Professional Documents
Culture Documents
Cardiovascular Monitor
Specification
Standard Configuration: NIBP Recall: 2000 groups
12 Lead ‐ECG,RESP, Dual‐Temp, Comen SPO2,NIBP Trend Graph: 150 hours
ISO 13485:2003 approved, CE marking according to MDD93/42/EEC card)
EDG diagnostic report 3000 groups
Physical Characteristics
Interface
Product Size: 305mm×168mm×308.4mm
Connectors: USB interface, and PCL printer can be
Net Weight: 3.5kg(standard configuration)
connected
Display:
RJ45 Ethernet interface
Type : 12” color TFT touch screen
Resolution: 800×600 Nurse Call interface。
Accuracy: ±3bpm
ST Mode: 0.05‐40HZ
Alarm range: 20‐250bpm
Defib. Protection: Withstand 4000VAC/50Hz voltage in
Masimo SPO2
isolation ,Again electrosurgical
Measurement & alarm 1~100%
interference and defibrillation
range
Recovery time : 3s
Resolution: 1%
ST SEGMENT detection: ‐2.0mV ‐ +2.0Mv
Accuracy: ±2% (70~100%,Ped, non‐motion+
Arrhythmia Analysis : Yes
±3% (70‐100%, Neo, non‐motion);
pacemaker detection: detectable
0‐69% unspecified
Alarm range 0~100%
Alarm: Yes, audible and visual alarm, alarm
Pulse rate: Range: 25~240bpm
events recallable
Resolution:1bpm
12 lead ECG Analysis: yes
Accuracy: ±3bpm(non‐motion)
Heart Rate
±5bpm(motion)
Range: Adult: 15‐300bpm
Alarm range:25~240bpm
Pediatric:15‐350bpm
Comen SPO2
Neonate:15‐350bpm
Measurement & alarm 1~100%
Resolution: 1 bpm
range
Accuracy: ±1% or ±1bpm (whichever is greater)
Resolution: 1%
Respiration
Accuracy: ±2% (70~100%,Ped, non‐motion+
Method: impedance method
±3% (70‐100%, Neo, non‐motion);
RR measurement range: Adult: 6‐120bpm
0‐69% unspecified
Pediatric/Neonate: 6 ‐150bpm
Alarm range 0~100%
Resolution: ±1 bpm Pulse rate: Range: 20~300bpm
Accuracy: 6‐150rpm:±1rpm or 1%, whichever is
Resolution:1bpm
greater
Accuracy: ±3bpm(non‐motion)
0‐6rpm: unspecified
±5bpm(motion)
RESP Apnea Alarm: Adult :10s‐60s
Alarm range:20~300bpm
Child and Neonate: 10s‐20s
NIBP
Alarm: Audible and visual alarm; alarm
Method Automatic oscillometric
events recallable
Work mode: Manual / Automatic/Continuous
Sweep Speed: 6.25,12.5,25 mm/s
Measurement Unit: mmHg / Kpa selectable
Gain Selection: X0.25, X0.5, X1, X2,X3,X4,X5
Measurement types: Systolic, Diastolic, Mean
Nellcor SPO2
Range of systolic pressure: Adult Mode:40‐270mmHg
Measurement & alarm 0~100%
Pediatric Mode:40‐200mmHg
Neonate Mode 40‐135mmHg Multi‐Gas/O2
Range of diastolic Adult Mode:10‐215mmHg Method: Infrared Absorption
pressure: Pediatric Mode:10‐150mmHg Gas sorts: CO2,O2,N2O,Des,Iso,Enf,Sev,Hal,
Neonate Mode 10‐100mmHg (optional paramagnetic sensor)
Range of mean pressure: Adult Mode:20‐235mmHg Measurement range: CO2 : 0‐15 vol% ±(0.2 vol% + 2 % of
Pediatric Mode:20‐165mmHg the reading)
Neonate Mode 20‐110mmHg N2O: 0‐100 vol% ±(2 vol% + 2 % of
Over‐pressure protection: both Hardware and software over the reading)
pressure protection HAL, ISO, ENF: 0‐8 vol% ±(0.15
Accuracy: Less than ±5mmHg vol% + 5 % of the reading)
Resolution 1mmHg SEV : 0‐10 vol% ±(0.15 vol% + 5 %
PR from NIBP: 40‐240bpm of the reading)
Temperature (Dual Channel) DES: 0‐22 vol% ±(0.15 vol% + 5 %
Measurement range: 0‐50°C (32 to 122F) of the reading)
Resolution: 0.1°C O2: 0‐100vol% ±(1 vol% + 2 % of the
Accuracy: ±0.1°C (without probe) reading)
Channel: 2 channels / T1, T2, TD Data output: Fi and Et values
(Temperature Difference) Respiratory rate: 0‐150rpm
IBP Preheating time: ISA CO2:< 10s,ISA OR+/AX+:< 20s
Channel 4 channels Alarm: User selectable alarm limits for all
Measured Pressure ART,PA,CVP,RAP,LAP,ICP,P1,P2 measurement
Measurement Range: ‐10mmHg~300mmHg others: Up to 4 waveforms displayed
Alarm Range: ‐50mmHg~300mmHg Agent mixture detection
Accuracy: ±2% or ±1mmHg, whichever is greater MAC value displayed
EtCO2(Sidestream) Cardiac Output(C.O.)
.
industry.
Add: No.2 of FIYTA Timepiece Building, Nanhuan
COMEN design and produce medical devices for OR, NICU, and ICU.
Avenue, Gongming Sub-district, Gongming New
The product range covers Anesthesia devices, Monitoring &diagnostic
District, Shenzhen, P.R.China
products, Infant incubators, ventilators, medical pendants, operation
Tel: +86-755-2640 8879 2641 0713
light, all designed with specific patient data managing system to link-in
Fax: +86-755-2643 1232
to multiple medical database.
Website: en.comen.com
Based in Shenzhen - the hi-tech “headquarter” of China, COMEN has
E-mail: info@szcomen.com
30 branch offices operating business in China and 8 international
Specifications subject to change without notice
branch offices in: USA, Russia, Turkey, Mexico, Thailand, India,
With all over 50,000 install bases in hospitals and medical centers
globally, COMEN products are accepted and favored with its friendly