Professional Documents
Culture Documents
AQUATIC
Users Manual
AquaCiser III
Disclaimer
Proprietary Notice
Retain this manual for future reference. Include it with the product
in the event of transfer to new users. Additional free copies are
available upon request from Customer Relations.
234-3280
234-3278
Ratings Plate
AquaCiser III
TABLE OF CONTENTS
Section Page
Section Page
Safety Information
AquaCiser III
1 - SAFETY INFORMATION
1.1 Danger
Danger notices indicate an immediate hazardous situation
which, if not avoided, will result in injury or death.
DANGER
Electric shock can cause death or serious injury. Install,
operate, and maintain the AquaCiser only as described in
this manual.
230/240 VAC
Hazardous voltage inside. Follow
procedure before servicing.
WARNING
Improper water conditioning can cause injury and
damage the AquaCiser. Maintain the water condition as
directed in this manual and keep a Water-Condition Log.
Improper parts and service can cause injury. Use only
Hudson parts and Hudson-approved service on the
AquaCiser.
Modifying the AquaCiser can cause injury and damage.
lockout/tagout
1.3 Important
Important notices emphasize important usage or maintenance
information.
Important
1.2 Warning
Warning notices indicate a potentially hazardous situation
which, if not avoided, could result in injury or death.
WARNING
Untrained operators can cause injury or be injured. Permit
only trained personnel to operate the AquaCiser.
Certain medical conditions can place a patient at risk. Require
all patients to have a physicians order for treatment.
Improper use of the AquaCiser can cause injury. Use the
AquaCiser only for the purpose described in this manual.
Contaminated water can spread disease. Discard
contaminated water and clean and disinfect the AquaCiser.
Improper operation of the AquaCiser can cause injury.
Operate the AquaCiser only as described in this manual.
An unattended patient can be injured. Stay with the
patient at all times.
Hot water can cause injury. Water temperature may
increase during AquaCiser use. Monitor the water
temperature and keep the temperature below 105F
(40.5 C).
Clothing, footwear and medical accessories can come loose
and cause injury. Operators are responsible for determining
the appropriateness of items worn in the AquaCiser.
Overflowing the reservoir can cause injury due to electric
shock, and can damage the AquaCiser. Never leave the
AquaCiser unattended while filling the reservoir.
Improper maintenance can cause injury. Maintain the
AquaCiser only as described in this manual and the
Service Manual provided.
Improper handling or use of water-treatment chemicals
can cause injury. Follow the chemical manufacturers
directions for use.
1.5 Recycling
The battery inside the AquaCiser contains lead, which must
be disposed of properly through recycling.
U.S.A.
In the United States of America, recycling is regulated by
the Environmental Protection Agency (EPA). Contact your
state EPA for recycling information in your area.
WEEE DIRECTIVE (E.U.)
European Union directive 2006/66/EC Waste
Electrical and Electronic Equipment (WEEE)
limits the disposal of electrical and electronic
equipment in landfills. The AquaCiser battery
must be separated from regular municipal
waste and disposed of properly according to
this standard.
Hudson Aquatic Systems, LLC, Inc. 234-3253-02 July 2009
AquaCiser III
Safety Information
Electrical Shock
Warning (9)
(Location: electrical
components)
Restart Label
(Location: top of
command console)
Moving Parts
Warning
(Location: treadmill
access panel)
Safety Information
AquaCiser III
General Warning of
Potential Injury
Emergency Stop
Emergency Drain
Operation Requires
One Trained Operator
Wear Gloves
Wear Eye
Protection
Wear
Dust Mask
Wear
Protective Apron
Main Power
Breaker Switch
WEEE Directive:
Dispose of Battery Properly
250 lb
113 kg
Belt Pinch Point:
Keep Hands Clear
Power ON/
RESTART
Protective Earth
(Ground)
Lockout/Tagout
Before Performing
Service
ON
OFF
Unlocked
Locked
Seat Load
Limit (Dry)
AquaCiser III
2.2 Training
Operator trainees need to:
WARNING
Untrained operators can cause injury or be injured.
Permit only trained personnel to operate the
AquaCiser.
Patient Evaluation
AquaCiser III
3 - PATIENT EVALUATION
3.1 Require Each Patient
To Be Evaluated
Require all patients to have a physician evaluate their
medical condition and order treatment before permitting
them to begin therapy in the AquaCiser.
The patient populations listed in Major Indications chart
may benefit from therapeutic exercise in the AquaCiser.
Patients with certain medical, mental or physical conditions
should not use the AquaCiser. A patient experiencing any of
the Major Contraindications should not use the AquaCiser.
Also see the Bloodborne Disease Notice on page 4.
WARNING
Certain medical conditions can place a patient
at risk. Require all patients to have a physicians
order for treatment.
Major Indications
(Potential Patient Populations)
ORTHOPEDIC
Gait analysis
SPORTS MEDICINE
Gait analysis
General conditioning
General conditioning
General weakness
Loss of endurance
Major Contraindications
(Should NOT Use AquaCiser)
Light-headedness
Fever
Non-ambulatory
Incontinence
Open wounds
History of seizures
Urinary infections
Senility or confusion
Infectious disease
AquaCiser III
4.2 Certification
WARNING
Improper use of the AquaCiser can cause injury. Use
the AquaCiser only for the purpose described in this
manual.
AQUACISER III RATINGS PLATE
The ratings plate (below) affixed to the AquaCiser provides
specific power consumption information. The plate affixed
to your AquaCiser will have the serial number in the
space indicated S/N and a punch-mark indicating that it
operates using 50 Hz or 60 Hz electricity.
DIMENSIONS
WEIGHT
Therapy pool
Midsection
Reservoir
Steps
Total weight (without water)
Total weight (with water)
ADDITIONAL SPECIFICATIONS
105F (40.5C)
0-9.9 mph (0-15.9 kph)
365 gal (1382 L)
48 in (122 cm)
3-4 minutes
3-4 minutes
approx. 20 minutes
AC MOTOR SPECIFICATIONS*
ADDITIONAL SPECIFICATIONS
AquaCiser III
4.4 Components
CONTROLS BOX
Command
Console
Emergency Drain
Buttons
Door Lever
Electrical Panel Access Door
Lint Basket Access Panel
Plumbing Access Panel
THERAPY POOL
WATER RESERVOIR
Handrails
(2 outside, 4 inside)
Chlorine Dispenser
Access Panel
CENTER
PLATFORM
Power Indicator
Touch
Screen
Handrails (4)
(2 each end)
Resting Seat
Restart Button
Emergency Stop
(side of console)
Lamps (4)
(2 each end)
Command Console
Therapy Jets (12)
(6 each end)
Therapy Pool (Looking Forward)
10
AquaCiser III
4.4.1 Reservoir
The water reservoir (see Components, page 10) holds and filters the water that is used in the therapy pool. Pumps move the
water back and forth between the reservoir, filter, and the therapy pool. The flow of water is controlled from the command
console and the EMERGENCY DRAIN buttons.
Emergency
Stop Symbol
Push EMERGENCY STOP to cut the electric power to the pump and
treadmill. Pressing the button will activate an alarm that sounds until the
RESTART button is pressed. Any water in the therapy pool will remain
in the chamber.
Restart
Symbol
The AquaCiser will not operate after the EMERGENCY STOP button
has been pressed. To reset the system and restore power, twist the
EMERGENCY STOP button clockwise to release the button lock, then
press the green RESTART button on the command console (See Restart
Button below).
RESTART BUTTON
The green RESTART button is located on the command console (Figure
2). To reset the system after the EMERGENCY STOP button has been
pressed, first twist the EMERGENCY STOP button clockwise to release
the button lock, then press the green RESTART button to reset the
computer and allow the AquaCiser to be used.
ELECTRICAL PANEL
Main Power/
Breaker
Symbol
The white, electrical panel access door (see Components, page 10)
allows access to the electrical panel (Figure 3). The Main Power breaker
switch can be accessed without opening the panel (See Main Power
Switch, below). The electrical panel itself has an electrical interlock
feature, which requires power to the AquaCiser to be shut off before it
can be opened. Only a qualified electrician should open the electrical
panel. See the Service Manual for access instructions.
MAIN POWER SWITCH
The Main Power breaker switch controls electrical power to the
AquaCiser. The switch is located inside the electrical panel but can be
accessed from a plastic enclosure (Figure 3A).
Turn the main power switch OFF before opening the lint basket,
performing maintenance, or making repairs to the AquaCiser. Leave the
main power switch ON at all other times.
Figure 3A - Main
Power Switch
(GFCI Breaker)
11
AquaCiser III
Control Knob
Figure 4 Chlorine
Dispenser
Rocker Switch
Important
Flooding can occur if the therapy pool door is not
securely fastened. Always close and secure the door by
pulling the lever down after the patient enters or exits
the therapy pool.
LAMPS
The AquaCiser has one light and a rotating color wheel.
Fiber-optic lines direct the colored light to four lamps in the
therapy chamber. (Two are visible in Figure 6, page 13). Use
the command console to turn the lamps ON and OFF (see
Components, page 10). When ON, the lights change color
periodically.
12
Closed
Position
AquaCiser III
JETS
The AquaCiser has two types of water jet: therapy and turbo
(Figure 6). The jets are controlled from the command console (see
Components, page 10). Only one type of jet can be used at a time.
Turning one type of jet ON automatically turns the other type of
jets OFF. All jets can be aimed as needed.
Note:
The filter operates when the jets are used.
Therapy Jets (12 total, 6 each end) - Therapy jets inject a mixture
of water and air into the therapy pool. Six jets are installed at each
end of the therapy chamber. They provide resistance and massaging
pressure for the patient from multiple angles. The water level must
be above the jets for the jets to be used.
Sensors in the therapy pool read the water level and turn on only
the jets that are below water level. For example, to use only the
lowest therapy jets (2 front and 2 rear), stop filling the therapy pool
when the water level covers the lowest jets.
Turbo Jets (2) - The turbo jets are located at the front (command
console end) of the therapy pool. They forcefully push streams of
water (without air) through the therapy pool and provide strong
pressure or resistance for the patient, from one direction. The water
level must be above the turbo jets for use.
Examples: If the patient is walking or jogging while facing the
therapy chamber doors, the turbo jets provide assistance by pushing
water toward the back of the patients legs. If the patient is facing
the command console end, the turbo jets provide resistance by
pushing water against the front of the legs. Follow your therapy
protocols for appropriate use of the turbo jets.
RESTING SEAT
250 lb
113 kg
The clear acrylic plastic resting seat is located at the front (command
console end) of the therapy pool. Lower the seat to provide a place
for the patient to sit if he/she tires during the therapy session.
Also use the seat to allow the patient to sit and perform stretching
exercises with or without therapy jets in use.
Pull the seat down to lower it for use. Push the seat up and into
its locking clamps (Figure 7) to store it. The load limit of the seat
when dry (not under water) is 250 pounds (113 kg).
Seat Load
Limit (Dry)
Emergency
Drain Symbol
13
AquaCiser III
14
AquaCiser III
OPERATOR ASSISTANCE
TREADMILL FUNCTIONS
MAINTENANCE FUNCTIONS
15
AquaCiser III
Important
After pressing the FILL or DRAIN TO RESERVOIR buttons,
there is a normal pause of 2-5 seconds before the
pump begins moving water between the reservoir and
therapy pool. Be patient and allow time for the pump
to begin operating.
Note
The temperature can be displayed in either degrees
Fahrenheit or degrees Celsius (Figure 10). To change the
units displayed, see Maintenance Functions, pages 28-29.
Display
in Degrees Fahrenheit
or Degrees Celsius
16
AquaCiser III
SETTING THE WATER TEMPERATURE
To set the water temperature, press SETPOINT in the Water
Temperature area of the screen. This will bring up a small
numeric keypad screen.
Enter the desired temperature on the
keypad and press the enter key (an arrow
pointing down and to the left).
Enter Symbol
The command console software will not
accept a Setpoint above 105F (40.5 C). An alarm will sound
and the AquaCiser will stop if the actual temperature exceeds
106 F (41.1 C).
Changing the Setpoint does not, by itself, change the water
temperature. To warm the water, use either of the following
methods to move water over the heater:
Important
After pressing the FILL or DRAIN TO RESERVOIR buttons,
there is a normal pause of 2-5 seconds before the
pump begins moving water between the reservoir and
therapy pool. Be patient and allow time for the pump
to begin operating.
FILLING AND DRAINING THE THERAPY POOL
Important
Flooding can occur if the therapy pool door is not
fastened. Always close and secure the door by pulling
the lever down after the patient enters or exits the
therapy pool.
AquaCiser III
Important
Move all water to the reservoir and press FILTER
ENABLE whenever the AquaCiser will be idle. The filter
will run 10 minutes every 8 hours to cycle the water
(minimum), or it will turn on as needed to keep the
water at the Setpoint temperature.
18
AquaCiser III
WARNING
Contaminated water can spread disease. Discard
contaminated water and clean and disinfect the
AquaCiser.
DISCARDING WATER
If there is water in the therapy pool and the water level
is higher than the first sensor above the treadmill, press
DRAIN DISCARD once to discard only the water in the
therapy pool to the waste drain. Press DRAIN DISCARD a
second time to drain all water from the AquaCiser reservoir
and therapy pool.
DRAIN DISCARD
VIA EMERGENCY DRAIN BUTTON
If the electrical power to the AquaCiser is interrupted
(for example, during a power outage), press either
EMERGENCY DRAIN button (Figure 15) to discard the
water from the therapy pool to the waste drain.
This allows a patient to exit the therapy pool when a power
outage occurs.
Note: If the AquaCiser has electrical power, pressing
either EMERGENCY DRAIN button stops the treadmill
and activates the pump to move the water from the therapy
pool to the reservoir. The EMERGENCY DRAIN button only
discards water to the waste drain when there is no electrical
power to the AquaCiser.
19
AquaCiser III
Note
The water temperature is displayed on the Treadmill
Functions screen but cannot be adjusted here. To adjust
the water temperature, go to the Therapy Pool Functions
screen. See Setting the Water Temperature, page 17.
The temperature is displayed in either degrees Fahrenheit
or degrees Celsius (Figure 16). To change the units
displayed, see the subsection, Units Selection under
the Maintenance Function Screen, page 29.
Display in
Degrees Fahrenheit
or Degrees Celsius
Display in
Miles or Kilometers
Display in
Miles per Hour or
Kilometers per Hour
20
AquaCiser III
TREADMILL DIRECTION
The treadmill runs in either direction. If the patients mobility is
limited, the reversible treadmill can be used to have the patient
walk forward and backward during the same therapy session
without requiring the patient to turn around.
To set the treadmill direction:
Press FORWARD to set the treadmill to move toward the
therapy pool door. This is the appropriate direction for a
patient to walk or jog while facing the Command Console
end of the AquaCiser (Figure 17).
Press REVERSE to set the treadmill to move toward the
Command Console (Figure 17). This is the appropriate
direction for a patient to walk or jog while facing the therapy
pool door (Figure 18).
If you press FORWARD or REVERSE while the treadmill
belt is moving, the treadmill will slow to a stop and then
begin moving in the new direction.
The therapist controlling the AquaCiser should face the
patient in the proper direction before starting the therapy
session. Tell the patient when the treadmill is to start, and in
which direction it will move. The patient can face either end
of the therapy pool as desired (Figures 17 and 18).
The two turbo jets are located at the command console-end
of the AquaCiser.
To have the turbo jets provide resistance for the patient,
the patient should face the jets (toward the command
console end of the AquaCiser, as in Figure 17).
To have the turbo jets provide pressure on the back of the
patients legs, face the patient toward the therapy pool
door during the therapy session, as shown in Figure 18.
USING THE TREADMILL AND ADJUSTING SPEED
The Treadmill Speed area displays the speed you have set for the
treadmill. The treadmill always begins at 0.1 miles per hour (or 0.1
kilometers per hour) and will ramp up to the speed that you have set.
Press START to start the treadmill.
Press the INC (increase) and DEC (decrease) buttons
repeatedly to increase or decrease the treadmill speed. When
the treadmill has adjusted to the new speed you have set, a
yellow AT SPEED indicator will light.
Press the red STOP button to stop the treadmill. The treadmill
will ramp down in speed until it stops.
Treadmill Direction:
REVERSE
Treadmill Direction:
FORWARD
Figure 17 - Reversible Treadmill
Important
Communicate with the patient at all times. Tell
the patient when the treadmill will start or stop,
when the treadmill speed will change, or when
the jets will be activated or stopped.
21
AquaCiser III
Three options are available on the Personal Program Menu
screen (Figure 19):
AquaCiser III
PERSONAL PROGRAM GUIDELINES
23
AquaCiser III
TREADMILL FUNCTIONS
Important
Communicate with the patient at all times. Tell the
patient when the treadmill will start or stop, when the
treadmill speed will change, or when the jets will be
activated or stopped.
AquaCiser III
5.7 Fault Indication Notices
Alarm Indicator
25
AquaCiser III
Note: You can also set the date and time from this screen.
Important
Several hours may be needed to warm the water to
the Setpoint after the AquaCiser has been idle.
Important
An operator must be present to turn the filter OFF after
an automatic heat start-up begins.
26
AquaCiser III
Important
Before setting an Automatic Heat Start-Up, verify that
the Temperature Setpoint is the desired temperature.
To set the date and time for the AquaCiser, press the SET
SYSTEM TIME AND DATE button on the Heat Start-Up
Settings Menu screen (Figure 26, page 26).
On the Date and Time Setup screen (Figure 28), enter the
information for:
When you have entered the date and time, press SET CLOCK.
27
AquaCiser III
28
AquaCiser III
PLC INPUTS - PLC OUTPUTS STATUS
Press the PLC Inputs or PLC Outputs buttons to display
screens that report the status of the Programmable Logic
Controls. Each logic control is preceded by an indicator
that will be either yellow or black, depending on the status
of the control (Figure 30).
Which indicators are yellow or black at a given time is
dependent on if there is a fault in the system or if the lights,
treadmill, jets, pumps, or other AquaCiser components
are running. See the Service Manual provided for detailed
information regarding the logic controls and their status.
TEMPERATURE CALIBRATION OFFSET
The Temperature Calibration Offset is a programmed
setting that matches the AquaCisers thermocouple to a
thermometer. The calibration is set by Hudson when the
AquaCiser is constructed, and should not require service by
the end-user.
Note: This feature is password-protected but is unlocked
when the password is entered (See Password Protection at
right). Do not change the temperature calibration setting.
FILTER HOURS
Total Filter Run Hours displays the total hours the
AquaCiser has been used since the filter was last changed.
The Change Filter fault indicator will be displayed when
the total run hours matches or exceeds the setpoint. After
changing the filter during the Maintenance Sequence, press
RESET HOURS to reset the time to zero.
Password Protection
Portions of the Maintenance Functions screen are
password-protected. When you press any of these areas,
you will be asked for the system password:
To unlock this screen, type: Hudson <enter>.
Pressing Main Menu to leave the Maintenance screen
will re-lock the password-protected areas.
29
AquaCiser III
30
DANGER
Electric shock can cause death or serious injury.
Install, operate, and maintain the AquaCiser only as
described in this manual.
WARNING
Improper operation of the AquaCiser can cause
injury. Operate the AquaCiser only as described in
this manual.
An unattended patient can be injured. Stay with the
patient at all times.
Hot water can cause injury. Water temperature may
increase during AquaCiser use. Monitor the water
temperature and keep the temperature below
105F (40.5 C).
Certain medical conditions can place a patient at
risk. Require all patients to have a physicians order
for treatment.
Clothing, footwear and medical accessories can come
loose and cause injury. Operators are responsible for
determining the appropriateness of items worn in
the AquaCiser.
Important
Communicate with the patient at all times. Tell the patient
when the treadmill will start or stop, when the jets will be
activated or when the treadmill speed will change.
Important
Lanolin and mineral oil can clog the filter. Do not use
products that contain lanolin or mineral oil on the
patient or in the AquaCiser.
AquaCiser III
6.3 Daily Start-up Routine
Follow the Daily Start-up routine at right before
beginning therapy sessions. Refer to the section, Using
the Command Console, pages 14-29 for instructions on
controlling the AquaCiser.
1 - CHECK THE WATER LEVEL
Open the filter access panel or reservoir access panel (see
Components, page 10) and check the water level in the
reservoir. Use a garden hose (not supplied) to add water as
needed to keep the water at the fill line shown in Figure 31.
Note: Be attentive when adding water to the reservoir
and do not leave the AquaCiser unattended during filling.
Overflowing the reservoir can cause injury and damage.
2 - TEST THE WATER CONDITION
Test the water condition two times each day: before therapy
sessions begin, and after therapy sessions are finished.
Treat the water (see Water Condition, pages 48 and 49) if
it fails to meet any of the following levels:
Daily Start-up
1.
2.
3.
4.
WARNING
Overflowing the reservoir can cause injury due
to electric shock, and can damage the AquaCiser.
Never leave the AquaCiser unattended while filling
the reservoir.
Water Hardness
250-500 ppm
Total Chlorine
less than 2 ppm
Free Chlorine
1-3 ppm
pH 7.2-7.8
Alkalinity
80-120 ppm
Use and maintain records of the water condition. Sample
Water-Condition logs are provided on pages 56 and 57.
Important
Important
The treadmill belt will stretch with use over time.
As a result, tensioning or tracking adjustments will
be needed. When adjustments are needed or when
the treadmill does not operate properly, remove the
AquaCiser from service and contact Hudson (page 2)
or your Hudson distributor.
Important
Improper repairs can damage the treadmill. Do not
attempt treadmill repairs. Contact Hudson Customer
Relations (page 2) or your Hudson distributor.
31
AquaCiser III
WARNING
Hot water can cause injury. Water temperature may
increase during AquaCiser use. Monitor the water
temperature and keep the temperature below
105F (40.5 C).
32
AquaCiser III
6.4 Before Each Therapy Session
Before each therapy session, follow the instructions in the
Before Each Therapy Session routine at right.
WARNING
Hot water can cause injury. Water temperature may
increase during AquaCiser use. Monitor the water
temperature and keep the temperature below
105F (40.5 C).
WARNING
Certain medical conditions can place a patient at
risk. Require all patients to have a physicians order
for treatment.
WARNING
Clothing, footwear and medical accessories
can come loose and cause injury. Operators are
responsible for determining the appropriateness of
items worn in the AquaCiser.
Important
Lanolin and mineral oil can clog the filter. Do not use
products that contain lanolin or mineral oil on the
patient or in the AquaCiser.
33
AquaCiser III
Therapy Session
1. Help the Patient Enter the Therapy Pool
2. Fill the Therapy Pool
3. Begin the Therapy Session
Important
Flooding can occur if the therapy pool door is not
fastened. Always close and secure the door by pulling
the lever down after the patient enters or exits the
therapy pool.
AquaCiser III
35
AquaCiser III
250 lb
113 kg
Seat Load
Limit (Dry)
AquaCiser III
5 - DRAIN THE THERAPY POOL
1. Keep the therapy pool door closed and fastened until
the water is completely drained from the chamber.
2. Go to the Therapy Pool Functions screen.
Important
Do not use ammonia-based cleaners (such as Windex)
to clean the AquaCiser. Hudson recommends using
isopropyl alcohol to clean the therapy pool windows
and door.
Important
Flooding can occur if the therapy pool door is not
fastened. Always close and secure the door by pulling
the lever down after the patient enters or exits the
therapy pool.
37
AquaCiser III
End of Day
Note:
Leave the AquaCiser Main Power ON.
4. Verify that the water temperature Setpoint is
the desired temperature (see Setting the Water
Temperature, page 17).
In this state, the AquaCiser will periodically cycle and keep
the water warmed to the Setpoint temperature.
2 - TEST THE WATER CONDITION
Important
Leave the AquaCiser Main Power ON.
ONLY turn the Main Power OFF when performing
maintenance, accessing the lint basket or filter, or plan to
leave the AquaCiser inactive for a long period of time.
Test the water condition two times each day - before therapy
sessions begin, and after therapy sessions are finished. Treat
the water (see Water Condition, pages 48-49) if it fails to
meet any of the following levels:
Water Hardness
250-500 ppm
Total Chlorine
less than 2 ppm
Free Chlorine
1-3 ppm
pH 7.2-7.8
Alkalinity
80-120 ppm
Use and maintain records of the water condition. Sample
Water-Condition Logs are provided on pages 56 and 57.
38
AquaCiser III
3 - CHECK THE FILTER CARTRIDGE
With the Filter Enabled and running (ON), open the filter
access panel (see Components, page 10), and read the pressure
gauge (Figure 38). If the pressure is 5-10 psi higher than
normal, remove and clean the filter as instructed in Change
the Filter Cartridge, page 45. If the pressure remains higher
than normal after changing the filter, replace the filter.
NOTE ON NORMAL PRESSURE GAUGE READING:
The pressure gauge will read differently depending on
many factors, including the water level in the therapy pool,
if the jets are in use, etc. To establish a normal pressure
for your location, check the pressure regularly when the
AquaCiser is in the same mode (for example, with all the
water in the reservoir and the filter enabled and running as
indicated above).
Figure
38 - Filter
Pressure
Gauge
Important
Flooding can occur if the therapy pool door is not
fastened. Always close and secure the door by pulling
the lever down after the patient enters or exits the
therapy pool.
39
AquaCiser III
DANGER
Maintenance Schedule
Wipe Down the Therapy Pool
(page 37)
Test the Water Condition
(pages 48-49)
Filter the Water (page 38)
Each Use
WARNING
Improper maintenance can cause injury. Maintain
the AquaCiser only as described in this manual and
the Service Manual provided.
Each Day
Each
Week
Each
Month
Every
6 months
As
Needed
SCHEDULE NOTES
40
AquaCiser III
Read the
Service Manual
Wear Gloves
Lubrication Point
41
AquaCiser III
1.
2.
3.
4.
5.
6.
42
DANGER
230/240 VAC
Hazardous voltage inside. Follow
lockout/tagout procedure before
servicing.
Maintenance Sequence
1. Discard the Water
2. Clean the Treadmill
3. Disinfect and Clean the Therapy Pool
4. Disinfect and Clean the Reservoir
5. Lockout/Tagout to Verify Power is OFF
6. Clean the Lint Basket
7. Change the Filter Cartridge
8. Refill the Reservoir
9. Lubricate the Door Seal
10. Test the Water Condition
11. Wipe Down the AquaCiser
Important
Personal Protective Equipment may be required. Follow
the manufacturers instructions and refer to the MSDS
sheets for all cleaning chemicals used.
Important
Lanolin and mineral oil can clog the filter. Do not use
products that contain lanolin or mineral oil on the
patient or in the AquaCiser.
Important
Abrasive cleaners can scratch. Follow the maintenance
instructions in this manual when cleaning the
AquaCiser.
AquaCiser III
3 - DISINFECT AND CLEAN THERAPY POOL
1. To disinfect, wipe all surfaces except the windows
and door with a hard-surface disinfectant such
as SaniZene and a soft-bristle brush. Follow the
disinfectant manufacturers directions for application
method and contact time.
2. Clean the windows and door using a spray bottle filled
with isopropyl alcohol, or use a mild disinfectant and a
clean, soft cloth. Do not use an ammonia-based cleaner.
3. To clean, apply a cleaner designed for use on
gelcoat fiberglass, such as Babes Boat Bubbles,
Meguires One-Step Cleaner-Wax or similar
boat wash/cleaner, available from a marine supply
store. Follow the manufacturers directions, and use
a chamois cloth to wipe down the fiberglass panels
inside and outside of the therapy pool.
4. Rinse the therapy pool thoroughly. Using a hose, run
water down between the treadmill and the chamber
wall and rinse all around the treadmill. Run the water
long enough to completely flush the area under the
treadmill.
5. If desired, the fiberglass surfaces may be waxed
to enhance the shine. Apply a boat/marine wax as
instructed by the wax manufacturer to the gelcoat
fiberglass. Do not use an automotive wax.
Important
Do not use ammonia-based cleaners (such as Windex)
to clean the AquaCiser. Hudson recommends using
isopropyl alcohol to clean the therapy pool windows
and door.
Important
When cleaning or waxing the AquaCiser, use only
products designed for use on gelcoat. Chemicals
designed for use on gelcoat may be purchased from a
boat/marine supply store.
Gelcoat
Surfaces
43
DANGER
AquaCiser III
DANGER
230/240 VAC
Hazardous voltage inside. Follow
lockout/tagout procedure before
servicing.
AquaCiser III
1.
2.
3.
4.
5.
6.
Wing Nut
Pressure
Gauge
Bleeder
Valve
Not Shown:
Reducer Nut
Filter Cover
O-Ring
Not Shown:
Shaft Guide Cap
Inside Cover
Filter
Shaft
Filter
Housing
O-Ring
Guide Cap
Important
Use filter-cleaning chemicals only as directed and
only for the contact time required by the chemical
manufacturer. To prepare the AquaCiser for service
more quickly, install the clean, dry spare filter (provided)
while the dirty filter is being cleaned.
Wet-Dry Vac and Shop-Vac are trademarks of Shop-Vac
Corporation, Williamsport, PA, USA.
Hudson Aquatic Systems, LLC, Inc. 234-3253-02 July 2009
AquaCiser III
WARNING
Overflowing the reservoir can cause injury due
to electric shock, and can damage the AquaCiser.
Never leave the AquaCiser unattended while filling
the reservoir.
Fill Line
Figure 48
- Reservoir
Fill Line
46
Important
Do not use ammonia-based cleaners (such as Windex)
to clean the AquaCiser. Hudson recommends using
isopropyl alcohol to clean the therapy pool windows
and door.
AquaCiser III
Important
Do not use a galvanized steel, aluminum, or porcelain
container to clean the filter. Filter-Cleanse can corrode
these materials. Use only a plastic or fiberglass
container.
Important
When cleaning the filter, wear Personal Protective
Equipment as recommended by the cleaning products
Material Safety Data Sheet.
The Filter-Cleanse manufacturer recommends gloves,
safety goggles and a dust mask be worn while using
the chemical. Use only in a well-ventilated area.
When using Filter-Fresh, gloves, safety goggles and
a plastic or rubber apron must be worn. Use only in a
well-ventilated area.
8. Allow the filter to air dry. The filter must be dry before
it can be used in the AquaCiser.
9. Reuse the filter until it begins to show wear. Replace
worn filters (See Supplies, page 52).
Note: Remember to reset the Filter Hours timer on the
Maintenance Functions Screen (Figure 50).
Wear Gloves
Wear Eye
Protection
Wear
Dust Mask
Wear
Apron
47
Water Condition
AquaCiser III
8 - WATER CONDITION
8.1 General Guidelines
Maintaining clean and proper water condition is important
for the patients health, for controlling algae and bacteria,
and for maintaining the AquaCiser.
Each facility is unique. The time it takes to condition the
water in your AquaCiser will vary based on the water supply
and the patient load.
In-service training and maintaining a proper Water-Condition
Log (included) are the best ways to learn how to condition
your water. See Using the Water-Condition Log, page 49.
Sample Water-Condition Logs are provided on pages 56 and
57, or you can make your own Water-Condition Log.
WARNING
Improper handling or use of water-treatment
chemicals can cause injury. Follow the chemical
manufacturers directions for use.
WARNING
Improper water conditioning can cause injury
and damage the AquaCiser. Maintain the water
condition as directed in this manual and keep a
Water-Condition Log.
48
AquaCiser III
Water Condition
For additional chlorine dispenser information, see the RolaChem Chlorinator manual included with the AquaCiser.
Also see the Chlorine Dispenser information on page 11. If
problems persist, see Basic Troubleshooting, pages 50-51.
Free
Chlorine
1-3 ppm
7.2-7.8
Total
Alkalinity
80-120 ppm
350
1/Hi
7.3
100
350
1/Lo
7.3
100
350
1/2
7.3
100
350
1/2
7.3
100
pH
COMMENTS
(Filter Changed,
Water Added, etc.)
Water
Hardness
250-500 ppm
PL
6/20
In-service
6/21
8 am, DB
6/21
11:30, DB
6/21
5 pm, DB
NUMBER
OF
PATIENTS
SA
M
DATE,
TIME,
INITIAL
49
Basic Troubleshooting
AquaCiser III
9 - BASIC TROUBLESHOOTING
9.1 Basic Troubleshooting
This section contains a few basic problems and solutions. If
the problem you are experiencing is listed on this page, try
these solutions first.
Important
If these Basic Troubleshooting solutions
do not correct the problem:
AquaCiser III
PROBLEM: Chlorine Levels too low/levels do not increase.
SOLUTIONS:
The chlorine container is empty. Change the chlorine
container (See Replacing the Chlorine Container, page 41).
The rocker power switch on the side of the chlorine
dispenser has been turned to OFF. Move the rocker to the
ON or TIMER position (Figure 51).
The chlorine dispenser is set too low. Verify the chlorine
dispenser rocker switch is on TIMER, then adjust the
chlorine dispenser control knob (See the chlorine
dispenser information on page 12 and Adjusting the
Chlorine Levels, page 49).
PROBLEM: When filling the therapy pool, the AquaCiser
stops and a Low Pressure message is displayed.
SOLUTIONS:
There is too little water in the AquaCiser, causing the heater
to lose pressure. Press DRAIN TO RESERVOIR to pump all the
water to the reservoir, then add water until the water level
reaches the Fill Line (see Check the Water Level, page 31).
Remember: test the water condition if water is added.
If the water level is okay, check the heater pressure switch.
(Refer to the Service Manual) or call for assistance.
PROBLEM: Possible Leak Detected message displayed.
SOLUTIONS:
If a patient was undergoing vigorous exercise in the therapy
pool, water may have been splashed down inside the
AquaCiser, setting off this alarm. Allow the sensor to dry.
Water may be leaking under the AquaCiser. See
Troubleshooting Water Leaks in the Service Manual.
Basic Troubleshooting
Heater ON
Indicator Light
51
AquaCiser III
10 - SUPPLIES
10.1 AquaCiser Supplies
Description
Part #
078-9730
078-9731
078-9546
078-9458
078-9459
078-9828
078-9456
078-9460
078-9461
078-9462
078-9800
078-9214
52
AquaCiser III
Phone 888.206-7802
WARNING
Improper parts and service can cause injury. Use
only Hudson parts and Hudson-approved service
on the AquaCiser.
WARNING
E-Mail info@hudsonaquactic.com
Description
Seat cushion
Upper platform floor mats (3)
Steps floor mats (right-side, 2)
Steps floor mats (center, 2)
Steps floor mats (left-side, 2)
Chlorine dispenser tubing (3'/1 m)
Therapy pool light bulb
Arm pad, right
Arm pad, left
AquaCiser Logo (1)
AquaCiser Labels*
Part #
095-1424
095-1425
095-1426
095-1427
095-1428
095-1429
095-1430
095-1431
095-1432
195-1146
195-1147
* This kit includes the water depth gauge label and all warning
and information labels.
53
Training Record
AquaCiser III
TRAINING RECORD
Date
54
Name
Training Method
Maintenance Record
AquaCiser III
MAINTENANCE RECORD
Date
Maintenance Performed
By
55
Water-Condition Log
AquaCiser III
WATER-CONDITION LOG
DATE,
TIME,
INITIAL
56
NUMBER
OF
PATIENTS
WATER CONDITION
Water
Hardness
250-500 ppm
Total
Chlorine
<= 2 ppm
Free
Chlorine
1-3 ppm
pH
7.2-7.8
Total
Alkalinity
80-120 ppm
COMMENTS
(Filter Changed,
Water Added, etc.)
Water-Condition Log
AquaCiser III
WATER-CONDITION LOG
DATE,
TIME,
INITIAL
NUMBER
OF
PATIENTS
WATER CONDITION
Water
Hardness
250-500 ppm
Total
Chlorine
<= 2 ppm
Free
Chlorine
1-3 ppm
pH
7.2-7.8
Total
Alkalinity
80-120 ppm
COMMENTS
(Filter Changed,
Water Added, etc.)
57
Index
AquaCiser III
INDEX
A
Basic Troubleshooting
Beeps (End of Personal Program Segment)
Bloodborne Disease Notice
50, 51
23
4
CE Certification
9
Chlorine Dispenser
10, 12, 41, 49-51
Chlorine Levels
(see Water)
Cleaning 40-47
Command Console
10, 11, 14-29
Components 10-13
Contaminated Water
(see Water)
Chemicals
For Filter
47
For Water Condition
49
Ordering
52
Customer Relations
2
Danger Notices
Daily Start-Up
Date and Time
Door Lever
Door Seal
Definitions
Dimensions (Specifications)
Direction (Treadmill)
Discarding Water
Disinfect
Lint Basket
Therapy Pool
Treadmill
Reservoir
Wipe Down After Each Session
Distance Meter
Drain Discard (Button On Touch-screen)
Drain to Reservoir
58
4
31, 32
27
10, 12
39, 45
(See Terms)
9
21
11, 19, 42, 49
44
43
42
43
37
20, 21
19, 42
16, 17, 19, 37
5, 6, 10, 11, 19
10, 11
8
General Guidelines
GFCI (Main Power) Switch
Heat Start-Up (Automatic)
Heater/Water Heating
Important Box
Inspecting the AquaCiser
Jets
H
I
J
L
25
2
17, 35
18
39
47
18, 38
45
28, 29
30
43
26-27, 39
18, 26, 27, 32, 33, 47, 51
4
40, 41
10, 12, 16, 18
Labels 6
Lamps (Lights)
7, 10, 12, 18
Lanolin (Not Permitted)
30, 33, 42
Lead Battery (Recycling Information)
4
Lint Basket
44
Lockout/Tagout
42, 44
Lubrication
Door Seal
39, 46
Chlorine Dispenser
41
O-Ring
45
13, 43
14, 15
40
42-46
30, 33, 42
See Jets
Index
AquaCiser III
P
Parts
53
Password 29
Patient
Evaluation
8
Facing Direction
21
Help Enter
34
Help Exit
37
Prepare for Therapy Session
33
Resting
36
Personal Programs
22, 23
Personal Protective Equipment (PPE)
41, 42, 47
pH
(see Water)
Physician Evaluation
8
Power Switch
13, 43
PLC (Programmable Logic Controls)
28, 29
Problems
(See Faults)
Ratings Plate
2, 9
RCD (Main Power) Switch
43
Recycling 4
Repairs 53
Reservoir
10, 13
Filling
46
Water Level
31, 37, 46
Clean and Disinfect
43
Reset Filter Hours
47
Restart Button
10, 13, 46
Resting Seat
13, 36
Routines
Before Each Session
33
Daily Start-Up
31, 32
End of Day
38, 39
Maintenance
40, 42-46
Therapy Session
34-37
Water Condition
48, 49
Run Timer
21
Seat
13, 36
Service 53
Settings
Automatic Heat Start-Up
27
Date and Time
27
Filter Hours
29
Units (Metric or Imperial)
29
Water Temperature
16, 17, 47
Set-Up Before Use
30
Specifications
9
Speed
20, 21
Supplies 52
Symbols Glossary
6
Table of Contents
3
Temperature
16, 17, 32, 33, 47
Terms
7
Therapy Jets
10, 12, 16, 18
Therapy Pool
Clean and Disinfect
43
Component
10
Function Screen
16-19
Wipe-Down
37
Therapy Session
33-37
Timer
(see Run Timer)
Time and Date
27
Touch-Sensitive Screen (Command Console)
14
Training, Operator
7, 24
Treadmill
Cleaning
42
Treadmill Direction
21, 35
Treadmill Functions Screen
20, 21
Treadmill Operation (Checking)
32
Treadmill Speed
20, 21
Troubleshooting (Basic)
50, 51
Turbo Jets
10, 12, 18
SaniZene
42, 43, 52
Safety Information
5-7
Safety Labels
5
Screens 14-29
Automatic Heat Start-Up
26-27
Main Menu Screen
14-15
Maintenance Functions
28-29
Operator Assistance
24-25
Personal Programs
22-23
Therapy Pool Functions
16-19
Treadmill Functions
20-21
Warning Notices
4
Warranty Statement
2
Water
Adjusting Chlorine, pH, Alkalinity
49
Condition
31, 38, 46, 48, 49
Condition Log
49, 56, 57
Contaminated Water (Dealing With)
19
Discarding
11, 19, 42, 49
Temperature 17, 18, 20, 21, 26, 27, 32, 33, 47, 51
Water Level (Reservoir)
31, 37, 46, 51
Water Level (Therapy Pool Fill/Drain) 16, 17, 34
WEEE Directive
4
Windex (Dont Use)
37, 43, 46
59