Professional Documents
Culture Documents
UNITED STATES
Plainview, NY
Tuscon, AZ
San Diego, CA
San Jose, CA
Santa Barbara, CA
Tustin, CA
Ft. Collins, CO
Minneapolis, MN
Chadds Ford, PA
Austin, TX
EUROPE
Cambridge, UK
Dourdan, France
Londonderry, N. Ireland
Mannheim, Germany
Munich, Germany
JAPAN
Osaka, Japan
Tokyo, Japan
ASIA PACIFIC
Beijing, China
Hsinchu, Taiwan
Penang, Malaysia
Singapore
Technology Guide
X-Ray Fluorescence Metrology Tools
1 14
Software and User Interface XRF System Overview
Primary Amplifier
Beam
Shutter
Analog
Collimator
to Digital
Resultant Converter
Beam Detector
and Preamp
Spectrum
Sample Generation
Stage Fluoresced
X-Rays
3 12
XRF Tool Performance X-Ray Source
a.
between these terms. dependent upon the precision, or repeatability, of the A typical X-ray source consists of an evacuated glass tube
measurements. Repeatability is an indication of how well containing a filament cathode (typically tungsten) and a
Accuracy a series of measurements taken on the exact same spot quantity of target anode material. The tube is sealed in
of a sample yield the same or nearly the same readings. an oil-filled metal (typically brass) enclosure that effec- Figure 4:
Accuracy is an expression of the deviation of the mean Repeatability is commonly expressed in terms of relative Filament/Target
tively shields X-rays.
measured value from the true value of the sample. standard deviation (RSD) from a mean value. Interaction
Assume, for example, that you have a sample with a When a current is applied to the filament, it emits negative-
coating thickness of 100 microinches. If, after a statisti- Repeatability is highly dependent upon the total number Interaction between
ly charged electrons which are accelerated toward the tar- accelerated filament
cally significant set of measurements, you obtained a of X-rays counted during measurement. X-ray counting is get material by a high voltage potential. These electrons electrons and target mate-
mean measurement value of 99 microinches, you could maximized by generating an incident beam of the highest interact with the target material to produce a broad spec- rial atoms produces fluores-
say that the system is accurate within one percent. intensity, using the largest beam size appropriate to the trum of X-rays. These primary beam X-rays exit through a cence by means of two processes:
sample geometry, and by using the longest possible “window” in the tube enclosure and are directed down the
;;;;;
Measurement accuracy is primarily a function of calibra- measurement duration. a. Photoelectric Effect: Filament electrons
generation conduit into the collimation mechanics.
tion accuracy, which is in turn dependent upon the transfer energy to target atom electrons, causing
accuracy of the calibration algorithm, the quality of the Reproducibility ejection of those electrons from their orbitals and
standards used to calibrate, and the calibration measure- a resultant photoelectric emission (fluorescence).
ment time. Reproducibility refers to an instrument’s ability to obtain
consistent measurement results when measuring the b. Bremstrahlung: Literally, “braking radiation”.
;;;;;
same sample at different times, with different operators, Filament electrons are attracted to the strong
positive charge of the nuclei, causing deceleration
and possibly with different instruments of the same type.
and release of energy.
Filament Target
11 4
X-Ray Collimation Quantitative Analysis
collimation. There are two principal methods of collima- These models are called calibrations. An XRF tool can be Exploded view shows Gold (Au)
tion, mechanical and optical. Shutter calibrated using a variety of empirical and fundamental and Nickel (Ni) foil interlayers
parameters techniques. on Copper (Cu) base layer.
Mechanical Collimation
Empirical Calibration
A mechanical collimator is essentially a pin-hole aperture. A
typical mechanical collimator assembly consists of a metal By measuring the X-ray count rates from a coating/base
Collimator
block featuring collimators of various dimensions (figure 5). combination of known thickness and/or composition (stan-
Block
To form the collimated beam, a single collimator is aligned dard), we obtain the reference data needed to calculate the
directly beneath the path of the primary beam X-rays. X-rays thickness and/or composition of unknown quantities of the
pass through the collimator and emerge in a resultant beam same coating and base combinations. In practice, a series
whose initial dimensions are equal to the dimensions of the of standards with different values are measured to obtain the
collimator and whose final dimensions fan out to a larger Resultant necessary reference countrates. Statistical methods are used
geometry. This resultant beam is targeted at the sample. Beam to correlate the standard values to the measured XRF count
rates.
Optical Collimation Sample Accurate calibration of the XRF tool is dependent on the
Optically collimated XRF tools employ an optical element to availability of high-quality standards for the specific applica-
shape and direct primary beam X-rays. Primary beam X-rays Figure 5: Mechanical Collimation tion under analysis.
incident upon the surface section of the optical element
(figure 6) propagate through the system and emerge in a Fundamental Parameters Calibration
convergent beam of extremely high intensity. Highly accurate mathematical techniques based on physical
The major practical drawbacks to empirical calibration
Primary Beam X-Rays methods are: properties have been developed for a wide range of appli-
Advantages and Disadvantages cations, enabling the XRF tool to be calibrated without (or
Mechanical collimators form the incident beam by blocking • The time required to measure a series of calibration with a reduced number of) standards.
all but a small percentage of the primary beam X-rays (those standards
traveling in a path coinciding with the orientation of the Shutter In a fundamental parameters (FP) calibration, an exten-
collimator tube). Mechanical systems are currently less • The difficulty of fabricating standard sets for intricate sive range of instrument parameters are first quantified. This
expensive and provide more than sufficient incident beam applications. information is incorporated into a mathematical model of
intensity for a wide range of analytical tasks where small the fundamental physics of X-ray interaction with the specific
beams (under 4 mils [100 microns]) are not required. materials under analysis.
Optically collimated systems capture and transmit a substan- Focusing Thickness The mathematical model is often enhanced through the
tially higher quantity of the primary beam X-rays produced Element use of physical standards (the Standard-FP method). The
(100 to 1000 times that of mechanically collimated systems). improvement in test process throughput enabled by reducing
A higher intensity beam produces significantly higher X-ray or eliminating the need for physical standards can be sub-
countrates from the sample, with commensurate benefits to stantial.
precision. XRF tools utilizing optical collimators are referred
to as Microbeam XRF (MXRF) tools. 200
FP techniques offer versatility and often enhanced capability.
min For certain applications, however, empirical calibration
Resultant Beam offers a more accurate, simple, flexible, and reliable solu-
1000 cps Countrate
Sample
tion. An XRF tool should offer both calibration strategies.
5 10
Spectrum Processing X-Ray Detection
High Voltage
Counts
Counts
Photon +
of Interest (ROI). Typically, the lower and upper limits of This is sometimes referred to energy dispersive X-ray - +
- Pulse
Processing
the ROI define a band of channels where pulses collected detection. There are two commonly used types of energy Section
(counts) are significantly higher than in the channels sur- dispersive detectors, the proportional counter and the p i n
rounding the peak; such background channels are charac- semiconductor detector.
Channel Channel
terized by low count levels caused by X-ray scattering from Figure 8: Semiconductor Detector
the sample and artifacts of the pulse processing electronics. Figure 13: Background Removal Proportional Counter Detectors
(L) Spectrum with background; (R) Spectrum with background removed. The proportional counter is a sealed, gas filled hollow cylin- region is increased through the use of a process known as
While accurate quantitative determinations can often be "lithium drifting". Incident X-rays produce ionizations of Si in
der with a central coaxial wire. An electric field is established
made by simply integrating over the ROI limits, it is some- Background Correction: Spectral peaks are initially super- the sensitive region of these detectors. The charge carriers
between the outside shell and the inner central wire using an
times necessary to further process the spectrum. A combina- imposed over a background level of counts (figure 13). are negative electrons and positive “holes”, which are drawn
external high voltage supply. X-rays enter through a window
tion of mechanical and mathematical methods are used to These readings originate from X-ray scattering, electronic to opposite ends of the detector due to the voltage bias
(usually thin beryllium) in the detector shell and ionize the
refine spectral data, including: noise, and errors in pulse processing. Various mathematical applied across the silicon chip. Like the proportional counter,
gases inside. Electrons are attracted to the positively charged
methods are employed to characterize the background and wire, producing a charge at the detector output. This charge total charge collected within the semiconductor detector is
Peak Deconvolution (Numerical Filtering): Peak over- subtract it from the pure spectrum. is collected through a pre-amplifier and converted into a directly proportional to the energy of the incident X-ray, and
lap is observed when the characteristic X-rays detected from
voltage pulse, which is then further amplified and shaped by is converted to a corresponding voltage amplitude through
two or more elements are close in energy. The degree of Mechanical Filters: Mechanical filters are thin foils which an amplifier. The resultant pulse amplitude is directly related the use of a preamplifier and amplifier.
overlap depends on the similarity of the energies and on the are placed between the sample and the detector (or in the
resolution of the detector used. Peak deconvolution is to the number of ion pairs created by the ionizing X-ray pho-
case of primary filters, between the primary beam X-rays ton, which in turn is directly related to the energy of the inci- To minimize electron/hole creation due to thermal effects
employed to mathematically decompose two or more peaks and the sample). The X-ray “absorption" characteristics of (insuring that pulses are produced in response to X-ray ion-
which overlap and superimpose to produce a single, wider dent photon.
the filter are unique to the foil material used. The filter will izations only) semiconductor detectors often employ cooling
composite peak (figure 12). A variety of techniques are absorb or block some units to maintain very low operating temperatures. Typical
Semiconductor Detectors
available for peak deconvolution, with advantages and limi- X-ray energies while cooling schemes use liquid nitrogen or electric heat
tations inherent in each method depending on the applica- allowing other energies Co Foil The two most common semiconductor detectors are the PIN exchange.
Absorption Edge
tion being analyzed. to pass through. By 7.7 keV
Diode and Si(Li) types. Both detectors incorporate silicon
selecting an appropriate chips which respond to X-rays in the same manner as the gas Relative Advantages of Each Detector Type
filter material for a given in the proportional counter.
Proportional counters are less expensive, require no mainte-
application, peak over-
a b
Measured peak of two elements Ni The pin diode is maintained under a reverse voltage bias, nance, and are very robust detectors. Their large volumes
lap can be substantially 7.4 keV
leaving a central p-n junction where detection of X-rays can and window surface areas provide a large solid angle for
eliminated. Cu
8.0 keV occur. The Si(Li) semiconductor also provides a p-n junction X-ray capture. The result is higher capture rates, providing
Additional spectrum pro- for X-ray detection, except that the thickness of its sensitive higher intensity measurements (resulting in better measure-
Counts
frequency distribution (histogram of energy vs. intensity), • The resulting analog signal is superimposed over a
referred to as the spectrum (figure 9d). randomly varying level of electronic noise, or
background (see figure 13).
At all stages in the pulse processing chain, proportionality
between the detected X-ray energy, the analog pulse ampli-
Channel
Figure 9: Processing the Detector Output Figure 11 (right): Pulse Height Histogram (Spectrum)
Multi-Channel Analyzer Spectral peaks are produced by high X-ray counts at specific
v v v energies. In the pulse height histogram (spectrum), X-ray counts
are displayed on the vertical (Y) axis. Energy level ranges (chan-
counts
frequency distribution (histogram of energy vs. intensity), • The resulting analog signal is superimposed over a
referred to as the spectrum (figure 9d). randomly varying level of electronic noise, or
background (see figure 13).
At all stages in the pulse processing chain, proportionality
between the detected X-ray energy, the analog pulse ampli-
Channel
Figure 9: Processing the Detector Output Figure 11 (right): Pulse Height Histogram (Spectrum)
Multi-Channel Analyzer Spectral peaks are produced by high X-ray counts at specific
v v v energies. In the pulse height histogram (spectrum), X-ray counts
are displayed on the vertical (Y) axis. Energy level ranges (chan-
counts
Counts
Counts
Photon +
of Interest (ROI). Typically, the lower and upper limits of This is sometimes referred to energy dispersive X-ray - +
- Pulse
Processing
the ROI define a band of channels where pulses collected detection. There are two commonly used types of energy Section
(counts) are significantly higher than in the channels sur- dispersive detectors, the proportional counter and the p i n
rounding the peak; such background channels are charac- semiconductor detector.
Channel Channel
terized by low count levels caused by X-ray scattering from Figure 8: Semiconductor Detector
the sample and artifacts of the pulse processing electronics. Figure 13: Background Removal Proportional Counter Detectors
(L) Spectrum with background; (R) Spectrum with background removed. The proportional counter is a sealed, gas filled hollow cylin- region is increased through the use of a process known as
While accurate quantitative determinations can often be "lithium drifting". Incident X-rays produce ionizations of Si in
der with a central coaxial wire. An electric field is established
made by simply integrating over the ROI limits, it is some- Background Correction: Spectral peaks are initially super- the sensitive region of these detectors. The charge carriers
between the outside shell and the inner central wire using an
times necessary to further process the spectrum. A combina- imposed over a background level of counts (figure 13). are negative electrons and positive “holes”, which are drawn
external high voltage supply. X-rays enter through a window
tion of mechanical and mathematical methods are used to These readings originate from X-ray scattering, electronic to opposite ends of the detector due to the voltage bias
(usually thin beryllium) in the detector shell and ionize the
refine spectral data, including: noise, and errors in pulse processing. Various mathematical applied across the silicon chip. Like the proportional counter,
gases inside. Electrons are attracted to the positively charged
methods are employed to characterize the background and wire, producing a charge at the detector output. This charge total charge collected within the semiconductor detector is
Peak Deconvolution (Numerical Filtering): Peak over- subtract it from the pure spectrum. is collected through a pre-amplifier and converted into a directly proportional to the energy of the incident X-ray, and
lap is observed when the characteristic X-rays detected from
voltage pulse, which is then further amplified and shaped by is converted to a corresponding voltage amplitude through
two or more elements are close in energy. The degree of Mechanical Filters: Mechanical filters are thin foils which an amplifier. The resultant pulse amplitude is directly related the use of a preamplifier and amplifier.
overlap depends on the similarity of the energies and on the are placed between the sample and the detector (or in the
resolution of the detector used. Peak deconvolution is to the number of ion pairs created by the ionizing X-ray pho-
case of primary filters, between the primary beam X-rays ton, which in turn is directly related to the energy of the inci- To minimize electron/hole creation due to thermal effects
employed to mathematically decompose two or more peaks and the sample). The X-ray “absorption" characteristics of (insuring that pulses are produced in response to X-ray ion-
which overlap and superimpose to produce a single, wider dent photon.
the filter are unique to the foil material used. The filter will izations only) semiconductor detectors often employ cooling
composite peak (figure 12). A variety of techniques are absorb or block some units to maintain very low operating temperatures. Typical
Semiconductor Detectors
available for peak deconvolution, with advantages and limi- X-ray energies while cooling schemes use liquid nitrogen or electric heat
tations inherent in each method depending on the applica- allowing other energies Co Foil The two most common semiconductor detectors are the PIN exchange.
Absorption Edge
tion being analyzed. to pass through. By 7.7 keV
Diode and Si(Li) types. Both detectors incorporate silicon
selecting an appropriate chips which respond to X-rays in the same manner as the gas Relative Advantages of Each Detector Type
filter material for a given in the proportional counter.
Proportional counters are less expensive, require no mainte-
application, peak over-
a b
Measured peak of two elements Ni The pin diode is maintained under a reverse voltage bias, nance, and are very robust detectors. Their large volumes
lap can be substantially 7.4 keV
leaving a central p-n junction where detection of X-rays can and window surface areas provide a large solid angle for
eliminated. Cu
8.0 keV occur. The Si(Li) semiconductor also provides a p-n junction X-ray capture. The result is higher capture rates, providing
Additional spectrum pro- for X-ray detection, except that the thickness of its sensitive higher intensity measurements (resulting in better measure-
Counts
collimation. There are two principal methods of collima- These models are called calibrations. An XRF tool can be Exploded view shows Gold (Au)
tion, mechanical and optical. Shutter calibrated using a variety of empirical and fundamental and Nickel (Ni) foil interlayers
parameters techniques. on Copper (Cu) base layer.
Mechanical Collimation
Empirical Calibration
A mechanical collimator is essentially a pin-hole aperture. A
typical mechanical collimator assembly consists of a metal By measuring the X-ray count rates from a coating/base
Collimator
block featuring collimators of various dimensions (figure 5). combination of known thickness and/or composition (stan-
Block
To form the collimated beam, a single collimator is aligned dard), we obtain the reference data needed to calculate the
directly beneath the path of the primary beam X-rays. X-rays thickness and/or composition of unknown quantities of the
pass through the collimator and emerge in a resultant beam same coating and base combinations. In practice, a series
whose initial dimensions are equal to the dimensions of the of standards with different values are measured to obtain the
collimator and whose final dimensions fan out to a larger Resultant necessary reference countrates. Statistical methods are used
geometry. This resultant beam is targeted at the sample. Beam to correlate the standard values to the measured XRF count
rates.
Optical Collimation Sample Accurate calibration of the XRF tool is dependent on the
Optically collimated XRF tools employ an optical element to availability of high-quality standards for the specific applica-
shape and direct primary beam X-rays. Primary beam X-rays Figure 5: Mechanical Collimation tion under analysis.
incident upon the surface section of the optical element
(figure 6) propagate through the system and emerge in a Fundamental Parameters Calibration
convergent beam of extremely high intensity. Highly accurate mathematical techniques based on physical
The major practical drawbacks to empirical calibration
Primary Beam X-Rays methods are: properties have been developed for a wide range of appli-
Advantages and Disadvantages cations, enabling the XRF tool to be calibrated without (or
Mechanical collimators form the incident beam by blocking • The time required to measure a series of calibration with a reduced number of) standards.
all but a small percentage of the primary beam X-rays (those standards
traveling in a path coinciding with the orientation of the Shutter In a fundamental parameters (FP) calibration, an exten-
collimator tube). Mechanical systems are currently less • The difficulty of fabricating standard sets for intricate sive range of instrument parameters are first quantified. This
expensive and provide more than sufficient incident beam applications. information is incorporated into a mathematical model of
intensity for a wide range of analytical tasks where small the fundamental physics of X-ray interaction with the specific
beams (under 4 mils [100 microns]) are not required. materials under analysis.
Optically collimated systems capture and transmit a substan- Focusing Thickness The mathematical model is often enhanced through the
tially higher quantity of the primary beam X-rays produced Element use of physical standards (the Standard-FP method). The
(100 to 1000 times that of mechanically collimated systems). improvement in test process throughput enabled by reducing
A higher intensity beam produces significantly higher X-ray or eliminating the need for physical standards can be sub-
countrates from the sample, with commensurate benefits to stantial.
precision. XRF tools utilizing optical collimators are referred
to as Microbeam XRF (MXRF) tools. 200
FP techniques offer versatility and often enhanced capability.
min For certain applications, however, empirical calibration
Resultant Beam offers a more accurate, simple, flexible, and reliable solu-
1000 cps Countrate
Sample
tion. An XRF tool should offer both calibration strategies.
5 10
XRF Tool Performance X-Ray Source
a.
between these terms. dependent upon the precision, or repeatability, of the A typical X-ray source consists of an evacuated glass tube
measurements. Repeatability is an indication of how well containing a filament cathode (typically tungsten) and a
Accuracy a series of measurements taken on the exact same spot quantity of target anode material. The tube is sealed in
of a sample yield the same or nearly the same readings. an oil-filled metal (typically brass) enclosure that effec-
Accuracy is an expression of the deviation of the mean Repeatability is commonly expressed in terms of relative tively shields X-rays.
measured value from the true value of the sample. standard deviation (RSD) from a mean value.
Assume, for example, that you have a sample with a When a current is applied to the filament, it emits negative-
coating thickness of 100 microinches. If, after a statisti- Repeatability is highly dependent upon the total number ly charged electrons which are accelerated toward the tar-
cally significant set of measurements, you obtained a of X-rays counted during measurement. X-ray counting is get material by a high voltage potential. These electrons
mean measurement value of 99 microinches, you could maximized by generating an incident beam of the highest interact with the target material to produce a broad spec-
say that the system is accurate within one percent. intensity, using the largest beam size appropriate to the trum of X-rays. These primary beam X-rays exit through a
sample geometry, and by using the longest possible “window” in the tube enclosure and are directed down the
;;;;;
Measurement accuracy is primarily a function of calibra- measurement duration. generation conduit into the collimation mechanics.
tion accuracy, which is in turn dependent upon the
accuracy of the calibration algorithm, the quality of the Reproducibility
standards used to calibrate, and the calibration measure-
ment time. Reproducibility refers to an instrument’s ability to obtain
consistent measurement results when measuring the
;;;;;
same sample at different times, with different operators,
and possibly with different instruments of the same type.
Energy (keV)
Composite Output
Output due to Photoelectric Effect
Output due to Bremstrahlung
11 4
X-Ray Fluorescence X-Ray Targeting
3 12
Software and User Interface
operators with little or no previous experience in XRF LANtastic , NT , and others); built-in security features for
TM TM
metrology can be testingwithin minutes. multi-user environments; optional remote control software
available.
Standard software features include: Ergonomic Design: Integrated design of GUI, chamber,
and input devices minimizes repetitive strain. Customizable
Automated Stage Controls: Stored sample handling interface streamlines measurement tasks.
recipes insure repeatable, reproducible X-ray beam
targeting. Optional pattern recognition configuration Data Export: Measurement data can be exported and
enables instant, hands-free correction of sample positioning modeled using such popular programs such as Word, Excel,
variations. and Lotus.
Primary Amplifier
Beam
Shutter
Analog
Collimator
to Digital
Resultant Converter
Beam Detector
and Preamp
Spectrum
Sample Generation
Stage Fluoresced
X-Rays
1 14
Matrix Metrologies Inc.
101-5 Colin Drive
Holbrook, NY, USA 11741
T 631.472.2400
F 631.472.2424
Technology Guide
X-Ray Fluorescence Metrology Tools