1. Standard
GB/T16927.1 Part One General Definition and Test Requirements, High Voltage Test Technique
GB/T16927.2 Part Two Measurement System, High Voltage Test Technique
GB/T7354 Partial Discharge Measurement
DL/T846.4 Part Four Partial Discharge Detector by Pulse Current Method, Terms of High Voltage
Test Equipment General Technique
GB/T311.1 Part One Definition, Principle and Rules, Insulation Coordination
2. Products List
No Products Name Products Description Q`ty Remarks
JF-A04D Partial Discharge Detector
1 JF-A04D 1pc
(Touch Screen)
2 Digital external calibrator JZF-10N 1pc
3 Input unit 4# 3pcs
4 Partial Discharge Signal Filter Adapter PDLB-1 3pcs
5 Coupling capacitor OWF-100kV/1000pF 3pcs
6 Three-phase frequency doubling power filter LBS-60 1pc
7 The power cable -- 1pc
8 High frequency line -- 1pc
9 Partial discharge signal 8m 3pcs
10 Red, black signal wirings -- 1pc
11 Charger -- 1pc
- 1 -
Main Technical Data:
Model: JF-A04D
Channels: 4 Channel
Capacitance of test product: 6pF--250μF
Measuring range: 0.1pC-10000nC
Display unit: 10.1 inch capacitive touch screen
Measuring sensitivity: 0.01pC
Accuracy: ±5%
With 10.1 inch touch screen, one button automatic calibration, high
Function: voltage test voltage divider ratio can be set, serial communication upload
test data
On low frequency :10kHz、20kHz、40kHz
Amplifier:
On high frequency :80kHz、200kHz、300kHz
6 grades of gross adjustment. Gain difference between grades is 20±1dB,
Amplifier Gain adjusting:
fine adjusting range >20dB
(1) Window width: adjustable, 5 °to 170 °at 50 Hz.
Time window: (2) window position: each window can be rotated 0 °to 170 °
(3) two time windows can be opened separately or simultaneously.
Frequency: 50/60Hz、100Hz、150Hz、200Hz、400Hz
Time base of ellipse scan: Turn: 30º per grade, revolution≥120°
display ways: Ellipse -- Expand -- Line
Display: Digital meter with 3.5 bits LED digital display; precision better
Measuring voltmeter:
than ±5%(with full graduation)
Zero mark : The zero flag is consistent with all ellipse scan frequencies
Power Source: 220V/60Hz
Function: To measure the PD value of tested object
- 2 -
Main Technical Data:
Model: JZF-10N
The ranges to output electric
5pC、10pC、20pC、50pC、100pC、200pC、500pC
charge:
Polarity: Positive and Negative alternative
Repeat frequency: 50、100、200、500、1200Hz
Frequency accuracy: <1%
Pulse rise edge time: <60nS
Pulse fall edge time: >100uS
Injection capacitance: 10pF,100pF
Correction of charge error: Eq≤±5%
Charging adapter: 220VAC input,12VDC output
Size: 150×65×35mm3
Weight: 0.5kg
Function: To correct when PD test,Built in rechargeable lithium battery
3. 4# Input Unit
- 3 -
Main Technical Data
Model: 4#
Model: OWF-100kV/1000pF
- 4 -
Operating Frequency: 60Hz
Allowed running time: 100%UN, IN running 60min; 70% UN, IN continuous running
Model: LBS-100
Circuit Number: 3
- 6 -