H

2
S & TOTAL SULPHUR SPECIFICATIONS
WWW.GALVANIC.COM
RAPID RESPONSE
SUPERIOR ACCURACY
REDUCED COST AND MAINTENANCE
OVER-RANGE & MULTI-RANGE CAPABILITY
WWW. GA L VA NI C . C OM E - MA I L : I NF O@ GA L VA NI C . C OM
Galvanic Applied Sciences Inc.
7000 Fisher Road S.E.
Calgary, Alberta T2H 0W3
Telephone: 403-252-8470
Fax: 403-255-6287
Houston, Texas, USA
13211K Windfern Rd
Houston, TX 77064
Telephone : 281-477-9002
Fax: 281-477-9004
Toll Free
1-866-252-8470
(US & Canada only)
WWW.GALVANIC.COM
PERFORMANCE
Ranges: Up to 0-2000ppm with no dilution;
up to 0-100% with dilution
Low Range: 0-1ppm or less (Please consult factory)
Linearity: 1% of full scale
Repeatability: 1% of full scale > 1ppm
2.5% of full scale <1ppm
2.5% of full scale>200ppm
2% of full scale for total sulfur
Cycle Time: Depends on range and gas concentration;
Maximum cycle time user configurable;
20 second response to alarm
FUNCTIONS
Streams : 4 Streams
Each stream includes: H
2
S analysis run,
zero run,reference 1 run, reference
2 run, calibration run and optional total
sulphur run
Each stream can have separate calibration
coefficient and range of measurement
Tape Life: Typical tape life given 3 minute cycle time is
5 weeks; can be extended to 15 weeks
Tape Analyzer measures exact amount of
Measurement: tape used
Auto/Manual In AUTO mode analyzer runs in predefined
Mode Of sequence that is user programmable
Operation: In MANUAL mode user can force any
stream to run
COMMUNICATION/INTERFACE
Dual Processor Ensures accuracy
Design:
Outputs: 6 - isolated 4-20mA user scalable and
programmable (16 Bits of resolution)
6 - Relay contact SPDT (form C) 8 Amps
220VDC max. user programmable
8 - Solenoid drivers 3A max. User
programmable
1 - USB or RS-232 Serial Port
1 - RS-232 Serial Port (Modem)
1 - RS-485 Serial Port (Isolated)
1 - High Speed LAN connection at 2.5M
Bits per second
Ethernet Option
All serial ports are MODBUS compatible
10 - Visual indicators (LED) for quick status
Large graphics display for easy viewing
Inputs: 4 - Dry contact inputs - user programmable
4 - Wet inputs for remote calibration etc.
user programmable
Two isolated 4-20ma analog inputs to allow
903 to be communication hub for other
devices that do not include that capability
Intrinsically Safe Keypad has 20 keys for
easy menu navigation
INSTRUMENT SPECIFICATIONS
Ambient: 10-50C (without enclosure)
Power 10W @ 24VDC
Consumption:
Power Universal AC 90-240V and 10-36V DC
Input range: input
Electrical: All systems are CSA C/US or field certified
Class 1,Div I, Group B,C,D (pending)
Class 1,Div II, Group B,C,D (pending)
Electronics: Dual processor design using intelligent
colorimetric sensor
Sensor is using 24 Bit AD converter to
achieve greatest sensitivity and
incorporates temperature sensor
Real Time Clock for accurate event time
stamping
512 Kbytes non-volatile memory for Data
Logging and Audit Trail (Up to 10 months of
H
2
S concentration data logs)
Field upgradeable firmware
Tape Length sensor to accurately measure
tape movement and usage
Independent Low Tape sensor is optional
Dimensions: 28 H x 16 W x 16 D
Weight: 903 Div II, 40 lbs
903 Div 1, 80 lbs
OPTIONS
Dilution System
High Pressure Sampling System
Solar Power
Diagnostic - Low Pressure
Diagnostic - Low Hydrogen
Enclosure / Cabinet
Total Sulphur
Back Panel
ISO 9001: 2000
0020422:800260
M
a
r
c
h
-
2
0
0
8
903 H
2
S & TOTAL SULPHUR
ANALYZER
PATENT PENDING
The Galvanic Applied
Sciences Leadership continues
Galvanic is the leading supplier of
H
2
S / Total Sulphur analyzers in
the world. Our foundation is built on
years of innovation, leadership and
customer satisfaction.
Experienced
Over 25 years of experience.
Established
Galvanic is a reliable and trusted
name in H
2
S analysis.
Customer Service
Our experienced field support
team is always there to meet your
requirements.
APPLICATIONS:
Natural Gas Processing
Natural Gas Transmission
Liquified Petroleum Gas
Liquified Natural Gas
Petrochemical Feedstock
Fuel Cell
Refining
Emission Monitoring
903 H
2
S TAPE ANALYZER
Galvanic's 903 H2S Tape Analyzer is designed to measure up to 4 streams of hydrogen
sulphide ( H2S) and total sulphur in gaseous streams. The new 903 H2S Tape Analyzer has the
same rapid 20 second response time of the traditional lead acetate tape, while making
unprecedented advances in accuracy and linearity, and reduced operation cost.
These advances are based on a patent pending algorithm and advanced digital sensor
technology developed by Galvanic.
The 903 H
2
S Tape Analyzer new capabilities include:
0-2000 ppm range without dilution - at least 50 times greater than traditional
tape analyzers;
Linearity and repeatability of just 1% of full scale;
5-15 week tape life - up to 2 times greater than that of traditional tape analyzers; and
Over-range and multi-range measurement
Operators can now measure the extent to which H2S exceeds pre-set alarm levels as well as
use the quick response H2S analyzer in new, higher ppm applications.
In addition, the 903 H2S Tape Analyzer can read samples from different ranges. This multi-
range capability replaces the need for separate analyzers calibrated each for separate streams.
Other new features of the 903 H2S Tape Analyzer include built-in modbus communications /
RTU capabilities, a greater number of outputs and inputs, sensor auto-calibration on each
analysis and dual redundant power supply.
The 903 H2S Tape Analyzer also comes with a total sulphur option. Using the multi-range
capability 1 unit can measure up to 4 separate H2S and total sulphur streams of varying
concentrations.
903 IS THE MOST ADVANCED TAPE
ANALYZER ON THE MARKET OFFERING:
Rapid Alarm
20 seconds to alarm
Reduced Cost and Maintenance
Up to 2 times greater tape life than other analyzers
Auto-calibration of digital colorimetric sensor
Tape measurement output
Greater Accuracy
Dynamic Range: 0-2000 ppm with no dilution; 0-100% with dilution
Linearity: 1% of full scale
Repeatability: 1% of full scale at > 1ppm
Enhanced Communications
Built-in RTU capabilities: MODBUS on RS-232,
RS-485, USB, Ethernet & Arcnet Increased I/O
Analog: 6 outputs & 4 inputs
Digital: 8 inputs, 6 relays & 8 solenoid drivers
Uninterrupted Operation
Dual redundant power supply
Up to 4 Streams
Over-range and Multi-range Measurement
903 COLORIMETRIC SENSOR
The 903 Sensor Block achieves cutting precise by making accurate measurements at the source of the
chemical reaction. All measurements, diagnostics, and calibrations are performed by an independent
microprocessor on the sensor block.
MULTI-STREAM MULTI-RANGE
Galvanics 903 has the ability to measure more than one stream of varying concentration. Traditionally
two analyzers were required when the concentration of two streams were significantly different. The
patent pending 903 algorithm allows one analyzer to measure 2 - and up to 4 - streams that have
concentrations differing by up to 2 orders of magnitude. For example, one analyzer could be used to
measure H
2
S in the 0-10 ppm range and Total Sulphur in the 0- 500 ppm range.
CONFIGURABLE SOFTWARE
The 903 comes with a Windows application which can be used to configure and
monitor its operation. All aspects of the 903 can be configured including alarms,
sequence of analysis operation, digital inputs, digital outputs, analog inputs, analog
outputs, communications, and logging. Monitoring is achieved using the Analysis
window, with further break downs of alarms on the Alarm window, and IO using the
Status window. In addition it is possible to download logs of analysis results, and an audit
trail showing alarms and configuration changes.
COMMUNICATIONS
With built-in MODBUS communications / RTU capabilities, The 903 is equipped with
RS - 485 & RS - 232 communication ports. The 903 supports Modicon and Enron
MODBUS protocols. The 903 can also take analog signals from other analyzers and
output the data via MODBUS communications.
TOTAL SULPHUR OPTIONS
The hydrogenation reaction furnace allows a standard Galvanic Applied Sciences H2S analyzer
tomeasure total sulphur concentration.
The sample is mixed with a hydrogen stream and heated to approximately 900C. At this temperature
in the presence of hydrogen all sulphur compounds will be converted to H2S. Also, any hydrocarbon
heavier than methane will be cracked to form methane.
Examples of the sulphur reaction are:
900C
COS + 4H
2
CH
4
+ H
2
O + H
2
S
Carbonyl Sulphide
900C
(CH
3
)
2
S + 2H
2
2CH
4
+ H
2
S
Dimethyl Sulphide
LEAD ACETATE TAPES
Galvanics controlled plant environment ensures top quality in our lead acetate tapes.
Our automated process and constant quality checks ensure consistent product. All our
tapes are cleaned after treatment to minimize problems caused by paper dust. We can provide tape
lengths up to three hundred and fifty feet. Analyzers run cleaner and smoother with Galvanics tape.
No interferences Colourless reactants
Sensitivity to ppb Linear respon
900C
(CH
3
CH
2
)
2
S + 4H
2
4CH
4
+ H
2
S
Diethyl Sulphide
900C
C
4
H
10
+ 3H
2
4CH
4
Butane: Example of Cracking
Microprocessor controlled sensor
compensates for drift, aging and
tape variances
Automatic zero set
Automatic span set
Remote diagnostics
Digital signal processing for best
accuracy and repeatability
Automatic light emitter intensity
compensation
24 Bit ADC converter for
measurement in situ
The Galvanic Applied
Sciences Leadership continues
Galvanic is the leading supplier of
H
2
S / Total Sulphur analyzers in
the world. Our foundation is built on
years of innovation, leadership and
customer satisfaction.
Experienced
Over 25 years of experience.
Established
Galvanic is a reliable and trusted
name in H
2
S analysis.
Customer Service
Our experienced field support
team is always there to meet your
requirements.
APPLICATIONS:
Natural Gas Processing
Natural Gas Transmission
Liquified Petroleum Gas
Liquified Natural Gas
Petrochemical Feedstock
Fuel Cell
Refining
Emission Monitoring
903 H
2
S TAPE ANALYZER
Galvanic's 903 H2S Tape Analyzer is designed to measure up to 4 streams of hydrogen
sulphide ( H2S) and total sulphur in gaseous streams. The new 903 H2S Tape Analyzer has the
same rapid 20 second response time of the traditional lead acetate tape, while making
unprecedented advances in accuracy and linearity, and reduced operation cost.
These advances are based on a patent pending algorithm and advanced digital sensor
technology developed by Galvanic.
The 903 H
2
S Tape Analyzer new capabilities include:
0-2000 ppm range without dilution - at least 50 times greater than traditional
tape analyzers;
Linearity and repeatability of just 1% of full scale;
5-15 week tape life - up to 2 times greater than that of traditional tape analyzers; and
Over-range and multi-range measurement
Operators can now measure the extent to which H2S exceeds pre-set alarm levels as well as
use the quick response H2S analyzer in new, higher ppm applications.
In addition, the 903 H2S Tape Analyzer can read samples from different ranges. This multi-
range capability replaces the need for separate analyzers calibrated each for separate streams.
Other new features of the 903 H2S Tape Analyzer include built-in modbus communications /
RTU capabilities, a greater number of outputs and inputs, sensor auto-calibration on each
analysis and dual redundant power supply.
The 903 H2S Tape Analyzer also comes with a total sulphur option. Using the multi-range
capability 1 unit can measure up to 4 separate H2S and total sulphur streams of varying
concentrations.
903 IS THE MOST ADVANCED TAPE
ANALYZER ON THE MARKET OFFERING:
Rapid Alarm
20 seconds to alarm
Reduced Cost and Maintenance
Up to 2 times greater tape life than other analyzers
Auto-calibration of digital colorimetric sensor
Tape measurement output
Greater Accuracy
Dynamic Range: 0-2000 ppm with no dilution; 0-100% with dilution
Linearity: 1% of full scale
Repeatability: 1% of full scale at > 1ppm
Enhanced Communications
Built-in RTU capabilities: MODBUS on RS-232,
RS-485, USB, Ethernet & Arcnet Increased I/O
Analog: 6 outputs & 4 inputs
Digital: 8 inputs, 6 relays & 8 solenoid drivers
Uninterrupted Operation
Dual redundant power supply
Up to 4 Streams
Over-range and Multi-range Measurement
903 COLORIMETRIC SENSOR
The 903 Sensor Block achieves precise results by making accurate measurements at the source of the
chemical reaction. All measurements, diagnostics, and calibrations are performed by an independent
microprocessor on the sensor block.
MULTI-STREAM MULTI-RANGE
Galvanics 903 has the ability to measure more than one stream of varying concentration. Traditionally
two analyzers were required when the concentration of two streams were significantly different. The
patent pending 903 algorithm allows one analyzer to measure 2 - and up to 4 - streams that have
concentrations differing by up to 2 orders of magnitude. For example, one analyzer could be used to
measure H
2
S in the 0-10 ppm range and Total Sulphur in the 0- 500 ppm range.
CONFIGURABLE SOFTWARE
The 903 comes with a Windows application which can be used to configure and
monitor its operation. All aspects of the 903 can be configured including alarms,
sequence of analysis operation, digital inputs, digital outputs, analog inputs, analog
outputs, communications, and logging. Monitoring is achieved using the Analysis
window, with further break downs of alarms on the Alarm window, and IO using the
Status window. In addition it is possible to download logs of analysis results, and an event
log showing alarms and configuration changes.
COMMUNICATIONS
With built-in MODBUS communications / RTU capabilities, The 903 is equipped with
RS - 485 & RS - 232 communication ports. The 903 supports Modicon and Enron
MODBUS protocols. The 903 can also take analog signals from other analyzers and
output the data via MODBUS communications.
TOTAL SULPHUR OPTIONS
The hydrogenation reaction furnace allows a standard Galvanic Applied Sciences H2S analyzer to
measure total sulphur concentration.
The sample is mixed with a hydrogen stream and heated to approximately 900C. At this temperature
in the presence of hydrogen all sulphur compounds will be converted to H2S. Also, any hydrocarbon
heavier than methane will be cracked to form methane.
Examples of the sulphur reaction are:
900C
COS + 4H
2
CH
4
+ H
2
O + H
2
S
Carbonyl Sulphide
900C
(CH
3
)
2
S + 2H
2
2CH
4
+ H
2
S
Dimethyl Sulphide
LEAD ACETATE TAPES
Galvanics controlled plant environment ensures top quality in our lead acetate tapes.
Our automated process and constant quality checks ensure consistent product. All our
tapes are cleaned after treatment to minimize problems caused by paper dust. We can provide tape
lengths up to three hundred and fifty feet. Analyzers run cleaner and smoother with Galvanics tape.
No interferences Colourless reactants
Sensitivity to ppb Linear response
900C
(CH
3
CH
2
)
2
S + 4H
2
4CH
4
+ H
2
S
Diethyl Sulphide
900C
C
4
H
10
+ 3H
2
4CH
4
Butane: Example of Cracking
Microprocessor controlled sensor
compensates for drift, aging and
tape variances
Automatic zero set
Automatic span set
Remote diagnostics
Digital signal processing for best
accuracy and repeatability
Automatic light emitter intensity
compensation
24 Bit ADC converter for
measurement in situ
H
2
S & TOTAL SULPHUR SPECIFICATIONS
WWW.GALVANIC.COM
RAPID RESPONSE
SUPERIOR ACCURACY
REDUCED COST AND MAINTENANCE
OVER-RANGE & MULTI-RANGE CAPABILITY
WWW. GA L VA NI C . C OM E - MA I L : I NF O@ GA L VA NI C . C OM
Galvanic Applied Sciences Inc.
7000 Fisher Road S.E.
Calgary, Alberta T2H 0W3
Telephone: 403-252-8470
Fax: 403-255-6287
Houston, Texas, USA
13211K Windfern Rd
Houston, TX 77064
Telephone : 281-477-9002
Fax: 281-477-9004
Toll Free
1-866-252-8470
(US & Canada only)
WWW.GALVANIC.COM
PERFORMANCE
Ranges: Up to 0-2000ppm with no dilution;
up to 0-100% with dilution
Low Range: 0-1ppm or less (Please consult factory)
Linearity: 1% of full scale
Repeatability: 1% of full scale > 1ppm
2.5% of full scale <1ppm
2.5% of full scale>200ppm
2% of full scale for total sulfur
Cycle Time: Depends on range and gas concentration;
Maximum cycle time user configurable;
20 second response to alarm
FUNCTIONS
Streams : 4 Streams
Each stream includes: H
2
S analysis run,
zero run,reference 1 run, reference
2 run, calibration run and total sulphur run
Each stream can have separate calibration
coefficient and range of measurement
Tape Life: Typical tape life given 3 minute cycle time is
5 weeks; can be extended to 15 weeks
Tape Analyzer measures exact amount of
Measurement: tape used
Auto/Manual In AUTO mode analyzer runs in predefined
Mode Of sequence that is user programmable
Operation: In MANUAL mode user can force any
stream to run
COMMUNICATION/INTERFACE
Dual Processor Ensures accuracy
Design:
Outputs: 6 - isolated 4-20mA user scalable and
programmable (16 Bits of resolution)
6 - Relay contact SPDT (form C) 8 Amps
220VDC max. user programmable
8 - Solenoid drivers 3A max. User
programmable
1 - USB or RS-232 Serial Port
1 - RS-232 Serial Port (Modem)
1 - RS-485 Serial Port (Isolated)
1 - High Speed LAN connection at 2.5M
Bits per second
Ethernet Option
All serial ports are MODBUS compatible
10 - Visual indicators (LED) for quick status
Large graphics display for easy viewing
Inputs: 4 - Dry contact inputs - user programmable
4 - Wet inputs for remote calibration etc.
user programmable
Two isolated 4-20ma analog inputs to allow
903 to be communication hub for other
devices that do not include that capability
Intrinsically Safe Keypad has 20 keys for
easy menu navigation
INSTRUMENT SPECIFICATIONS
Ambient: 10-50C (without enclosure)
Power 10W @ 24VDC
Consumption:
Power Universal AC 90-240V and 10-36V DC
Input range: input
Electrical: All systems are CSA C/US or field certified
Class 1,Div I, Group B,C,D (pending)
Class 1,Div II, Group B,C,D (pending)
Electronics: Dual processor design using intelligent
colorimetric sensor
Sensor is using 24 Bit AD converter to
achieve greatest sensitivity and
incorporates temperature sensor
Real Time Clock for accurate event time
stamping
512 Kbytes non-volatile memory for Data
Logging and Audit Trail (Up to 10 months of
H
2
S concentration data logs)
Field upgradeable firmware
Tape Length sensor to accurately measure
tape movement and usage
Independent Low Tape sensor is optional
Dimensions: 28 H x 16 W x 16 D
Weight: 903 Div II, 40 lbs
903 Div 1, 80 lbs
OPTIONS
Dilution System
High Pressure Sampling System
Solar Power
Diagnostic - Low Pressure
Diagnostic - Low Hydrogen
Enclosure / Cabinet
Total Sulphur
Back Panel
ISO 9001: 2000
0020422:800260
M
a
r
c
h
-
2
0
0
8
903 H
2
S & TOTAL SULPHUR
ANALYZER
PATENT PENDING
WWW.GALVANIC.COM
TYTRONICS

SENTINEL
THE WATCHDOG OF CHEMICAL
ANALYSIS AND WATER MONITORING
151909_Galvanic_BRO 3/30/04 10:37 AM Page 1
UNIQUE BATCH SAMPLING CAPTURE TECHNIQUE
The Tytronics

patented sample capture technique is incorporated into the Tytronics

Sentinel. This means of sample capture provides the simplest and most reliable method
of capturing a reproducible volume of sample that is available. The sample inlet uses
1/4" tubing, thus minimizing blockages and allowing even high turbidity samples to be
readily handled.
The sample inlet system uses an internal fast loop ensuring that a fresh and representative
sample is always available for analysis. Sample capture volume can range from 5 to 30
ml. This allows the Tytronics

Sentinel to be used for ppb as well as ppm levels of


measurement.
Mixing is accomplished through the use of a magnetic Teflon

coated stir bar. Two types


of measurement sensors, colorimetric and potentiometric, can be mounted in the reaction
cell.
Peristaltic pumps are
used to introduce
calibrating and cleaning
solutions.
Reagents are introduced
through high precision
positive displacement
piston pumps which
require minimum
maintenance.
The Tytronics

Sentinel is the first on-line analyzer capable of both process chemical


analysis and the monitoring of potable water and wastewater. Offering an unmatched
performance to price ratio, the Tytronics

Sentinel provides measurement qualities and


accuracies simply not seen before. There is built-in reliability, thanks to a simple and
patented sample capture system and wide-bore tubing. This is an analyzer that can
handle difficult and dirty samples.
Whether you are currently analyzing or monitoring liquids, or looking to do so, the
Tytronics

Sentinel is your first choice in on-line analysis or monitoring.


ON-LINE ANALYSIS AND MONITORING
EASE OF OPERATION
Simple menu-driven operation and a modular design make the Tytronics

Sentinel easy
to use and maintain. A person with little chemical background can use the analyzer with
minimal training. Simple step by step instructions are displayed to guide the operator
through the procedure to complete a calibration and to set the unit in the automatic
analysis mode.
THE BEST METHODOLOGY FOR THE APPLICATION
One of the keys to Tytronics

leadership in on-line analysis is the commitment to provide


the best possible methodology (colorimetry, ISE, colorimetric or potentiometric titrations)
for the application at hand. Thus, you can be assured that your measurements of acids,
ammonia, bases, metals, nitrate, phosphate, etc., are done in the very best way.
All parameters in the Tytronics

Sentinel are user programmable. Multiple analysis


methods can be configured to tackle different concentration ranges. Multiple calibration
setups can be used to define different low and high range calibration curves. A cleaning
cycle can be programmed to clean out or dissolve any buildup on the measurement
sensor or reaction cell. Analysis, calibration and cleaning cycles can be executed at
periodic (user-programmable) intervals to provide hands- free automation of the
analyzer.
APPLICATIONS:
Acids
Alkalinity
Aluminum
Ammonia
Caustic & Carbonate
Chlorine
Chrome
Color
Copper
Fluoride
Hardness
Hardness in Brine
Iron
Manganese
Nickel
Nitrate
Nitrite
Peroxide
Phosphate
Silver
Sulfide
Zinc
...and many more...
Sensor Types:
Colorimetric/Potentiometric
Sample cleans the cell Flow is stopped and sample is
captured by siphon
Reagents are added directly
to the cell
Sample In To Water
To Water
151909_Galvanic_BRO 3/30/04 10:37 AM Page 2
Colorimeter
The colorimetric system uses dual wavelength optical configuration and an autoblank feature
which compensate for background color and turbidity. Colorimetric sensors are used to make
absorbance measurements for direct colorimetric measurements (Beers Law). A submerged
colorimetric probe measures the absorbance of the solution after reaction of sample and
reagents. The probe is cleaned and rinsed between analyses, thus eliminating the problems
of optics fouling. The probe is made of Kynar

and the reflector of Hastalloy C, which


provides excellent chemical resistance. The probe can be supplied in two different optical
pathlengths, 2 and 4 cm, determined by concentration
CALIBRATION: True multi-featured two point calibration technique allows trace analysis to be
routinely checked by simply using distilled water. Periodic checking of span standard can be
easily performed.
Titrator
Tytronics

Sentinel on-line colorimetric and potentiometric (pH and Redox) Titrations deliver
simple, reliable and highly cost-effective on-line analysis. Acids, Alkalinity, Caustic,
Carbonate, Cyanide, Hardness, Iron, Peroxide, Silver and many other analyses are common
applications across a wide variety of process industries. Tytronics

Sentinel Titrators have the


capability of performing single or dual end-point titrations. The Titrator is equipped with three
detection methods (fixed, relative and derivative endpoint) to insure flexibility and accuracy.
Expanded diagnostic capabilities include programmable learn cycles to automatically detect
chemistry endpoints. Users may program several methods and the system allows for multiple
reagents to be added. Analyses are fully automatic in all cases. Users can easily change the
frequency of analysis and calibration, as well as most of the default values, to improve or
customize a setup; three different access levels provide system security. The Tytronics

Sentinel
also features a full auto-cleaning capability for reliability and reduced maintenance. In sum,
this results in an extremely reliable, flexible, and user-friendly general purpose Titrator.
CALIBRATION: Automatic calibration is user selectable and configurable with calibration
repeat and alarm limits.
ISE
Potentiometric sensors are used for ion selective measurements. The single known addition
(SKA) technique compensates for electrode drift on every analysis. Tytronics

Sentinel ISE
analyzer is fitted as standard with an RTD-100 temperature sensor which permits automatic
temperature compensation of the analysis. Direct Method is used for high concentration
analysis.
CALIBRATION: Automatic calibration is user-selectable and configurable with calibration repeat
and alarm limits. The calibration is performed using the SKA method on a calibration standard.
TYTRONICS

SENTINEL: DESIGNED FOR OPTIMUM
PERFORMANCE
SAMPLING OF TYTRONICS

SENTINEL ANALYSIS METHODOLOGIES


Measure Technique Typical Range(s) Remarks
Acids Titration varies Potentiometric titration; various acids
Alkalinity Titration Varies Colorimetric titration for total alkalinity and potentiometric for P and M values.
Aluminum Colorimetric 0 to 100 ppb to 0 to 2 ppm Lower levels measured on potable water.
Ammonia Colorimetric 0 to 100 ppb to 0 to 25 ppm Uses the modified Berthelot method.
Ammonia ISE 0 to 2 ppm to 0 to 100 ppm Direct or using SKA method.
Caustic & Carbonate Titration varies Caustic scrubber
Chlorine Colorimetric 0 to 500 ppb to 0 to 3 ppm Free and/or total available chlorine may be measured.
Color Direct Spectrometry 0 to 10 to 0 to 500 Hazen Hazen or APHA color units.
Copper Colorimetric 0 to 500 ppb to 0 to 10 ppm Lower levels for potable water, higher levels usually on industrial waste streams.
Cyanide Colorimetric 0 to 200 ppb to 0 to 1 ppm Used on purified waste water streams.
Cyanide ISE 0.5 ppm and higher Waste water streams.
Fluoride ISE 0 to 2 ppm and higher Potable water using SKA technique.
Hardness Colorimetric 0 to 1 ppm to Hardness in water or brine
0 to 10 ppm
Hardness Titration 0 to 20 to 0 to 500 ppm Complexometric colorimetric titration.
Iron Colorimetric 0 to 100 ppb to 0 to 2 ppm Potable water.
Manganese Colorimetric 0 to 200 ppb Potable water.
Nitrate ISE 0 to 10 ppm to 0 to 100 ppm Direct or SKA method used.
Nitrate Direct UV Spectrometry 0 to 1000 ppm Consult factroy for suitability.
Nitrite Colorimetric 0 to 1 ppm Potable water derived from deep wells.
Phosphate Colorimetric 0 to 500 ppb to 0 to 3 ppm Used for corrosion prevention in potable, cooling tower and boiler water supplies.
Phosphate Colorimetric 0 to 10 ppm and higher Mainly used on waste water streams.
And many more...
151909_Galvanic_BRO 3/30/04 10:37 AM Page 3
S
E
N
T
I
N
E
L


0
3

-

0
4


P
r
i
n
t
e
d

i
n

C
a
n
a
d
a
WWW. GALVANI C . C OM E - MAI L : I NF O@GALVANI C . C OM
Toll Free 1-800-370-2345 (US only)
Galvanic Applied Sciences Inc.
Bay 6, 6325-12 Street S.E.
Calgary, Alberta T2H 2K1
Telephone: 403-252-8470
Fax: 403-255-6287
Calgary, Alberta
1200, 715-5th Avenue S.W.
Calgary, Alberta T2P 2X6
Telephone: 403-252-8470
Fax: 403-571-2684
Edmonton, Alberta
Telephone: 780-430-0142
Fax: 780-430-0172
Houston, Texas
Telephone: 281-477-9002
Fax: 281-477-9004
TYTRONICS

Galvanic Applied Sciences USA manufactures the complete brand of Tytronics

on-line wet
chemical and gas analyzers. The Tytronics

products include the Sentinel On-line Analyzers.


These analyzers use "state of the art" technology to perform colorimetric, titrimetric, ISE,
and spectrophotometric analyses. These analyzers can be configured to perform many
EPA and ASTM wet chemical methods, as well as many Standard Methods. The Sentinel
analyzers use a modular design that can be configured to meet the specific requirements of
the customer's application.
The Sentinel Analyzers are capable of both process chemical analysis and the monitoring
of potable water and wastewater. This instrument can analyze up to six streams for the same
chemical parameter, at different concentration ranges. The Sentinel can measure the
chemical composition of liquid processes from A (aluminum) to Z (zinc).
FEATURES
MULTI-STREAMING: The instruments can be expanded to analyze up to 6 sample streams per
analyzer. The frequency and order of stream analysis is fully programmable. Each stream has a
separate alarm and analog output with multi-range capability. The ability to optimize the method
used for each sample stream further proves the superiority of the advanced capabilities of the
Tytronics

Sentinel.
SAMPLE STREAM: Flow rate of 50 to 900 ml/min; pressures from 0 to 50 psig; temperature
from 5 to 70 Celsius and 1/4 sample inlet connection.
OPERATING TEMP: 5 to 45 Celsius; vortex air cooler recommended above 45 Celsius
AUTO CLEANING: An automatic wash pump to do periodic (user-programmable) cleaning
cycles; a suitable wash solution is introduced using the wash P-Pump to clean out or dissolve any
buildup on the electrode or reaction cell.
ANALOG OUTPUTS: 1 output per stream; isolated 0/4 to 20 mA; fully range-programmable;
maximum load 1000 Ohms; result values are held until the reading is updated.
DIGITAL OUPTUTS: Dual-isolated RS-232C for printer and computer interface at user-
selected baud rates, with remote control capability.
DIGITAL INPUTS: 8 digital inputs with fully programmable functions allow interface to a PLC and/or
interface to voltage-free contact from sensors; expandable to 16 inputs with multiplexor board option.
RELAY CONTACTS: Programmable voltage-free contacts (minimum 8 relays configurable to
N.C. or N.O.), expandable to 14 relays with multiplexor board option; contact rated at 5A at
240VAC maximum with a non-inductive load.
ALARMS: Stream-specific High and Low set point alarms; loss of sample flow alarm; fail-safe
power failure alarm; general, hardware and stream-specific fault signals,; result-ready, test-in-
progress, calibration-in-progress and other alarm and interface functions (consult Manual); alarms
can be assigned in software to voltage-free contact relays; grouping of alarms to relays is fully
programmable.
DATA STORAGE: Non-volatile memory saves the last 120 analysis results and last 25
calibration results; data retention is approximately 5 years.
POWER SUPPLY: Universal power supply 100 to 240 VAC, 50-60 Hz; consumption 200 watts
maximum.
ENCLOSURE TYPE: IP65, NEMA 4X, fiberglass wall-mounted industrial enclosure; modular design
of the fluidic and electronic sections results in easy maintenance and simple field retrofits or upgrades.
DIMENSIONS: (W x H x D) 24 x 25 x 13 or 61cm x 64cm x 33cm
WEIGHT: Approximately 80lb (36kg) fully equipped.
Galvanic Applied Sciences USA Inc.
41 Wellman Street
Lowell, MA. 01851
Telephone: 978-848-2701
Fax: 978-848-2713
151909_Galvanic_BRO 3/30/04 10:37 AM Page 4
WWW.GALVANIC.COM
Model BRM 941-LS
LiquidSulphurAnalyzerSystem
Analyzer H
2
S and H
2
S
x
in Liquid Sulphur
ResidualH
2
SinClausproducedsulphurcontainsbothH
2
SandH
2
S
x
,whichexistin
equilibriumwithelementalsulfur(S
8
).

Analysescanbeconductedin2minutes
(wetchemicalmethodrequiresseveralhours)
Analysesindependentofoperatorskill

TheuseofstateoftheartFouriertransforminfraredspectrometersallowsthe
resolutionoftheH
2
SandH
2
S
x
absorptionsgivingaccurate(10%ofthemeasure
mentvalue)determinationsoftotalH
2
Sinthe0350ppmwrangewithquantification
downto2ppmw.

Key Component
FTIRAnalyzer
Calibratedheatedcell
Sulphursamplingdevice(hotpot)
Proprietarysoftwarepackage

Can be used for:


Verificationthatrundownpitliquidsulphurmeetsspecification
Verificationthatdegassingprocessisproducingonspecproduct
Monitoringandcontrolofdegassingprocess
(energyconsumptionmanagement)
ProvidingdocumentedQualityAssuranceformarketingofliquidsulfur
product
Allowingassessmentofsafetyissuesinthehandlingoftheliquidsulphur
MonitoringofperformanceofcatalyticconvertorsinClaustrain.
(highH
2
Sresidualmayindicatelowcatalystactivity.)
DeterminationofH
2
Scontentinfrontendfurnacecondenserstream

Analysis time: 2 minutes
Accuracy: 10% of total residual H
2
S
Calibration : H
2
S
x
= 150 ppmw
Limits: H
2
S = 200 ppmw
Total = 350 ppmw
Headoffice:7000FisherRoadS.E.Calgary,AlbertaCanadaT2H0W3Tel:403.252.8470Fax:403.255.06287
GalvanicAppliedSciencesUSAInc.41WellmanStreet,LowellMA01851Tel:978.848.2701Fax:978.848.2713
HoustonOffice13211KWindfernRoad,HoustonTexas,77064USATel:281.477.9002Fax:281.477.9004
info@galvanic.comwww.galvanic.com
WWW.GALVANIC.COM
PLGCII
GAS CHROMATOGRAPH
GAS COMPOSITION INCLUDING
BTU/HEATING VALUE CALCULATION
SPECIFIC GRAVITY CALCULATION
pg g
The PLGC II is a compact and low cost gas chromatograph equipped with a universal
thermal conductivity detector. The unit offers a wide range of applications where
hydrocarbon gas or fixed gas analysis is required, such as in BTU measurement for the
natural gas industry.
Remember when field chromatography was considered too complex or expensive for
many applications. The PLGC II removes these drawbacks. The Galvanic Applied
Sciences PLGC II analyzer is a low cost, low maintenance chromatograph. The PLGC II
can be used in many measurement applications where an isothermal TCD scheme is
suitable. The PLGC II is fully automated and designed to perform on-line, real time
analysis. The Windows

based configuration program allows the user to view most


recent or saved chromatograms as well as configure the analyzer.
The PLGC II calculates the following for up to 8 streams:
PLGCII GAS CHROMATOGRAPH FOR
PIPELINE AND PROCESS APPLICATIONS
THE PLGC II LCD
Displays a choice of:
Heating valve/specific gravity
Component values
Chromatograms
Temperature control
Output alarms
Status
Heating value
Specific gravity
Wobbe index
Compressibility
Specific heat ratios
Component concentrations
Features:
24 Volt DC operation
Division II electrical classification
standard (Div I optional)
Thermistor TCD detector
No instrument air required
Low helium consumption
Reliability
Low maintenance
6
3 2
5
4
1
G
R
N B
L
U
R
E
D R
E
D
T T
DETECTOR
The Detector is a thermistor type TCD selected for high sensitivity at ambient
temperatures and short response time. It consists of a heated filament contained in a
stainless steel block. Gas flowing through the block loses heat to the block depending
on the concentration and the thermal conductivity of the gas.
The detector is a temperature controlled enclosure to minimize the effect of ambient
temperature changes. To further reduce temperature and flow effects, a reference
filament is used in a wheatstone bridge configuration.
Analyzer Detector
TCD Detection
FILAMENT SUPPORT 525
Filament Support 525
A-A Mount
8 Milliamp Constant Current
Output to Pre-amp
500 Ohm 500 Ohm
The micro volume diaphram valve is
actuated by carrier gas to perform the
sample injection.
The LCD display is capable of displaying
all information about the sample or status
of the PLGC II.
Solenoids electromechanically switch the
flow of gas to inject sample gas into the
system.
Customer Connections for Helium carrier
gas, Calibration gas and Sample gas.
The TCD measures the concentration of the
components of the sample gas based on
changes in the thermal conductivity of the
components.
The individual components are separated in
the column by process of absorption and
diffusion. Up to two columns are supported.
The computations, setup and
communications for the PLGC II are
performed by the Motherboard.
The Excitation Board provides power for
the thermal conductivity detector.
Optional Extended box to allow such
options as: AC Power and multiple
streams.
Optional AC Power Adapter
1
2
3
4
5
6
7
8
9
10
2
3
1
5
6
7
10
4
8
9
6
pg g
PLGC II SOFTWARE
The PLGC II software has been designed to receive, interpret, and plot data from the
PLGC II gas chromatograph. Peaks in the resulting chromatogram identify the
compounds by retention time and the concentration by the area of the peak. Various
parameters, including heating value, specific gravity, and compressibility factor, can be
displayed with the software. The Graphical User Interface (GUI) for the software is a
user friendly Windows

based system.
Computer Requirements:
Operating System:
An IBM compatible PC with Windows

98,
NT or ME operating system will be required.
Memory:
The PC must have a minimum of 64 MB of
RAM and Windows

98.
Disk Drives:
The PLGC II configuration program is
provided on CD Rom, and requires at least
20MB on the disk drive. Extra space will be
required to save data.
Serial Port:
To run the PLGC II configuration program a
serial cable is required to connect the
personal computer to the PLGC II. Most
computers will have multiple COM ports to
select from. The COM port selected cannot
be shared with any other device.
DIMAC Gas Chromatogram
SAMPLE INJECTION VALVE
The chromatograph is composed of
a 10-port micro volume diaphragm
valve and generally 2 columns
(some applications may only
require one). The DV22 performs
the injection and the backflush. It is
specifically designed for heavy-duty
applications and is rated at
1,000,000 injections before
requiring service. In two column
systems, column 1 passes all
components except C6+, trapping
C6+ for a quick backflush out.
Column 2 separates the remaining
components. Total analysis time is
about 15 minutes and is run iso-
thermally at 40C. The temperature
of the chromatography enclosure is
controlled to minimize retention
time shifts due to ambient
temperature changes.
PLGC II Sample Handling Equipment
Notes:
1 - Sample loop volume =.025 cu cm
Sample
Vent
Carrier
col 2
col 1
Detector
1
3
10
4
5
6
7
8
9
2
Sample
Vent
Carrier
col 2
col 1
Detector
1
3
10
4
5
6
7
8
9
2
Valve On
Valve Off
pg g
ISO 9001: 2000
Registered QMS
P
L
G
C

I
I


1
0

-

0
3


P
r
i
n
t
e
d

i
n

C
a
n
a
d
a
WWW. GA LVA NI C . C OM E - MA I L : I NF O@ GA LVA NI C . C OM
Toll Free 1-866-252-8470 (US & Canada only)
Galvanic Applied Sciences Inc.
Bay 6, 6325-12 Street S.E.
Calgary, Alberta T2H 2K1
Telephone: 403-252-8470
Fax: 403-255-6287
Calgary, Alberta
1200, 715-5th Avenue S.W.
Calgary, Alberta T2P 2X6
Telephone: 403-252-8470
Fax: 403-571-2684
Edmonton, Alberta
Telephone: 780-430-0142
Fax: 780-430-0172
Houston, Texas
Telephone: 281-477-9002
Fax: 281-477-9004
SPECIFICATIONS
Range:
0-2000 ppm to 100%
for individual components
Components:
16 maximum
Streams:
8 plus calibration
Response:
15 minutes to C6+ (5 minutes optional),
2 minutes nominal for single components
depending on background composition
Linearity:
1% for individual components
Repeatability:
0.5 BTU/1000 0.05%
Temperature Ambient:
10 to 40 C
Humidity:
0-95% non-condensing
Electrical:
24 volts DC @ 50 watts
Class 1 Div 2 Grp B, C, D
Temp code T3
Inputs:
6 dry contact 2-24 VAC
Outputs:
240 x 64 backlit LCD
Relays:
Four user selectable for alarm or fault
(SPDT, 5 amps, 250 VAC)
Analog:
Four isolated 4-20 mA (user selectable)
Serial:
Two RS-232,
One RS-485 modbus
Data logging:
128 days hourly averages
12 months daily averages
1 year monthly averages
Last 64 calibrations
2 user configurations
1 factory configuration
Input A/D:
Three channels 24 bit sigma delta
Real time clock:
Time and date stamp,
1 ms resolution
Sampling Rate:
60 HZ
Chromatograph Valve:
Maximum eight
Columns:
One to four 1/16 stainless steel
(application dependent)
Dimensions:
14 H X 12 W X 8 D
Weight:
35 pounds
Utilities:
UHP helium @ 10 cc/min @ 60 PSI
Sample:
10 to100 PSI clean,
Dry gas @ 95 cc/min
Electrical:
120/240 VAC @ 50 watts
Class 1 Div 1 Grp C,D
Temp Code T3
Dimensions:
29 H X 13 W X 9 D
Weight:
50 pounds
MODIFIED SPECIFICATIONS FOR THE PLGCII EXPLOSION PROOF
Cabinet may vary in size according to the number of streaming of values.
12
16
17
9
PLGCII - DIVISION II
(Optional Cabinet - 28x28x15)
DIMENSIONS
PLGCII - DIVISION I
(Optional Cabinet - 30x36x15)
27
33
10
25
25
pg g