You are on page 1of 4

Digital Multimeter Measurement Errors Series

AC Voltage Measurement
Errors in Digital Multimeters
Application Note AN 1389-3

Introduction
When you make measurements with a digital
multimeter (DMM), common errors will crop up.
The following discussion will help you eliminate
potential measurement errors and achieve the
greatest accuracy with a DMM. This paper covers
ac voltage measurement errors. For an overview of
system cabling errors and dc voltage measurement
errors, see Application Note 1389-1. For a discussion
of resistance; dc current; ac current; and frequency
and period measurement errors, see Application
Note 1389-2. (NOTE: The Agilent 34401A, a 6-1/2-
digit, high-performance DMM with both benchtop
and system features, will be used as an example
throughout this article).
AC Voltage Measurement Errors

Many of the errors associated with shielded, responds differently The multimeter’s ac voltage and
dc voltage measurements also in the backward input case ac current functions measure the
apply to ac voltage measurements. due to slight differences in ac-coupled true RMS value. This
This paper covers errors that are stray capacitance to earth. The is in contrast to the ac+dc true
unique to ac voltage measurements. multimeter’s errors are greatest RMS value shown above. Only
For information about dc voltage for high-voltage, high-frequency the “heating value” of the ac
measurement errors, see inputs. Typically, the multimeter components of the input waveform
Application Note 1389-1. will exhibit about 0.06% additional are measured (dc is rejected).
error for a 100 V, 100 kHz reverse For sinewaves, triangle waves
Common Mode Errors — Errors are
input. You can use the grounding and square waves, the ac and
generated when the multimeter’s
techniques described for dc common ac+dc values are equal since
input LO terminal is driven with
mode problems to minimize ac these waveforms do not contain
an ac voltage relative to earth.
common mode voltages (see a dc offset. Non-symmetrical
The most common situation
Application Note 1389-1). waveforms such as pulse trains
where unnecessary common
contain dc voltages, which are
mode voltages are created is True RMS AC Measurements.
rejected by ac-coupled true RMS
when the output of an ac True RMS- (root mean square)
measurements.
calibrator is connected to the responding multimeters like the
multimeter "backwards." Ideally, Agilent 34401A measure the Crest Factor Errors — A common
a multimeter reads the same “heating” potential of an applied misconception is that “since an
regardless of how the source is voltage. Unlike an “average- ac multimeter is true RMS, its
connected. However, both source responding” measurement, a sinewave accuracy specifications
and multimeter effects can true RMS measurement is used apply to all waveforms.” Actually,
degrade this ideal situation. to determine the power dissipated the shape of the input signal can
in a resistor. The power is dramatically affect measurement
Because of the capacitance
proportional to the square of accuracy. A common way to
between the input LO terminal
the measured true RMS voltage, describe signal waveshapes is
and earth (approximately 200 pF
independent of waveshape. An crest factor. Crest factor is the
for the Agilent 34401A), the
average-responding ac multimeter ratio of the peak value to the RMS
source will experience different
is calibrated to read the same as value of a waveform.
loading, depending on how the
a true RMS meter for sinewave
input is applied. The magnitude of For example, a pulse train’s crest
inputs only. For other waveform
the error is dependent on the factor is approximately equal to
shapes, an average responding
source’s response to this loading. the square root of the inverse of
meter will exhibit substantial
The multimeter’s measurement the duty cycle as shown in Figure
errors (see Figure 1).
circuitry, while extensively 9. In general, the greater the crest
factor, the greater the energy
contained in higher frequency
harmonics. All multimeters
exhibit measurement errors that
are crest-factor-dependent.

Figure 1.

2
The following equation shows how For Low Frequencies: capacitor in parallel with the
to estimate the measurement multimeter’s input terminals. There
–100 x R s
error due to signal crest factor: Error (%) = may be some experimentation
R s + 1MΩ
Total Error = Errorsine + Errorcrest factor + involved to determine the correct
Errorbandwidth Additional error for high capacitor value for the particular
frequencies: application.
Where:
Where: Most extraneous noise is not
Errorsine = DMM’s Sinewave Accuracy
Errorcrest = DMM’s Crest Factor
correlated with the input signal.
1 The equation below shows how to
Errorbandwidth = Estimated Bandwidth Error Error (%) = 100 x ––––––––––––––––– – 1
(see below): √1 + (2π x F x R s x Cin)2 determine the error:
–––––––––––
Errorbandwidth =
–C.F.2 x F Voltage Measured = √ Vin + Noise
2 2

4π x BW R s = Source Resistance
Where: F = Input Frequency Correlated noise, while rare, is
especially detrimental because it
C.F. = Signal Crest Factor C in = Input Capacitance will always add directly to the
F = Fundamental Input Signal (100 pF) plus Cable Capacitance
Frequency input signal. Measuring a low-level
BW = DMM’s –3 dB Bandwidth (1 MHz Note: Be sure to use low-capaci- signal with the same frequency as
for the Agilent 34401A) tance cable when measuring the local power line is a common
Example: Calculating Crest Factor Error high-frequency signals. situation where this error is likely
to occur.
Calculate the approximate Low-Level AC Measurement Temperature Coefficient and
measurement error for a pulse Errors — When measuring ac Overload Errors — The Agilent
train input with a crest factor of 3 voltages less than 100 mV, be 34401A uses an ac measurement
and a fundamental frequency of aware that these measurements technique that measures and
20 kHz. For this example, assume are especially susceptible to removes internal offset voltages
the multimeter’s 90-day accuracy errors introduced by extraneous when you select a different
specifications: ±(0.05% + 0.03%) noise sources. An exposed test function or range. If you leave
and the error for a 2-3 crest factor lead will act as an antenna and a the multimeter in the same range
is specified as 0.15% of reading. properly functioning multimeter for an extended period of time,
Total Error = (0.05+0.03)% + 0.15% + will measure the signals received. and the ambient temperature
((3^2*20kHz)/(4π*1000kHz))% The entire measurement path, changes significantly (or if the
including the power line, acts as a multimeter is not fully warmed
Total Error = 0.08% + 0.15% + 1.4% = 1.6%i
loop antenna. Circulating currents up), the internal offsets may
AC Loading Errors — In the ac in the loop will create error change. For the Agilent 34401A,
voltage function, the input of the voltages across any impedances in this temperature coefficient is
Agilent 34401A appears as a 1MW series with the multimeter’s input. typically 0.002% of range per °C.
resistance in parallel with 100 pF For this reason, apply low-level ac The coefficent is automatically
of capacitance. The cabling used voltages to the multimeter through removed when you change
to connect signals to the multimeter shielded cables, and connect the functions or ranges.
will also add additional capacitance shield to the input LO terminal.
When manual ranging to a new
and loading. Figure 2 shows the Make sure the multimeter and the range in an overload condition,
multimeter’s approximate input ac source are connected to the the internal offset measurement
resistance at various frequencies. same electrical outlet whenever may be degraded for the selected
Input Frequency Input Resistance
possible, and also minimize the area range. Typically, an additional
of any ground loops that cannot 0.01% of range error may be
100 Hz 1 MW be avoided. A high-impedance introduced. This additional error
1 kHz 850 kW source is more susceptible to is automatically removed when
noise pickup than a low-impedance you remove the overload
10 kHz 160 kW
source. To reduce the high-frequency condition and then change
100 kHz 16 kW impedance of a source, place a functions or ranges.
Figure 2.
3
Conclusion

When making high frequency or Agilent Technologies' Test and Measurement By internet, phone, or fax, get assistance
low voltage AC measurements it Support, Services, and Assistance with all your test & measurement needs
Agilent Technologies aims to maximize the value Online assistance:
is important to minimize error you receive, while minimizing your risk and www.agilent.com/find/assist
mechanisms. When practical, use problems. We strive to ensure that you get the
a low-impedance source, use test and measurement capabilities you paid for Phone or Fax
proper cabling and minimize loops and obtain the support you need. Our extensive United States:
support resources and services can help you (tel) 1 800 452 4844
between cables. To determine AC choose the right Agilent products for your Canada:
measurement errors, it is important applications and apply them successfully. Every (tel) 1 877 894 4414
to include errors due to signal instrument and system we sell has a global (fax) (905) 282 6495
shape, noise and frequency. warranty. Support is available for at least five China:
years beyond the production life of the product. (tel) 800 810 0189
Two concepts underlie Agilent's overall support (fax) 800 820 2816
policy: "Our Promise" and "Your Advantage." Europe:
For more information about the Our Promise (tel) (31 20) 547 2323
Agilent 34401A DMM, go to Our Promise means your Agilent test and (fax) (31 20) 547 2390
www.agilent.com/find/34401a measurement equipment will meet its advertised Japan:
performance and functionality. When you are (tel) (81) 426 56 7832
choosing new equipment, we will help you with (fax) (81) 426 56 7840
product information, including realistic Korea:
performance specifications and practical (tel) (82 2) 2004 5004
recommendations from experienced test (fax) (82 2) 2004 5115
engineers. When you use Agilent equipment, we
Latin America:
can verify that it works properly, help with
(tel) (305) 269 7500
product operation, and provide basic (fax) (305) 269 7599
measurement assistance for the use of specified
Taiwan:
capabilities, at no extra cost upon request. Many
(tel) 080 004 7866
self-help tools are available.
(fax) 080 028 6331
Your Advantage
Your Advantage means that Agilent offers Other Asia Pacific Countries:
(tel) 1 800 375 8100
a wide range of additional expert test and
(fax) (65) 836 0252
measurement services, which you can purchase
(e-mail) tm_asia@agilent.com
according to your unique technical and business
needs. Solve problems efficiently and gain a Product specifications and descriptions in this
competitive edge by contracting with us for document subject to change without notice.
calibration, extra-cost upgrades, out-of-warranty © Agilent Technologies, Inc. 2002
repairs, and on-site education and training, as Printed in the USA January 31, 2002
well as design, system integration, project 5988-5513EN
management, and other professional engineering
services. Experienced Agilent engineers and
technicians worldwide can help you maximize
your productivity, optimize the return on
investment of your Agilent instruments and
systems, and obtain dependable measurement
accuracy for the life of those products.

You might also like