Professional Documents
Culture Documents
In accordance with European Directive 2002/96/EC on Waste Electrical and Electronic Equipment, this symbol
indicates that the product must not be disposed of as unsorted municipal waste, but should be collected separately.
Refer to your local Olympus distributor for return and/or collection systems available in your country.
Warranty
Warranty
The Sonic 134D/DL Ultrasonic Thickness Gauge has been designed and manufactured as
a high quality instrument.
Inspect the unit thoroughly upon receipt for evidence of external or internal damage that
may have occurred during shipment. Notify the carrier making the delivery immediately
of any damage, since the carrier is normally liable for damage in shipment. Preserve
packing materials, waybills, and other shipping documentation in order to establish
damage claims. After notifying the carrier, contact Olympus NDTTM so that we may assist
in the damage claims, and provide replacement equipment, if necessary.
Olympus NDT guarantees the Sonic 134D/DL Ultrasonic Thickness Gauge to be free
from defects in materials and workmanship for a period of one year (twelve months) from
date of shipment. This warranty only covers equipment that has been used in a proper
manner as described in this instruction manual and has not been subjected to excessive
abuse, attempted unauthorized repair, or modification. DURING THIS WARRANTY
PERIOD, Olympus NDT LIABILITY IS STRICTLY LIMITED TO REPAIR OR
REPLACEMENT OF A DEFECTIVE UNIT AT ITS OPTION. Olympus NDT does not
warrant the Sonic 134D/DL Ultrasonic Thickness Gauge to be suitable of intended use, or
fitness for any particular application or purpose. Olympus NDT accepts no liability for
consequential or incidental damages including damage to property and/or personal injury.
In addition to our standard one year warranty, Olympus NDT also offers an optional two
year warranty (call for further details).
This warranty does not include transducers, transducer cables, or battery. The customer
will pay shipping expense to the Olympus NDT plant for warranty repair; Olympus NDT
will pay for the return of the repaired equipment. (For instruments not under warranty, the
customer will pay shipping expenses both ways.)
In this manual, we have attempted to teach the proper operation of the Sonic 134D/DL
Ultrasonic Thickness Gauge consistent with accepted flaw detection techniques. We
believe the procedures and examples given are accurate. However, the information
contained herein is intended solely as a teaching aid and should not be used in any
particular application without independent testing and/or verification by the operator or
the supervisor. Such independent verification of procedures become more important as
the criticality of the application increases.
For these reasons, we make no warranty, expressed or implied, that the techniques,
examples, or procedures described herein are consistent with industry standards nor that
they will meet the requirements of any particular application. Olympus NDT expressly
disclaims all implied warranties of merchantability and of fitness for any particular
application.
Olympus NDT reserves the right to modify all products without incurring the
responsibility for modifying previously manufactured products. Olympus NDT does not
assume any liability for the results of particular installations, as these circumstances are
not within our control.
THE WARRANTIES SET FORTH HEREIN ARE EXCLUSIVE AND ARE IN LIEU
OF ALL OTHER WARRANTIES WHETHER STATUTORY, EXPRESS, OR
IMPLIED (INCLUDING WARRANTIES OF MERCHANTABILITY AND FITNESS
FOR A PARTICULAR PURPOSE, AND WARRANTIES ARISING FROM COURSE
OF DEALING OR USAGE OR TRADE).
PowerLink Introduction
The PowerLink feature enables the Sonic 134D instrument family to recognize
Nortec PowerLink probes and scanners, configuring the instrument according to the
parameters programmed into the ID chip.
Each PowerLink probe is programmed at the factory to identify itself by model number,
pre-selected operating frequency, and serial number.
PowerLink Connection
The PowerLink connection is automatically made when a PowerLink probe is
connected into the Sonic 134D instrument. When the PowerLink probe is connected to
the instrument, the screen displays:
Serial Number
Operating Mode
To proceed without using the PowerLink feature, press the ENTER key followed by the
MAIN key.
To use the PowerLink feature, rotate the SmartKnob clockwise to change SET INST?
from CANCEL to CONFIRM. Pressing the ENTER key will download the information
from the ID chip to the instrument. A new screen displays:
Probe Description
Operating Frequency
Serial Number
Coil Type
Operating Mode
Date
Pre-Stored Programs
After the PowerLink screen appears, you can recall any program that has been saved in
the program storage menu. By rotating the SmartKnob clockwise, the list of pre-stored
programs and the date stored will appear at the bottom of the display.
For every probe, PowerLink can be bypassed to allow the probe to be used for any
applicable test.
PowerLink is an additional capability when used on Eddy Current instruments.
PowerLink Introduction
Note:
Probe operation may be erratic if a probe is being driven at less than 1/10th of
its rated frequency in High Probe Drive. To avoid this problem, do NOT
drive a probe at less than 1/10th of its rated frequency.
Table of Contents
Table of Contents
Warranty
Technical Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
Control Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.1 Instrument Controls . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.2 Power Button . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.3 MODE Key. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.4 +/- Keys . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.5 Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.6 Basic Operations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.7 134D Control Descriptions. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
3.8 134DL Control Descriptions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
19
19
20
20
20
21
22
22
32
39
39
39
39
41
41
42
43
GaugeMaster Software . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.1 Introduction . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.2 Minimum System Requirements . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.3 Installing GaugeMaster . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.4 Starting GaugeMaster. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.5 GaugeMaster Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
7.6 Quick Configuring the Instrument . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .
49
49
50
50
50
51
54
1.1
1.2
Tilt Bail
The Battery Door module is attached in much the same way with a sturdy, but easily
removable screw through the center of the module to the bottom of the Base unit. The
PowerLink transducers use the dual coax system on the top of the instrument. The Tilt
Bail is attached to the back of the instrument, either in the up or down position.
Note:
1.3
For safety, do not attach a lanyard to the Tilt Bail to support the weight of the
instrument.
10
Note:
1.4
Other transducers for special applications are available. Contact your sales
representative.
11
1.5
SONIC 134D
THK
ALARM
MODE
ON/
OFF
12
CAL
GAIN
SONIC 134DL
DATA
FILE
THK
ALARM
CAL
GAIN
MODE
ON/
OFF
1.6
Power Requirements
As a fully portable inspection instrument, the Sonic 134D family of gauges rely on four (4)
AA alkaline batteries as the primary source of power. Four alkaline batteries provide
power up to 100 hours of continuous operation in the Sonic 134D, DL models.
The battery compartment is located on the bottom of the unit and is conveniently accessed
by loosening the quick release screw and removing the battery cover.
Batteries are easily replaced, however, it should be noted that all four batteries need to be
replaced as a group. It is equally important to compare the polarity of the batteries with the
battery door module. If the batteries are installed incorrectly, the Sonic 134D instrument
will not operate.
The low battery icon on the Base Unit LCD will illuminate when the batteries require
changing. When this indicator turns on there is approximately 30 minutes of battery life
remaining and the instrument will automatically shut off in 20 minutes.
13
In order to extend battery life, the Sonic 134D instrument will automatically shut itself off
after approximately 10 minutes of non-use (no key presses or thickness readings). If the
instrument will be off for an extended period of time, remove the batteries to extend their
storage life.
Note:
1.7
Operating Environment
The Sonic 134D/DL is designed as a portable instrument and requires no special site
preparation. When in operation, the instrument should be firmly supported to prevent
damage due to a fall. The unit should be protected from water and chemical spills, and
rapid temperature changes. The unit should be operated away from large electrical
equipment that may interfere with the operation of internal circuitry.
14
2.1
2.1.1
Pulser
Pulser Type Spike Pulser
Pulse Voltage 130 Volts
Damping Fixed 150 Ohms 5%
Measurement Modes Dual, Initial Pulse to First Echo only
2.1.2
Receiver
Receiver Bandwidth 0.3 MHz to 13 MHz
Bandpass Filter Low (0.3 MHz to 4 MHz); High (3 MHz to 13 MHz)
Gain Automatically set through PowerLink, manual +3 dB and -3 dB settings, 40 to 80
dB, accuracy better than 2dB
2.1.3
Timebase
Range 0.020" - 25.00"
Velocity 0.060 to 0.600"/s (1.5 to 15.2 mm/s)
Resolution 0.001" from 0.020" to 9.999" 0.01" from 10.00" to 25.00"
Accuracy Measurement Accuracy (in steel)
0.020-9.999" to 0.002": with custom V-Path programmed transducers.
0.020-9.999" to 0.003", + 1 count: with standard V-Path programmed V-path
transducers.
10.00-25.00" to 0.004", + 1 count
Repetition Rate Varies with thickness and resolution. Up to 1 kHz max.
15
2.1.4
General
Instrument Shutoff Automatic-10 minutes with no usage or manual ON /OFF
Measurement Selectable via GaugeMaster -hold last reading or blank display on
uncoupled
Alarm Limits Operator adjustable high and low set points with visual and operator
controlled audible alarm
Differential Mode Difference between operator entered value and actual value with
visual and operator controlled audible alarms.
Display Update Rate 4 Hz (20 Hz in fast scan mode)
Thickness 4- Digit LCD
Display Character Height - 0.5", LED Backlight with Manual ON/OFF
Power Supply 4 each 1.5 V AA Alkaline cells
Battery Life 100 hours with backlight on 1/3 of the time typical at 75F.
Low Battery Front panel display, an icon indicates less than 1/2 hour operating time
remaining.
Dimensions 6.5" L x 3.1" W x 1.6" D
Weight 13.5 oz with batteries
Operating Temp +10F to 120F (-10C to 50C)
16
2.2
17
18
Control Descriptions
Control Descriptions
Whats in this section?
3.1 Instrument Controls
3.2 Power Button
3.3 MODE Key
3.4 +/- Keys
3.5 Display
3.6 Basic Operations
3.7 134D Control Descriptions
3.8 134DL Control Descriptions
3.1
Instrument Controls
This section covers the operation al controls of the Sonic 134D/DL.
SONIC 134DL
DATA
FILE
THK
LOG
CAL
Thickness
Display
LOG Key
ALARM
MODE Key
GAIN
MODE
+/- Keys
POWER Key
ON/
OFF
Battery
Compartment
19
To attach probe:
Align pins on probe
plug with receptacle
on 134D/DL, and
push straight in.
(do NOT force!)
SONIC 134D
THK
ALARM
CAL
SONIC 134D
THK
GAIN
ALARM
CAL
GAIN
3.2
Power Button
The Power button toggles power ON/OFF. Immediately after power is turned ON,
initialization begins and the instrument determines configuration, performs power-up
diagnostics and calibration, and readies itself for operation.
3.3
MODE Key
The MODE key toggles the instrument menus and selects parameters for editing,
depending on the test.
3.4
+/- Keys
The +/- keys are used adjust parameters and select test types.
20
Control Descriptions
3.5
Display
3.2Power Button
The Power button toggles instrument power on and off. Immediately after power is turned ON,
an initialization routine is started. The instrument microprocessor determines the instrument
configuration, performs power up diagnostics and calibration, and readies the instrument for
operation.
3.3MODE Key
The MODE key toggles the instrument menus and selects parameters for editing, depending on
the particular test.
3.4+/- Keys
The +/- keys are used to adjust parameters and select test types.
3.5Display
Immediately after power is turned on, the LCD display will turn on the entire display to
allow verification of the alphanumeric and iconic segments.
After a few moments, the LCD display will change to display a numeric FPGA version
code. At this point, pressing the ( + ) and ( - ) keys simultaneously will perform an
instrument reset to factory default parameters. Following the FPGA code is the firmware
version.
Note:
Placing the LCD display in direct sunlight results in temporary fade due to
the elevation of internal liquid temperature. Shade the display to avoid this.
21
3.6
Basic Operations
The Base Unit ON/OFF key is used to turn the gauge on and off.
With a transducer connected, the Sonic 134D instrument will display the transducer
PowerLink ID number and perform an auto zero.
Then, the instrument changes to the thickness measurement mode that is defined by the
flashing indicator immediately below the THK text.
From the fully ON display notice that there are several alphabetic and three (3) iconic
symbols around the edge of the numeric display. These represent different modes or
features that the instrument is capable of performing. Access these modes by using the
MODE key.
Pressing the MODE key once will step one feature at a time through the backlight, high
calibration, low calibration, velocity calibration, and inch/mm selections. Use the ( + / - )
keys to make adjustments to the selected feature.
To access the other menu functions (high/low alarm, etc.), the Sonic 134D/DL must be
turned OFF and while holding the MODE key down, turned back ON (the MODE key
only needs to be pressed until the software revision has been displayed). This will allow
the Sonic134D/DL to access all of the control functions when the MODE button is
pressed. The next time the Sonic 134D/DL is turned OFF and then back ON, the MODE
key will access the short list of menu items described above.
Indicators of features that are adjusted to be ON will remain ON when the instrument is
returned to the Thickness Measurement mode. This provides the status of the instrument
with just a quick glance at the numeric display.
A limited number of the features provide extra information or adjustments when the LOG
key is pressed. But, the typical use for the LOG key is to store thickness readings in the
data logger on the Sonic 134DL. The following section describes all the features of the
Sonic 134D and how these features are used and adjusted.
3.7
22
Control Descriptions
Inch/millimeter selection
Tri-level gain adjustment
Minimum capture mode
Differential mode
Adjustable high/low alarms
The Sonic134D Thickness gauge is very flexible and can be factory upgraded to a data
logger (134DL).
3.7.1
3.7.2
Low Battery
The nearly empty battery icon is the low battery indicator. This indicator displays
automatically when the instrument has approx. 30 min. of power remaining. In order to
23
safeguard the instrument memory, the Sonic 134D will automatically shutoff 20 min. after
the low battery indicator displays. It is recommended that the batteries be replaced as soon
as possible when the low battery indicator displays.
3.7.3
Backlight
The Backlight status is indicated by the lightbulb icon on the LCD display. When the
backlight has been selected with the MODE key, the numeric display will show either ON/
OFF, or AUTO. The ( + ) and ( - ) keys change the setting. When set to ON, the backlight
is ON continuously, indicated by the lightbulb staying on when in the THK mode. When
set to AUTO, the backlight comes ON only when the instrument detects that it is coupled.
This is indicated by the lightbulb being on continuously and AUTO is indicated when in
this adjustment mode.
3.7.4
HLV - CALibration
The Sonic 134D allows three modes of thickness measurement calibration: single point,
two point, and velocity. We recommend that single point be used for steel only.
24
Control Descriptions
Note:
3.7.5
25
3.7.6
26
Control Descriptions
3.7.7
3.7.8
Units (in/mm)
27
To change the present setting, press the ( + ) key for inches or the ( - ) key for millimeters.
To adjust for another feature, simply press the MODE key to move on. To return to the
Thickness Measurement mode either wait 20-30 seconds for the instrument to
automatically return to the THK mode or use the MODE key to continue around the
display until the THK indicator is reached.
3.7.9
Gain (HML)
28
When using the J3L (High Temp) transducer, the H, M, and L settings can
also be considered High (600 - 1000F), Medium (to 300 - 600F), and Low
(to 300F) temperature settings.
Control Descriptions
3.7.10
3.7.11
29
Differential Measurement mode allows the operator to enter a given thickness and the
instrument will display the difference between the preset thickness and the measured
thickness.
The operator selects the Differential Measurement mode by pressing the MODE key until
the DIF indicator is displayed. The numeric display will show either OFF or ON to
indicate the current state of the feature. Using the ( + / - ) keys, the DIF mode is turned
either OFF or ON as desired.
To change the preset thickness, press the LOG key. The numeric display will now show
the current preset thickness. The ( + / - ) keys are used to adjust the current preset
thickness. The LOG key may be used to return to the DIF state or the MODE key may be
used to move on to other features.
To return to the Thickness Measurement mode either wait for the instrument to
automatically return to the THK mode or use the MODE key to continue around the
display until the THK indicator is reached.
A few of the other features on the Sonic 134D and Sonic 134DL are affected when the
instrument is in Differential Measurement mode. While DIF is ON, both the High and
Low Alarm set points are converted to differential measurement values.
3.7.12
Alarms (HI/LO)
30
Control Descriptions
3.7.13
3.7.14
31
To change the High Alarm Set Point, press the LOG key. The numeric display shows the
current high alarm set point, and the ( + / - ) keys are used to adjust the high alarm set point.
The LOG key may be used to return to the HI state or the MODE key may be used to move
on to other features. To return to the Thickness Measurement mode either wait for the
instrument to automatically return to the THK mode or use the MODE key to continue
around the display until the THK indicator is reached.
3.8
If the transducer is uncoupled when the log key is pressed, the location is
marked as obstructed (obr).
After logging a thickness reading into a location, the location number is automatically
incremented and momentarily displayed. The gauge is immediately ready to log again.
Note:
The following sections contain the information required to understand the operation of the
data logging features of the Sonic 134DL including a detailed description of all controls
available.
3.8.1
File (#EPS)
The File features of the Sonic134DL allow the user to control and adjust the parameters
relating to files that are in the data logger. These file features consist of File Set up,
Change File Number, Erase File, and Print File.
32
Control Descriptions
3.8.2
33
3.8.3
3.8.4
Erase File ( E )
At times, erasing the thickness readings in a file will be necessary, so that, new thickness
readings can be taken. Occasionally, erasing the data in all the files will be necessary,
either to clean the slate or to allow the file sizes to be altered.
34
Control Descriptions
The + in front of the numeric display will be displayed if the file contains one or more
thickness readings.
To erase all the thickness readings in this file, press the LOG key. To erase the data in all
the files use the ( + or - ) keys to adjust the file number to F001, and then press the ( - ) key
once.
The numeric display will show ALL indicating that all files will be erased when the LOG
key is pressed. When the desired file or files have been erased, the MODE key may be
used to move on to other features.
To return to the Thickness Measurement mode, either wait for the instrument to
automatically return to the THK mode or use the MODE key to continue around the
display until the THK indicator is reached.
3.8.5
Print File ( P )
The Print File mode is used to send file information either to a serial printer or to a
personal computer using the RS-232 connector on the back of the instrument.
35
To print a file, press the MODE key until P is displayed under the FILE text. The
Measurement Display will show the current file number for printing, such as, F121 on the
display. A + on the display to the left of the file number indicates that there is at least one
reading in that file. Press the LOG key at this point to print the indicated file.
A different file number is selected by pressing the ( + / - ) keys until the desired file number
has been reached. When the desired file or files have been printed, the MODE key may be
used to move on to other features. To return to the thickness measurement mode either
wait for the instrument to automatically return to the THK mode or use the MODE key to
continue around the display until the THK indicator is reached.
3.8.6
Data ( #RE )
The Data feature of the Sonic 134DL allows the user to control and adjust the parameters
relating to individual data locations in the data logger. The three data features are
independent of each other.
3.8.7
36
Control Descriptions
indicates that the location is already full. When the desired location has been set, the
MODE key may be used to move on to other features.
To return to the Thickness Measurement mode either wait for the instrument to
automatically return to the THK mode or use the MODE key to continue around the
display until the THK indicator is reached. Pressing the LOG key will now store a
thickness reading at the selected location and automatically increment to the next location.
3.8.8
Data Recall ( R )
Data Recall allows the operator to examine the thickness reading stored in any location of
the current data storage file (see Section Change File Number to change files).
3.8.9
Data Erase ( E )
Data Erase allows the operator to erase the thickness reading stored in any location of the
current data storage file (see Section Change File Number to change files).
37
38
4.1
Surfaces
Surface patterns such as machining, extrusion marks, or casting irregularities can cause
false readings. Some of these readings can be corrected by rotating the transducer head or
applying sufficient couplant, however, it may be necessary to smooth the surface further
to eliminate the ill effects on the measurements.
Curved surfaces can also pose difficulties. When measuring these types of surfaces be
sure to use adequate couplant, keep the transducer stable and at right angles to the surface
of the part under test. Typically, a smaller diameter transducer will minimize stability
problems, but developing a steady technique is very important.
4.2
Materials
The material under test can also cause variations in the readings. The presence of material
grain noise or a defect in the material, such as slag pockets, inclusions, voids or cracks may
cause a false thickness reading.
4.3
Elevated Temperatures
The temperature of the material under test can also influence the readings on thickness
gauges. Elevated temperatures typically cause two main effects: 1) material expansion,
and 2) signal attenuation. Material expansion will cause the reading to increase as the
temperature is increased. For maximum accuracy at high temperatures, calibration should
be performed at the same temperature that measurements will be made.
Signal attenuation can be compensated for by increasing the gain used during testing. Be
sure that the temperature of the part under test does not exceed the temperature rating of
either the transducer or couplant being used. Make measurements quickly and allow the
transducer to cool between readings, as continuous exposure to high temperatures will
degrade and eventually destroy the transducer.
39
40
5.1
Introduction
Although the Sonic 134D family of gauges are designed as portable, hand held devices,
they also have the capability to be controlled or monitored by an external computer or
terminal. The RS-232 communication port allows an external device to send commands
to or obtain information from the instrument. With the exception of a few controls, all
operations available from the front panel are also available from the RS-232 interface.
Serial communication is performed using an ASCII command string. In general, the
command string consists of a three character command code, a command operator, a
command value, and a carriage return (Enter).
The command code is a three character code that identifies a unique instrument parameter
(see Section Command Code List). The command operator is a one character code that
indicates whether the command in question is a read command ?, a write command = or
an execute command Enter.
The command value is a sequence of characters, either numeric or alphabetic that has
meaning for the specific command code. The carriage return or Enter is used to signify that
the command is ready to process. The Sonic 134D family of gauges will not process a
command string until the carriage return or Enter has been received. The 134D family also
checks each incoming command string for errors. If an error is encountered an error
prompt is returned to the external device.
Two types of error prompts exist, one for invalid command strings (C?) and one for invalid
command values (V?). For example, suppose you wish to set the gain 4 dB higher than the
current setting. The command string sequence might read:
41
Command String
Instrument Response
Description
gain?
C?
gan?
63
Instrument response to
read
gan=95
V?
Invalid Command,
Value=80 maximum
gan=67
>
Table 1
5.2
Baud Rate
Word Size
Parity
None
Stop Bits
Table 2
Configuration
The RS-232 serial interface communicates with either a serial interface printer or an
externally connected computer. The 9-pin D sub-miniature male connector connects to an
industry standard serial port.
42
9-pin D Sub-miniature
Male Connector
Function
Signal Name
pin 2
output
TXD
pin 3
input
RXD
pin 4
input
CTS
pin 5
ground
GND
pin 8
output
RTS
5.3
Function
RWX
Code
Range
Description
Alarm High
State
RW
AHS
OFF, ON
Alarm High
Value
RW
AHV
AlarmLoVal
< value <
maxRange
Alarm Low
State
RW
ALS
Off, On
Alarm Low
Value
RW
ALV
minRange
Backlight
RW
LIT
OFF, ON
Battery status R
BAT
OK, LOW
Serial Baud
Rate
RW
BPS
1200,2400,
Cal Hi Value
RW
CLH
Cal Lo Value RW
CLL
minRange < +
value
Calibration
Mode
RW
CLM
1-pt 2-pt
Function
RWX
Code
Range
Description
Differential
mode
RW
DFS
OFF, ON
43
Differential
value
RW
DFV
Receiver filter RW
FLT
LOW HIGH
Gain setting
RW
GAN
File Erase
RW
FER
Max file
number
RW
FMN
File number
RW
FNM
Print file
number
RW
FPR
Receiver filter RW
FLT
LOW HIGH
Gain setting
RW
GAN
GND
1 - 20
Function
RWX
Code
Range
Description
Header state
RW
HDS
OFF ON
44
Horn
RW
HRN
OFF ON
Language
RW
LAN
ENG, FRCH,
Sets language of reports.
ESP, DEU, ITAL
Parameter lock.RW
LOK
OFF ON
Maximum gainRW
MAG
minGain+2 <=
value <= 80
Maximum
Range
RW
MAR
Measure hold RW
MHD
Minimum
Acquire
RW
ACQ
OFF, 0 - 15
(integer)
Minimum GainRW
MIG
Minimum
Range
RW
MIR
Printer
RW
PRT
45
Function
RWX
Code
Range
Description
Resolution
RW
RES
OFF ON
Statistics State RW
STS
OFF ON
Units
RW
UNT
IN MM
Velocity
RW
VEL
Version
VER
46
Error Codes
Error Codes
The Sonic 134D family of gauges has several built-in self check procedures that can
generate error codes to be displayed on the numeric display. In the event that you should
see such a code, consult the following error code list and take appropriate action.
E000
Cause: Data location is full.
Correction: Erase the data storage location or select a new data storage location.
E001
Cause: Data file is full.
Correction: Erase the data storage file or select another data storage file.
E002
Cause: Error in the base memory.
Correction: Restart the instrument. If error persists, return to service center for repair.
E004
Cause: Printer error.
Correction: Be sure the printer is connected, turned on, and on line.
E005
Cause: Invalid data from the program key.
Correction: Reconnect the program key and try again. If error persists, return to service
center for repair.
E006
Cause: Could not auto zero the transducer.
Correction: Be sure the transducer is not coupled and the lens does not have couplant on it.
E007
Cause: Power down due to inactivity or low battery time out.
Correction: Turn the instrument back on and continue operation.
E008
Cause: Power down requested by the user.
E009
Cause: Power down caused by inappropriate configuration.
Correction: Be sure the top bumper is correctly installed.
E010
Cause: Previous instrument configuration could not be recovered.
47
Correction: Reconfigure the instrument as needed. If the error persists, return to service
center for repair.
1.
Cause: Measurement detected outside of transducer specified range.
Correction: Select another transducer appropriate for thickness range under inspection.
48
GaugeMaster Software
GaugeMaster Software
Whats in this section?
7.1 Introduction
7.2 Minimum System Requirements
7.3 Installing GaugeMaster
7.4 Starting GaugeMaster
7.5 GaugeMaster Operation
7.6 Quick Configuring the System
7.1
Introduction
The GaugeMaster software application is designed to assist in personal computer (PC)based acquisition on the Sonic 134DL thickness gauge. GaugeMaster is designed to
operate under Windows 98, ME, NT, 2000 and XP operating systems.
The GaugeMaster software will enable the users to achieve the following:
7.2
IBM-compatible computer
49
7.3
Mouse
Installing GaugeMaster
To install GaugeMaster, perform the following steps:
7.4
Starting GaugeMaster
The GaugeMaster application will typically be used with the Sonic134DL Thickness
Gauge, since the primary purpose of the application is to upload data (instrument settings,
thickness, programs, reports) from the instrument and download them from the computer.
However, there may be circumstances where the user may wish to review the ultrasonic
data stored in the computer without connecting to the instruments. The application has
also been designed to operate under these special circumstances.
For operation with the Sonic 134DL, Figure 7-1 shows how the instrument must be
connected to the host PC. A straight-through RS-232 (9 to 9, or 9 to 25 pin) cable is
50
GaugeMaster Software
required for this connection, details of which are described in Section 6. Do not use a Null
Modem cable.
COMPUTER
KBD
MOUSE
USB
PORTS
PRINTER
SERIAL
Back of
Instrument
VGA
RS-232
7.5
GaugeMaster Operation
This section covers how to upload thickness data in blocks or files from the Sonic 134DL
instrument into GaugeMaster.
In addition, it shows how to:
Apply thresholds
51
7.5.1
Basic Concepts
Upon installation, the GaugeMaster application creates two sub-directories within the
Main directory (c:\gaugemstr).
7.5.2
Files: the default directory for storing thickness data uploaded from the Sonic DL
instrument.
Main Window
To start the program, double-click on the GaugeMaster icon in the Gagemstr folder. The
GaugeMaster application contains four Main Menu fields:
7.5.3
File: This menu is used to setup the default printer and to load standard thickness data
from disk or from the Sonic 134 instrument.
Configure: This menu is used to configure the COM port settings or to Quick
Configure the instrument settings.
7.5.4
52
GaugeMaster Software
7.5.5
7.5.6
7.5.7
7.5.8
7.5.9
The first time FROM DISK is selected, the root directory is displayed. Select
c:\gaugemstr\files and then select the appropriate file.
53
7.6
54