Professional Documents
Culture Documents
Introduction
¤ FDA 1994
What is radiation?
X-ray radiation contains more energy than ultraviolet, infrared, radio waves,
microwaves or visible light. X-ray radiation has sufficient energy (>30 eV) to
cause ionizations. An ionization is a process whereby the radiation removes an
outer shell electron from an atom (Figure 1-2).
X-ray Production
X-rays are produced when high velocity electrons are decelerated (slowed or
stopped) or by a nucleus of an atom especially by high atomic number material,
such as the tungsten target (anode) in a X-ray tube. An electrically heated
filament (cathode) within the X-ray tube generates electrons that are accelerated
from the filament to the tungsten target by the application of a high voltage to the
tube. The energy gained by the electron is equal to the potential difference
(voltage) between the anode and cathode. This electron energy is typically
expressed in kilovolts (kV). The accelerated electron interacts with the target
(anode) nucleus. As the electric field of the electron interacts with nucleus, the
electron releases energy in the form of X-rays. This method of of x-ray
production is called bremsstrahlung or braking radiation (Figure 1-4).
FIgure 1-4: X-ray Production (Bremstrahlung)
Courtesy of the University of Michigan Student Chapter of the Health Physics Society
Since the degree of interaction of the accelerated electron with the target nucleus
can vary, the energy spectrum, or distribution of energy, of the X-rays produced
by the bremsstrahlung process is continuous.
The lower energy X-rays are absorbed within the X-ray tube. This reduces the
number of lower energy X-rays in the resultant spectrum since the lower energy
X-ray are less penetrating. The beam is considered "harder" when there is more
filtration. Most X-ray manufacturers add filtration, commonly consisting of
aluminum, since lower energy X-rays do not contribute to images and add to
patient dose.
The resulting x-ray spectrum energy (Figure 1-5) is a mixture of the characteristic
and bremsstrahlung radiation, less the primarily low energy X-rays absorbed by
the X-ray tube (and added filtration). The maximum energy of the X-ray
produced is equal to the maximum potential applied across the x-ray tube. This
peak X-ray energy is usually described with the unit kVp (kilovolt peak or kilovolt
potential). The type of target anode, potential (kVp) and added filtration produce
a beam of a given "quality" which implies specific shape of an X-ray spectrum.
The quantity of electron flow (current) in the X-ray tube is described in units of
milliamperes (mA). The rate of X-ray production is directly proportional to the X-
ray tube current. Higher mA values indicate more electrons are striking the
tungsten target, thereby producing more X-rays. The voltage (kVp) primarily
determines the maximum X-ray energy produced but also influences the number
of X-rays produced. Increasing the kVp attracts more electrons from the filament
increasing the rate of X-ray production. However, this relationship is not directly
proportional but higher kVp setting will result in a substantial increase in the
number of X-rays produced. The total number of X-rays produced at a set kVp
depends directly on the product of the mA and exposure time and is typically
described in terms of mA-s or mAs. Fluoroscopy is usually performed using 2 to
5 mA current at a peak electrical potential of 75 to 125 kVp.
Once generated, the X-rays are emitted in all directions in a uniform manner
(isotropically). The lead housing surrounding the X-ray tube limits X-ray
emission through a small opening or port in the X-ray tube. The resulting primary
beam of useful radiation is shaped by additional lead shutters, or collimators, that
can be adjusted to provide different beam shapes or sizes.
Since the initial beam travels in straight but divergent directions, geometry in a
three dimensional world dictates that the radiation intensity will decrease with the
inverse square of the distance. Consequently, the number of X-rays traveling
through a unit area decreases with increasing distance. Likewise, radiation level
decreases with increasing distance since exposure is directly proportional to the
number of X-rays interacting in a unit area. The intensity of the radiation is
described by the inverse square law equation:
This relationship indicates that doubling the distance from a radiation source
decreases the radiation level by a factor of four. Conversely, halving the
distance, increases the radiation level by a factor of four. Intelligent application
of inverse square law principles can yield significant reductions in both patient
and operator radiation exposures.
Example 1:
Solution 1:
The original exposure was 30 mrem (15 mrem/min for 2 min). The new exposure
would be:
The concept of contrast and electron density X-ray interaction can be shown in
Figure 1-9.
Assume 1,000 X-rays strike the following body portions. The number of X-rays
reaching the recording media (film, TV monitor) directly effect the image's
brightness.
Tissue Thickness
As tissue thickness increases, the probability of X-ray interaction increases.
Thicker body portions remove more X-rays from the useful beam compared to
thinner portions (Figure 1-10). In fluoroscopy, this effect must be compensated
for while panning across variable tissue thickness to provide consistent
information to the image-recording device.
Note that on average, only 1 percent of the X-rays reach the image-recording
device (e.g., image intensifier, film), yielding useful information. Thus, 99 percent
of the X-rays generated are either absorbed within the patient (patient radiation
exposure) or are scattered throughout the examination room (staff radiation
exposure).
Energy
Higher kVp X-rays are less likely to interact with tissue and are described as
more "penetrating." Increasing kVp, thereby generating more penetrating
radiation, reduces the relative image contrast (or visible difference) between
dense and less dense tissue. Conversely, less radiation dose results to the
patient since less X-rays are absorbed. Figure 1-11 illustrates this effect. The X-
rays that do not reach the image recording device are either absorbed in the
patient (patient radiation dose) or are scattered throughout the exam room (staff
radiation dose)
The fundamental tie between Physics and fluoroscopy Radiation Safety occurs
during the sequence of steps which lead to radiation biological effects. These
steps are deposition of energy (Figure 1-13) and biochemical changes caused by
X-rays (Figure 1-14).
The unit Roentgen is only defined for air and can not be used to describe dose to
tissue. Radiation dose is the energy (joules) imparted per unit mass of tissue and
has the US units of rad (radiation absorbed dose). Patient exposures, particularly
in Radiation Oncology are described in units of radiation dose. There is an
international (SI) unit for dose termed the Gray (Gy). The conversion between
the units is: 100 rad = 1 Gy.
By 1906 Bergonie and Tribondeau realized that cells were most sensitive to
radiation when they are:
Rapidly dividing
Undifferentiated
Have a long mitotic future
Author Note: When DNA, which had not been discovered at this time, has no backup (single strand).
Radiation sensitivity
Radiosensitivity is a function of the cell cycle with late S phase being the most
radioresistant and G1, G2, and especially mitosis being more radiosensitive.
Radiation can directly interact with a molecule and damage it directly. Because of
the abundance of water in the body, radiation is more likely to interact with
water. When radiation interacts with water, it produces labile chemical species
(free radicals) such as hydronium (H.) and hydroyxls (.OH). Free radicals can
produce compounds such as hydrogen peroxide (H2O2) which subsequently exert
chemical toxicity. The body has sophisticated protections against this type of
chemical damage. For example, this is why hydrogen peroxide foams when it is
poured on a cut. The hydrogen peroxide is being destroyed by an extremely
rapid enzyme, hydrogen peroxidase.
Nonetheless, while DNA and other repair methods are extremely capable of
protecting the body, some fraction of the damage is not repaired or may be
repaired incorrectly. In either case, if the DNA is damaged, several things can
happen. The most likely is that the damage will be repaired before the end of the
cell’s growth cycle. If not, the cell will probably die. There is some chance that the
cell will survive and behave differently because of the damaged DNA.
Photomicrograph showing examples of radiation-induced chromosome damage in cancer cells following radiotherapy treatment
(Bushong 1980). Courtesy of Scott Sorenson.
When DNA is damaged, the harm can be magnified by the cellular machinery.
One example of a possible consequence is a cell which loses control over
replication-this mutation is better known as cancer. Ionizing radiation is thought
to initiate (start), but not promote (help grow) mutations.
Very large radiation exposures can kill humans. The lethal dose(LD) for half the
population (50%) within 60 days is termed the LD50/60d. The LD50/60d in humans
from acute, whole body radiation exposure is approximately 400 to 500 rads (4-5
Gy). The temperature elevation in tissue caused by the energy imparted is much
less than 1° C. The severe biological response is due to ionizing nature of X-ray
radiation, causing the removal of electrons, and therby chemical changes in
molecular structures.
A number of ionizing radiation effects occur at high doses. These all seem to
appear only above a threshold dose. While the threshold may vary from one
person to another, these effects can be eliminated by keeping doses below 100
rad. The severity of these effects increases with increasing dose above the
threshold. These so-called deterministic (non-stochastic) effects are usually
divided into tissue-specific local changes and whole body effects, which lead to
acute radiation syndrome (Table 2-1).
Sorenson, 2000.
These whole body (to entire body) doses are very unlikely for patients and staff from fluoroscopy
or any diagnostic radiology study.
Several factors, such as total dose, dose rate, fractionation scheme, volume of
irradiated tissue and radiation sensitivity all affect a given organ’s response to
radiation. Radiation is more effective at causing damage when the dose is higher
and delivered over a short period of time. Fractionating the dose (i.e. spreading
the dose out over time) reduces the total damage since it allows the body time
for repair. Patient exposures are higher than attending staff but they occur over
short periods of time whereas staff exposures are normally low and occur over
several years.
The Table 2-2 provides examples of possible radiation effects to skin caused by
typical fluoroscopy exposures. Note that patient and technique factors can
substantially increase exposure rates, significantly reducing the time necessary
for the subsequent effect.
[Specific case studies of radiation-induced skin injury are presented in the next section]
Please note these localized effects will not be seen immediately (in the clinic).
These effects take time to develop and the minor effects may not be noticed and
are often attributed to other causes. This effectively results in a lack of warning
for serious effects such as dermal necrosis. This lack of warning has led the
FDA, HFH Radiation Safety Committee, and HFH Hospital Medical Executive
Committee (HMEC) to have concerns about fluoroscopy utilization.
Consequently, fluoroscopy safety training and monitoring of fluoroscopy times
were mandated.
Cataractogenesis
Since the discovery of radiation by Roentgen, there have been many groups in
which radiation effects have been studied (Bushong 1980) (Table 2-3):
The LNT hypothesis implies that any amount of radiation exposure will increase
an individual's risk of cancer. Thus, all radiation doses should be mimimized or
kept As Low As Reasonably Achievable (ALARA).
Radiation Risk
The cancer risk coefficient derived from the LNT model for radiation exposure is
approximately 4.8 * 10-4 per rem. To put this number into perspective, imagine
that you have two similar groups of 10,000 people; exposed and not exposed.
Each member of the exposed group receive 1 rem 1,000 mrem) of radiation
exposure. The control group does not recieve any additional radiation exposure
beyond that received by natural sources. Since the natural incidence of cancer is
about 30%, the control group would expect about 3000 people to die from
cancer. In comparison, the exposed group would expect that about 3005 people
would die from cancer. Thus, the exposed group should expect about 5
additional cancers from the 1 rem exposure.
One way better understand radiation risks is to compare radiation risks with other
commonly accepted risks (Figures 2-4 and 2-5).
Different mutations differ significantly in the rate at which they are produced by
a given dose.
There is a substantial dose rate effect with no threshold for mutation
production.
The male was more radiosensitive than the female. The males carried most of
the radiation induced genetic burden.
The genetic consequences of a radiation dose can be greatly reduced by
extending the time interval between irradiation and conception. Six months to a
year is
recommended.
The amount of radiation required to double the natural and spontaneous
mutation rate is between 20 to 200 rads.
Radiation apparently does not cause unique types of mutations, but simply
increases the mutations rate above their natural rate of occurrence. Controlled
studies of genetic effects are only available from animal models. The risk
coefficient for serious genetic disorders from radiation exposure is approximately
8 * 10-5 per rem (NCRP 116). This is less than the cancer risk coefficient (4.8 *
10-4 per rem). Thus, if you protect against cancer, you are simultaneously
protecting against genetic effects.
Once conception has occurred (mother is pregnant), the unborn child (fetus) can
be harmed by radiation. Certainly, the unborn child can have the same health
problems that an adult might have such as cancer and genetic defects. The Law
of Bergonie and Tribondeau predicts that a fetus would be exquisitely sensitive to
radiation since they are:
1. Rapidly dividing;
2. Undifferentiated; and
In addition to the health effects which are a concern for an adult, there is also a
serious concern about the possibility of developmental errors (teratogenesis)
which can occur. There are three general prenatal effects which have been
observed:
1. Lethality;
2. Congenital abnormalities at birth; and
3. Delayed effects, not visible at birth, but manifested later in life.
The expression of effects are dependent upon the dose and stage of fetal
development (Figure 2-6):
Medical X-ray procedures are performed to directly benefit the patient. Note that
limits have not been established for how much dose a patient may receive. The
regulations properly leave this matter up to the discretion of physicians. The
purpose of this course is to educate physicians about how to properly make that
decision for their patients. When in doubt, contact a Board Certified Radiologist.
Modern fluoroscopy imaging systems (Figure 3-1) consist of: the X-ray tube
which produces X-rays; an Image Intensifier (I-I) that captures or stops the X-
rays and converts the X-ray energy into light; a closed-circuit video system which
ultimately producing a "live" image on a monitor. On some systems the light
output can also be distributed to a spot film or cinematography recording
systems, though the X-ray output must be greater for these imaging modalities.
Generator Control
The X-ray generator controls the quantity (number) and quality (spectrum of
energy) of the X-rays produced. There are three basic controls to the generator:
From a safety perspective, the beam- on- time is the most important control.
Beam-on-Time (Foot Pedal/Switch)
The X-ray tube current (mA setting) essentially controls the quantity of X-rays
produced per unit time. If you double the mA, you double the patient and
operator exposure. The higher current mode (High Level Control or "boost"
mode) dramatically increases the number of X-rays produced and thereby
improves image quality. The boost mode is activated in some systems when the
foot switch is firmly depressed (a light push activates the normal, low dose,
mode). An audible alarm will indicate when the high level mode is being used.
HFH policy limits the rate for the boost mode to 20 R/minute (compared with
maximum of 10 R/minute for the normal, Automatic Brigtness Control (ABC)
mode and 5 R/minute for the manual mode). Both the patient and fluoroscopy
operator dose are proportional to the total amount of current used by the
machine. Thus, a light touch on the pedal (educated use of the boost mode) will
minimize both patient and worker doses.
During normal mode fluoroscopy, the average patient Entrance Skin Exposure
(ESE) is approximately 2 R/min. The level of radiation exposure falls off
exponentially with increasing tissue depth due to attenuation and inverse-square
effects. Only approximately 1% of the original radiation beam reaches the I-I for
image generation.
The ESE exposure rate can be as high as 20 R/min under certain conditions
using a high dose rate or "boost" mode if the patient’s skin is close to the X-ray
tube. During cineangiography, ESE may exceed 90 R/min.
Fluoroscopy Timer
Fluoroscopy machines are equipped with a timer and an alarm which sounds at
the end of 5 minutes of fluoroscopy use. This system is designed alert the user
when the usage is becoming significant and provide additional warnings 5
minutes after each reset. Fluoroscopy systems also display the total fluoroscopy
time for a procedure. There are internal requirements for recording fluoroscopy
times discussed
Fluoroscopy Monitoring
The Radiation Safety Committee requires that the total elapsed fluoroscopy time
be recorded for every procedure that exceeds 10 minutes of fluoroscopy. Certain
areas, primarily in the Henry Ford Hospital, are also required to record all
fluoroscopy times. Sheets used to record fluoroscopy times are available from
the Radiation Safety Office.
Imaging Modes
Modern fluoroscopy machines produce images from an I-I that captures the
radiation exiting the patient. The machine can be operated in either a manual
mode or in an Automatic Brightness Control (ABC) mode (Figure 3-2). When the
ABC mode is selected, the ABC circuitry controls the X-ray intensity measured at
the I-I so that a proper image can be displayed on the monitor.
The radiation exposure rate is independent of the patient size, body part imaged
and tissue type when the manual mode is used. However, the image quality and
brightness are greatly affected (often adversely) by these factors when the
operator "pans" across tissues with different thickness and composition. For this
reason, most fluoroscopic examinations are performed using ABC.
Obese patients drive the X-ray machine output up considerably. Thus, obese
patients have the greatest risk of skin injury.
Magnification Modes
Under Normal mode, there is little magnification with the whole beam used to
generate a bright image. The "Normal" mode is used in the majority of
fluoroscopy procedures. The radiation output is sufficient to provide images for
guiding most procedures or observing dynamic functions. The typical exposure
rate at the X-ray beam entrance into the patient Entrance Skin Exposure (ESE) is
2 R/min.
Under Mag 1 mode, a smaller beam area is projected to the same I-I output. The
resulting object size is larger, but the image is dimmer due to the less beam
input.
The Food and Drug Administration (FDA) regulates the construction of all
fluoroscopy systems. For routine fluoroscopy applications, the FDA limits the
maximum ESE to 10 R/min. The use of higher radiation rates ("High Level
Control" or "boost" modes) are useful in situations requiring high video image
resolution. ESE of up to 20 R/min is permitted for short duration. Special operator
reminders, such as audible alarms, are activated during "boost" modes.
Cineangiography
The maximum useful area of the X-ray beam, or field size, is machine specific.
Most fluoroscopy systems allow the operator to reduce the field size through the
use of lead shutters or collimators. Figure 3-5 shows a diagram of an X-ray tube
and collimator system.
Prudent use of collimators can also improve image quality by blocking-out "bright
areas," such as lung or other low density regions, allowing better resolution of
other tissues.
1. The patient receives less total radiation exposure since less tissue is in
the radiation beam (Figure 3-7).
2. The workers in the area receive less radiation exposure since there is less
radiation available to scatter toward staff.
3. The image is improved since scatter contributes noise to the image.
The tube voltage (kV) controls the maximum energy of electrons produced by the
X-ray tube and the maximum energy of the resultant X-ray spectrum (kilovolt-
peak energy or kVp). Use of high kVp techniques can reduce patient dose but
contrast between differing tissues is also reduced. Experienced operators can
optimize the choice between image quality and patient dose through careful
adjustment. The ABC on most X-ray systems can change the kVp and mA used
to optimize imaging.
No portion of the operator's body should be in the primary beam during imaging.
Thus, the majority of the radiation dose received by the operator is due to
scattered radiation from the patient. After interacting with the patient, radiation is
scattered more or less uniformly in all directions (Figure 3-9).
It is important to note that the patient does not uniformly emit this scattered
radiation since some of the scattered radiation is absorbed or reduced in intensity
by passing through the patient. The intensity of scatter decreases with increasing
distance, due to inverse square law effects (see Chapter 1). Consequently,
scatter radiation is highest near its source (i.e. the X-ray beam entry point on the
patient). Because radiation scattered in the forward direction (into the patient) is
subject the most tissue attenuation, radiation levels are significantly lower on the
I-I side than the X-ray tube side.
Highest scatter radiation levels are often where the operator stands. Radiation
levels increase with decreasing distance from the point of X-ray entry (Figure 3-
10). In general, an operator positioned 3 feet from the X-ray beam entrance area
will receive 0.1% of the patient’s ESE. Staff members positioned further away
receive much less exposure due to inverse square law effects. In almost all
cases, the tableside operator will receive the highest occupational radiation
exposure during the fluoroscopic procedure. Tableside fluoroscopy receive
among the highest occupational radiation exposures within the health system.
Radiation levels are highest beneath the table (when the X-ray tube is below the
patient) because the patient provides an effective beam stop (Figure 3-11).
Highest levels are directed at the operator's waist (See bar chart on figure).
Effect of rotating X-ray system. Images taken with the I-I away (Figure 3-12)
result in higher radiation exposure to the operator's eyes compared to images
with the I-I towards the operator (Figure 3-13).
A basic principle in Radiology is that you collect dose along with image
information. The clearest, least jittery, images produce the highest doses. Thus,
it is very important for clinicians to judiciously use appropriate judgment when
increasing the X-ray beam output and to learn to work with most amount of
(necessary) imaging imperfections as possible which still allow the needed
clinical outcome.
The above skin changes were in areas not visible to the patients and were only
identified upon physical examination.
These case studies indicate that extensive use of fluoroscopy can induce severe
skin damage, even under the most favorable geometries.
The temporal progressions of these effects are consistent with high levels of
acute exposure to x-ray radiation. The temporal differences in the responses are
due in part to the levels of radiation received, but are also likely due to variations
in radiation sensitivity amongst the patients.
Multiple Procedures
Although intervals between procedures should permit the skin to recover, healing
might not be complete. This may lower the tolerance of the skin for further
procedures (Wagner 1999). Example cases are given below:
1. A patient underwent two PTCA procedures about one year apart. Skin
changes appeared approximately three weeks after the second procedure.
At seven weeks a cutaneous ulcer had developed over the right scapula
and healed without grafting.
2. A patient underwent two unsuccessful cardiac ablations involving
approximately 100 minutes of fluoroscopy in a lateral oblique orientation.
Approximately 12 hours after the second attempt, an
erythema developed in the right axilla. At one month the area was red and
blistering. At two years the area was described as a 10 x 5cm atrophic
indurated plaque with lineal edges, hyper- and hypopigmentation, and
telangiectasia. The patient was described as having difficulty raising her
right
arm.
3. Three PTCAs were performed on the patient, the last two completed on
the same day approximately 6 months after the first procedure. The total
fluoroscopy time was approximately 51 minutes. Erythema was noted
immediately after the last procedure. This progressed from a prolonged
erythema with poor healing into a deep dermal necrosis. The patient
underwent a successful split thickness skin graft two years after the last
procedure.
4. Past treatment of pulmonary tuberculosis often resulted in many patients
undergoing extensive exposure to fluoroscopy. These patients had a
demonstrated high incidence of breast cancer.
Previous procedures can lower the skin’s tolerance for future irradiation. Prior to
commencing any lengthy fluoroscopic procedure, the patient’s medical history
should be reviewed. The skin of the patient should be examined to ascertain if
any skin damage is apparent should the patient have a history of lengthy
fluoroscopic examinations. Direct irradiation of damaged areas should be
avoided when possible.
Positions of arms
Keeping arms out of the x-ray beam during some procedures can be a difficult.
Careful attention must be given to providing the arms with a resting position that
will not restrict circulation but will at the same time maintain the arms in an area
that is outside the radiation field (Archer 2000).
A middle-aged woman had a history of progressively worsening episodes of
arrhythmia. A radiofrequency electrophysiological cardiac catheter ablation was
scheduled to treat the condition. The procedure employed 20 min of beam-on
time for each plane of a bi-plane fluoroscope. Prior to the procedure the
separator or spacer cone was removed so that the fluoroscopic c-arms could be
easily rotated around the patient. The spacer cone is a spacer attached to the
tube housing designed to keep the patient at a reasonable distance from the x-
ray source. This is done specifically to avoid the high skin-dose rates that can be
encountered near the tube port.
The patient’s arms were originally placed at the patient’s side but the right arm
later fell into a lower position directly in front of this x-ray tube. However,
personnel were not aware of this change because sterile covers were draped
over the patient and did not correctly interpret the image (Figure 4-4). The right
humerus was directly in the beam at the port. Because the separator cones were
removed, the arm was only about 20–30 cm from the focal spot. With the soft
tissue and bone of the arm directly in the beam, the automatic brightness control
drove the output to high levels at the surface of the arm. The cumulative dose
probably exceeded 25 Gy (2500 rad). This procedure was not performed at
Henry Ford Health System.
The patient was released from the hospital the day after the procedure. At the
time there were no complaints regarding her arm and no indication of erythema.
About three weeks after the procedure, a bright erythema was demonstrated
(Figure 4-5).
The condition worsened and at five months a large ulcer the size of the
collimated x-ray port developed.
The separator or spacer cone ensures that a minimal distance between the X-
ray source and the patient is maintained (inverse square law effects). For some
X-ray machines, the spacer cone is designed to be removable in order to provide
more flexibility in positioning for some special surgical procedures (e.g., portable
C-arms). There is a risk of very high dose rates to the skin surface when it is
removed.
Skin Sensitivity
Erythema developed after diagnostic angiography and liver biopsy. Skin necrosis
requiring rib resection evolved in the same patient after a TIPS procedure. The
wound remained open for five years before a successful cover was put in place.
Investigation into the events revealed that the patient suffered from multiple
problems, including Sjøgren’s syndrome and mixed connective tissue disease.
Injuries to personnel
The following are modern-day examples of how improper use of the fluoroscope
can lead to injuries in personnel (Wagner 1999).
Doses accumulated to hands and eyes from frequently using the fluoroscope
with the tube above the patient can be extremely high. Only routine application of
proper radiation management techniques will be effective at avoiding such high
doses.
Go to Chapter 5: Protection Methods
Maximize Distance
A small increase in the operator's distance from the patient can significantly
reduce the operator's exposure. Standing one step further away from the patient
can cut the physician's exposure rate by a factor of 4 (AAPM 1998) (Figure 5-1).
You should periodically self-evaluate you personal technique to identify whether
opportunities to increase distance exist.
Many procedures require staff to intermittently interact with the patient near the
fluoroscopy system. The operator can reduce staff exposure by delaying
fluoroscopy until these activities are completed and/or by alerting these
personnel when imaging; especially during high dose rate modes like
cineangiography (Figure 5-3).
Provisions should be made to eliminate extraneous light that can interfere with
the fluoroscopic examination. Room lighting should be dim to enhance
visualization of the image. Excessive light can decrease the ability of the eye to
resolve detail. Measures taken to improve detail often involve increasing
patient/staff exposure.
Fluoroscopy examinations have the smallest operator exposure when the X-ray
tube is underneath the examination table (Figure 5-3). Whenever possible, the
operator should avoid the X-ray tube side of the table when imaging oblique or
lateral images.
Note: The benefit is exaggerated-some operator dose occurs on the I-I side.
Figure 5-4: Benefit of Reducing the Air Gap (I-I Close to Patient)
Courtesy of Sorenson, 2000.
Care should be taken whenever the image view angle is changed during the
procedure (e.g, changing from an ANT to a steep LAO). The I-I is often moved
away from the patient while changing X-ray tube position. Large air gaps can
result if the table or I-I height remains unadjusted.
After changing views, a 10-cm air gap between I-I and patient is inadvertently
maintained. What is the increase in radiation exposure to a 20-cm thick patient
positioned with the table 30 cm away from the X-ray source, assuming the ABC
compensates by increasing mA only?
Solution:
Assuming the air gap could have been eliminated by moving the I-I closer, and
that the brightness loss follows the inverse square law:
The brightness level with the air gap is only 69% of the zero air gap brightness.
The ABC system compensates for brightness loss by producing 31% more X-
rays. The exposure rate to the patient and staff is subsequently increased by
31%.
Reducing air gaps between patient and I-I also reduces image blur. Blurring of
the image is caused by geometric magnification caused by air gaps. Gaps
between patient and I-I enhance geometric magnification. The objects will appear
larger with increasing gap size. However, note that image edges are more fuzzy
(Figure 5-4). The degree of "fuzziness" will increase with increasing air gap.
Collimating the primary beam to view only tissue regions of interest reduces
unnecessary tissue exposure and improves the patient’s overall benefit-to-risk
ratio. Optimal collimation also reduces image noise caused by scatter radiation
originating from outside the region of interest (See Chapter 3). A good rule of
thumb is that fluoroscopy images should not be totally "round" when collimators
are available for use, the collimator edges should always be visible in the image.
Steeply angled oblique images (e.g., LAO 50 with 30 cranial tilt) are typically
associated with increased radiation exposure since: X-rays must pass through
more tissue before reaching I-I. ABC compensates for X-ray loss caused by
increased attenuation by generating more X-rays; Steep oblique angles are
typically associated with increased X-ray tube to I-I distances. The ABC
compensates for brightness loss caused by inverse square law effects by
generating more X-rays. Oblique views may bring the X-ray tube closer to the
operator side of the table, increasing radiation
exposure from scatter.
When possible, use alternate views (e.g., ANT, LAO with no tilt) when similar
information can be obtained (Figure 5-5). The physician can reduce personal
exposure by re-locating himself when oblique views are taken. For example,
dose rates can be reduced by a factor of 5 when the physician stands on the I-I
side of the table (versus X-ray tube side) during a lateral projection (AAPM
1998).
Selection of an adequate kVp value will allow sufficient X-ray penetration while
reducing the patient’s dose rate. In general, the highest kVp should be used
which is consistent with the degree of contrast required (high kVp decreases
image contrast).
Henry Ford Hospital has many resources available (e.g., Staff Radiologists,
Medical Physicists) to assist the operator in optimizing the fluoroscopy image
while minimizing patient exposure.
Ceiling-mounted lead acrylic face shields should be used whenever these units
are available, especially during cardiac procedures. Correct positioning is
obtained when the operator can view the patient, especially the beam entrance
location, through the shield.
Lead aprons do not stop all the x-rays. Typically at least a 80% reduction in
radiation exposure is obtained by wearing a lead apron (Figure 5-7). It should be
noted that the apron's effectiveness is reduced when more penetrating radiation
is employed (e.g., the ABC boost's kVp for thick patients). Two piece lead apron
systems are recommended for most users since they provide "wrap-around
protection" and distribute weight more evenly on the user. Some aprons contain
an internal frame that distributes some of the weight from the shoulders onto the
hips much like a backpack frame. So called "light" aprons should be scrutinized
to ensure that adequate levels of shielding are provided. State of Michigan law
requires the use of 0.5 mm lead equivalent aprons.
Note that higher tube voltages sharply reduces the shielding benefits of lead aprons. Higher tube
voltages will occur when imaging large patients or thick body portions. Also note that light aprons
(0.25 to 0.35 mm Pb) provide less protection compared to the recommended 0.5 mm thickness.
Thyroid shields provide similar levels of protection to the individual’s neck region.
Thyroid shield use is required for operators who use fluoroscopy extensively
during their practice.
Optically clear lead glasses are available that can reduce the operator's eye
exposure by 85-90% (Siefert 1996). However, due to the relatively high threshold
for cataract development, leaded glasses are only recommended for personnel
with very high fluoroscopy work loads (e.g., busy Radiology and Cardiology
Interventionists). Glasses selected should be "wrap-around" in design to protect
the eye lens from side angle exposures. Leaded glasses also provide the
additional benefit of providing splash protection. Progressive style lenses for
bifocal prescriptions are available from a limited number of manufacturers.
The latex leaded gloves provide extremely limited protection. Standard (0.5 mm
lead equivalent) leaded gloves provide useful protection to the user’s hands.
However, trade-offs associated with use of 0.5 mm leaded gloves include loss in
tactile feel, increased encumbrance and sterility. For these reasons, use of
leaded gloves is left to the operator’s discretion. To minimize radiation exposure
to the hands, the operator should:
The requirements for dosimetry has been determined by the Radiation Safety
Committee for each work area. These specify the types of dosimeters issued as
well as the collection frequency.
Some workers are issued a single whole body badge (black figure icon). This
whole body dosimeter should be worn on the collar outside of any protective
equipment worn (lead aprons). Readings from this position provide an estimate
of the radiation exposure to the eyes. Dose estimates to the individual’s whole
body are made using the appropriate algorithm. Other workers are issued
multiple dosimeters. These are designed to be worn as shown (Figure 5-8):
Infection Control has evaluated the use of ring badges in surgical arenas. For
open surgical theaters, ring badges are contraindicated. Catheter procedures
may be performed with ring badges.
Dosimetry Practices
Absent dosimeters are taken very seriously by the institution. Reports of which
individuals have failed to properly return dosimeters (who did not report the loss
of the dosimeter to the RSO) are sent to: the Radiation Safety Committee; the
Department chairs; the Hospital Medical Executive Committee; and the Board of
the institution. To avoid this negative attention, turn your dosimeter in on time
and promptly report the loss of a dosimeter to the Radiation Safety Office. A new
dosimeter will be issued at no cost and your good name will be preserved.
ALARA Philosophy
Regulatory dose limits should be viewed as the maximum tolerable levels. Since
stochastic radiation effects, such as carcinogenesis, can not be ruled-out at low
levels of exposure, it is prudent to minimize radiation exposure whenever
possible. This concept leads to the As-Low-As-Reasonably-Achievable (ALARA)
philosophy.
Simply stated, the ALARA philosophy requires that all reasonable measures to
reduce radiation exposure be taken. Typically, the operator defines what is
reasonable. The principles discussed in this manual are intended to assist the
operator in evaluating what constitutes ALARA for his/her fluoroscopy usage.
End of Manual!
Acknowledgements:
Select from one of the following exams, but do not repeat any exam you have
tested with previously: