Professional Documents
Culture Documents
Operation Manual
Version:V2.4
II
Symbol of production
Manufacture’s Trademark
Date of manufacture
Manufacture
FRAGILE
with care.
THIS WAY UP
1.1 Introduction
1
1.2 Characteristics
1.3 Functions
2
17) Automatically air released when treatment finished.
18) Low-pass digital filter eliminates 50 Hz and high-frequency interference.
19) Ways of inflated pressure:arms, legs and buttocks—3 sequence;legs and
buttocks—3 sequence;legs—2 sequence(heart ratio: 120/min).
20) Range of inflated pressure: 0~59kPa.
21) Protective function of limited pressure.
22) An emergency stop.
23) Working hours: 8 hours. Oxygen saturation LED light with wavelength of
660nm red light and 940nm infrared red light, the rated current of 30mA.
24) Oxygen saturation LED light with wavelength of 660nm red light and 940nm
infrared red light, the rated current of 30mA.
25) Keep patient quiet and remain him stationary, especially measured finger, while
operator measures the oxygen saturation.
26) Measurement range of SpO2: 70%-l00%.
27) SpO2 update time: a heart rate cycle--5 seconds
28) The time when patient wears the probe on the same finger do not exceed 2 hours.
29) The probe of the equipment has biocompatibility test report.
30) The probe uses Remo quick connector.
31) The probe is used for adult, worn on the finger—the best site monitored is index
finger, considering other fingers when the index finger is not available. Operator
cleans the measured finger and ensures it dry before wear the sensor to patient.
32) Elements affecting the measurement accuracy of SpO2: Cleanliness of
measurement site and stationary.
33) SpO2formula: :
Oxygen andhemeglobin
Spo2% 100%
Oxygen andhemeglobin Deoxy-hemoglobin
34) Data updated time:5 seconds
35) If the data of SpO2 is less than 95%, the data shows in red, alarm indication.
36) The device is no pollution to the environment. Dispose scrap processing measures
of equipment and components in accordance with local laws and regulations.
37) When the acquisition signal is too weak, it will affect the SPO2 measurement
accuracy.
38) The equipment only uses the probe provided by our company. Operator ensures
that the probe is connected correctly before he uses.
39) Ground connection must be connected and reliable.
40) ECG electrode cable can not contact with other conductors, not with earth
contact.
41) Use medical electrodes.
42) The device can not be directly used for the heart.
43) The device can not be directly connected with other devices.
44) The device can be used for pacemaker patients.
45) Heart rate display range: Adult 30-125 ±2 bpm; Children 30-165 ±2 bpm.
46) The device has the tall T wave suppression capability.
3
47) Heart rate calculation accuracy and speed of response: heart rate calculation error
of ±2 per cardiac cycle response time.
48) No heart rate over speed alarm.
49) Display scan speed: 25mm / s and 50mm / s.
50) The device with analog ECG feature can automatically charge and exhaust
counterpulsation. Charging exhaust timing can be adjusted automatically and
display in real time at the interface.
51) Treatment time of 1-60 minutes is continuously adjustable.
52) Counterpulsation ratio of 1:1 (one cardiac cycle counterpulsation 1) and 1:2 (two
cardiac cycles counterpulsation 1) two tranches.
53) Treatment time or the treatment remaining time is displayed in real time on the
interface.
1.4.2 Convalescence
1) Exertional disease
2) Mental fatigue
4
1.5 Contraindication
5
It’s necessary to administrate glycerol trinitrate and monitor hemodynamics in
patients with acute myocardial infarction accompanied by left heart insufficiency.
ECP might provoke lung edema in patients with coronary heart disease accompanied
by left heart insufficiency, but some cases showed heart function improved more than
drugs could achieve.
4) Patient with Acute Myocardial Infarction and cardiac functional insufficiency
5) It depends on physical fitness when the patient is too old ,or over the age of 80.
6) Patients with blood pressure higher than 160/100 mmHg or a heart rate more than
120 bpm should be controlled prior to treatment with enhanced external counter
pulsation.
7) It depends on physical fitness when the patient is too old, or over the age of 80
1.8 Explanations
6
a. Temperature: 10 ℃~30℃ , when temperature is over 28 ℃, An air
conditioner should be installed to protect patients from skin hurt.
b. Relative Humidity: <70%.
c. Atmospheric Pressure: ordinary pressure
d. Dust-free, non-corrosive gases, and a well-ventilated, no strong
electromagnetic interference.
3) Power conditions: single-phase AC: 120V±12V 60Hz Power consumption:
2600VA
4) Safe and protective function:
a. ECG electrodes ≥ 4000V isolation, the pulse probe ≥ 2500V
isolation.
b. Function of pressure protection, working pressure≤59kPa。
c. Set up 2nd level inflated & deflated device when heart ration >120
d. Heart rate automatically stop counterpulsation in the following states,
tolerance of 2 times / min:
a) Adult: <35 times/min, >125times/min;
b) Child ( under 10 years old) <35 times/min, >165times/min
e. Automatically set up deflated protection when premature beat happened.
f. A function of emergency stop for counterpulsation.
1.9 Precautions
7
with the marking on the instrument.
10) The oxygen probe optical probe should be demanded to register with the FDA.
11) Product warranty period of one year, the service life of 10 years; the oxygen
probe, ECG lines consumables according to manufacturer requirements.
12) Products and components over the life of the period,our company does not
undertake any responsibility.
13) Product maintenance need specially trained personnel to do.
14) Please run the device at least 30 minutes a week in order to protect the normal use
of the product.
8
CHAPTER 2 INTRODUCTION OF INSTALLATION
SpO2
Explanation of Symbols :
1.bed 2.indicator light 3.the total power jack 4.casters 5.air switch 6.refers to the pulse
line 7.the heart wires 8.fixed base 9.fixed trestle 10.touch screen monitor 11.backrest
12.emergency stop switch 13.power switch 14.bed surface 15.the leg pipe joints
16.the thigh joint trachea 17.the hip joints trachea 18.lower arm pipe joints 19.upper
arm pipe joints
Caution:
Please ensure the working environment reaches the requirements of The P-ECP /TI External
Counterpulsation with SpO2 before installation.
Caution:
Read the contents of this part carefully before installation.
Step 1:
Open the packing box of treatment bed and then push to its right position. Please
9
(Explanation of Symbols :4 ).
Step 3:
Fixed base(Explanation of Symbols :8) with four screws is fixed on the bed (1). Pierce
the laptop (10)power cord and USB cable from the fixed trestle(9) and install it in a
fixed base (8).
Step 4:
Connect the power cord and USB to serial cable to the laptop, and the laptop is fixed
to the fixed trestle 8, according to the operator habits to choose USB mouse or the
computer touch keyboard.
Step 5:
Crush capsule set, thigh the capsule set, buttocks the capsule set and Forearm capsule
set are inserted respectively from 15 to 17 fast interface. Note that the capsule set
have left side and right side, to install the set should inward bandage. (Note: when
installing the set may need tools: beak clamp)
Step 6:
Take out the line of the finger pulse probe and plug it into the finger pulse
probe socket.
Step 7:
Take out the ECG cable and plug it into the Nihon Konden socket.
Step 8:
Insert the total power cord into the total power jack ( 3).
Step 9:
Make sure that the emergency stop switch 12 is in the unlocked position.
Step 10:
Open air switch, power indicator 2 lit.
Step 11:
Press the power switch (13), open the laptop power supply, and wait for the system
start-up.
Step 12:
Adjustable backrest comfort demand by the patient before the treatment can also be
adjusted by remote control or handle.
10
CHAPTER 3 OPERATING PROCEDURES
Caution:
The Demonstration mode is only applied for demonstration and debug. It can not
to be used for real treatment. Or there should be bad effects on patients.In that
case ,hospitals or users should take all the responsibilities for it.
12
the patient’s ECG signal. During the treatment (Fig. 3.3), the beginning of reading line
displays the inflation point. The end of reading line displays the deflation point. SpO2
signal (third row): display the patient’s SpO2 signal.
“Stop” means the system is in Stop-Status and the curve is green as well as both
of the compressor and solenoid valves are stopped.
Double clicking this button under different 3 situations: 1. if system is
originally in Stop-status, then nothing happen; 2. if system is in Standby-Status, it will
cause system return to Stop-Status; 3. if system is in Treatment-Status, it will stop the
treatment and return to Stop-Status.
“Standby” means double clicking this button will change system into
Standby-Status with the solenoid valves start working, but the compressor is NOT
working. If system is originally in Standby-Status, double clicking this nothing will
happen
If system is in Treatment-Status, the text on this button will change to “Pause”.
Double clicking this button now will stop the work of compressor, but system is still
in Treatment-Status, and then the text on the button will change to “Come back”.
Double click this button again, the compressor will restart working, and the text will
change to “Pause” again.
When system is in Standby-Status, double click this button and system will
change to Treatment-Status. Both of the compressor and solenoid valves are start
working, and this means system start ECP treatment. In this case, this button becomes
13
to show a clock of treatment time, just like: .
When double clicking Stop button or the treatment is completed, system returns to
Stop-Status with the compressor and solenoid valves stop working.Then the text of
this button returns back to “Start”.
1:1 or 1:2 stands for the immediate counterpulsation-rate during the ECP
treatment. (System default rate 1:1) “1:1” means one complete counterpulsation will
be triggered by one cardiac cycle (identified R-wave). “1:2” means one complete
counterpulsation will be triggered by two cardiac cycles (identified R-wave).
After system startup, the display area is unfrozen and system keep drawing ECG
curve, Inflation-Deflation curve and SPO2 curve with keep changing all parameters.
After double clicking this button, the system stop drawing curves and parameters
changing stop. Double click this button again, the screen will return to unfrozen
status.
This button will be enabled when system works with database and system is in
the treatment-status. It will start the save after double click [Save] button (usually,
each saving lasts for 10 seconds) with the button disabled. After the save, the button
will turn back to be enabled again and you can repeat the save.You can save data
many times in each treatment, but just need to save a typical data is recommended.
14
When system is stopped, double click [Exit] button to exit this main screen and
enter into start interface. If system is in Standby or Treatment status, a dialog like
Fig3.4 will be displayed when double click [Exit] button. It will remind the operator
to stop the system first and then double click to exit.
Caution:
R-I and R-D is automatically set and safe range is fixed. So it will
not exceed safe limits.
Use the time adjust key to set the beginning of the red line
Use the time adjust key to set the beginning of the red line
15
ECG parameter displayed and adjusted control area
clicking the button or while observing the results on the screen. The
R-wave of the QRS complex should be about 1/2 to 1-inch in height.
the button or while observing the results on the screen. The amplitude
should be about 1/2 to 1-inch in height.
16
Fig. 3.9 Current time and working hour of air compressor displayed area
In this area, the first line displays the current date, the second line is the time.
The last line in this area shows the accumulative working hour of air compressor.
Caution:
1. Before the operating, please ensure both of the P-ECP/TI hardware and soft
ware installed
2. The therapy strictly rules to the EECP’s indication and contraindication.
3. During the therapy, do not keep away from the patient. The emergency-switch
button is beside the bed.
4. Before the operating, please read this manual.
Step1:
Laid the patient in bed, then adjusting the bed for the best comfortable and
convenient position.
Caution:
Avoiding hurt skin, we demand the patient wear cotton underwear.
Step2:
ECG electrodes positions: The red electrode is placed at the standard V4 position,
the fifth intercostals space on the left mid-clavicle line. The White electrode is placed
between the left clavicle and the first rib at the right sternal border. The Black
electrode is placed at the standard V4R position, on the right side of the chest at the
same height as the red electrode.
Check reliable link of ECG Cable and interface every time.
17
Caution:
Place the electrodes on a flat site over bone to minimize muscle artifact on the
ECG. Placement of the electrodes may be varied if the ECG waveform is weak or
motion artifact is encountered. Prominent R-waves are necessary to trigger
Inflation/Deflation. Dry shaving is recommended, as it removes excess hair and
surface residue from the skin.
Step3:
Apply (clip) the SpO2 Sensor Probe to the patient’s forefinger or middle finger
tip, placing the diodes against the finger pad. Position the patient’s arm pointing
upward in a resting position while keeping the elbow and wrist bent and the finger
curled in order to minimize motion artifact.
Caution:
The patient should be positioned supine in the centre of the Treatment Table.
Step4:
Before treatment, select the Cuff Set size that will fit the patient properly.
Selection of the Cuff Set size should be based on the patient’s waist-to-ankle
measurement.
Step5:
Cuffs should be wrapped in the following order: calf cuffs, lower thigh cuffs and
upper thigh/buttocks cuffs (see Fig. 3.3) and should be wrapped tightly to obtain
optimal diastolic augmentation with minimal skin irritation. However, do not wrap the
cuffs so tightly as to harm the patient
Caution:
Ask a male patient to hold his scrotum up against the lower abdomen, clear of
the lower thigh cuff, which must fit snugly high into the joint.
Step6:
Put the patient’s forefinger into the spO2 Sensor Probe placing the diodes against
the finger pad. If the red light is flashing, please adjust the probe or change another
finger.
Step1:Set-up
18
1-1:Press the System Power Button in the Console, the computer will start
automatically and enter the Main Menu Screen.
1-2:According to the methods in Item 3.1.2,start the hardware of EECP.(Before
the therapy, please click the music tag to choose music as you want.)
1-3:After log in, choosing the type of the therapy.
Step2:ECP Therapy
2-1:Observe the ECG waveform on the top of screen. If a good quality ECG
signal is obtained you should see the patient’s heart rate shown on the screen,
generated by the System in synchronization with the patient’s ECG. If the ECG
R-wave height (amplitude) is very low or very high, the red and/or white electrodes
may be relocated to yield the best ECG signal possible to satisfy stable triggering. If
excessive artifact is degrading the ECG signal, the heart rate will not be indicated on
the screen.
2-2:Observe the Sop2 waveform on the top of screen. Apply (clip) the SpO2
Sensor Probe to the patient’s forefinger or middle finger tip, placing the diodes against
the finger pad. Position the patient’s arm pointing upward in a resting position while
keeping the elbow and wrist bent and the finger curled in order to minimize motion
artifact.
2-3:Click the key to start the valve. The key will become
highlighted. The sound of the valves should be rhythmic and have a characteristic
pattern: three short beats-one longer, three short beats-one longer, etc.
Click the numeric keys on the right side of dialog box to input the Patient-ID.
Click button to clear Patient-ID. Click button to delete the last digit of
19
Patient-ID. Click the button on the right side of Treatment Time to adjust the
When operator inputs a wrong Patient-ID, a warning dialog like Fig 3.12 will be
displayed.
If there is no use or setting of patient database, a dialog box like Fig 3.13 will
pop up. Input or choose the Therapy time and then click to treatment. Click
20
2-5:When treatment is finished or operator stop the machine, automatically
system is back to stop.
Step3:Power off
3-1:If there’s no further patient scheduled for the day, rotate the System Power
Switch to OFF position.
3-2:If the System Power Switch indicator light goes off, it indicates that the
System is shut-down.
3-3:Power supply is off.
Caution:
If you clicked the emergency-switch button when some incidents appeared,
please stop the therapy by clicking the , back to the beginning and
restart it.
Step1:
Take off patient’s sensor cable, clean and reserve it well.
Step2:
In sequence take off the red, black and white electrodes from patient .Clean the
coupling agent which is from electrodes on the body of patient with a small amount of
alcohol.
Step3:
In sequence take off the buttocks cuffs, thigh cuffs and shank cuffs from patient.
Step4:
Help the patient get off the bed.
Step5:
The equipment is not in direct contact with the body, so it need not disinfection.
Equipment maintenance personnel need only routine maintenance - sweep the dust.
21
Step two:
Swab the coupling agent on the body which electrodes come with a small amount
of.
Step three:
Use alcohol spray to sprinkle to the area where electrodes pasted.
Step four:
Swab with a clean dry tissue on the area where electrodes pasted.
Step five:
Only link new disposable electrodes to the ECG cables for the next use.
Step six:
Use alcohol spray to sprinkle SpO2 sensors and wipe away the sebum from the
finger of patient who took treatment before with a cotton swab.
22
Fig. 3.14 Starting Screen of Patient Database
In this system, whenever or whatever you want to input, a soft-keyboard will pop
up (Fig 3.15. For example, when you input Patient-ID (Fig 3.16), or input patient
information(Fig 3.17), or input patient name in a popup dialog(Fig 3.18), the
soft-keyboard will be helpful.
Patient ID
23
Fig 3.17 Insert Patient Screen
Browse Patient
24
Fig. 3.19 Browsing Screen of Patient Database
Search Patient
Click button to find out the patient’s information by his
name as shown in Fig. 3.20. In the text box, enter all or part of the patient's name (not
case-sensitive), then click button to start searching. If there is no result, no
information will be displayed on the main screen. If there is only one result,
information of this patient will be displayed. If there are more than one result, a dialog
similar to Fig. 3.21 will be popped up. Select one patient and click button
to display information of this selected patient.
25
Fig. 3.21 Search and Select Patient Dialog
3.22. Edit whatever you want to edit, then click button to save the
26
3.3.5 Insert Patient
Fig 3.23. Input information of the patient, then click button to save the
27
Fig.3.24 Print Report
28
Delete Treatment Record
If there has selected a patient, the treatment records should be listed in the lower
part of the main screen. Click the gray block on the left of treatment records to select
a record. The selected record will be highlighted in blue, and a right triangle will
appear. Click the button on the bottom of main screen, a confirm
dialog will appear. Click button to delete this record, and click
button to cancel this operation.
4.1 Warnings
Due to violation of the the following precautions, our company does not
undertake any responsibility.
1. Without permission Do not disassemble P-ECP/TI.
2. You cannot install the irrelevant software and hardware in the treatment
system.
3. Bed, mattress surface as well as the capsule sets of surface dirt can be wiped
off with furniture care dedicated solvent, but never on the monitor screen.
4. Minimize the number of the starting up and shutdown,ensure that its interval is
more than 5 minutes.
29
working under the contact of Serial Port Line/ indicator light and the
condition of standby Bad contact of 24 v power special valves
supply
Bad contact between
Rejoin
electrodes and ECG’ s wires
No ECG’ s Bad contact between patient’s Clean electrodes and put it in
display ,or ECG’ s skin and electrodes accurate place.
display trouble Electrodes are in wrong places Put it in accurate place
ECG’ s wires are not
Rejoin
connected with control panel
All /Some of the
Check the control panel and
Cuffs have No Bad contact of the gas circuit
the gas circuit
Inflation & Deflation and the control panel
in the EECP Therapy
Some of cuffs have
Bad contact between cuffs and
no pressure in the Rejoin
hoses
EECP Therapy
All of cuffs have no
Bad contact between pump
pressure in the EECP Rejoin
and hoses
Therapy
RD point exceeds P wave’s
Adjust RD point by using the
Troubles in deflations peak and close to R
laptop/Clean it.
wave/Block in the deflation
30
CHAPTER 5 APPENDIX
Dimension:
Length×Width×Height: 2050mm×760mm×600mm
Sort out the used Consumables such as: Electrode Slice, Broken Cuffs and Hoses.
Broken ECG and Finger wires should get back to our company.
31