You are on page 1of 5

f(x) = 50 x − 100

Graph Dose versus 8 bits Pixel


50
dose ( cGy)

Dose (cGy)
Linear (Dose (cGy))

0
50 60 70 80 90 100

8 bits pixel

Graph: Calibration of Gafchromic EBT2 Film by LINAC


Graph Dose versus 8 bits pixel
550

500

450

f(x) = − 6.69 x + 776.68


400

350

300
Dose ( cGy)

Linear ()
250

200

150

100

50

0
50 60 70 80 90 100 110 120 130 140

8 bits pixel

Graph : Calibration of Gafchromic EBT2 Film by Ir-192 source


f(x) = 0
50 x − 100
Calibration Gafchromic EBT 2 film
50

Ir-192
dose ( cGy)

Linear (Ir-
192)
LINAC
Linear
(LINAC)

0
50 60 70 80 90 100

8 bits pixel
In air measurement

Cylinde Dose from


r Size Position Reading TPS Ir-Calibration LINAC calibration
Dose % Dose %
1 2 3 4 Average ( cGy) difference (cGy) difference
Surface (1.25cm) 52 52 52 53 52.25 405.54 426.86 4.8 389.15 4.03
2.5cm Bolus ( 1.75cm) 70 70 70 71 70.75 305.04 303.07 0.64 277.4 9
Surface ( 1.50cm) 53 53 53 53 53 501.69 421.87 5.24 384.62 30.43
3.0cm Bolus ( 2.0cm) 69 69 69 69 68.75 325.44 318.78 2.9 289.43 11.04
Surface ( 1.75cm) 54 53 53 53 53.25 466.41 420.19 6.7 383.11 17.8
3.5 cm Bolus( 2.25cm) 74 75 72 72 73.25 303.51 286.33 6 262.3 15.6

In water measurement

Cylinde Dose from


r Size Position Reading TPS Ir-Calibration LINAC calibration
Dose % Dose %
1 2 3 4 Average ( cGy) difference (cGy) difference
2.5cm Surface ( 1.25cm) 50 50 50 49 49.75 405.54 427.96 5.24 404.26 0.31
Bolus ( 1.75cm) 68 68 68 68 68 305.04 310.84 1.865 294.01 4
3.0cm Surface ( 1.50cm) 52 52 52 52 52 501.69 443.1 13.24 390.67 28.41
Bolus ( 2.0cm) 70 69 69 69 69.5 325.44 320.88 1.42 294.95 12.44
3.5 cm surface ( 1.75cm) 50 50 50 50 50 466.41 441.35 5.61 402.75 13.6
Bolus( 2.25cm) 65 65 65 64 65 303.51 340.96 10.98 312.14 8.4