Troubleshooting
Signal quality Transducer wrongly positioned. Reposition transducer until signal quality
indicator is red indicator is green.
continuously. FHR less than 50 bpm. None.
FHR exceeds 300 bpm. FHR is half-counted (for example 320 bpm is
recorded as 160 bpm).
FHR not recorded. FHR is less than 50 bmp or over 300. None.
Light or no trace. Wrong paper or dirty printhead. Use recommended paper or clean printhead.
End of paper Bad paper feed or wrong paper. Check paper feed and use recommended paper.
noted when pack
not finished.
If you suspect the transducer. Carry out the Parameter Test as described on page
152.
If you suspect the recorder or display. Carry out the Quick Test as described on page
151.