Vito Fernicola
IMGC-CNR
Impedance-based sensor
Direct output in RH unit
(capacitance, etc.)
Wet-bulb e¢s (t w ) - A¢ × P × (t - t w )
f=
psychrometers e¢s (t )
PRT Aspiration
Flow
Air
temperature
f (t s , Ps ) × es (t s ) Pc
f= ×
f (t c , Pc ) × es (t c ) Ps
Relative Temperature
humidity 10 °C 20 °C 30 °C 50 °C
10 %rh 0.7 %rh 0.6 %rh 0.6 %rh 0.5 %rh
50 %rh 3.5 %rh 3.2 %rh 3.0 %rh 2.6 %rh
90 %rh 6.3 %rh 5.7 %rh 5.4 %rh 4.6 %rh
2.0
% difference (Magnus-Sonntag eq.)
1.6
1.2
0.8
0.4 +0.31%
0.0
-0.31%
-0.4
-0.8
-1.2
-40 -20 0 20 40 60 80 100
Temperature / °C
2.0
1.5
1.0
RHref - RHread / %
0.5
0.0
-0.5
-1.0 T = 60 °C
-1.5
T = 25 °C
-2.0
0 10 20 30 40 50 60 70 80 90 100
RHread / %
1 T = 65 °C
T = 40 °C
0
T = 10 °C
RHref - RHread / %
-1
-2
-3
-4
-5
-6
0 10 20 30 40 50 60 70 80 90 100
RHread / %
0.8
dRH/dT
0.6 [%rh/°C]
0.4
0.2
0.0
-0.2
-0.4
-0.6
RH [%]
-0.8
0 20 40 60 80 100
3
2nd order
RHref - RHread / %
1 3rd order
0 1st order
-1
T = 25 °C
-2
0 10 20 30 40 50 60 70 80 90 100
RHread / %
Ø which order ?
Ø extrapolation is allowed ?
Istituto di Metrologia "G. Colonnetti" EUROMET Workshop on Uncertainty in Humidity Measurements
Vito C. Fernicola 15
Temperature effect on the electronics
(e.g. partially-immersed probe)
-1
RHref - RHread / %
-2
-3
-4
yr. 1998 yr. 2002
-5 yr. 2001 yr. 2004
-6
0 10 20 30 40 50 60 70 80 90 100
RHread / %
p p p
Y = X1 1 × X 2 2 × × × X NN
2 N 2
é u ( y) ù é u ( xi ) ù
u r2 ( y ) = ê c
y
ú = åê i x ú
p × with pi = ±1
ë û i =1 ë i û
u c (f )
= u r2 [es (t s )] + u r2 [(es (t c )] + u r2 ( f ) + u r2 ( P)
f
Air temperature / °C
%rh 5 10 20 30 40 50 60 70
10 0.60 0.59 0.59 0.59 0.59 0.59
20 0.66 0.66 0.61 0.60 0.60 0.60 0.60 0.60
30 0.74 0.73 0.64 0.63 0.62 0.63 0.62 0.66
40 0.83 0.69 0.67 0.66 0.65 0.66 0.72 0.70
50 0.75 0.74 0.72 0.70 0.69 0.69 0.77 0.75
60 0.81 0.79 0.76 0.74 0.72 0.87 0.84 0.81
70 0.87 0.85 0.82 0.79 0.76 0.94 0.90 0.87
80 0.93 0.91 0.87 0.84 0.81 1.02 0.97
90 1.00 0.97 0.93 0.89 0.85 1.09 1.04