Professional Documents
Culture Documents
Principles of Biomedical Systems & Devices
Principles of Biomedical Systems & Devices
Devices
0909.504.04 / 0909.402.02
WEEK 10:
Flow & Volume of Blood
Measurement of
Flow & Volume of Blood
Change in [] due to dm dt
continuously added C
indicator m to volume V dV dt
O 2 consumption(mL/ min)
Cardiac Output
[O 2 ]a [O 2 ]v
250 mL/ min
5 L/ min
190mL / L 140mL/L
m
F t
C (t )dt
0
Indicator – Dilution Curve
After the bolus is injected at time A, there is a transportation delay before the
concentration begins rising at time B. After the peak is passed, the curve enters
an exponential decay region between C and D, which would continue decaying
alone the dotted curve to t1 if there were no recirculation. However, recirculation
causes a second peak at E before the indicator becomes thoroughly mixed in the
blood at F. The dashed curve indicates the rapid recirculation that occurs when
there is a hole between the left and right sides of the heart.
An Example
Dye Dilution
The indicator is cold – saline, injected into the right atrium using a
catheter
Temperature change in the blood is measured in the pulmonary
artery using a thermistor
The temperature change is inversely proportional to the amount of
blood flowing through the pulmonary artery
Measuring Cardiac Output
Several methods of measuring cardiac output In the Fick method, the indicator is O 2;
consumption is measured by a spirometer. The arterial-venous concentration difference is
measure by drawing simples through catheters placed in an artery and in the pulmonary artery.
In the dye-dilution method, dye is injected into the pulmonary artery and samples are taken
from an artery. In the thermodilution method, cold saline is injected into the right atrium and
temperature is measured in the pulmonary artery.
Electromagnetic
Flowmeters
Based on Faraday’s law of induction that a conductor that moves
through a uniform magnetic field, or a stationary conductor placed
in a varying magnetic field generates emf on the conductor:
Zero-crossing
detector / LPF
Determine
direction
High acoustic
impedance material
Transit time flowmeters
Effective velocity of sound in blood: velocity of sound (c) +
velocity of flow of blood averaged along the path of the ultrasound (û)
u
F f s (cos cos )
c
Problems Associated with
Doppler Flowmeters
There are two major issues with Doppler flowmeters
Unlike what the equations may suggest, obtaining direction information is
not easy due to very small changes in frequency shift that when not in
baseband, removing the carrier signal without affecting the shift frequency
becomes very difficult
Also unlike what the equation may suggest, the Doppler shift is not a single
frequency, but rather a band of frequencies because
• Not all cells are moving at the same velocity (velocity profile is not
uniform)
• A cell remains within the UT beam for a very short period of time; the
obtained signal needs to be gated, creating side lobes in the frequency
shift
• Acoustic energy traveling within the beam, but at an angle from the bam
axis create an effective ∆θ, causing variations in Doppler shift
• Tumbling and collision of cells cause various Doppler shifts
Directional Doppler
(a) Quadrature-phase detector. Sine and cosine signals at the carrier frequency are summed with
the RF output before detection. The output C from the cosine channel then leads (or lags) the
output S from the sine channel if the flow is away from (or toward) the transducer. (b) Logic
circuits route one-shot pulses through the top (or bottom) AND gate when the flow is away from
(or toward) the transducer. The differential amplifier provides bi-directional output pulses that
are then filtered.