You are on page 1of 5

CPE554

Universiti Teknologi MARA

A. General Start-Up Procedures


1. Ensure all valves are closed.
2. Turn on the power for the control panel.
3. Prepare a 3040 L mixture containing ethanol and filtered water at the desired composition.
Charge the unit as follows:
i.
Open the charge port cap at the reboiler B1, and fill the reboiler up to the level of the
internal overflow height.
ii.
Pour the remaining ethanol-water mixture into the feed vessel B2 through the feed charge
port.
4. Set the reflux divider KFS-101 to total reflux. This is to prevent any distillate from escaping when
starting up the distillation column.
5. Let the cooling water flow into the condenser W2 and product coolers W3 & W4 by opening
valves V13, V14 and V15.
6. Ensure that the tip of the level switch is located below the reboiler equator, which is at the
liquid inlet from the cartridge heaters.
7. Switch on the cartridge heaters at W1. Allow the liquid in the reboiler to boil.
8. Observe the vapor rising from the boiling liquid into the distillation column. As the vapor
reaches the condenser, it will condense and flow back into the column as liquid. Mass transfer
will take place between this liquid and vapor phase in the column.
9. Check that the condenser has sufficient cooling water flow to minimize escape of vapor into the
vent.
10. Let the distillation column run until a stable condition is observed. The stable condition is
reached when all temperature indicators on the column give a constant reading with an
allowable fluctuation of 0.2C.

B. General Shut-Down Procedures


1. Set the reflux divider KFS-101 to total reflux.
2. Keep the cooling water running through the condenser W2 and product coolers W3 & W4.
3. Switch off the cartridge heaters W1. Monitor the temperature drop in the reboiler. Wait until it
has cooled down to less than 50C.
4. Close the cooling water valves V13, V14 and V15.
5. Drain all liquid from the reboiler B1 and product vessels B3 and B4 by opening valves V6, V9 and
V12. Also drain the liquid in the pipelines by opening valves V5, V8, V10 and V11. The liquid can
be re-mixed and recycled for future use.
6. Turn off the power for the control panel.

Exp/Distillation Column

CPE554

Universiti Teknologi MARA

C. Sampling Procedures
Samples can be taken from the distillate and the bottom product for analysis:
Distillate sampling:
i.
Drain any remaining liquid in the sampling valve V11.
ii.
Place a vial below valve V11.
iii.
Set the reflux divider to total distillate offtake for a few seconds.
iv.
Open valve V11, collect enough sample and close the valve.
v.
Switch the reflux divider back to the original setting.
Bottom sampling:
i.
Drain any stagnated liquid above valve V8.
ii.
Place a vial below valve V8.
iii.
Open valve V8, collect enough sample and close the valve. Be careful of the hot liquid
while collecting a sample.
iv.
Cool the sample by immersing the vial in cold water.
Note: Always re-cap the vials quickly after collecting samples to avoid excessive evaporation of ethanol
before analysis.
D. Preparation of Calibration Curve
Refractive index (RI) readings are used to determine the composition of ethanol-water mixtures in the
samples.
1) Prepare a set of mixtures containing ethanol and water within a specified range of composition
between pure water and pure ethanol. Refer to the sample table in Appendix A.
2) For each mixture, obtain their refractive index readings using a refractometer.
3) Plot the calibration curve of RI vs. composition for the whole range of mixtures.
4) Examine the calibration curve. Note that a strong relationship exists between the RI and
composition of ethanol-water mixtures at low concentrations. However, observe that the curve
will flatten at the mid range concentrations and starts to slope downwards at high
concentrations. At this range, it is difficult to obtain an accurate value for ethanol composition.

Exp/Distillation Column

CPE554

Universiti Teknologi MARA

E. Composition Analysis of Ethanol in Concentrated Samples


Measure the refractive index of a sample and determine its composition from the calibration curve. If
the composition value is difficult to be determined accurately due to the calibration curves flat slope,
dilute the sample so that a more accurate reading can be obtained. Perform the dilution as follows:
1) Place a 10 mL measuring cylinder on a weighing scale (accurate to 0.01 g) and zero the scale.
2) Using a dropper, transfer 1 to 2 g of sample into the measuring cylinder. Record the sample
weight.
3) Add water into the measuring cylinder to dilute the sample. Obtain a total solution weight in the
range of 5 to 10 g. Record the solution weight.
4) Ensure that the diluted solution is well mixed before measuring its refractive index (RI).
5) Obtain the diluted solution's composition from the calibration curve.
6) Calculate the actual sample's composition as folows:

7) Use the sample table in Appendix C for data recording and calculation.

EXPERIMENT : Batch Distillation At Total Reflux


OBJECTIVE:
To determine the number of theoretical plates at total reflux.
PROCEDURES:
1. Perform the general start-up procedures (Section A).
2. Once the distillation column has reached stable condition, record the trays, top and bottom
temperature.
3. Collect samples from the distillate and bottom product (Section C).
4. Measure the refractive index for all samples and determine their compositions (Section E).

ANALYSIS & DISCUSSION:


1. Use the sample table in Appendix B for data collection.
2. Using the X-Y equilibrium diagram for ethanol-water system, mark the trays, top and bottom
compositions on the diagram and determine the number of theoretical plates in the distillation
column using the McCabe-Thiele method.

Exp/Distillation Column

CPE554

Exp/Distillation Column

Universiti Teknologi MARA

CPE554

Exp/Distillation Column

Universiti Teknologi MARA

You might also like