Product
Options
Invasive Blood Pressure (IBP)
Carbon dioxide ( CO2) Module
Thermal Printer
Touch Screen
Three display output: TTL / CMOS LCD, LVDS LCD and standard VGA output
Built-in Wifi
Product Specifications
ECG
Lead Mode 5 Leads ( R, L, F, N, C or RA, LA, LL, RL, V)
Lead selection I, II, III, avR, avL, avF, V,
Waveform 2 ch
Lead mode 3 Leads ( R, L, F or RA, LA, LL)
Lead selection I, II, III,
Waveform 1 ch
Gain x2.5mm/mV, x5.0mm/mV, x10mm/mV, x20mm/mV, auto
HR and Alarm
Range
Adult 15 ~ 300 bpm
Neo/Ped 15 ~ 350 bpm
Accuracy ± 1% or ± 1bpm, which great
Resolution 1 bpm
Sensitivity > 200 (uV )
Differential Input Impedance > 5 MΩ
CMRR
Monitor > 105 dB
Operation > 105 dB
Diagnosis > 85 dB
SYS 40 ~ 200 mmHg Electrode offset potential
DIA 10 ~ 150 mmHg ±300mV
MEAN 20 ~ 165 mmHg Leakage Current < 10 uA
Neonatal Mode Baseline Recovery < 3 S
SYS 40 ~ 135 mmHg After Defi.
DIA 10 ~ 100 mmHg ECG Signal Range ±8 m V
MEAN 20 ~ 110 mmHg ( Vp-p )
Resolution Bandwidth
Pressure 1mmHg Surgery 1 ~ 15 Hz
Accuracy Monitor 0.5 ~ 35 Hz
Pressure Diagnostic 0.05 ~ 100
Maximum Mean error ±5mmHg Hz
Maximum Standard deviation ±8mmHg Calibration Signal 1 (mV ),
Overpressure Protection Accuracy : ±5%
Adult Mode 297±3 mmHg ST Segment Monitoring
Pediatric Mode 240±3 mmHg Range
Neonatal Mode 147±3 mmHg Measure and Alarm -2.0 ~
+2.0 mV
SpO2 ARR Detecting
Measuring Range 0 ~ 100 % Type ASYSTOLE,
Alarm Range 0 ~ 100 % VFIB/VTAC, COUPLET,
Resolution 1 % BIGEMINY, TRIGEMINY,
Accuracy 70% ~ 100% ±2 % R ON T, VT>2, PVC,
0% ~ 69% unspecified TACHY, BRADY, MISSED
Actualization interval about 1 Sec. BEATS, PNP, PNC
Alarm Delay 10 Sec. Alarm Available
Pulse Rate Review Available
Measuring and Alarm Range 0~254bpm
Resolution 1bpm RESPARATION
Accuracy ±2bpm Method Impedance
between RF (RA-LL)
TEMPERATURE Differential Input
Channel 1 Impedance >2.5 MΩ
Measuring and Alarm Range 0 ~ 50 C Measuring Impedance
Resolution 0.1~C Range: 0.3~5.0Ω
Accuracy ±0.1~C Base line Impedance
Actualization interval about 1 Sec. Range: 0 – 2.5 KΩ
Average Time Constant < 10 Sec. Bandwidth 0.3 ~ 2.5 Hz
Resp. Rate
IBP Measuring and Alarm
Label ART, PA, CVP, RAP, LAP, ICP, P1, P2 Range
Measuring and alarm range Adult 0 ~ 120 rpm
ART 0 ~ 300 mmHg Neo/Ped 0 ~ 150 rpm
PA -6 ~ 120 mmHg Resolution 1 rpm
CVP/RAP/LAP/ICP -10 ~ 40 mmHg Accuracy ±2 rpm
P1/P2 -10 ~ 300 mmHg Apean Alarm 10 ~ 40 S
Press Sensor
Sensitivity 5 uV/V/mmHg NIBP
Impedance 300-3000Ω Method Oscillometric
Resolution 1 mmHg Mode Manual, Auto, STAT
Accuracy ±2% or ±1mmHg, which great about 1 Sec. Measuring Interval in
AUTO Mode 1, 2, 3, 4, 5,
10, 15, 30, 60, 90, 120,
180, 240,480 (Min)
Measuring Period in STAT
Mode 5 Min
Pulse Rate Range 40 ~
Meditech Equipment Co.,Ltd (Meditech Group) 240 bpm
Alarm Type SYS, DIA,
Tel: (+86) 532 85832673 MEAN
Fax: (+86) 532 81705331 Measuring and alarm
range
Email: sales@meditech.cn
Adult Mode
SYS 40 ~ 270
Address: 89 Laoshan Road ,Building 69, Qingdao, Shandong mmHg
Province, P. R. China DIA 10 ~ 215
Manufacturer:140 Zhuzhou Rd, Qingdao , Shandong Province, P. R. mmHg
China 266035 MEAN 20 ~ 235
mmHg
Pediatric Mode
Tel: (+86) 532 88702078
Middle East:
Email: Eg@meditech.cn
To serve you well please kindly CC: every mail you send to meditech
Customer Care Team: cct@meditech.cn