Brand

Type
Origin

Dimension
PxLxT
Total Weight

Monitor Display
Type
Size
Resolution

Triggering method
External Trigger
External Trigger Ratio
ECG Trigger range

Operating Environment
Temperature
Relative humidity
Atmospheric pressure
Atmospheric free of corrosive gas

Power Requirements
AC Single Phase
Maximum Power

Noise
Pressure Range

ECG Amplifier
Input impedance
Common mode rejection ratio

Treatment Session Time


Adjustable treatment time
System automatically stop when set time expires

Cuffs
EECP
(Enhanced External Counterpulsation)

PSK
TI MODEL
CHINA

2070 mm x 1120 mm x 1100 mm


204 kg

Color LCD Touchscreen


12.1” (31 cm) diagonal
1024 Horizontal x 768 Vertical

R-wave of patient ECG

< 35 bpm, >125 bpm (adult)


< 35 bpm, > 165 bpm (children)

10 °C - 30 °C
≤ 70%
86 kPa to 106 kPa

220 V, 16 A, 50Hz/60Hz
2.6 KVA

≤60db
0 - 440 mmHg or 0 - kPa

> 2 MΩ
> 80db

1 min to 60 min (max)


Brand
Type
Origin

Dimension
RENEW NCP-4 Unit ( P x L x T )
Treatment Table ( P x L x T )
Total Weight NCP-4 Unit
Total Weight Treatment Table

Triggering method
External Trigger
External Trigger Ratio
ECG Trigger range

Operating Environment
Temperature
Relative humidity
Atmospheric pressure
Atmospheric free of corrosive gas

Power Requirements
AC Single Phase
Maximum Power

Noise

Pressure Range

ECG Amplifier
Input impedance
Common mode rejection ratio
Treatment Session Time
Adjustable treatment time
System automatically stop when set time expires

Cuffs
6 sizes fit all patients types
Calf, lower thigh, upper thigh, buttocks are intregrated

Backrest
Manual
Safe working load
Position adjustable

Software Safety Features


System will be alerted and suspension of inflation / deflation in any of these
events :

Electrical and Mechanical Safety Features


Pressure relief valve is set to default factory setup to prevent exposure to
excessive cuff pressures.
Accessible Patient / Operator top-mounted Emergency Shut Off swiches
Pneumatic cuffs automatically deflate with power loss or system shutdown
Electtrical isolation circuitry for patient protection
ON/OFF switch on the machine is fused
EECP
(Enhanced External Counterpulsation)

RENEW
TM NCP-4 MODEL
USA

740 mm x 510 mm x 1500 mm


2170 mm x 840 mm x 950 mm
90 kg
96 kg

R-wave of patient ECG

< 35 bpm, >125 bpm (adult)


< 35 bpm, > 165 bpm (children)

16 °C - 38 °C
10% to 95% RH (non-condensing)
80 kPa to 100kPa

220 V, 6A, 50Hz/60Hz


1.3 KVA

≤60db

Up to 6.0 psi

> 2 MΩ
> 80db
10 min to 60 min (max)

XS, S, M, L, XL, XXL (Short and Long)

0 - 35 degrees adjustable backrest


150 Kg

a) if heart rate exceeds the normal range of operation ( B57< 35


BPM or > 125 BPM )
b) when an external (patient) ECG signal is not available or
reliable
c) if the computer is unable to display patient or treatment data
Brand
Type
Origin

Dimension
Pole Cart ( L x T )
Treatment Table ( P x L x T )
Total Weight Pole Cart
Total Weight Treatment Table

Operating Environment
Temperature
Air conditioning capacity: at least
Relative Humidity
Absence of electromagnetic interference
Free of flammable anesthetics

Power Requirements
Standard or hospital grade outlet
Dedicated, twist and lock outlet
Compatibility upon request

International specifications
Standard (min) outlet
Single Phase Dedicated

Treatment pressure range

Treatment duration timer

Treatment cuffs
Durable & cleanable high denier Nylon with Velcro closure
Buttock and thigh cuffs are separate for better fit

Table
Motorized, continuously adjustable backrest
Safe Working Load
Easy to clean, fluid resistant fabric
Quick connect hoses attach without tools

System protection
Bed Base ON/OFF switch is fused
Air/Vacuum Pump relay is overload protected
Air/Vacuum Pump is thermally protected

Printer
Wireless, Laser*

Patient protection
Auditory alarms :

Automatic vacuum deflation of cuffs on early/extra systole


Over pressure limit protection
Treatment is suspended during episodes of high or low heart
rate
Emergency stop button
Air to cuffs is actively cooled

Patient care package


Initial supply of electrodes, treatment pants,
Cuffs and bladders
EECP
(Enhanced External Counterpulsation)

SCOTTCARE
TM MODEL
USA

750 mm x 1680 mm
2080 mm x 750 mm x 1680 mm
16 kg
159 kg

65-75ºF/18-24º C
8800 BTU per hour
35-65%
in the area (e.g.: MRI device)

1 – 110 VAC 60Hz 15 A (1.65 KVA)


1 – 220 VAC 60Hz 20 A (4400 KVA)
220VAC 50Hz

220 VAC 10 A
220 VAC 15 A outlet (NEMA L6-20R)

0 – 6 PSI/310 mmHg

10-60 minutes; 5 minute increments


Treatment and air/vacuum pump stop when set time expires
3 sizes fit all patient types

– 0 to 45 degrees continuous
350 lbs. /160 kg

Heart rate out of limits


Oxygen saturation out of limits
Lack of an adequate triggering signal
Loss of communication between computer and ECP device