Modules with Patient Isolation

Our modules that record the EKG trace transmit the trace data as well as the pulse value to the host system by a serial interface. The EKG cable is connected to the isolated side of the module. All data processing, including 50Hz and 60Hz filtering is done on the module. A galvanic patient isolation according to class CF is integrated on the modules.

Selection Guide
EG01000 One Channel, depending on placement of electrodes, Einthoven I, II or III Software R-wave trigger Three lead cable, Right arm, Left arm, Left leg No respiration measurement EG01010 Three Channels, one out of three, Einthoven I or II or III, selectable by user command EG05000 Five channels synchronuously. Einthoven I and II and III and aVL and aVR and aVF and C EG12000 Twelve channels synchronuously. Einthoven I and II and III and aVL and aVR and aVF and C1-C6.

Software and high accuracy hardware Software and high accuracy hardware Software and high accuracy hardware R-wave trigger R-wave trigger R-wave trigger Three lead cable, Right arm, Left arm, Five lead cable, Right arm, Left arm, Left leg Left leg, Right leg, Chest Respiration with optional piggy backPCB Respiration with optional piggy backPCB Ten lead cable, Right arm, Left arm, Left leg, Right leg, 6 Chest leads Respiration with optional piggy backPCB

EG01000 One Channel Module EG01010 Module with three channels (I,II,III) and optional respiration measurement over the ECG electrodes EG05000 Module with channels I, II, III, aVR, aVL, aVF, V1 and optional respiration measurement over the ECG electrodes EG12000 Module with channels I, II, III, aVR, aVL, aVF, C1-C6 and optional respiration measurement over the ECG electrodes