within-run, intra-batch
precision or repeatability, which assesses precision
during a single analytical run, and between-run, interbatch precision or repeatability, which measures precision
with time, and may involve different analysts, equipment,
reagents, and laboratories
17 |
18 |
Recovery
The recovery of an analyte in an assay is the detector
response obtained from an amount of the analyte added
to and extracted from the biological matrix, compared to
the detector response obtained for the true concentration
of the pure authentic standard
Recovery pertains to the extraction efficiency of an
analytical method within the limits of variability
19 |
Recovery
Recovery of the analyte need not be 100%, but the extent
of recovery of an analyte and of the internal standard
should be consistent, precise, and reproducible
Recovery experiments should be performed by
comparing the analytical results for extracted samples at
three concentrations (low, medium, and high) with
unextracted standards that represent 100% recovery.
20 |
23 |
24 |
25 |
Matrix effect
In the case of LC-MS-MS-based procedures, appropriate
steps should be taken to ensure the lack of matrix effects
throughout the application of the method, especially if the
nature of the matrix changes from the matrix used during
method validation
Whenever possible, the same biological matrix as the
matrix in the intended samples should be used for
validation purposes
For tissues of limited availability, such as bone marrow,
physiologically appropriate proxy matrices can be
substituted
26 |
Dilution of samples
27 |
Stability
Drug stability in a biological fluid is a function of the
storage conditions, the chemical properties of the drug,
the matrix, and the container system
The stability of an analyte in a particular matrix and
container system is relevant only to that matrix and
container system and should not be extrapolated to other
matrices and container systems
The stability of the analyte should be established
preferably prior to sample analysis
28 |
Stability
Stability procedures should evaluate the stability of the
analytes during sample collection and handling, after
long-term (frozen at the intended storage temperature)
and short-term (bench top, room temperature) storage,
and after going through freeze and thaw cycles and the
analytical process
Conditions used in stability experiments should reflect
situations likely to be encountered during actual sample
handling and analysis
29 |
Stability
The procedure should also include an evaluation of
analyte stability in stock solution.
All stability determinations should use a set of samples
prepared from a freshly made stock solution of the
analyte in the appropriate analyte-free, interference-free
biological matrix
Stock solutions of the analyte for stability evaluation
should be prepared in an appropriate solvent at known
concentrations
30 |
32 |
3. Long-Term Stability
The storage time in a long-term stability evaluation should
exceed the time between the date of first sample
collection and the date of last sample analysis
Long-term stability should be determined by storing at
least three aliquots of each of the low and high
concentrations under the same conditions as the study
samples
33 |
3. Long-Term Stability
The volume of samples should be sufficient for analysis
on three separate occasions
The concentrations of all the stability samples should be
compared to the mean of back-calculated values for the
standards at the appropriate concentrations from the first
day of long-term stability testing
34 |
35 |
5. Post-Preparative Stability
The stability of processed samples, including the resident
time in the autosampler, should be determined
The stability of the drug and the internal standard should
be assessed over the anticipated run time for the batch
size in validation samples by determining concentrations
on the basis of original calibration standards
36 |
APPLICATION TO
ROUTINE DRUG ANALYSIS
Assays of all samples of an analyte in a biological matrix
should be completed within the time period for which
stability data are available
Samples can be analyzed with a single determination
without duplicate or replicate analysis if the assay method
has acceptable variability
Where high precision and accuracy may be difficult to
achieve, duplicate or even triplicate analyses can be
performed for a better estimate of analyte
37 |
Calibration curve
A calibration curve should be generated for each analyte
to assay samples in each analytical run and should be
used to calculate the concentration of the analyte in the
unknown samples in the run
An analytical run can consist of QC samples, calibration
standards, and either (1) all the processed samples to be
analyzed as one batch or (2) a batch composed of
processed unknown samples of one or more volunteers in
a study
QC
38 |
QC
QC
QC
QC
QC
Calibration curve
Standards & QC samples can be prepared from the same
spiking stock solution, if stability and accuracy is verified
A single source of matrix may be used, if selectivity is
verified
Standard curve samples, blanks, QCs, and study samples
can be arranged as considered appropriate within the run.
Placement of standards and QC samples within a run
should be designed to detect assay drift over the run
QC
39 |
QC
QC
QC
QC
QC
Calibration curve
The calibration (standard) curve should cover the
expected unknown sample concentration range in
addition to a calibrator sample at LLOQ
Estimation of concentration in unknown samples by
extrapolation of standard curves below LLOQ or above
the highest standard is not recommended
Instead, the standard curve should be redefined or
samples with higher concentration should be diluted and
reassayed
40 |
Calibration curve
A matrix-based standard curve should consist of a
minimum of six standard points, excluding blanks (either
single or replicate), covering the entire range.
The same curve fitting, weighting, and goodness of fit
determined during pre-study validation should be used for
the standard curve within the study
Changes in the response function relationship between
pre-study validation and routine run validation indicate
potential problems
41 |
42 |
43 |
44 |
45 |
46 |
47 |
Repeat analysis
SOP or guideline for repeat analysis, reasons for repeating and
acceptance criteria
Reasons for repeat analyses could include:
48 |
49 |
DOCUMENTATION
Documentation of successful completion of validation
studies should be provided in the assay validation report
General and specific SOPs and good record keeping are
an essential part of a validated analytical method
The data generated for bioanalytical method
establishment and the QCs should be documented and
available for data audit and inspection
50 |
Documentation to submit
Documentation for submission to the Agency should
include:
(1) summary information,
(2) method development and establishment,
(3) bioanalytical reports of the application of any methods to
routine sample analysis, and
(4) other information applicable to method development and
establishment and/or to routine sample analysis.
51 |
52 |