You are on page 1of 7

UN806 Six-parameter module Specs

manual

Add: 14F,Wan'anguoji,Gexin Road #568,


Wujiashan, Dongxihu District,Wuhan,China
Postcode: 430040
Phone: (86) 027-83083133
Email: info@unmedtech.com
Website: www.unmedteh.com
Chapter 1. UN806 Specifications
 Six-parameter module: ECG, NIBP,SPO2, TEMP, PR
 Provide three/five ECG lead system
 Three patient modes, Adult/PED/NEO, the default mode is adult
 Five operating modes: Manual measurement mode, Automatic
measurement mode, Continuous measurement mode, Calibration mode
and Leakage check mode
 Small in size, flexible in installment, highly reliable and accurate in testing. Rapid,
accurate testing. Continuously work.

Chapter 2. Host System Requirements


2.1. Serial port communication

ECG & SPO2: ECG communicate through asynchronous serial communication


interface with the host, the following parameters of serial communication;
Transmission rate:
The baud rate of ECG & Spo2 part is: 57600 Baud

NIBP: NIBP communicates with the host via serial port, receiving outside
commands, executing operations, returning system status and data.
Baud rate: 4800bps

2.2. Interface definition


2.2.1.The plate interface

1) ECG input socket


P1
1 2 3 4 5 6
RA LA LL V1 SH RF

2) TEMP input socket


P2
1 2 3 4
T1+ T1- T2+ T2-
Note: T1+&T1-,T2+&T2-can not connect together.
3) SPO2 input socket
1 2 3 4 5 6
OUT SVDD RED+ RED- GND Probe oupt

4) ECG/SPO2 communication socket

5) NIBP communication socket

6) Pump and valve socket

2.3. Mounting dimensions


Appendix A Performance specification

1. Performance Specification

1.1 NIBP
NIBP Measurement Range
Adult PED NEO
SYS 40~270mmHg 40~200mmHg 40~135mmHg
MEAN 20~230mmHg 20~165mmHg 20~105mmHg
DIA 10~210mmHg 10~150mmHg 10~95mmHg

NIBP Measurement Precision


Average deviation: <=±5mmHg, standard deviation: <=8mmHg

Protection
software protect pressure Hardware protect pressure
Adult 300mmHg 320-330mmHg
PED 240mmHg 265~275mmHg
NEO 150mmHg 160-165mmHg
Pressure precision
±3mmHg, resolution 1mmHg

Standard
[EN 1060-1]
[EN 1060-3]
[IEC 60601-1]
[IEC 60601-2-30]
[ANSI/AAMI SP 10 ]
perform the test in the monitor.

2. SPO2

SPO2 Measurement Range


SPO2 Pulse Rate
0~100% 30-250

SPO2 Resolution
Pulse rate:30~250bpm
±2bpm normal
±3bpn motion or low perfusion

SPO2:70%~100%
±2bpm normal
±3bpn motion or low perfusion
Below 70% not specified

3. ECG
1.1 ECG specifications

 3-channel ECG waveform


 ST Segment Analysis
 Heat rate calculation
 Arrhythmia analysis
 Lead off status indication

1.2 Respiration
 1-channel respiration waveform(R-F/RA-LL)
 Respiration rate calculation
1.3 Temperature
 2-channel temperature datum

3. Function Description

2.1 ECG
Lead mode 5 leads (R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection I, II, III, aVR, aVL, aVF, Vx, CAL
Waveform 3 ch

Lead mode 3 leads ( R, L, F or RA, LA, LL)


Lead selection I, II, III, CAL
Waveform 1 ch
Gain 0.25V/mV, 0.5V/mV, 1V/mV, 2V/mV
Heart rate(HR)
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ±1% or ±1bpm
Resolution 1 bpm
Sensitivity > 200 uVpp
Differential Input Impedance > 5 MΩ
CMRR > 100 dB
Electrode offset potential 300mV
Insulation Intensity 1500VACrms
Baseline Recovery < 3 s ,after Defi.
ECG Signal Range >5 mVpp
Bandwidth
Diagnostic 0.05 ~ 100 Hz
Monitor 0.5 ~ 40 Hz
Surgery 1 ~ 20 Hz
Strong filter 5 ~ 20Hz
Calibration Signal 1mVpp,Accuracy 3%
ST segment monitoring range
Range:-2.0mV ~2.0mV

Arrhythmia analysis
ASY、VF/VTA、CPT、RUN、BGM、TGM、R on T(ROT)、VPB、TAC、BRD、MIS
、PNC、PNP
2.2 Respiration
Waveform 1 ch
Respiration Rate
Range 7-120bpm
Resolution 1bpm
Accuracy 2bpm
Baseline Impedance 500 - 4000Ω
Measuring Range 0.5 - 3Ω

2.3 Temperature
Channel Number 2 ch
Range 0-50C
Resolution 0.1C
Accuracy 0.2C

Operation environment
Operating temperature 0~45℃
Operating humidity 15%~90%, non-condensing
Storage Temperature -20~60℃
Storage humidity 0 ~ 95 %, non-condensing
Operating altitude -500 meters ~ 5000 meters

You might also like