CTS-10566

Seal, Head Plunger, 2pk Equivalent to: Waters WAT270938

Technical Data Sheet


1.0 Description
These Seals are designed to be compatible with Waters 2690, 2690D, 2695, 2695D, 2790, 2795, and GPC/V 2000 HPLC systems. They are equivalent to Waters part number WAT270938.

2.0 Specifications
2.1 Wetted Materials
UHMW-PE 7. Reinstall the plunger & seal wash assembly into the pump head. 8. Install the assembly into the pump. 9. Finish the diagnostic for that pump. 10. Repeat for the other pump.

3.0 Replacement Frequency


Replacement may be indicated by leaks at the plunger head or if you start to see poor chromatography.

5.0 Warranty
We view your business with us as a partnership based on mutual trust. We extend a Lifetime Warranty against any manufacturing defects on all CTS parts. On the rare occasion that you experience a problem, we invite you to contact your Account Manager directly. Returns can only be accepted if a Return Material Authorization Number (RMA) is obtained prior to shipment. Please call (1) 952-895-8292 or email: ctssales@ctshplc.com to get an RMA.

4.0 Installation Information


CTSs parts are designed to be compatible and interchangeable with the OEM parts 1. Initiate the Head Removal & Replacement diagnostic. 2. Step through the procedure until the pump head, plunger, & seal wash system are removed as an assembly from the pump. 3. Remove the plunger & seal wash assembly from the pump head. 4. Remove the plunger seal from the pump head. 5. Wet the new seal and seal opening with 100% alcohol before installing. 6. Insert the new seal into the pump head with the spring side into the pump head. Press firmly into place.