Professional Documents
Culture Documents
Outline
Brief history of thermal analysis Theory of thermal analysis techniques
Thermal Gravimetric Analysis (TGA) Differential Scanning Calorimetry (DSC)
Recent advances
EYP 2006
Calorimetry
Calorimetry
The study of heat transfer during physical and chemical processes
Calorimeter
A device for measuring the heat transferred Lavoisier and Laplace (1782-1784): oil was burned in a lamp (Fig 9) held in a bucket (Fig. 8) held in a wire mesh cage (f) surrounded by ice in spaces b and a of the double walled container a foot in diameter lid (F) was topped with ice, as was a mesh lid (not shown) beneath it that covered the inner volume b
EYP 2006
EYP 2006
Sensors:
temperature sensors usually thermocouples
sample pan reference pan
Furnace:
one block for both sample and reference cells
inert gas vacuum thermocouples
Temperature controller:
temperature difference between the sample and reference is measured
EYP 2006
Sensors:
Pt resistance thermocouples. Separate sensors and heaters for the sample and reference
differential thermal power is supplied to the heaters to maintain the temperature of the sample and reference at the program value
EYP 2006
Outline
Brief history of thermal analysis Theory of thermal analysis techniques
Thermal Gravimetric Analysis (TGA) Differential Scanning Calorimetry (DSC)
Recent advances
EYP 2006
DSC Calibration
Baseline Calibration
evaluation of the thermal resistance of the sample and reference sensors measurements over the temperature range of interest
EYP 2006
DSC Calibration
Temperature
match melting onset temperatures to the known melting points of standards analyzed by DSC should be calibrated as close to desired temperature range as possible
Heat flow
use calibration standards of known heat capacity, slow accurate heating rates (0.52.0 C/min), and similar sample and reference pan weights metals Indium 156.6 C; 28.45 J/g Zinc 419.47C, 108.17 J/g inorganics KNO3 128.7 C KClO4 299.4 C organics polystyrene 105 C benzoic acid 122.3 C; 147.3 J/g
EYP 2006
calibrants
high purity accurately known enthalpies thermally stable light stable not hygroscopic do not react (pan, atmosphere)
Sample Preparation
accurately-weighed samples (~3-20 mg, usually 3-5 mg for simple powders) small sample pans (0.1 mL) of inert or treated metals (Al, Pt, stainless) several pan configurations, e.g., open , pinhole, or hermetically-sealed pans same material and configuration should be used for the sample and the reference material should completely cover the bottom of the pan to ensure good thermal contact avoid overfilling the pan to minimize thermal lag from the bulk of the material to the sensor * small sample masses and low heating rates increase resolution, but at the expense of sensitivity
Al
EYP 2006
Pt
alumina
Ni
Cu
quartz
Purge Gases
Sample may react with air - oxidising or burning Control moisture content of atmosphere Use inert gas e.g. nitrogen or argon Flowing purge gas In some cases deliberately choose reactive gas, e.g.
hydrogen to reduce an oxide to metal carbon dioxide which affects decomposition of metal carbonate
Outline
Brief history of thermal analysis Theory of thermal analysis techniques
Thermal Gravimetric Analysis (TGA) Differential Scanning Calorimetry (DSC)
Recent advances
EYP 2006
sulphapyridine
EYP 2006
Impure substances Broad, asymmetric melting peak melting characterized peak maxima at
0.0
Exo
-0.5
-1.0
-1.5
-2.0
-2.5 140
Exo Up
145
150
155
160
165
170
175
Universal V3.3B TA Instruments
Temperature (C)
EYP 2006
Enthalpy of Fusion
0.5
156.50C 28.87J/g
0.0
Exo
-0.5
-1.0
-1.5
-2.0
157.81C
-2.5 140
Exo Up
145
150
155
160
165
170
175
Universal V3.3B TA Instruments
Temperature (C)
EYP 2006
Endo
Purity by DSC
1-3 mg samples in hermeticallysealed pans are recommended Peak width a valuable measure of purity: impurities lower the melting point Less pure (non-perfect) crystals melt first followed by purer larger crystals polymorphism interferes with purity determination, especially when a transition occurs in the middle of the melting peak Accurate measurement of Hf needs pure samples of polymorphs
EYP 2006 benzoic acid
Exo
97%
99%
99.9%
Plato, C.; Glasgow, Jr., A.R. Anal. Chem., 1969, 41(2), 330-336.
Glass Transitions
characterized by change in heat capacity (no heat absorbed or evolved) transition from a disordered solid to a liquid appears as a step (endothermic direction) in the DSC curve gradual enthalpy change may occur, producing an endothermic peak superimposed on the glass transition Exo
EYP 2006
Recognizing Artefacts
Sample movement in pan sample pan distortion mechanical Pan shock / knock moves in bench furnace cool air entry into cell
sensor contamination
thermal stability melting crystallization solid-state transformations desolvation glass transition Exo sublimation decomposition heat flow heat of fusion heat of transition heat capacity
-0.5
-1.0
-1.5
Form II II Form II Variable Form III Hydrate Dihydrate Acetic acid solvate
-2.0
-2.5
50
100
150
200
250
300
350
Exo Up
Temperature (C)
EYP 2006
Lot A - pure
-1
-2
-3
Lot B - seeds
-4
-5
Exo Up
80
130
180
230
280
Universal V3.3B TA Instruments
Temperature (C)
Lot A: pure low melting polymorph melting observed Lot B: seeds of high melting polymorph induce solid-state transition below the melting temperature of the low melting polymorph
EYP 2006
Outline
Brief history of thermal analysis Theory of thermal analysis techniques
Thermal Gravimetric Analysis (TGA) Differential Scanning Calorimetry (DSC)
Recent advances
EYP 2006
Microcalorimetry
High sensitivity DSC Solutions Scan range typically 0-120 C Scanning rate of 0-120 C/hr Reverse scan rate 0-45 C/hr (depending on efficiency of cooling tank) Useful for looking at low energy modifications e.g. protein relaxation and refolding, polymer characterisation
EYP 2006
trehlose
Lot A
Lot A
-0.2
Lot B
-0.4
Lot B
-0.6
DSC010622b.1 483518 HCL (POLYMORPH 1) DSC010622d.1 483518 HCL
-0.50 110
Exo Up
120
130
140
150
160
170
180
Universal V3.3B TA
-0.8 110
Exo Up
120
130
140
150
160
170
180
Universal V3.3B TA
Temperature (C)
Temperature (C)
EYP 2006
Hyper DSC
Fast scanning DSC Only possible with power compensated Normal equipment 100 C/min Specialised up to 500 C/min Increased sensitivity, loss of resolution e.g. amorphous content in mainly crystalline sample
change of specific heat at Tg is linear relationship to the amorphous content Conventional DSC 10% amorphous limit of detection Hyper DSC <1% amorphous easily detected Lappalainen, M., I. Pitkanen, et al. (2006). International Journal of Pharmaceutics 307(2): 150-155.
EYP 2006
Caution
It is a bulk tool
Analysing the gross average of events in a sample Conversely, small powder sample in DSC may not represent packing of powder bulk in decomposition studies Instrument error in DSC typically 0.5 - 1C In Scanning modes, thermal events may be smeared by a thermal lag Sample temperature not keeping up with instrument Bigger effect at higher heating rates Typically 1C at 10C/min
EYP 2006
Acknowledgements
Professor Adrian Williams, University of Reading Dr Ian Grimsey, University of Bradford Dr Peter Timmins, Bristol Myers Squibb Dr Wendy Hulse, University of Bradford Luciana DeMatos, University of Sheffield
EYP 2006
Questions
EYP 2006
e.g. see Pharmaceutical Research: 17 (6): 696-700, June 2000 Craig, DQM et al.
EYP 2006
temperature calibration based on ferromagnetic transition of Curie point standards (e.g., Ni)
120 100
12.15%
80
19.32%
larger sample masses, lower temperature gradients, and higher purge rates minimize undesirable buoyancy effects
60
29.99%
Weight (%)
40
20
20
40
60
80
100
120
140
160
Time (min)
EYP 2006
Hyphenated Techniques
thermal techniques alone are insufficient to prove the existence of polymorphs and solvates other techniques should be used, e.g., microscopy, diffraction, and spectroscopy development of hyphenated techniques for simultaneous analysis TG-DTA TG-DSC
80 120
15.55% (0.9513mg) 24.80C 100.0% 179.95C 84.45%
4.2
3.2
Weight (%)
TG-FTIR TG-MS
2.2
40
1.2
0.2 0 -0.8
-40
Exo Up
20
70
120
170
220
270
-1.8
Temperature (C)