AGENDA
• Beer law
• Spectrophotometer block diagram
• Source of drift
• Conventional Spectrophotometer vs. double
beam spectrophotometer
• Beam splitter and chopper
• Spectrophotometers instruments
Light Transmission Dependence on
Concentration
Beer’s Law
The Beer-Bouguer-Lambert Law
I0 I
A b c
molecular absorptivity distribution
curve
molecular absorptivity distribution
curve
molecular absorptivity distribution
curve
Run spectrophotometer
Absorbance
1. Obtain a monochromatic
wavelength for the maximum 2.0
absorption wavelength.
0.0
Wavelength (nm)
2. Make a dilution series of a known quantity of
analyte and measure the Absorbance
3. Plot concentration vs. Absorbance.
Absorbance at 540 nm
1.2
0.8
0.4
1 2 3 4
Concentration (g/l) glucose
Ideal: Absorbance determined using Beer Law
Observed: Absorbance determined using instrument
4. From the figure determine the
concentration of unknown solution from
knowing the absorbance.
What concentration do you think the unknown
sample is?
Warning:
Absorbance at 280 nm
1.0
A
0.5 Slope of Standard Curve =
C
1 2 3 4 5
Concentration (mg/ml)
Absorbance at 540 nm
1.2
0.8
0.4
1 2 3 4
Concentration (g/l) glucose
Absorption
Spectra Plot of
Absorbance vs.
wavelength
called
absorption
spectrum.
Emission Spectra
Emission
Spectra Plot of
emission
intensity vs.
wavelength
called
emission
spectrum.
Light Transmission Dependence on
Concentration
Beer’s Law
Transmittance and Path Length:
Beer’s Law
Concentration
ConstConcentration
T I / I0 e
Transmittance and Concentration
The Bouguer-Lambert Law
ConstPathlength
T I / I0 e
BEER LAMBERT LAW
Light
I0 I