You are on page 1of 4

1)Beds Monitoring Facilities (Cardiac Monitors)

N.B: The following specifications or equivalent should be met

FEATURES

 Non fade TFT LCD color display


 Electro-surgical interference suppression/protection
 Defibrillator protector
 Freeze and cascade facility
 Waveform traces speed; 25 & 50mm/sec.
 Screen size: min. 12” color display
 Monitor must have capability to monitor vital signs of
 Adults , pediatrics & infants/neonates patients

TECHNICAL SPECIFICATION:

ECG:

 Numeric: heart rate


 Waveform six wave forms minimum, real time and
 Freeze ECG trace

NON-INVASIVE BLOOD PRESSURE(NIBP):

 Method : oscillometric principle


 Numeric : systolic , diastolic and mean pressure
 Selectable auto inflate interval settings
 Rising cuff/ continuous pressure display

TEMRATURE:
 Numeric: temptrature selectable in OC/OF
PULSE OXIMETRY:

 Numeric : 0-100% oxygen saturation measuring range.


 Waveform-plethysmograph pulse with pulse strength
 Indication
 Reusable sensor electrode
 Reusable cuff all sizes
 2xprobes newborn
 2xprobes peads

ARRHYTHMIA ANALYSIS:

 ARRHYTHMIA analysis and st analysis


 Rapid learning : rapid learning should be only 20 sec for recognition of dominant
rhythm
 Storage: real time arrhythmia detection should have storage of atleast the latest 80ARR
events including waveforms and parameter

IBP:

 Channels: should have 2 channels


 Measurements ranges : SYS 50-300mmHg
 DIA 50-300mmHg
 MAP 50-300mmHg
 Pressure sensor sensitivity : 5µ V/V/mmHg
 Pressure impedance: 300-2500 ohms
 Filters should be 8, 16 ,22 Hz and adjustable

CO2 Main Stream(ETCO2):

 Methods: should be infrared absorption


 Measuring parameters: EtCO2 and AWRR
 Measuring range: CO2 (0-100%)
 AWRR: 30-100bpm

RESPIRATION:

 Breath rate display and settable apnea alarms


 Sweep speed ; 6.25 , 12.5 mm/sec

OTHER FEATURES:

 Trend data; graphical and tabular : 160 Hours

ALARMS:

 High & low ( settable) on all parameters


 Visual and audible indications of alarms

OPERATING REUQIREMENTS:

 AC 220V/50 HZ
 Built-in rechargeable battery for at least 2 hours operation in case of power failure at
full parameters

STANDARD ACCESSORIES:

 1 x ECG cable with 10-leadwires


 1 x NBP Hose
 3 x NBP Cuff for adult, XL , Peads & neonates
 1 x SPO2 Finger probe for adult, peads & neonates
 1 x skin temperature probes
 1 x operating manual

Mn ,.QAII quoted cardiac monitors should comply:

1. Electrical protection class 1


2. Applied parts protection class B /BD/CF (Defib protected)
3. Rail /Rod mounted platform with monitor clamps
4. Fire protection class A/B
5. Ingree protection (IP) CLASS 66 and above
6. Option for multiple window display with algorithms for arrhythmia management.
7. Alarms and alerts
8. All monitors should be DICOM enabled with pull and push ro srever
9. Outer cases and screens of all monitors should be resistant to fading on application of
detergent and disinfectants
10. Future upgradation with hemodynamic and respiratory parameters can be achieved.
FDA / CE.

4) Defibrillator with crash cart

N.B : The following specifications or equivalent should be met.

 Biphasic transthpracic (external) defibrillator with high resolution LCD display


 Synchronized output with ECG
 Energy selection on control panel and paddies for external defibrillation
 Energy delivery on paddies for external defibrillator
 Energy selection and delivery in control panel for internal defibrillation. Charging
 Indicator
 The energy range should be adjustable for peads and adults up to 200 (360) joules
 Charging time for full energy will be less than 07 sec.
 Screen size of approx. 5 inch with high resolution display
 Display of HR. ECG through paddies and lead I. II & III patient cable
 Built in recorder for printing of full summery on standard 80mm paper
 Alarms for high and low heart rate , low battery warning
 Built-in rechargeable battery with charger for 3 hours of continuous ECG monitoring.
 Auto tester /self check
 Real time CPR feedback with display
 External paddies (adult & paeds)
 AED facility with cable
 Pacing facility with reusable pads
 AC 220V / 50Hz operated
 BP apparatus table top

Accessories:

Complete win standard accessories, including reusable defibrillation

Pads for adult & pedriatic with original Trolley

FDA/CE marked