according to the LipoClear® Product Insert Sheet. Upon obtaining results, the values obtained on the treated samples were subsequently multiplied by a 1.2 dilution factor to account for the dilution of the actual reagent. (StatSpin® LipoClear) • LipoClear Tubes are 1.5ml microcentrifuge tubes pre-filled with LipoClear reagent. LipoClear is a convenient, non-toxic, non- carcinogenic liquid polymer reagent which clears lipaemic serum or plasma after mixing and centrifugation. Simple Procedure
• The simple procedure involves pipetting 0.5ml
of sample into a LipoClear Tube, mixing and leaving for 5 minutes at room temperature, and then centrifuging. Ideally, centrifuges spinning at 10,000G or more can be used and will allow separation and clearing in 2 minutes. Low speed centrifuge for example 2000G will need 20 minutes for spinning LipoClear Tubes. Tabung lipoclear 1,5 ml • After centrifuging, the cleared sample must be removed for analysis. However, please note that all results must be multiplied by 1.2 to compensate for the sample dilution.