Reference Manual

Section 2: Configuration

00809-0100-4100, Rev AA

2.10.1

March 2013

Master reset
The master reset function will reset the device electronics. To perform a master reset:

Performing master reset using a Field Communicator


From the HOME screen, enter the fast key sequence

Device Dashboard Fast Keys

3, 5, 1, 2, 1

Performing master reset using AMS Device Manager

2.10.2

1.

From the Home screen, select 3: Service Tools.

2.

Select 5: Maintenance

3.

Select 1: Calibration

4.

Select 2: Factory Calibration

5.

Select 1: Restore to restore to factory presets.

Join status
Viewing join status using a Field Communicator
From the HOME screen, enter the fast key sequence

Device Dashboard Fast Keys

3, 4, 1

Viewing join status using AMS Device Manager


To view the join status of the device, perform the following procedure:
1.

From the Home screen, select 3: Service Tools.

2.

Select 4: Communications.

3.

Select 1: Join Status.

Wireless devices join the secure network through a four step process:

Configuration

Step 1. Network Found

Step 2. Network Security Clearance Granted

Step 3. Network Bandwidth Allocated

Step 4. Network Join Complete

25