Professional Documents
Culture Documents
Electroencephalograph Machine
Electroencephalograph Machine
Machine
Kousha
Talebian
February
25,
2013
BMEG
550
Project
Report
Executive
Summary
An
electroencephalography
machine
is
an
instrument
that
measures
the
neural
activity
of
the
brain.
The
machine
consists
of
electrodes,
an
amplifier,
filters,
analog
to
digital
convertor,
a
computer
module,
and
a
monitoring
device.
The
amplifiers,
filters,
analog
to
digital
convertor
are
all
part
of
the
same
circuitry.
The
biggest
challenge
to
the
EEG
machine
is
not
the
data
collection,
but
rather
the
physiological
interpretation
of
the
data.
Since
the
collected
signal
contains
artifacts
from
different
biopotentials,
isolating
the
EEG
signal
is
to
some
measures
a
difficult
task.
Measuring
these
biopotentials
simultaneously
using
extra
electrodes,
and
then
removing
them
from
the
signal
is
common
practice.
The
specification
of
an
EEG
machine
is
not
universal,
as
different
applications
require
slightly
modified
EEG
machine.
The
frequency
of
an
EEG
falls
between
0.01
to
70Hz,
and
an
EEG
machine
should
at
least
respond
to
this
band.
The
higher
cutoff
frequency
may
be
larger
if
fast
sampling
rate
is
needed.
The
electrodes
themselves
require
a
very
inform
low
resistance
for
no
signal
distortion.
The
amplifiers
input
resistance
on
the
other
hand,
should
be
as
high
as
possible
with
a
common-mode
rejection-ratio
of
at
least
100dB.
An
EEG
signal
consists
of
4
wavelets,
labeled
as
alpha,
beta,
delta
and
theta,
covering
the
frequency
range
of
0.01
to
70Hz.
Each
has
its
own
characteristics
and
is
correlated
with
a
certain
physiological
feature.
The
digitalized
data
is
often
then
decomposed
into
these
wavelets
and
the
doctor
can
interpret
the
results
and
determine
the
cause
of
problem
with
the
patients
brain.
Table
of
Contents
List
of
Figures
............................................................................................................................
4
List
of
Abbreviation
.................................................................................................................
5
1.
Introduction
.......................................................................................................................
6
1.1.
History
......................................................................................................................................
6
2.
Source
of
EEG
Activity
.....................................................................................................
8
2.1.
Alpha
Wave
.............................................................................................................................
9
2.2.
Beta
Wave
.............................................................................................................................
10
2.3.
Delta
Wave
...........................................................................................................................
10
2.4.
Theta
Wave
...........................................................................................................................
10
3.
Usages
of
EEG
..................................................................................................................
11
3.1.
Clinical
Usage
.......................................................................................................................
11
3.2.
Research
Usage
...................................................................................................................
12
4.
EEG
Machine
....................................................................................................................
13
4.1.
Electrode
...............................................................................................................................
13
4.2.
Amplifiers
and
Filters
.......................................................................................................
17
4.3.
A/D
Convertor
.....................................................................................................................
21
4.4.
Computer
Program/Monitoring
...................................................................................
21
4.5.
Calibration
............................................................................................................................
22
4.6.
Technical
Specification
.....................................................................................................
23
5.
EEG
Signal
........................................................................................................................
25
5.1.
Sensitivity
Distribution
of
EEG
Signal
.........................................................................
25
5.2.
Behavior
of
the
EEG
Signal
..............................................................................................
26
5.3.
Basic
Principles
of
EEG
Diagnostics
.............................................................................
28
6.
Risks
&
Hazards
.............................................................................................................
29
7.
Artifacts
............................................................................................................................
30
7.1.
Patient
Artifacts
..................................................................................................................
30
7.2.
Technical
Artifacts
.............................................................................................................
30
7.3.
Artifact
Correction
.............................................................................................................
30
Conclusion
...............................................................................................................................
31
List of Figures
Figure 1: Frequency and voltage range of some of the most common biopotential signals (Webster, 2010). ............................................................................................................... 6 Figure 2: First recording of EEG performed by Dr. Berger in 1924 (Wallis, 2007). ..... 7 Figure 3: The connection of two neurons. The EEG machine measures the SP produced (Adeli & Ghosh-Dastidar, 2010). .......................................................................... 8 Figure 4: The four major types of EEG waves (Webster, 2010). ........................................... 9 Figure 5: The effect of opening and closing the eye replaces the alpha waves by asynchronous higher frequency waves (Webster, 2010). .......................................... 10 Figure 6: A typical EEG machine. The instrument on the right is the amplifier/filter (taken from http://emgneedleelectrode.com) ................................................................ 13 Figure 7: A subdural electrode .......................................................................................................... 14 Figure 8: A surface electrode ............................................................................................................. 14 Figure 9: An electrode cap (Teplan, 2002) .................................................................................. 15 Figure 10: The Standard International 10/20 configuration (Kropotov, 2005) ......... 16 Figure 11: The Standard International 10/10 configuration (Kropotov, 2005) ......... 16 Figure 12: An EEG channel is the difference between two electrodes voltage (EBME.co.uk Ltd, 2008) .............................................................................................................. 17 Figure 13: Differential amplifier producing the EEG channel (EBME.co.uk Ltd, 2008) ............................................................................................................................................................... 18 Figure 14: Calculation showing why the difference between two channels is always the difference between the two electrodes, irrelevant to the manner the channels were produced. ........................................................................................................... 18 Figure 15: (A) Bipolar and (B) common reference. Note that the montage of the EEG channel depends on the measurement location (Malmivuo & Plonsey, 1995) . 19 Figure 16: The amplifier/filter component of the EEG machine. All electrodes plus the reference and the ground signals are inserted into this machine. The operator can use the computer to configure the differential manner and view the montage on the monitor (Picture courtesy of the MitSar Medical). ............... 19 Figure 17: A sample screenshot of the MitSar program used to monitor EEG. ........... 22 Figure 18: D179 Performance Checker used to calibrate an EEG machine ................... 23 Figure 19: The Rush-Discroll model. The sensitivity to the brain region decreases as the electrodes move closer to each other (Malmivuo & Plonsey, 1995) .............. 26 Figure 20: Summary of the signal decomposition showing how different waves are extracted from the EEG signal (Adeli & Ghosh-Dastidar, 2010). .............................. 27 Figure 21: An example of a decomposed EEG signal into wavelets (Malmivuo & Plonsey, 1995). ............................................................................................................................... 27 Figure 22: EEG signal of an awake, light sleep, REM sleep and deep sleep person along with a person in coma (Malmivuo & Plonsey, 1995). ....................................... 28
List
of
Abbreviation
A/D
AP
C
ECG
EEG
EMG
EP
F
O
P
Ref
SNR
SP
T
Pp.
Analog to Digital Action Potential Central Electrocardiography Electroencephalography Electromyography Evoked Potential Frontal Occipital Posterior Reference Signal to Noise Ratio Synaptic Potential Temporal Voltage peak-peak
1. Introduction
An
electroencephalogram
(EEG)
machine
is
a
device
used
to
create
a
diagram
of
the
electrical
activity
of
the
brain.
The
machine
is
used
both
for
medical
diagnosis
(see
section
3.1),
as
well
as
research
application
(see
section
3.2).
On
the
figure
below,
the
voltage
and
frequency
band
of
the
EEG
is
compared
with
the
other
biopotential
signals;
there
are
significant
overlaps
between
the
different
signals,
and
this
will
lead
to
bio-artifacts
artifacts
collected
by
the
sensor
from
a
biopotential
signal
that
is
not
intended
for.
Figure
1:
Frequency
and
voltage
range
of
some
of
the
most
common
biopotential
signals
(Webster,
2010).
1.1.History
By
the
end
of
the
nineteenth
century,
scientists
were
trying
to
study
and
measure
the
electrical
property
of
the
brain.
In
1875,
Richard
Catton,
a
British
surgeon
from
Liverpool,
measured
the
electrical
potential
of
the
exposed
cortex
of
rabbits
and
monkeys
brain.
In
1902,
Hans
Berger,
studied
the
correlation
between
the
brain
activity
and
the
physiological
phenomena
(such
as
sleep,
epilepsy,
anesthesia).
He
used
the
Lippmann
capillary
electrometer
and
Edelman
galvanometer
to
record
the
electrical
pulses.
However,
none
produced
a
satisfactory
result.
It
was
finally
in
1924
where
Dr.
Berger
made
the
first
EEG
recording
of
a
humans
brain.
Figure
2:
First
recording
of
EEG
performed
by
Dr.
Berger
in
1924
(Wallis,
2007).
It took another five-years of research by Edgar Douglas Adrian and B. C. H. Matthews, who reproduced Bergers results, before EEG was accepted. In 1936, Walter Gray used a large number of electrodes pasted onto the scalp to create a map of the brain activity. Areas around a tumor showed abnormal electrical activity.
Figure
3:
The
connection
of
two
neurons.
The
EEG
machine
measures
the
SP
produced
(Adeli
&
Ghosh-Dastidar,
2010).
Under normal conditions, actions of the chemical transmitters at the postsynaptic neurons contribute very little to the overall biopotential energy on the scalp. The firing of these SP are asynchronous and in random directions. Consequently, the net result of the spatial and temporal influence of the potential energy on the scalp is small and negligible. When a person performs a specific task, the part of brain that is associated with, fires the SP synchronously and uniformly. These potentials (called evoked potentials EP1) have relatively high amplitudes and can be detected with electrodes. It is important to note that the electrical activity monitored by the EEG is the postsynaptic pulse and the electrode collects the perpendicular component of the signal. 1 Evoke potentials are potentials due to a stimulus 8
The
EEG
is
described
as
either
rhythmic
or
transients.
The
rhythmic
activities
are
divided
into
bands
based
on
the
frequency
and
are
found
at
different
part
of
the
human
brain.
Table
below
summarizes
the
main
four
frequency
bands.
There
are
also
the
two
more
waves
known
as
gamma
and
mu
that
are
not
discussed
in
this
report.
Table
1:
Comparison
between
the
alpha,
beta,
delta,
and
theta
wave
group
of
the
EEG
(Webster,
2010)
Group Frequency (Hz) Location Delta Below 3.5 Frontal in adults, posterior in children Theta 4 - 7 Location not related to a task at hand Alpha Beta 8 - 13 13 - 40
Normally - Sleeping adults - Babies - Young children - Drowsiness and arousal in adults - Idling Posterior region of - Relaxing/reflecting head - Blinking - Inhibition control Evenly distributed - Alert on both sides, more - Active/busy/thinking towards the frontal
Pathologically - Diffuse lesion - Deep midline - Focal subcortical lesions - Some instances of hydrocephalus - Coma - Benzodiazepines (BZs)
Figure
4:
The
four
major
types
of
EEG
waves
(Webster,
2010).
can also be from the frontal and posterior of the scalp as well. The amplitude of these waves ranges form 20 to 200V. When a person sleeps, alpha waves completely disappear. The alpha waves are also replaced by asynchronous waves of higher frequency due to sensation or when the person is directed towards a specific activity. In the figure below, one can observe that the action of closing and opening the eye completely replaces the alpha waves.
Figure
5:
The
effect
of
opening
and
closing
the
eye
replaces
the
alpha
waves
by
asynchronous
higher
frequency
waves
(Webster,
2010).
2.2.Beta
Wave
Although
these
waves
normally
occur
between
13-30
Hz,
during
intense
mental
activity,
they
can
go
as
high
as
50Hz.
They
are
usually
recorded
from
the
frontal
and
parietal
region
of
the
scalp.
There
are
two
types
of
beta
waves:
beta
I
and
beta
II.
Beta
I
acts
just
as
alpha
waves.
Beta
II
on
the
other
hand,
appear
during
intense
CNS
activity.
Thus,
beta
I
is
elicited
by
mental
activity,
and
beta
II
is
inhibited
by
it.
10
3. Usages
of
EEG
According
to
R.D.
Bickford
(Bickford,
1987),
EEG
signals
are
used
to
1. Monitor
alertness,
coma
and
brain
death
2. Locate
areas
of
damage
following
head
injury,
stroke,
tumor,
etc.
3. Test
afferent
pathways
(by
evoked
potentials)
4. Monitor
cognitive
engagement
(alpha
rhythm)
5. Produce
biofeedback
situations,
alpha,
etc.
6. Control
anesthesia
depth
(servo
anesthesia)2
7. Investigate
epilepsy
and
locate
seizure
origin
8. Test
epilepsy
drug
effects
9. Assist
in
experimental
cortical
excision
of
epileptic
focus
10. Monitor
human
and
animal
brain
development
11. Test
drugs
for
convulsive
effects
12. Investigate
sleep
disorder
and
physiology.
3.1.Clinical
Usage
A
routine
EEG
recording
can
be
as
short
as
20min
to
as
long
as
couple
of
days.
It
usually
involves
the
attachment
of
8+
electrodes
to
the
scalp,
depending
on
the
test
and
the
level
of
resolution
required.
EEG
devices
were
the
first
method
for
detecting
tumors.
However,
MRI
and
CT
scans
have
replaced
the
EEG
machine
for
detection
of
tumors.
The
EEG
recording
is
also
used
to
distinguish
between
epileptic
seizures
from
the
other
types
(psychogenic
non-epileptic
seizures).
It
can
also
be
used
to
differentiate
between
delirium
and
catatonia.
Since
the
EEG
waves
are
a
measure
of
brain
activity,
then
it
can
be
used
to
assess
brain
death/coma.
A
patient
with
sleep
disorder
can
also
be
stationed
at
the
hospital
and
be
connected
to
an
EEG
for
up
to
a
week
to
record
and
analyze
their
sleep
pattern.
2
This
is
my
research
thesis
with
Dr.
Guy
Dumont.
The
BIS
index,
which
is
a
spectral
analysis
of
EEG,
is
used
as
the
depth
of
hypnosis
(DOH)
signal.
Aim
of
the
thesis
is
design
an
adaptive
closed
loop
control
of
the
(DOH).
11
3.2.Research
Usage
EEG
is
extensively
used
in
neuroscience,
cognitive
science,
and
anesthetic
studies.
Even
though
EEG
has
been
replaced
by
MRI
and
CT
scans,
it
is
still
used
for
cases
where
high
spatial
resolution
of
<1mm
is
required.
In
the
field
of
anesthetics,
EEG
is
used
as
a
control
objective
as
it
is
directly
related
to
the
patients
anesthetic
state.
The
signal
therefore
can
be
used
to
automatically
administrate
the
drugs
required
to
keep
the
patient
at
a
certain
level.
Researchers
are
still
trying
to
understands
all
the
connection
between
all
physiological
and
the
EEG
signal.
12
4. EEG
Machine
An
EEG
machine
can
simply
be
broken
down
into
four
components:
the
electrode,
amplifiers,
a
computer
control
module,
and
a
display
device.
One
of
the
advantages
of
the
EEG
is
that
for
its
simplicity,
it
offers
substantial
insight
into
brain
activity.
Generally,
the
electrodes
are
manufactured
using
German
silver:
an
alloy
made
up
of
copper,
nickel
and
zinc.
German
silver
is
soft
enough
to
grind,
and
hard
enough
not
to
break.
Alternatively,
stainless
steel
is
sometimes
used
for
its
corrosion
resistant
property.
However,
since
it
is
a
harder
material,
manufacturing
process
is
more
challenging
and
costly.
Figure
6:
A
typical
EEG
machine.
The
instrument
on
the
right
is
the
amplifier/filter
(taken
from
http://emgneedleelectrode.com)
4.1.Electrode
There
are
5
basics
types
of
electrodes:
1. Disposable
(pre-gelled)
2. Reusable
3. Headband
and
electrode
cap
4. Saline-based
electrode
5. Needle
electrode
(inserted
into
the
brain)
13
The first four electrodes are known as surface electrodes as they are attached to the scalp. The last one is a subdural electrode, as the scalp is punctured and the electrode is placed into the brain directly.
Figure
7:
A
subdural
electrode
Figure
8:
A
surface
electrode
For a multichannel montage3, the electrode cap is recommended as it reduces the setup time. Skin preparation differs slightly, but cleaning the oil and dead skin is recommended. For some electrode types (specially the reusable Germany silver electrodes), a conductive gel is needed.
3 A montage is the final result of filtration and decomposition of the EEG signal 14
Figure
9:
An
electrode
cap
(Teplan,
2002)
The EEG signal is simply the difference of voltage between two electrodes. To make the comparison of the signal between different institutes and hospitals easier, the Canadian Dr. Herpert Jasper has defined an international standard configuration. Called the Standard International 10/20, this configuration shows the placement of the electrodes on the patients head. Each electrode should be placed within 10% of the standard. There is also a standard sequence of measurements (Rahey, 2007). Before the introduction of this standard, EEG readings were limited in accuracy, as even the same doctor would place the electrodes at different locations on every patient. If the electrodes are not placed on the right position, then the perpendicular components of the signal to the electrodes could be minimized, leading to a weaker signal. The scalp is divided into regions: F (frontal), C (central), P (posterior), O (occipital), and T (temporal). The letters are followed by odd numbers for the left side of the brain, and even numbers for right side of the brain. Left and right side is denoted from the patients point of view.
15
Figure
10:
The
Standard
International
10/20
configuration
(Kropotov,
2005)
A higher resolution EEG can be obtained by using the Standard International 10/10 configuration shown below.
Figure
11:
The
Standard
International
10/10
configuration
(Kropotov,
2005)
High input impedance leads to increased signal distortion, which reduces SNR and therefore, the actual signal being lost. To prevent a small SNR, an electrode input impedance of < 5k and balanced within 1k of all electrodes must be used.
16
With modern EEG instruments, the choice of ground electrode is minimal. Forhead (Fpz) or near ear location is common practice, but legs and wrists are sometimes used as well. There are also several different reference electrode placement mentioned in literature. The reference electrode could be chosen as the linked ear, tip of nose, C7, cortex, and etc. The only constraint is that the reference electrode must be at an electrically neutral location. Each choice of the location has its own advantage and disadvantages.
Figure
12:
An
EEG
channel
is
the
difference
between
two
electrodes
voltage
(EBME.co.uk
Ltd,
2008)
Each of the channels is then subtracted from the ground electrode to remove the background noise and the low-frequency compartment (signal from the biopotential from other organs). When input 1 is more negative then input 2, the deflection is up.
17
Figure
13:
Differential
amplifier
producing
the
EEG
channel
(EBME.co.uk
Ltd,
2008)
There are three manners in which the electrodes are connected: common reference derivation, average reference, or bipolar. The difference between two channels is always the difference between the two electrodes; irrelevant to the manner the channels were produced. This can easily be seen as followed. Assume common-reference derivation is used. Denote channel 1 as F1-Ref and channel 2 as F4-Ref. Then channel 1 minus channel 2 has the reference signal cancelled and produces F1-F4.
Figure
14:
Calculation
showing
why
the
difference
between
two
channels
is
always
the
difference
between
the
two
electrodes,
irrelevant
to
the
manner
the
channels
were
produced.
In the common-reference derivation, each amplifier records the difference between each electrode and the common reference. The reference signal is usually the electrode by the ears.
18
Figure
15:
(A)
Bipolar
and
(B)
common
reference.
Note
that
the
montage
of
the
EEG
channel
depends
on
the
measurement
location
(Malmivuo
&
Plonsey,
1995)
In the average-reference derivation, the average-reference signal is denoted as the average of all electrodes. The amplifier then records the difference between each individual electrode and the average reference. The EEG machine allows the operator to remove any of the channels to not be included in the calculation of the average reference signal. In bipolar derivation, each amplifier records the difference between two electrodes that are opposite of each other. For instance, F3-C3 or C3-P3 are either opposite or transpose of each other. This will lead to maximum SNR, as it will be discussed in section 5.1.
Figure
16:
The
amplifier/filter
component
of
the
EEG
machine.
All
electrodes
plus
the
reference
and
the
ground
signals
are
inserted
into
this
machine.
The
operator
can
use
the
computer
to
configure
the
differential
manner
and
view
the
montage
on
the
monitor
(Picture
courtesy
of
the
MitSar
Medical).
The EEG amplifiers input is equipped with diodes in order to prevent the voltage to travel backwards, towards the scalp. This is a safety feature and prevents shock to the patient in case of a device malfunction.
19
The basic requirements that all biomedical amplifiers must have are (Nagel, 1995):
The
physiological
process
to
be
monitored
should
not
be
influenced
in
any
way
by
the
amplifier
The
measured
signal
should
not
be
distorted
The
amplifier
should
provide
the
best
possible
separation
of
signal
and
interferences
The
amplifier
has
to
offer
protection
of
the
patient
from
any
hazard
of
electrical
shock
The
amplifier
itself
has
to
be
protected
against
damages
that
might
result
from
high
input
voltages
as
they
occur
during
the
application
of
defibrillators
or
electrosurgical
instrumentation
The
EEG
signal
recorded
by
the
electrode
consists
of
five
signals:
the
desired
biopotential,
undesired
biopotential,
power
line
signal
interference
of
60Hz
(the
AC
power
line
frequency),
interference
produced
because
of
bad
connection
between
the
electrode
and
the
scalp,
and
noise.
For
isolating
the
electrically
noisy
environment,
a
high
amplifier
input
impedance
of
at
least
100M
is
implemented.
The
amplifier
also
requires
having
a
high
common-mode
rejection-ratio
(CMMR)
of
at
least
100dB
(Teplan,
2002).
The
amplification
gain
is
1,000
to
100,0004
(Nagel,
1995).
The
optimal
gain
is
one
that
maximizes
the
SNR.
The
raw
scalp
EEG
signal
is
about
10-100V
and
10-20mV
when
subdural
electrodes
are
used.
Filters
are
implemented
in
the
amplification-integrated
unit
to
remove
noise
and
unwanted
artifacts.
A
high
pass
filter
is
used
to
remove
the
low
frequency
noise
4
The
higher
the
gain
does
not
mean
the
better
the
signal
will
be.
There
are
a
few
parameters
(such
as
sampling
rate,
the
A/D
conversion,
the
noise)
that
will
determine
the
optimal
gain
20
that remains in the signal after the subtracting the voltage from the ground5 (Teplan, 2002). The cutoff frequency of this high-pass filter is between 0.1 0.7Hz (Bronzino, 1995). A low-pass filter is used to ensure the signal is band- limited. The cutoff frequency is the highest frequency of the signal, which falls between 40Hz (beta wave) and half the sampling rate (which is usually 512Hz) (Bronzino, 1995). If higher frequencies survive, then according to Nyquist theorem, aliasing could occur.
After the data is digitalized, they can go under further digital filtration, according to the objective of the test being performed. Linear filtering such as FIR or IIR to more novel non-linear filtering methods is used. Clinical machines have a 12bit A/D convertor (Teplan, 2002), but research convertors might be larger if higher resolution is required.
Figure
17:
A
sample
screenshot
of
the
MitSar
program
used
to
monitor
EEG.
4.5. Calibration
Routine
calibration
is
required
to
model
the
circuitry,
and
connection
wires
noise
as
well
as
determining
the
exact
amplification
gain.
Usually
a
known
impulse
(sine
or
triangle)
is
generated
on
all
of
the
inputs
of
the
main
EEG
amplifier.
Thus,
the
generated
signal
passes
through
the
entire
EEG
machine,
with
the
exception
of
the
electrodes.
The
output
can
be
recorded
and
compared
with
the
input
to
determine
the
amplification
gain,
and
to
model
the
noise.
This
noise
should
mainly
be
from
the
amplifier
circuit
and
the
A/D
converter.
Noise
value
should
be
consistent
with
the
manufacturer
spec,
and
is
usually
0.3-2VPP
(Teplan,
2002).
Noise
can
also
be
determined
by
short-circuiting
the
inputs
of
the
amplifier,
or
by
placing
the
electrodes
into
a
salt-bath,
and
then
measuring
the
output.
Since
input
signal
is
zero,
then
the
output
signal
is
the
noise
model.
Once
the
noise
is
modeled,
the
number
of
useful
bits
can
easily
be
computed.
Taking
the
ratio
of
the
average
generated
EEG
signal
amplitude
over
the
noise
amplitude,
the
nearest
exponent
of
2
is
the
number
of
useful
bits.
For
instance,
if
the
ratio
is
50V/1V,
then
there
are
5
useful
bits.
22
There are commercial calibrators (known as performance checker), which generate the required calibrating signal. More advanced EEG machines have a built-in calibrators. An example of such a device is the D179 Performance Checker by Digitimer. D179 has 32 outputs with 1 ground/common socket and 2 reference sockets. The output of the EEG machine is connected to the D179, and the Performance Checker will automatically compare the input to the output and model the noise and the amplification gain.
Figure
18:
D179
Performance
Checker
used
to
calibrate
an
EEG
machine
23
frequencies between 0.1-5Hz, a low-pass filter centered at 50-256Hz. There should also be a 50-60Hz AC filter to remove the power line noise. It requires a CMMR of at least 100dB with amplifier input impedance of at least 100M and electrode resistance less than 5k. Below is the specification of BWII EEG machine: 25 channel: 22 EEG channels, 2 Ref, 1 ground Calibration on board: 0.5Hz, 7Hz, 15Hz (square and sine wave) CMMR ~ 125dB Low/high pass filter 0.0016 to 70Hz A/D 12bit conversion As a comparison, the government of India requires the following minimum requirement from an EEG machine 32 channel Frequency response between 0.05 to 70Hz Low-pass filter of 15Hz, 30Hz, and 70Hz High-pass filter of 0.1Hz, 0.3Hz, 1.5Hz, 3Hz, 5Hz) CMMR > 100dB Amplifier input resistance > 10M
24
5. EEG
Signal
5.1. Sensitivity
Distribution
of
EEG
Signal
An
important
question
when
performing
EEG
is
the
sensitivity
of
the
signal;
in
other
words,
how
deep
into
the
brain
can
the
signals
be
generating
from
and
still
be
detected
by
the
EEG
instrument?
This
problem
is
of
no
significance
when
subdural
electrodes
are
used;
it
is
a
matter
of
discussion
for
surface
electrodes.
Rush
and
Driscoll
performed
the
first
such
studies
in
1969.
In
their
study,
the
patients
head
is
modeled
as
a
perfect
sphere
an
assumption
only
partially
valid.
They
place
two
electrodes
on
the
patients
head;
one
produces
a
current
while
the
other
measures
the
current.
Their
results,
while
pioneer
in
nature,
had
limited
practicality.
In
1987,
Puikkonen
and
Malvimuo
used
the
same
model
as
Rush
and
Driscoll,
but
presented
the
results
with
lead
field
current
flow
lines.
The
direction
of
the
sensitivity
is
then
simply
normal
to
the
field
current
flow
lines
and
the
magnitude
of
it
is
seen
from
the
density
of
these
field
lines
(Puikkonen
&
Malmivuo,
1987).
A
limitation
of
this
theorem
is
that
it
is
a
2D
model,
representing
a
3D
distribution.
The
field
lines
thus
break
to
denote
a
change
in
the
third
dimension
(see
Figure
19.
Note
how
the
blue
lines
suddenly
break,
denoting
a
movement
in
the
third
dimension).
In
the
following
diagrams,
the
two
electrodes
are
separated
by
angles
of
180,
120,
60,
40,
and
20.
In
the
diagrams,
the
solid
blue
lines
are
the
field
current
flow
lines,
the
dotted
black
lines
are
the
isosensitivity
lines,
and
the
green
shaded
region
is
known
as
Half-sensitivity
volume
the
volume
at
which
the
field
current
density
is
at
least
50%
of
its
maximum
value.
The
location
of
the
two
electrodes
is
visible
from
the
highly
dense
field
current
lines.
The
sphere
itself
has
three
shades:
the
outer
is
the
skin,
the
middle
is
the
skull
and
the
inner
25
is the cerebral region. As the electrodes move closer and closer to each other, the current flows more in the skin region, thus decreasing the access to the brain itself. The noise also increases as a result (Suihko, Malmivuo, & Eskola). The conclusion of their work then suggests using electrodes that are as far from each other as possible to achieve optimal quality (Malmivuo & Plonsey, 1995).
Figure 19: The Rush-Discroll model. The sensitivity to the brain region decreases as the electrodes move closer to each other (Malmivuo & Plonsey, 1995)
26
Figure
20:
Summary
of
the
signal
decomposition
showing
how
different
waves
are
extracted
from
the
EEG
signal
(Adeli
&
Ghosh-Dastidar,
2010).
With the separated EEG wavelets, it is possible to perform diagnostics. Below is a sample wavelet of an actual EEG signal.
Figure
21:
An
example
of
a
decomposed
EEG
signal
into
wavelets
(Malmivuo
&
Plonsey,
1995).
27
Figure
22:
EEG
signal
of
an
awake,
light
sleep,
REM
sleep
and
deep
sleep
person
along
with
a
person
in
coma
(Malmivuo
&
Plonsey,
1995).
28
29
7. Artifacts
Artifacts
are
signals
with
high
amplitudes
and
different
shapes
as
compared
to
the
desired
biopotential
signal.
They
can
either
be
patient
artifacts
(unwanted
physiological
biopotentials),
or
technical
artifacts
(power
line
noise,
and
interference
signals).
7.1.Patient
Artifacts
These
are
the
signals
that
are
picked
up
from
the
scalp,
but
are
originated
from
the
non-cerebral
compartment.
Many
physiological
signals
can
arise
from
the
patients
body.
Any
body
movements,
such
as
blinking
the
eye,
can
cause
a
spike
in
the
EEG
signal.
EMG
(signal
from
skeletal
muscle),
and
ECG
(signal
from
the
heart)
can
also
be
picked
by
the
EEG
electrode
and
can
contribute
to
the
artifact.
7.2.Technical
Artifacts
These
are
the
signals
that
have
originated
from
the
outside
of
the
patients
body.
Movement
by
the
patient
can
cause
the
electrode
to
move,
and
create
an
electrode-skin
interference.
The
electrode
impedance
can
sometimes
fluctuate
and
cause
signal
distortion.
Poor
grounding
can
also
cause
significant
60Hz
artifacts.
7.3.Artifact
Correction
To
correct
the
patient
artifacts,
additional
electrodes
to
monitor
eye
movement,
the
EMG
and
ECG
may
be
required.
Recently,
using
Fourier
analysis,
signals
have
been
decomposed
and
the
components
related
to
other
biopotentials
have
been
weighted
out.
For
the
technical
artifacts,
better
grounding,
insulation
and
higher
quality
electrodes
can
be
used
to
minimize
the
noise
picked
up
by
the
system.
30
Conclusion
Electroencephalography
belongs
to
a
bioelectrical
imaging
toolsets
used
widely
in
both
clinical
as
well
as
research
environments.
EEG
machine
measures
changes
in
the
electrical
potentials
created
as
the
results
of
excitation
of
the
postsynaptic
neurons.
The
signal
itself
is
composed
of
four
primarily
waves,
denoted
as
alpha,
beta,
theta,
and
delta.
EEG
machine
is
a
non-invasive
method
to
measure
the
EEG
signal
(with
the
rare
occasion
when
subdural
electrodes
are
used)
and
causes
no
pain
and
has
minimal
risk
to
the
patient.
The
EEG
machine
itself
consists
of
an
electrode
(from
either
German
silver
or
actual
silver)
coated
with
conductive
gel,
an
amplifier
with
a
gain
of
1,000
100,000
with
an
input
impedance
of
at
least
100dB,
analog
high-pass
filter
to
remove
low
frequency
noise
with
a
cutoff
frequency
centered
at
0.1-0.7Hz,
analog
low-pass
filter
to
band-limit
the
signal
and
remove
high
frequency
noise
with
a
cutoff
frequency
centered
at
50Hz
to
half
sampling
frequency,
at
least
12bit
A/D
convertor,
EEG
software
capable
to
perform
digital
filtering
and
wavelet
decomposition
to
create
the
required
montages.
Many
believe
that
EEG
machine
will
lead
to
a
wide
range
of
discoveries
in
basic
brain
functionality
as
well
as
better
understanding
of
neurological
diseases.
31
Bibliography
Adeli,
H.,
&
Ghosh-Dastidar,
S.
(2010).
Automated
EEG-Based
Diagnosis
of
Neurological
Disorders
-
Inventing
the
Future
of
Neurology.
Boca
Raton:
Taylor
&
Francis
Group.
Ananthi,
A.
(2006).
A
Textbook
of
Medical
Instruments.
New
Delhi:
New
Age
International
Pvt
Ltd
Publishers.
Bickford,
R.
D.
(1987).
Electroencephalography.
In
G.
Adelman,
Encyclopedia
of
Neuroscience
(pp.
371-373).
Birkhauser:
Cambridge.
Bronzino,
J.
D.
(1995).
Principles
of
Electroencephalography
.
In
J.
D.
Bronzino,
The
Biomedical
Engeneering
Handbook
(pp.
201-212).
Florida:
CRC
Press.
Brunet,
D.,
&
Young,
G.
(2000).
Electroencephalography,
Guidelines
for
Clinical
Practice
and
Facility
Standards.
Ontario:
College
of
Physicians
and
Surgeons.
EBME.co.uk
Ltd.
(2008).
Introduction
to
EEG.
Retrieved
2013,
from
EBME:
http://www.ebme.co.uk/arts/eegintro/eeg2.php
Fisch,
B.
(1999).
Fisch
and
Spehlmann's
EEG
Primer:
Basic
Principles
of
Digital
and
Analog
EEG.
Elsevier.
Hirsch,
L.
J.,
&
Brenner,
R.
P.
(2006).
Atlas
of
EEG
in
Critical
Care.
Wiley-Blackwell.
Kropotov,
J.
D.
(2005).
Quantitative
EEG,
Event-Related
Potentials
and
Neurotherapy.
New
York:
Elsevier
.
Malmivuo,
J.,
&
Plonsey,
R.
(1995).
Bioelectromagnetism.
New
York:
Oxford
University
Press.
Nagel,
H.
N.
(1995).
Biomedical
Amplifiers.
In
J.
D.
Bronzino,
The
Biomedical
Engineering
Handbook
(pp.
1185-1195).
Florida:
CRC
Press.
Puikkonen,
J.,
&
Malmivuo,
J.
A.
(1987).
Theoretical
investigation
of
the
sensitivity
distribution
of
point
EEG-electrodes
on
the
three
concentric
spheres
model
of
a
human
head
-
An
application
of
the
reciprocity
theorem.
Tampere
Univ.
Techn.,
Inst.
Biomed.
Eng.,
Reports.
Rahey,
S.
R.
(2007).
Fundamentals
of
EEG
Technology.
Halifax:
Capital
Health.
Rush,
S.,
&
Driscoll,
D.
A.
(1969).
EEG-electrode
sensitivity
-
An
application
of
reciprocity.
EEE
Trans.
Biomed.
Eng
,
15-22.
32
Suihko, V., Malmivuo, J., & Eskola, H. Distribution of sensitivity of electric leads in an inhomogeneous spherical head model. Tampere Univ. Techn., Ragnar Granit Inst., Rep. 7. 1993. Tatum, W. O., Kaplan, P. W., Benbadis, S. R., & Husain, A. M. (2008). Handbook of EEG Interpretation. Demos Medical Publishing. Teplan, M. (2002). Fundamental of EEG Measurement. Retrieved 2013, from Measurement Science Review: http://www.ccs.fau.edu/eeg/teplan2002.pdf Wallis, C. (2007, September 20). Electroencephalogram Lab. Retrieved February 16, 2013, from California State University Long Beach: http://www.csulb.edu/~cwallis/482/eeg/eeg.html Webster, J. G. (2010). Medical Instrumentation Application and Design. Wiley.
33