Professional Documents
Culture Documents
IN LABORATORY SERVICES
GROUP MEMBERS
MS.ARSHIA ANJUM
MS.BAKHTAWAR SIDDIQUE
MS.LUBNA FATIMA
QUALITY CONTROL MANAGEMENT
IN LABORATORY SERVICES
IFTIKHAR’S MEMORIAL
HOSPITAL
IFTIKHAR’S MEMORIAL HOSPITAL
1.QUALITY
2.CONTROL MATERIAL
3. STATISTICAL PRESENTATION
4. PROBLEM SOLVING
QUALITY
DEFINE AS “ To MEET THE DEMAND OF CUSTOMER”
In a Laboratory services, quality is a cyclic process from
Physician order to physician demands.
Includes
1.Physician order
2. Phelbotmy
Sample labelling
Sample collection (e.g avoid hemolysis,takecare about ca,
mg,hco3)
3. Instrument maintenance , callibration, control.
4. Statistical presentation with the help of Major Central
Tendency & Normal Distribution.
What are Standards
30
Westgard Rules- 12s
One control exceeds the mean ±SD ,warning
sign, this rule is used as a warning rule to
trigger careful inspection of the analyte
WARNING RULE – not cause for rejecting a
run
One of two control results falls outside ±2SD
Alerts technologist to possible problems
Must then evaluate the 13S rule
Westgard – 12S Rule
32
Westgard Rules- 13s
One control exceeds the mean ±3
SD ,rejection, excessive random error
A run is rejected when a single control
measurement exceeds the mean plus 3s or the
mean minus 3s control limit.
Westgard Rules- 13s
100
90
80
Antibody Units
70 +3 sd
60 +2 sd
+1 sd
50 Target value
40 -1 sd
30 -2 sd
20 -3 sd
10
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Assay Run
Westgard Rules- 22s
22s - Reject when 2 consecutive control
measurements exceed the same mean plus 2s
or the same mean minus 2s control limit.
Two consecutive control exceed the same ± SD
limits, rejection, sensitive to random error
Patient results cannot be reported.
Requires corrective action
Westgard Rules-R4s
Reject when 1 control measurement in a group
exceeds the mean plus 2s and another exceeds
the mean minus 2s.
One control exceeds the mean + 2SD and the
other exceeds the mean -2 SD , rejection ,
random error
Westgard Rules-41s
Reject when 4 consecutive control
measurements exceed the same mean plus 1s
or the same mean minus 1s control limit.
Four consecutive observations exceed the
mean + or – 1SD,rejection,systematic error
Westgard Rules-10x
Reject when 10 consecutive control
measurements fall on one side of the mean.
Ten consecutive observation fall on one side of
the mean ,rejection, systematic error
Westgard Rules-10x
100
90
80
Antibody Units
70 +3 sd
60 +2 sd
+1 sd
50 Target value
40 -1 sd
30 -2 sd
20 -3 sd
10
0
1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 16
Assay Run
VZV IgG ELISA: Target Value = 49 U/ml
Types of error Control rule that’s detects it
Random Error 13s,22S,R4s
Systemic Error 10x,41s,
Detection and Resolution of Quality
Problems
When these problems are identified the following corrective actions
can be taken.
Check expiration date of the control.
Check expiration date of the reagent.
If a new control was used, make sure it was reconstituted
properly.
Retest the control. If the new value is within acceptable limits,
record both values and proceed with patient testing.
If the repeat value is still out of range, run a new vial of control. If
the new control value is within acceptable limits, record the
values and proceed with patient testing. The problem with the first
set of controls was probably specimen deterioration.
Detection and Resolution of Quality Problems
When analytical problem are found it is best to have a clear plan of action
that is execute sequentially until the problem is resolved, it is necessary to
document the problem, investigation, resolution and any data that indicate
that the problem as actually resolved.
Action taken to bring system back in action (Summery)
Repeat assay with fresh aliquot.
Repeat assay with fresh new set of control (old control may be mislabeled,
possible enzyme deterioration, evaporation, precipitation or other error)
Look for obvious problem such as clot, reagent levels and mechanical
fault.
Recalibrate the instrument for the out of control analyte then assay
control.
Out of Control System
control,
Recalibration
Two levels of control run, results within + 2sd.
The method are use must be informed and authorized
by the supervisor and must entered in to problem log
book and formally sign and dated.
Delta Check
Reducing manpower.
Increasing service quality
Simplifying processes
Decreasing the report release turnaround time
(TAT).
In addition to the introduction of automated
equipment and the development of a laboratory
information system (LIS) technology, another
way to raise working efficiency is to build an
auto verification (AV) system
GENERAL CHECKLIST: STARTING THE INVESTIGATION
PRE-ANALYTICAL VARIATION: POST-COLLECTION
Sample transport:
Issues that may affect analyte levels
Timing:
delayed centrifugation, WBC glucose utilization, leakage of RBC contents
Temperature:
Arterial blood gases, cryo globulin, lactic acid, ammonia
Light exposure:
bilirubin, vitamins, Porphyrins
Centrifugation:
Timely separation of serum and cells (within 2 hrs.) Delayed separation affects
glucose, K+, LD, ammonia, phosphate
Short spins: keep cellular components in the serum: K+, enzymes affected
Excessive spins: may cause hemolysis due to RBC membrane damage
Storage :
For frozen sample , avoid excessive freeze-thaw cycles
ANALYTICAL VARIATION