Professional Documents
Culture Documents
Article
Article
www.chromatographyonline.com
500
www.chromatographyonline.com
Injector Column
HPLC pump
0.25 mL/min
1.0
1.5
Time (min)
2.0
Phenacetin
signal
(%)
Circle 54
10-4
10-3
10-2
10-7
10-8
10-5
100
ESI-MS
5 L/min
Syringe pump
(phenacetin)
1.0
10-6
www.chromatographyonline.com
100
50
Intensity for
urapidil (%)
502
100
APCI Interface
1.0
2.0
3.0
50
ESI Interface
1.0
2.0
3.0
Time (min)
Figure 3: Comparing ion suppression experienced using APCI (top) and ESI (bottom). The
chromatograms show the effect of sample components after protein precipitation, eluted
from the analytical column on the signal intensity of a postcolumn infusion of 10 M urapidil
(compare the experimental setup in Figure 1). (Reproduced from reference 14 by courtesy of
Elsevier Science.)
tially very high basicities and surface activities, the limit concentration of 105 M of
Circle 56
ions is reached quickly, and ion suppression occurs with such samples (10).
Another theory for signal suppression
in ESI considers the effects of an increase
of viscosity and surface tension of the
droplets from the high concentrations of
interfering compounds, thus, reducing
solvent evaporation and the ability of the
analyte to reach the gas phase (11,12).
Finally, the presence of nonvolatile material also has been evaluated as the cause for
ion suppression. Nonvolatile materials
can decrease the efficiency of droplet formation through coprecipitation of analyte
or preventing droplets from reaching their
critical radius required for gas phase ions
to be emitted (8). In addition to the
described condensed-phase processes,
analyte ions also can be neutralized in the
gas phase via deprotonation reactions
with high gas-phase basicity substances,
thus, leading to suppression of their
response signal, similar to the reactions
seen in APCI (12).
APCI frequently gives rise to less ion
suppression than ESI (Figure 3), which is
related to the different mechanisms of
ionization (9,10). As a result, ion suppression in the presence of coeluted compounds often is different for ESI and
APCI. Unlike ESI, there is no competition between analytes to enter the gas
phase, because neutral analytes are transferred into the gas phase by vaporizing the
liquid in a heated gas stream. Also, ion
suppression is not related directly to
charge saturation, because the maximum
number of ions formed by gas-phase ionization is much higher, as reagent ions are
redundantly formed (10). Nonetheless,
APCI does experience ion suppression,
which has been explained by considering
the effect of sample composition on the
efficiency of charge transfer from the
corona discharge needle (13). In addition,
because there is very little chance for analytes to pass through the vaporization
region and remain in solution, another
mechanism of ion suppression in APCI is
solid formation, either as pure analyte or
as a solid coprecipitate with other nonvolatile sample components (14).
Although the actual mechanisms of ion
suppression in the different API interfaces
are still under investigation, the consequences must be considered carefully. Ion
suppression usually results in reduced
detection capability, possibly even to the
extent of a false negative for an existing
analyte. The other extreme, a false positive, also can occur for applications where
maximum residue limits are monitored, if
the internal standard experiences ion suppression. Due to the natural variation of
endogenous compounds in biological
samples, varying levels of ion suppression
often result as well. This variation in turn
can lead to both systematic and random
errors in the signal response, effecting ion
intensity ratios and linearities.
Detecting Ion Suppression
www.chromatographyonline.com
80.3
Absorbance (mAU, 265 nm)
504
60.0
Quercetin
40.0
2
20.0
-12.0
0
10
20
30
40
50
60
70
80
Time (min)
Figure 4: Chromatograms obtained using sample preparation via molecular-imprinted polymer extraction (lower trace) and no sample preparation (upper trace). (Reproduced from reference 26 by courtesy of Elsevier Science.)
Reducing
Ion
506
www.chromatographyonline.com
preparation and chromatographic selectivity are the two most effective ways of
circumventing ion suppression. Often, it
is easiest to adjust the chromatographic
conditions so that the analyte peaks are
not eluting in regions of suppression (5).
This protocol requires a chromatographic
profile to determine where the interferences are eluted and then a method to
effectively resolve the analyte from those
interferences. The two areas of the chromatographic elution that are most
affected by interferences are the solvent
front, where unretained compounds are
eluted, and the end of the elution gradient, where the strongly retained compounds are eluted. So it is recommended
to adjust the capacity factors of the analytes to elute them between these two
regions and from other areas that can be
affected by ion suppression.
An easy and effective way to change
chromatographic selectivity is by modifying mobile phase strength or gradient
conditions. A change in the mobile
phases organic solvent can provide dramatic selectivity changes in HPLC. As
well, related to mobile phase chemistry, is
the use of additives and buffers to aid in
Circle 59
1.8
80
60
40
20
% Intensity
1.0
% Intensity
1.0
% Intensity
% Intensity
2.5
1.2
1.0
0.8
0.6
1.5
2.0
2.5
Echo
peak
Sample
peak
0.2
0.0
10 12 14 16 18 20 22 24
Time (min)
1.0
1.5
2.0
2.5
80
60
40
20
1.0
1.5
2.0
80
60
40
20
% Intensity
(f)
2.0
80
60
40
20
0.5
(e)
17.92
1.4
0.4
0.5
(d)
1.5
80
60
40
20
0.5
(c)
17.07
1.6
0.5
(b)
www.chromatographyonline.com
(a)
% Intensity
508
1.0
1.5
2.0
80
60
40
20
0.5
1.0
1.5
2.0
Time (min)
Figure 5: The effect of different sample preparation techniques on the amount of remaining endogenous plasma components and the extent of ion suppression, respectively, for
phenacetin. Shown are MRM chromatograms of a postcolumn infusion of phenacetin, after
injecting 10 L of a blank plasma sample on-column, prepared by each of the following sample preparation methods: (a) protein precipitation blank, (b) Oasis SPE (Waters Corp., Milford,
Massachusetts), (c) methyl-tert-butyl ether LLE, (d) Empore C2 disk SPE (3M, St. Paul, Minnesota), (e) Empore C8 disk SPE, and (f) Plasma Empore C18 disk SPE. (Reproduced from reference 25 by courtesy of John Wiley and Sons.)
Calibration
510
(6).
The most widely used technique
involves internal standards. An internal
standard allows the response of a given
analyte to be normalized, thus, compensating for possible variations during sample preparation, injection, chromatography, matrix effects, and so forth. To
compensate for matrix effects, the internal
standard must have ionization properties
very similar to those of the analyte. The
internal standard and analyte also must
have identical or very close retention
times to ensure that they are exposed to
the same coeluted compounds. By modifying the LC conditions, the internal standard and analyte often can be eluted at the
same time. If a stable isotope-labeled analog is used as the internal standard, which
has identical chemical and structural
properties to those of the analyte, the analyte and internal standard will behave
identically not only during chromatography but also during sample preparation.
Isotopic analogs are therefore the best
choice of internal standard to reduce signal variability and improve precision. In
fact, Freitas and coworkers (6) have illustrated nicely that the use of isotopic internal standard is a prerequisite for achieving reliable quantification and high
precision. However, ion suppression also
has been found to occur between analyte
and their isotope-labeled internal standard (23). To avoid this, the internal standard concentration should not be too
high. Although an appropriate internal
standard concentration for the particular
investigated analytes and experimental
parameters can be determined by experiment, isotope-labeled internal standards
are not always available and often very
expensive, especially in a multicomponent
analysis, in which a separate internal standard for each analyte is required.
A new and interesting alternative to the
internal standard concept is the echo peak
technique. The technique simulates the
use of internal standard, without the
demand for an isotope-labeled analog of
the target analyte. It consists of two injections, the unknown sample, and a standard solution, within a short time period.
Thus, an echo peak forms, where the analyte peak is eluted in close proximity to
the peak from the sample (Figure 6).
Because both peaks are eluted so closely, it
is expected that they are affected some-
www.chromatographyonline.com
what equally by coeluted matrix components. Therefore, ion suppression experienced by the analyte will be compensated
for.
Summary