You are on page 1of 4

C100 

Cardiovascular Monitor   
Specification 

 
Standard Configuration:  NIBP Recall:  2000 groups 
12 Lead ‐ECG,RESP, Dual‐Temp, Comen SPO2,NIBP    Trend Graph:  150 hours 

Optional Configuration:  Trend Table:  150 hours 

  Nellcor/Masiom SPO2, 4‐IBP,EtCO2 module, ICG,AG,C.O. , C100A  Alarm Event Recall:(+ARR  500 groups 


Telemetery monitor ,built‐in WIFI module  recall ) 

Safety Standards  ECG recall  7×24 hours of 12 lead ECG (8G    SD 

ISO 13485:2003 approved, CE marking according to MDD93/42/EEC  card) 
EDG diagnostic report  3000 groups 
Physical Characteristics 
Interface 
Product Size:  305mm×168mm×308.4mm 
Connectors:  USB interface, and PCL printer can be 
Net Weight:  3.5kg(standard configuration) 
connected 
Display: 
RJ45 Ethernet interface 
Type :  12” color TFT touch screen 
Resolution:  800×600  Nurse Call interface。 

waveforms:  up to 12 waveforms  AC interface 

External display  1 display through VGA  Grounding pole   


DVI‐D interface 
Operation Environment 
Multi‐function interface to connect 
Temperature:    5‐40℃ 
with sub‐monitor and plug in slot 
Humidity:      lower than 80% 
IR module interface, it supports the 
Power requirement: 
following modules: CO2, dual IBP, AG, 
AC Volatage:  AC100‐250V, 50/60Hz 
ICG,C.O.   
Current:  1.9A~0.8A 
Recorder: 
Battery: 
Trace:  Waveform: 6 channels for patient 
Battery Type:      Lithium‐ion 
monitor , 12 channels for ECG 
Battery Capacity:  2600mAh    16.8V 
Type :  Built‐in; thermal array 
Battery working time    2 hours   
Speed:  Speed:25mm/s,50mm/s 
Battery Recharging Time:    Maximum 4 hours for charging; 

Date storage  Record width:  216 mm 


Alarm Event Recall:  500 groups  Printer:  External printer supported 
ECG  range: 
TM
Lead Type  Cardio Tec  12‐leads ECG Analysis, 5  Resolution:  1% 
Lead and 3 lead selectable  Accuracy:  ±2% (70‐100%, MAX‐A, MAX‐AL, 
Lead selection:    3‐lead: I; II; III  MAX‐N, MAX‐P, MAX‐I and MAX‐FAST 
5‐lead: I; II; III; aVR; aVL;aVF; V  sensors); 
12‐Lead I; II; III; aVR; aVL;aVF; V1‐V6.  ±3% (70‐100%, D‐YS, DS‐100A, 
Gain Selection:  2.5mm/mV ,5.0mm/mV,10mm/mV,  OXI‐A/N and OXI‐P/I sensors); 
20mm/mV, auto  0‐69% unspecified 
Sweep Speed:    6.25,12.5, 25, 50mm/s 
Bandwidth :  MON Mode: 0.5‐40Hz  PR Measurement Range:      20~300bpm 
DIA Mode: 0.05‐130Hz 
OPE Mode:1‐20Hz  Resolution:    1bpm 

Accuracy:  ±3bpm 
ST Mode: 0.05‐40HZ 
Alarm range:  20‐250bpm 
Defib. Protection:  Withstand 4000VAC/50Hz voltage in 
Masimo SPO2 
isolation ,Again electrosurgical 
Measurement & alarm  1~100% 
interference and defibrillation 
range 
Recovery time :  3s 
Resolution:  1% 
ST SEGMENT detection:  ‐2.0mV ‐ +2.0Mv   
Accuracy:  ±2% (70~100%,Ped, non‐motion+ 
Arrhythmia Analysis :    Yes   
±3% (70‐100%, Neo, non‐motion); 
pacemaker detection:  detectable 
0‐69% unspecified 
Alarm range  0~100% 
Alarm:  Yes, audible and visual alarm, alarm 
Pulse rate:  Range: 25~240bpm 
events recallable 
Resolution:1bpm 
12 lead ECG Analysis:  yes 
Accuracy: ±3bpm(non‐motion) 
Heart Rate               
±5bpm(motion) 
Range:  Adult: 15‐300bpm 
Alarm range:25~240bpm 
Pediatric:15‐350bpm 
Comen SPO2 
Neonate:15‐350bpm 
Measurement & alarm  1~100% 
Resolution:  1 bpm 
range 
Accuracy:  ±1% or ±1bpm (whichever is greater) 
Resolution:  1% 
Respiration 
Accuracy:  ±2% (70~100%,Ped, non‐motion+ 
Method:  impedance method 
±3% (70‐100%, Neo, non‐motion); 
RR measurement range:  Adult: 6‐120bpm 
0‐69% unspecified 
Pediatric/Neonate: 6 ‐150bpm 
Alarm range  0~100% 
Resolution:    ±1 bpm  Pulse rate:  Range: 20~300bpm 
Accuracy:  6‐150rpm:±1rpm or 1%, whichever is 
Resolution:1bpm 
greater   
Accuracy: ±3bpm(non‐motion) 
0‐6rpm: unspecified 
±5bpm(motion) 
RESP Apnea Alarm:  Adult :10s‐60s 
Alarm range:20~300bpm 
Child and Neonate: 10s‐20s 
NIBP 
Alarm:  Audible and visual alarm; alarm 
Method  Automatic oscillometric 
events recallable 
Work mode:  Manual / Automatic/Continuous 
Sweep Speed:  6.25,12.5,25 mm/s 
Measurement Unit:  mmHg / Kpa selectable 
Gain Selection:  X0.25, X0.5, X1, X2,X3,X4,X5 
Measurement types:    Systolic, Diastolic, Mean 
Nellcor SPO2 
Range of systolic pressure:  Adult Mode:40‐270mmHg 
Measurement & alarm  0~100% 
Pediatric Mode:40‐200mmHg 
Neonate Mode 40‐135mmHg  Multi‐Gas/O2 
Range of diastolic  Adult Mode:10‐215mmHg  Method:  Infrared Absorption 
pressure:  Pediatric Mode:10‐150mmHg  Gas sorts:  CO2,O2,N2O,Des,Iso,Enf,Sev,Hal, 
Neonate Mode 10‐100mmHg  (optional paramagnetic sensor) 
Range of mean pressure:  Adult Mode:20‐235mmHg  Measurement range:  CO2 : 0‐15 vol%      ±(0.2 vol% + 2 % of 
Pediatric Mode:20‐165mmHg  the reading)   
Neonate Mode 20‐110mmHg  N2O: 0‐100 vol%      ±(2 vol% + 2 % of 
Over‐pressure protection:  both Hardware and software over  the reading)   
pressure protection  HAL, ISO, ENF: 0‐8 vol%      ±(0.15 
Accuracy:    Less than ±5mmHg  vol% + 5 % of the reading)   
Resolution  1mmHg  SEV : 0‐10 vol%      ±(0.15 vol% + 5 % 
PR from NIBP:    40‐240bpm  of the reading)   
Temperature (Dual Channel)  DES: 0‐22 vol%      ±(0.15 vol% + 5 % 
Measurement range:    0‐50°C (32 to 122F)  of the reading)   
Resolution:    0.1°C  O2: 0‐100vol%      ±(1 vol% + 2 % of the 
Accuracy:    ±0.1°C (without probe)  reading) 
Channel:    2 channels /    T1, T2, TD  Data output:  Fi and Et values 
(Temperature Difference)  Respiratory rate:  0‐150rpm 
IBP  Preheating time:  ISA CO2:< 10s,ISA OR+/AX+:< 20s
Channel  4 channels  Alarm:  User selectable alarm limits for all 
Measured Pressure  ART,PA,CVP,RAP,LAP,ICP,P1,P2  measurement 
Measurement    Range:    ‐10mmHg~300mmHg  others:  Up to 4 waveforms displayed 
Alarm Range:  ‐50mmHg~300mmHg  Agent mixture detection 
Accuracy:    ±2% or ±1mmHg, whichever is greater MAC value displayed 

EtCO2(Sidestream)  Cardiac Output(C.O.) 

CO2 Measurement Range:    0 ‐150 mm Hg, 0 to 79%, 0 to 20kPa  Method:  Thermodilution 

(at 760mmHg)  Measurement range:  C.O.:0.1~20L/min 


Accuracy:  ± 2 mm Hg (0 – 40 mm Hg)                      BT:      25~43℃ 
± 5% of reading (41 – 70 mm Hg)  IT:        0~27℃ 
± 8% of reading (71 –100 mm Hg)  Resolution:  C.O.:0.1L/min 
± 10% of reading (101 –150 mm Hg)  BT、IT:0.1℃ 
awRR range    0‐150rpm  Accuracy:  C.O.:±5% or ±0.1 L/min,whichever is 
RR    accuracy:    ±2rpm (0‐70rpm)                                  greater 
±5rpm (>70rpm)  BT、IT: ±0.1℃(no sensor) 
Response time:    <240msec (10% to 90%) Alarm Range:  BT Hi limit: (LO limit +0.1)~43℃ 
Delay time:    <2s  (C.O.)  BT Lo limit:23.0~(Hi limit‐0.1)℃ 
EtCO2(mainstream)  Step: 0.1℃ 
CO2 Measurement Range:    0 ‐150 mm Hg, 0 to 79%, 0 to 20kPa  Impedance Cardiography (ICG)   
(at 760mmHg)  Method:  Thoracic Bio‐impedance 
Accuracy:  ± 2 mm Hg (0 – 40 mm Hg)                      measurement 
± 5% of reading (41 – 70 mm Hg)  Measurement range:  SV:5~250 ml/b 
± 8% of reading (71 –100 mm Hg)  HR:40~250 bpm 
± 10% of reading (101 –150 mm Hg)  C.O.:1.4~15L/min 
awRR range    0‐150rpm  Accuracy:  SV:unspecified 
awRR precision  ±2rpm (0‐70rpm)                                  HR:  ±2bpm 
±5rpm (>70rpm)  C.O:unspecified 
Delay time:    <2s  Alarm range:  CI: HI (LO+0.1)~15.0 L/min/m2 
LO 0~(HI-0.1)L/min/m2  Alarm Deviation  CI:±01. L/min/m2 
TFC: HI (LO+1)~150/KΩ  TFC:±1 KΩ 
LO 10~(HI-1) KΩ 

 
 
 
 

COMEN, one of the leading medical device manufacturers in China,


Shenzhen Comen Medical Instrument Co.,ltd
and admitted to be the quickest developing company in medical

industry.
Add: No.2 of FIYTA Timepiece Building, Nanhuan
COMEN design and produce medical devices for OR, NICU, and ICU.
Avenue, Gongming Sub-district, Gongming New
The product range covers Anesthesia devices, Monitoring &diagnostic
District, Shenzhen, P.R.China
products, Infant incubators, ventilators, medical pendants, operation
Tel: +86-755-2640 8879 2641 0713
light, all designed with specific patient data managing system to link-in
Fax: +86-755-2643 1232
to multiple medical database.
Website: en.comen.com
Based in Shenzhen - the hi-tech “headquarter” of China, COMEN has
E-mail: info@szcomen.com
30 branch offices operating business in China and 8 international
Specifications subject to change without notice
branch offices in: USA, Russia, Turkey, Mexico, Thailand, India,

Malaysia and Indonesia.

With all over 50,000 install bases in hospitals and medical centers

globally, COMEN products are accepted and favored with its friendly

design and excellent performance.