You are on page 1of 7

Ultrasound in Med. & Biol., Vol. 25, No. 8, pp.

1221–1227, 1999
Copyright © 1999 World Federation for Ultrasound in Medicine & Biology
Printed in the USA. All rights reserved
0301-5629/99/$–see front matter

PII S0301-5629(99)00083-6

● Original Contribution

IMPLEMENTATION OF SPECTRAL WIDTH DOPPLER IN PULSATILE


FLOW MEASUREMENTS

BOR-RAY LEE*, HUIHUA KENNY CHIANG*, YI-HONG CHOU†, CHENG-DENG KUO‡,


JIA-HORNG WANG‡ AND SAN-KAN LEE§
*Institute of Biomedical Engineering, National Yang-Ming University, Taipei, Taiwan; †Department of Radiology
General Veteran Hospital-Taipei/National Yang-Ming University, Taipei, Taiwan; ‡Department of Respiratory
Therapy, General Veteran Hospital-Taipei/National Yang-Ming University, Taipei, Taiwan; §Department of
Radiology, General Veteran Hospital-Taichung, Taichung, Taiwan

(Received 4 January 1999; in final form 7 June 1999)

Abstract—In this paper, we present an automatic beam-vector (Doppler) angle and flow velocity measurement
method and implement it in pulsatile flow measurements using a clinical Doppler ultrasound system. In current
clinical Doppler ultrasound flow velocity measurements, the axis of the blood vessel needs to be set manually on
the B-scan image to enable the estimation of the beam-vector angle and the beam-vector angle corrected flow
velocity (the actual flow velocity). In this study, an annular array transducer was used to generate a conical-
shaped and symmetrically focused ultrasound beam to measure the flow velocity vectors parallel and perpen-
dicular to the ultrasound beam axis. The beam-vector angle and flow velocity is calculated from the mode
frequency (fd) and the maximum Doppler frequency (fmax) of the Doppler spectrum. We develop a spectrum
normalization algorithm to enable the Doppler spectrum averaging using the spectra obtained within a single
cardiac cycle. The Doppler spectrum averaging process reduces the noise level in the Doppler spectrum and also
enables the calculation of the beam-vector angle and flow velocity for pulsatile flows to be measured. We have
verified the measurement method in vivo over a wide range of angles, from 52° to 80°, and the standard deviations
of the measured beam-vector angles and flow velocities in the carotid artery are lower than 2.2° and 12 cm/s
(about 13.3%), respectively. © 1999 World Federation for Ultrasound in Medicine & Biology.

Key Words: Flow velocity measurement, Doppler spectrum, Beam-vector angle, Spectrum averaging,
Annular array

INTRODUCTION axis. Hein (1997) used a three-transducer probe to mea-


sure the flow vectors in three-dimensional space. New-
Medical Doppler ultrasound is a powerful tool for de-
house et al. (1987) used lateral transit time broadening
tecting blood vessel disease in patients, and has been
(Andrew and Junru 1995; Censor et al. 1988; Chabria
widely used to measure the flow velocity in blood ves-
and Newhouse 1997; Tortoli et al. 1995) to measure the
sels. In spectral Doppler ultrasound, knowledge of the
flow velocity perpendicular to the beam axis, and they
beam-vector angle between the nominal beam direction
also used two transducers (Newhouse et al. 1994) to
and the direction of blood motion is required. The oper-
measure flow vectors in three-dimensional space. One
ator manually set the angle cursor along the blood vessel
on the B-scan image (Fish 1990). This process is incon- promising technique, an automatic Doppler angle and
venient, operator-dependent, and is inaccurate for mea- flow velocity estimation method (Lee and Chiang 1999)
suring the beam-vector angle. Recently, many research- that utilizes the combination of classic and transverse
ers have developed several flow velocity measurement Doppler effects, has been validated for constant flow
methods to solve this problem. Bohs et al. (1993, 1995) measurements. This technique uses only one annular
used a two-dimensional speckle tracking technique to array transducer to generate a symmetrically focused
measure the flow velocity perpendicular to the beam ultrasound beam to measure the Doppler spectrum,
which inherently contains the information of the trans-
verse and longitudinal components of the flow vectors
Address correspondence to: Dr. Huihua Kenny Chiang, National
Yang-Ming University, Institute of Biomedical Engineering, 155, Section (as seen in Fig. 1). Therefore, the beam-vector angle and
2, Li-Numg Street, Taipei 112, Taiwan. E-mail: chiang@bme.ym.edu.tw flow velocity could be measured. However, due to pres-

1221
1222 Ultrasound in Medicine and Biology Volume 25, Number 8, 1999

variant, so the Doppler spectra cannot be averaged di-


rectly, as in constant flow measurements.
Although many blood flow velocity measurement
techniques have been developed, a spectral Doppler-
based ultrasound flow velocity measurement method that
utilizes only one transducer and is capable of measuring
pulsatile flows has not yet been presented. In this paper,
we theoretically developed a Doppler spectrum normal-
ization algorithm, an effective and fast Doppler spectra
data-averaging process to be finished in a single cardiac
cycle, for pulsatile flow measurements, and experimen-
tally implemented this measurement method to in vivo
pulsatile flow measurements.

PRINCIPLE
Fig. 1. The schematic diagram of the beam-vector angle and
flow velocity measurement using an annular array transducer. In classic Doppler, the relationship between the
The transverse, longitudinal components of the flow velocity, flow velocity v and the mode frequency fd of the Doppler
and the beam-vector angle can be measured. VA and VB are two spectrum is:
arbitrary flow vectors with the same beam-vector angle.

2䡠v
fd ⫽ 䡠 cos␪ (1)

ence of noise and signal variations in the Doppler spec-
trum, the Doppler spectrum needs to be averaged many
times to reduce the spectral noise. Figure 2 shows the where ␭ is the wavelength of acoustic wave, and ␪ is the
variation in the calculated beam-vector angle vs. the measured beam-vector angle. Sometimes, fd is also re-
number of spectral lines in the averaging. The standard ferred to as the mean frequency of the Doppler spectrum
deviations of the calculated angles are less than 2°, when (Tortoli et al. 1995), rather than the mode frequency used
more than 15 spectra are used in the averaging. Because in this article. Theoretically, when the center of the
the flow velocity of a constant flow is time-invariant, the sample volume is much smaller than the diameter of a
Doppler spectra can be directly averaged, whereas in conduit and is positioned at the center of the conduit, the
pulsatile flow conditions, the flow velocities are time- mean frequency equals to the mode frequency (Law et al.
1991).
For a converging ultrasound beam crossed by a flow
line, a multiplicity of ray-to-flow line angles can be said
to exist. Instead of a single frequency shift, a broad-range
frequency shift around fd would be observed in a mea-
sured Doppler spectrum. This broadening effect has been
described as transverse Doppler theory by Newhouse et
al. (1987) and is described by the following equation:

2␯ W
Bd ⫽ 䡠 䡠 sin␪ (2)
␭ F

where Bd is the spectrum bandwidth, W is the aperture of


transducer and F is the focal length.
Tortoli et al. (1995) have shown that the maximum
frequency (fmax) of the Doppler spectrum could be rep-
resented as fd plus one-half of the Doppler spectrum
bandwidth. The Doppler spectrum bandwidth represents
Fig. 2. The calculated beam-vector angle vs. the number of the transverse component of the flow velocity measured
spectrum averaging times for a constant flow measured at 70°, by the transducer. The maximum frequency fmax of the
33 cm/s. Doppler spectrum is:
Spectral width Doppler in pulsatile flow ● B.-R. LEE et al. 1223

␯W Therefore, fmax only depends on the flow velocity v,


f max ⫽ f d ⫹ 䡠 sin␪ (3) and the parameters, ␭, W, ␪, and F, are fixed constants.
␭ F
For two spectra with different flow velocities measured
at the same beam-vector angle (denoted spectrum⫺1 and
The longitudinal and transverse components of the flow
spectrum⫺2), the scaling factor between spectrum⫺1
velocity vectors could be represented by the functions of
(flow velocity v1) and spectrum⫺2 (flow velocity v2) can
fd and (fmax ⫺ fd). Therefore, the flow velocity and the
be determined as:
beam-vector angle could be represented as (derived in
Appendix A):
f max⫺1 ␯ 1
R⫽ ⫽

冑冋 册 冋
(7)

␭ 2
␭䡠F 2 f max⫺2 ␯ 2
␯⫽ 䡠f ⫹ 䡠 共 f max ⫺ f d兲 (4)
2 d W
The scaling factor R can be also determined from the
ratio of the energy peak Doppler frequency fd⫺1 兾fd⫺2.
␪ ⫽ tan⫺1 冋 f max ⫺ f d 2 䡠 F
fd

W 册 (5) Because the maximum Doppler shift frequency fmax
would be less noisy, the ratio of fmax⫺1兾fmax⫺2 is used to
determine the scaling factor R. Thus, a set of Doppler
Equations (4) and (5) have been theoretically derived and spectra can be normalized to a specific Doppler spectrum
experimentally verified in constant flow measurements by a scaling factor R before the spectrum-averaging
(Lee and Chiang 1999). process.
The main problem associated with using eqns (4)
and (5) is that the Doppler spectrum needs averaging to METHODS
reduce the noise level, and then to determine the stable
values of fd and fmax. Averaging a consecutive set of Flow phantom measurement
Doppler spectra is straightforward in constant flow mea- Doppler systems. A color Doppler ultrasound sys-
surements, but requires ECG gating or some other means tem, GE-Vingmed CFM-750 (GE-VingMed Sound,
for pulsatile flow measurements. Figure 2 shows the Horten, Norway), with a four-element annular array
variation of the calculated beam-vector angle vs. the transducer, operating at 2.5 MHz in Doppler mode and
number of lines of spectral data averaged. The beam 3.25 MHz in B-scan, 1.47 cm in diameter and 7.8 cm in
vessel angle was set at 70° and the standard deviation focal length, was used in the flow phantom experiments.
was less than 2° after 15 spectra lines were averaged. The The length of sample volume was kept at 1.5 mm long
Doppler spectrum-averaging process reduces the noise and was set at the focus. The pulsed repetition frequency
level and fluctuations in the spectrum, and improves the was 10 kHz. Each Doppler spectrum calculation time
accuracy of the measurement. However, in pulsatile flow (including sampling and computation time) in the CFM-
measurements, the flow velocities vary with time; thus, 750 is less than 8 ms.
the spectrum cannot be averaged directly as in constant
flow measurements. Flow phantom. We used a UHDC (Shelley Medical
To reduce the noise level and fluctuations in the Imaging Technologies, London, Ontario, Canada) flow
Doppler spectrum obtained from a pulsatile flow mea- phantom to generate pulsatile flows. The flow velocity at
surements, a Doppler spectrum normalization and aver- the systolic peak of pulsatile flow was set at 33 cm/s in
aging algorithm is developed. The normalization of the a latex conduit with a 9.5-mm internal diameter. The
Doppler spectrum is based on the assumption that the conduit was 1 m long and was immersed horizontally in
beam-vector angle is unchanged during the fast systolic a water tank. The UHDC blood mimic fluid was used as
region of the cardiac cycle (about 0.12 s). Therefore, for the scattering source in the flow. The density of the fluid
pulsatile flows measured at a fixed beam-vector angle, was 1.01 g/mL, and the dynamic viscosity was 1.34 cp (1
the maximum Doppler shift frequency fmax can be sim- cp ⫽ 10⫺2 g/s 䡠 cm ).
plified from eqn (3): Signal processing. The maximum Doppler fre-
quency fmax is defined as the intersection of the Doppler
f max ⫽ ␯ 䡠 ⌬ (6) spectrum, and an empirically determined threshold is set
at ⫺6 dB. The setting of the threshold for fmax should be
where the constant empirically determined for each Doppler ultrasound sys-
tem, because each system has its own Doppler gain

⌬⫽
2
␭冉cos␪ ⫹
W
2䡠F
䡠 sin␪ 冊 curve. In this experiment, the threshold was determined
by calibrating a set of estimated beam-vector angles to a
1224 Ultrasound in Medicine and Biology Volume 25, Number 8, 1999

set of known angles (e.g., 50°, 60°, 70°) set in UHDC and Fig. 3D represents the beam-vector angle corrected
flow phantom. Different thresholds (including ⫺6 dB, flow velocities calculated by mode frequencies at the
⫺8 dB, ⫺10 dB, ⫺15 dB and ⫺20 dB) have been systolic peak. In Fig. 3B–D, the solid lines represent the
empirically tested by a controlled flow phantom system. mean values, and the error bars represent the standard
In measuring the flow velocity of the pulsatile flow, deviations in 12 trials. The standard deviations of calcu-
a consecutive set of the Doppler spectra was obtained in lated flow velocities in Fig. 3D are lower than 2.8 cm/s,
the systolic cycle (as shown in the white dotted rectan- and the standard deviations of calculated beam-vector
gular window in Fig. 3A). These measured spectra were angles in Fig. 3C are lower than 2.2°. In Fig. 3B, fmax and
normalized to the Doppler spectrum at the systolic peak fd at 85° are overestimated, because the measured Dopp-
and were averaged to obtained a less noisy Doppler ler spectrum is affected significantly by the Doppler
spectrum. The reasons why we chose the Doppler spectra system’s wall filter and most of the Doppler spectra are
in the systolic cycle were that (1) the spectra of fast flows filtered out. Therefore, fd are overestimated and (fmax ⫺
would be less affected by the wall filter of the Doppler fd) are underestimated, so that the beam-vector angle will
system, and (2) to insure the measured spectra to be in be underestimated, as shown in Fig. 3C.
the laminar flow.
Figure 3A shows the Doppler spectrum of a pulsa- In vivo measurements
tile flow generated by a UHDC flow phantom. The The measurements were conducted at different
systolic peak flow velocity was set at 33 cm/s. The beam-vessel angles between 52° and 80° in 4° incre-
beam-vector angle was set at 50°. A consecutive set of ments. Figure 4B and C show the calculated beam-vector
Doppler spectra obtained during the systolic cycle was angles and the angle corrected flow velocities of a 25-y-
normalized and averaged to estimate the beam-vector old male in six trials. In Fig. 4B and C, the solid lines
angle and the flow velocity at the systolic peak. represent the mean values, and error bars represent the
standard deviations. The standard deviations of the
In vivo measurements beam-vector angles are lower than 2.2°, as shown in Fig.
For in vivo measurements, a different type of annu- 4B, and standard deviations of flow velocities are lower
lar array transducer, four-element, operating at 3.5 MHz than 12 cm/s, as shown in Fig. 4C.
in Doppler mode and 4 MHz in B-scan mode, 1.15 cm in Figure 5A–C shows the traces of the flow velocities
diameter, and 4.7 cm in focal length, was used to mea- of a human subject’s carotid artery measured, respec-
sure the carotid artery of a 25-y-old male subject, from tively, at 56°, 68° and 76° beam-vessel angle settings.
52° to 80° beam-vessel angle. The beam-vector angle The calculated beam-vector angles are 57.5°, 69° and
and flow velocity were calculated in the chosen part in 72.8°, respectively. The three traces represent the flow
the measured spectra (as shown in Fig. 4A). The GE- velocities calculated without angle correction, with man-
VingMed 750 can provide 18 spectra within 0.12 s in the ually set beam-vessel angle correction, and with the
chosen part of the measured spectra. automatic beam-vector angle correction (our method).
In a human subject’s carotid artery measurements, The major limitation in spectral Doppler measure-
the longitudinal direction of the carotid artery was first ments is caused by the Doppler system’s wall filter. It
manually set using a vessel axis marker on the B-scan causes difficulties in measuring slow flow or at large
ultrasound image. The vessel axis and the boundary of beam-vector angles. The existence of the wall filter is
the blood vessel are highlighted by solid line in Fig. 4A. essential, but it inevitably filters the lower frequency
The Doppler sample volume was set at the center of the portion of the Doppler spectrum. We can observe, in Fig.
blood vessel and kept at the minimum 1.5 mm. The value 5C, that flows measured at large beam-vector angles or
of the reference beam-vessel angle is determined visually measured during their slow flow region would be af-
by physicians by aligning the vessel axis marker along fected by the wall filter. Therefore, the spectral Doppler
the blood vessel axis. This process may incur some method is not suitable for large beam-vector angles (larg-
human errors in determining the value of the reference er than 80° in this case) and slow flow velocity (lower
beam-vessel angle. than 18 cm/s) measurements.

RESULTS DISCUSSION
Flow phantom measurements The best measurement position is at the focus where
Figure 3B represents the fmax and fd determined the beam has the narrowest width and highest intensity;
from the spectrum normalization and averaging process the effect of the lateral transit time broadening will be
from 45° to 85° beam-vessel angles (reference angles). most salient, and the signal will also be most sensitive. In
Figure 3C represents the calculated beam-vector angles, modern ultrasound systems, the multiple zone focusing is
Spectral width Doppler in pulsatile flow ● B.-R. LEE et al. 1225

used in transmitting, and the dynamic focusing is used in


receiving. The measurements do not have to be con-
ducted at the focus. However, if the Doppler sample
volume is set at the focus, not set at the center of a certain
multiple focusing zone, then the lateral transit time
broadening effect will be negatively decreased. Also, the
sensitivity of the measurement will be affected, accord-
ing to the spectrum broadening effect discussed by Ne-
whouse et al. (1987).
The nonstationary effect caused by the flow velocity
acceleration is quite small in this measurement. For
example, in our experiment, the pulsatile flow velocity of
the carotid artery rose from 0.3 m/s to 0.95 m/s in 0.12 s,
corresponding to a less than 25-Hz nonstationary spectral
broadening in the higher frequency portion of spectral
width in our experiment (see Appendix B). In this mea-
surement, the difference between fd and fmax varied from
138 to 415 Hz, so we could neglect the effect caused by
the flow velocity acceleration.
The 1.5-mm sample volume still contained velocity
gradients in the sample volume. However, the fd and fmax
were least affected because the sample volume was po-
sitioned at the center of the vessel (Law et al. 1991).
Velocity gradients in the sample volume affect the spec-
tral broadening effect mostly in the lower frequency
portion of the Doppler spectrum, although the effect is
minimized in the upper frequency portion, as described
by Tortoli et al. (1995). This characteristic also leads to
the choice of using the maximum frequency and the
mode frequency in this measurement method.

CONCLUSIONS
The major assumption of the presented Doppler
spectrum normalization algorithm is that the beam-vec-
tor angle remains unchanged during the chosen part of
the cardiac cycle. Therefore, a consecutive set of Dopp-
ler spectra can be normalized and averaged, even though
these Doppler spectra are obtained from different flow
velocities. The time in the chosen part of the cardiac
cycle is less than 0.12 s. Thus, it could be easily imple-
mented in real time operation with appropriate imple-
mentation.

Fig. 3. (A) Real-time Doppler spectrum of a pulsatile flow in a


flow phantom, with a 33-cm/s systolic peak flow velocity,
measured at 50° beam-vector angle. The spectrum normaliza-
tion window is chosen in the fast flow region (systolic region)
marked by the white dotted bracket. (B) The maximum Doppler
frequency fmax and mode frequency fd measured over different
beam-vessel angle settings. (C) The calculated beam-vector
angles vs. the referenced beam-vessel angles. (D) The beam-
vector angle corrected flow velocity calculated by mode fre-
Fig. 3. quency at the systolic peak.
1226 Ultrasound in Medicine and Biology Volume 25, Number 8, 1999

Fig. 4. (A) The two-dimensional duplex scan of a carotid artery


measured at a 56° beam-vessel angle setting (AC ⫽ ⫺56°),
with a 1-m/s systolic peak flow velocity. The vessel axis and
the boundary of the blood vessel are highlighted by solid black Fig. 5. Time traces of the measured flow velocity in a carotid
lines. (B) The calculated beam-vector angles vs. the manually artery without or with beam-vector angle correction. The man-
set beam-vessel angles. (C) The beam-vector angle corrected ually set beam-vessel angles and the automatically calculated
systolic peak flow velocities calculated by mode frequency in beam-vector angles are: (A) 56° and 57.5°, (B) 68° and 69° and
the carotid artery. (C) 78° and 72.8°.
Spectral width Doppler in pulsatile flow ● B.-R. LEE et al. 1227

It is worth noting that this simple method only Hein IA. 3-D flow velocity vector estimation with a triple-beam lens
transducer experimental results. IEEE Trans Ultra Ferro Freq Con
measures the sign and magnitude of the beam-vector 1997;44:85– 89.
angle ␪; that is, the magnitude of the transverse flow Law YF, Bascom PAJ, Johnston KW, Vaitkus P, Cobbold RSC.
velocity vector ⱍv⬜ⱍ , and the longitudinal flow velocity Experimental study of the effects of pulsed Doppler sample volume
size and position on the Doppler spectrum. Ultrasonics 1991;29:
vector v//, as shown in Fig. 1. Thus, two arbitrary flows 404 – 410.
VA and VB with the same magnitude of beam-vector Lee BR, Chiang HK, Kuo CD, Lin WL, Lee SK. Doppler angle and
angle ␪ A and ␪ B cannot be distinguished by using this flow velocity estimations using the classic and transverse Doppler
effects. IEEE Trans Ultra Ferro Freq Con 1999;46:252–256.
method. However, this limitation does not affect mea- Newhouse VL, Censor D, Vontz T, Cisneros JA, Goldberg BB. Ultra-
suring the flow velocity in regular blood vessels, in sound Doppler probing of flows transverse with respect to beam
which the blood flows in only one direction. Therefore, axis. IEEE Trans Biomed Eng 1987;34:779 –789.
Newhouse VL, Dickerson K, Cathignol SD, Chapelon JY. Three-
the application of this simple method could successfully dimensional vector flow estimation using two transducers and spec-
measure the beam-vector angle and flow velocity in tral width. IEEE Trans Ultra Ferro Freq Con 1994;41:90 –95.
blood vessels. Tortoli P, Guidi G, Newhouse VL. Improved blood velocity estimation
using the maximum Doppler frequency. Ultrasound Med Biol
In summary, we present a convenient flow velocity 1995;21:527–532.
estimation method, which is accurate, and is also easy to
implement in existing Doppler ultrasound systems. One APPENDIX A
additional advantage of this method is that it can measure
Equation (3) minus eqn (1):
the blood flows not only in the longitudinal scan but also
in the cross-sectional scan. This characteristic enables
␯ W
the measurement of curved blood vessels in which the f max ⫺ f d ⫹ 䡠 䡠 sin ␪ (A1)
␭ F
longitudinal scan is hard to obtain. Future works will
apply this method in renal arteries, hepatic arteries, aorta,
Equation (A1) divided by eqn (1):
etc.
Acknowledgements—Parts of this work were supported by the National f max ⫺ f d W
⫽ 䡠 tan␪ (A2)
Science Council, R.O.C. fd 2䡠F

REFERENCES where W is the aperture of the transducer, and F is the focal length.
Thus, the beam-vector angle (Doppler angle) ␪ is derived:
Andrew JW, Junru W. Correction of intrinsic spectral broadening
errors in Doppler peak velocity measurement made with phased
sector and linear array transducer. Ultrasound Med Biol 1995;21:
1928 –1935.
Bohs LN, Freiemel BH, McDermott BA, Trahey GE. A real time
␪ ⫽ tan ⫺1 冋 f max ⫺ f d 2 䡠 F
fd

W 册 (A3)

system for quantifying and displaying two-dimensional velocities


using ultrasound. Ultrasound Med Biol 1993;19:751–761. APPENDIX B
Bohs LN, Freiemel BH, Trahey GE. Experimental velocity profiles and
volumetric flow via two-dimensional speckle tracking. Ultrasound In our experiment, the pulsatile flow velocity of carotid artery
Med Biol 1995;21:885– 898. rose from 0.3 to 0.95 m/s in 0.12 s. The measured mode frequency fd
Censor D, Newhouse VL, Vontz T. Theory of ultrasound Doppler- rose from 762 to 2287 Hz, and the maximum frequency fmax rose from
spectra velocimetry for arbitrary beam and flow configurations. 900 to 2702 Hz. The higher frequency portion of bandwidth (fmax ⫺ fd)
IEEE Trans Biomed Eng 1988;35:740 –751. rose from 138 to 415 Hz. The nonstationary parameter ␤ is:
Chabria Y, Newhouse VL. Estimation of axial blood velocity using the
Doppler equation corrected for broadening. Ultrasound Med Biol ␤ ⫽ ⌬f d/⌬t, (B1)
1997;23:967–968.
Fish P. Physics and instrumentation of diagnostic medical ultrasound.
Chichester: John Wiley and Sons; 1990. where ⌬fd ⫽ (2.287 ⫺ 0.762) ⫽ 1.525 kHz, and kHz, and ⌬t ⫽ 120 ms.
Fish PJ. Nonstationary broadening in pulsed Doppler spectrum mea- Therefore, the corresponding nonstationary broadening caused by the
surements. Ultrasound Med Biol 1991;17:147–155. flow velocity acceleration was less than 25 Hz (Fish 1991).

You might also like