Professional Documents
Culture Documents
Ca Cervix Part 4/6
Ca Cervix Part 4/6
Carcinoma Cervix
Moderater - Dr Rajeev Jain sir
Presented by - Dr Rajendra Patel
BRACHYTHERAPY
• Brachytherapy is a type of radiation treatment in
which small, encapsulated radioactive sources are
arranged in a geometric fashion in & around tumor.
• ADV.
– It delivers very high dose of radiation to tumor
– Sparing normal tissue
– Dose delivered in short duration as compared to External
beam RT.
BRACHYTHERAPY in Cervix
• Brachytherapy plays vital role in treatment of ca cx. & is
mainly applied as an intracavitary procedure in selected
cases complemented by interstitial implants.
– 4cm
– 6cm
• Packing helps to
– keep the applicators in position
– to reduce dose to bladder and anterior rectal wall.
RULES
• The point A should receive the same dose rate, irrespective of the
combination of applicators used.
• Not more than one third of the total dose to point A should
be
delivered by the vaginal ovoids. So that tolerance of vagina mucosa is
not exceeded
• Standard or ideal loading is 60-40 i.e. 60% of the dose to point A is
contributed by intrauterine sources while 40% is contributed
by ovoids.
• Total Dose to point A : 8000 R
– Total number of applications : 2
– Total time for each application : 72 hrs
– Total time : 144 hrs
– Dose rate desired : 55.5 R /hour to point A
• Amount of radium to be used was defined in terms of units.
• 1 unit = 2.5 mg of radium filtered by 1 mm platinum.
• The loadings were specified in terms of integral multiples of this unit.
LOADING PATTERN
• Total dose at point A using different combinations of
U. tube & ovoids :
– Large tube with large ovoid and washer : 57.5 R
– Large tube with large ovoid and spacer: 56.9 R
– Large tube with small ovoid and washer: 57.6 R
– Medium tube with small ovoids and spacer: 57.3 R
– applicator type
– Loading pattern
• A trapezoid is constructed in a
plane passing through transverse
line in pelvic brim plane and
midpoint of ant. aspect of body
of L4
PELVIC WALL REFERENCE POINTS
• The wall reference
represents absorbed dose at the distal part
pelvic
of the parametriumpoint,and at the
obturator
L.N.
• Reporting dose at reference points related
to well defined bony structures & L.N.
areas is particularly useful when I/C BT
is combined with EBRT
• On a AP radiograph, pelvic-wall reference
point is located at intersection of following
lines
– a horizontal line tangential to the highest
point of the acetabulum,
– a vertical line tangential to
the inner
aspect of the acetabulum.
• On a lat. radiograph, the highest points of
the right & left acetabulum, in
cranio -
caudal direction, are joined &
lateral
projection of the pelvic-wall reference point
REFERENCE VOLUME
• Volume encompassed by
reference theisodose, selected and
specified to compare treatments
performed in different centres using
different techniques.
• Pt is anaesthesitized.
• Patient is in lithotomy
position
• Perineal area is
disinfected
Anesthesi
• a
Potential options include general, spinal or iv
conscious sedation.
• Dilatation of the
cervix with standard
tooling.
Tandem
placement:
• Uterine sound
• Hegar uterine dilators are used to dilate the os to 6
mm.
• Tandem length is usually 6–8 cm but its
variable from patient to patient.
• For tandems >8 cm, avoid loading/active
source at end to protect small bowel.
• Tandem should be located centrally between the
ovoids on the AP view and bisect the ovoids on
the lateral view.
PROCEDURE
• Correct length of IU-tube
ovoids
& are selected
• Inserted by one and
attached
one to fixing
mechanism.
• To determine the rectal
on CT or radiograph a radio
wall
opaque marker is inserted
• After insertion of
gauze packing is done
applicator
behind the ovoids to push
rectum and bladder
reducing
away the dose to these
organs
• After procedure orthogonal radiographs are taken
to check applicator geometry.
Brachytherap
• y
Stage IA (microinvasive) tumors
– Treated with Brachytherapy alone
– LDR dose is approximately 60 Gy in one insertion or
75 to 80 Gy in two insertions to point A
OR
– with HDR an equivalent
dose, with one or two
fractions per week.
• A reasonable target is to
give to Point A
– 65-75 Gy for Stage IA
disease
– 75-85 Gy for IB - IIB
LIMITATIONS OF (2 D)RADIOGRAPHIC
IMAGING
For determination of target
• point based dosimetry
• point A may overestimate or underestimate the tumor dose based on
3D imaging*
• no optimization:
• tumor coverage relies on tumor volume at time of BT, larger tumors
requiring greater optimization to be adequately covered by the prescribed
isodose line
• Kim et al** found that dose to point A was significantly lower than
the D90 for HR-CTV calculated using 3D image-based optimization
• *Kim RY, Pareek P. Radiography-based treatment planning compared with computed tomography (CT)-based treatment planning for
intracavitary brachytherapy in cancer of the cervix: analysis of dose-volume histograms. Brachytherapy 2003;2:200–206.
• **Kim H, Beriwal S, Houser C, et al. Dosimetric analysis of 3D image-guided HDR brachytherapy planning for the treatment of
cervical cancer: is point A-based dose prescription still valid in image-guided brachytherapy? Med Dosim 2011;36:166–170.
AIM
to analyze dosimetric outcome of 3D IGBT &compare dose coverage of HRCTV
to traditional Point A dose.
• N=32patients (stage IA2-IIIB cervical cancer) treated with IGBT
• dose: 5.0-6.0 Gy/# ×5 fractions.
• delineation of CTV as per GYN GEC/ESTRO guidelines.
• D90 for HRCTV was 80-85 Gy,
• D2cc of bladder, rectum, and sigmoid was limited to 85 Gy, 75 Gy &75 Gy.
• RESULTS
• The mean D90 for HRCTV was 83.2 ± 4.3 Gy SD significantly higher (p
<0.0001) than mean value of Point A dose (78.6 ± 4.4 Gy).
• The dose levels of the OARs were within acceptable limits
• Dose to Point A was found to be significantly lower than the D90 for HRCTV
• rectal markers is used which tend to lie on posterior wall of rectum while the
anterior wall is at greater risk.
• Stiff markers can move rectum, flimsy ones are difficult to push deep.
• ICRU rectal point doesn’t usually represent the maximum rectal does,
which, again often is 2-4 cm cephalad.
• maximum does is up to 3 times the ICRU point
None of this localizes the superior bowel - an organ very much at risk.
To be continued
Thank you
c A C E R V I X PA R T 5
Contents
Pre-implant evaluation
Applicator selection & insertion
Imaging
Delineation
Prescription
Treatment planning
Target dose specification
Treatment delivery
American Brachytherapy society cervical cancer task group
Pre-implant evaluation
Advantages:-
verifies proper placement of applicator
reasonable estimate of the location of uterus
fairly good for visualizing bladder and rectum.
analyses 3D BT dose distribution
depicts changes in the OAR related to tumour shrinkage & filling status.
3D dose calculations & optimisation possible
OAR dosimetry based on CT is similar to that based on MRI when
optimized similarly*
long experience in treatment planning for EBRT
readily available in radiotherapy departments
*Eskander RN, Scanderbeg D, Saenz CC, et al. Comparison of computed tomography and
magnetic resonance imaging in cervical cancer brachytherapy target and normal tissue
contouring. Int J GynecolCancer 2010;20:47–53
Problems with CT treatment planning
Contouring of OARS
rectum : began 1 cm above the anus, ended at the sigmoid
flexure, and covered the outer wall of the organ.
sigmoid :begin at the level of the rectosigmoid flexure and
ended at the anterior crossing of the sigmoid by the pubic
symphysis.
bladder contour included the outer wall of the bladder and
ended at the beginning of the urethra.
Purpose: To compare contours and DVH of tumor & OAR
with CT vs. MRI in cervical cancer BT
A standardized approach to contouring on CT (CTStd) was
developed, implemented and compared with the MRI
contours
primary endpoint :assess the feasibility of using these CT-
standardized (CTStd) contours to approximate MRI-based
treatment parameters.
secondary endpoint: to determine whether CT and MRI
provide dosimetrically similar results for the organs at risk
(OARs)
CT Vs MRI CONTOURING
Vishwanathan et al. IJROBP 2007
10 patients (Stage IIA–IIIB)
underwent pelvic EBRT+ CT f/b tandem and ring BT
Planning CT and MRI were performed and contouring carried out
separately,no contrast used in study.
MRI contoured in accordance with the GEC-ESTRO
recommendations
The CT and MRI volumes were fused
CT contours of the HR-CTVCT and IR-CTVCT were adapted
based on the GEC-ESTRO recommendations for MRI.
The GTV could not be defined on CT. Also difficulty delineating the
superior border of cervix and lateral border of parametria (if
involved) and accurate delineation of the OARs
Volumetric and DVH values
A two-sided t test comparing mean values of height, thickness,&
volume showed no significant differences among three.
However, the tumor width was significantly different for HR-
CTVCTStd and compared with MRI values.
difference in width resulted in statistically significant differences in
D100 and D90.
the tumor width was significantly different for IR-
CTVCTStd compared with corresponding MRI values
resulting in statistically significant differences in D100,
and D90
No statistically significant differences in the dose to 0.1 cm3, 1
cm3, and 2 cm3 for the OARs were noted
RESULTS
CT significantly overestimated the width of the tumor and altered the
D90, D100
OAR DVH were the same
MRI remains standard for contouring
Planning : TPS
Target dose specification
OAR specification
Catheter reconstruction:
Loading pattern
dose specification & optimisation method, if applied
Radiotherapy and Oncology 78 (2006)
67–77
www.thegreenjournal.com
3D dose-volume parameters for BT of cervical carcinoma
Defined dose volume parameters for target & OAR
Cumulative dose volume histograms (DVH) are recommended for
evaluation
Target dose specification for
GTV,HRCTV & IRCTV with their uses
& limitation
Limitation:-
V100 is based on prescribed physical dose, only relevant
within a specific dose rate and fractionation, cannot be used
for intercomparison purposes, should be applied solely for
intra-patient plan comparison
The intercomparison problem is avoided when BED are used, e.g.
V(60 GyEQD2),V(85 GyEQD2).
For fractionated treatment, however, this parameter is only usable
for evaluation after last fraction, as it uses summed doses of
all #.
Dose volume parameters for OAR
Typical adverse effects from BT such as ulceration, necrosis and
fistulas occur mainly in limited volumes adjacent to applicator
irradiated with high doses (>80 Gy),
whole organ side effects like overall organ inflammation,
fibrosis or telangiectasia occur mainly after whole organ
irradiation with intermediate or high doses (60–70CGy). within
short time periods
When assessing late effects from BT , small organ (wall)
volumes irradiated to a high dose is of major interest.
As there is rapid dose fall-off near the sources, adjacent small
organ (wall) volumes, dose assessment has to refer to one (or
more) defined dose point(s) in these limited volumes.
OAR specification with
volumetric imaging
recommendations suggest three
quantities to characterize the dose
distribution for OAR. .
minimum dose in the most
irradiated tissue volume
adjacent to the applicator (0.1, 1
& 2 cc) is recommended for
recording
D2 cc can be useful during dose
planning and for evaluating
toxicities
D0.1 cc is indicative of the
maximum dose.
RECOMMENDED DOSE
PRESCRIPTION
TARGET
For HDR: cumulative dose from EBRT +BT
HR-CTV dose: EQD2=85-90Gy
The IR-CTV dose: EQD2=60 Gy.
Target coverage D90 should be equal to 100 % prescribed
dose
OAR
D2cc bladder ≤90 Gy EQD2
D2cc rectum & sigmoid ≤70-75Gy EQD2
Radiotherapy and Oncology (2010)
In IGBT :geometry of the applicator is extracted from patient 3D images
and introduced into TPS
Due to the steep dose gradients, reconstruction errors can lead to
major dose deviations
applicator commissioning and reconstruction methods must be
implemented in order to minimise errors.
Applicator commissioning verifies location of source positions in
relation to applicator
for optimal visualisation of applicator. Para-transverse imaging with
small slice thickness (≤5 mm) is recommended
contouring and reconstruction should be performed in same image
series in order to avoid fusion uncertainties.
Under well-controlled circumstances reconstruction uncertainties are in
general smaller than other brachytherapy uncertainties.
MR imaging criteria have to be fulfilled.
Technical requirements, patient preparation, as well as image
acquisition protocols have to be tailored to the needs of BT
pelvic MRI scanning to be performed prior to radiotherapy
(‘‘Pre-RT-MRI examination’’) and at time of BT (‘‘BT MRI
examination’’) with one MR imager.
Multiplanar (transversal, sagittal, coronal and oblique image
orientation) T2- weighted images obtained with pelvic surface
coils are considered as the golden standard for visualisation of
the tumour & OAR.
Radiotherapy and Oncology (2012)
Thank you
Radium 226
Half life – 1600 yr
Photon energy – 0.047-2.45
MeV (0.83 avg)
HVL – 12.0mm
Disintegrate to Ra 222+He 4
both are gas
49 gamma rays
Radium sulfate / radium
chloride
0.1-0.2 mm gold foil
Full intensity 0.66 mg/cm
Half intensity 0.33 mg/cm
55Cesium 137
Half life – 30 year
Photon energy 0.662 MeV
HVL - 5.5
Monoenergetic Gamma emitting
Insoluble powder or ceramic microsphere
Doubly encpsulated in stainless steel needles or tubes
27 Co - 60
Half life 5.26 year
Photon energy 1.17 & 1.33 MeV
HVL – 11 mm
High specific activity
Encpsulated in platinum iridium or stainless steel wire
77 Iridium 192
Half life – 73.8 days
Photon energy – 0.136-1.06 (0.38 avg) MeV
HVL – 2.5 mm
Alloy of 30% Ir + 70% Pt flexible wire or nylon ribbons
containing Ir seeds of 3mm long and 0.5mm diameter
53 Iodine 125
Half life - 59.4 days
Energy -0.028 MeV
HVL – 0.025 mm
Used for permanent implat
Model 6711 contain silver
wire with silver iodide
adsorbed on its surface.
Figure B
Depending upon loading Based on dose rates used
technology LDR - 0.4-2 Gy/ hr
Preloaded – radium tubes Ra 226, Cs 137
Afterloaded MDR - 2-12 Gy/hr
Manual – Cs 137 Cs 137
Remote – Cs 137, Co 60, HDR > 12 Gy/hr
Ir192 Co 60, Ir192
System of implant
dosimetry
STOCKHOLM SYSTEM
Fractionated (2-3 #s) course over a period of one month.
For a period of 22 hours each.
Separated by 1-3wks
This system used
Intravaginal boxes made up of silver or gold
The intrauterine tube made up of flexible rubber.
These were not fixed together
Unequal loading of Radium
30 to 90 mg of Radium was placed inside the uterus
While 60 - 80 mg were placed inside the vagina.
A total dose of 6500 -7100 mg -hrs was prescribed out of which
4500 mg Ra was contributed by the vaginal box.
dose rate-110R/hr
PARIS SYSTEM
25-100 ½
>100 1/3
2) The spacing of the needles should
not be more than 1 cm from each other
or from the crossing ends.
3) If the ends of the implant are
uncrossed (Fig. 15.15B or C), the
effective area of dose uniformity is
reduced.6 The area is, therefore,
reduced by 10% for each uncrossed
end for tablereading purposes.
4) In the case of multiple implant
planes, the radium should be arranged
as in rules 1 to 3, and the planes
should be parallel to each other.
Volume Implants
Some tumors are better implanted using three-dimensional shapes such
as cylinders, spheres, or cuboids.
1. The total amount of radium is divided into eight parts and distributed
as follows for the various shapes.
cylinder is composed of belt, four parts; core, two parts; and each
end, one part.
sphere is made up of shell, six parts, and core, two parts.
cuboid consists of each side, one part; each end, one part; and core,
two parts.
2. The needles should be spaced as uniformly as possible, not more
than 1 cm apart. There should be at least eight needles in the belt and
four in the core.
3. If the ends of the volume implant are uncrossed, 7.5% is deducted
from the volume for the uncrossed end for table-reading purposes.
For a volume implant, the prescribed dose is stated 10% higher than the
minimum dose within the implanted volume.
Figure Example of a volume
implant with one end uncrossed.
The implant has eight needles in the
belt, four in the core (not shown), and
four at one end. Whereas the needles
in the belt and core are 1 mg each, the
crossing needles at the end are 0.5 mg
each, thus satisfying the Paterson-
Parker rule of radium distribution.
The Quimby System
The Quimby system of interstitial implantation is characterized by
a uniform distribution of sources of equal linear activity.
this arrangement of sources results in a nonuniform dose
distribution, higher in the central region of treatment.
For planar implants, the Quimby table gives the milligram-hours
required to produce 1,000 R in the center of the treatment planes,
up to a 3-cm distance from the plane of implant.
The stated dose is thus the maximum dose in the plane of
treatment.
For volume implants, the stated dose is the minimum dose within
the implanted volume.
The original Quimby tables, like the Manchester tables, are based
on an exposure rate constant of 8.4 R-cm2/mg-h instead of the
currently accepted value of 8.25 R-cm2/mg-h.
The Paris System
The Paris system of dosimetry is intended primarily for removable
implants of long line sources, such as 192Ir wires.
The system prescribes wider spacing for longer sources or larger
treatment volumes.
the sources are of uniform linear activity and are implanted in
parallel lines.
In the Paris system the dose specification is based on an isodose
surface, called the reference isodose.
However, in practice, the value of the reference isodose is fixed at
85% of the “basal dose,” which is defined as the average of the
minimum dose between sources.
It has been shown that the reference isodose for a Paris implant
surrounds the implant within a few millimeters, and its value is
approximately equal to 85% of the basal dose.
Determination of basal dose
(BD) in implants using the Paris
system.
A: Line sources implanted in
patterns of (a) single
plane, (b) squares, and (c)
triangles.
B: Isodose curves in central plane
of a volume implant using the
Paris system. The isodose values
are normalized to the average
basal dose, which is given by
1/4(BD1 + BD2 + BD3 + BD4).
Computer system
The sources of uniform strength are implanted, spaced
uniformly (1.0 to 1.5 cm, with larger spacing for larger-size
implants), and cover the entire target volume.
Active length 30-40% longer than Target length as ends
uncrossed
Treated volume - sufficient safety margins; peripheral
sources implanted on outer surface
Dose specified by isodose surface that surrounds target
Whole planning with help of computers
Isodose curves for a volume implant using two parallel planes
containing five 192Ir line sources per plane.
A. Central cross-sectional plane.
B. Longitudinal plane through middle of implant. Prescription dose
is specified on the 45-cGy/h isodose curve, which just encloses
the implant in the central cross-sectional plane.
THANKS