Professional Documents
Culture Documents
1. Introduction
Modern countercurrent chromatography (CCC) has arisen only
over the last 30 or so years and offers the natural product chemist
a further separation tool with many unique advantages. It is inherently
the mildest form of chromatography with no solid support and
hence no chance of loss of substrate by binding to the stationary
phase. The only media encountered by the sample are solvent and,
typically, Teflon tubing. The former is common to all forms of
chromatography and the latter to most. It is true that many of the
solvents systems (SSs) have more components than most other
forms of chromatography, but these can be chosen from the most
nonreactive and innocuous solvents. Hence, the chromatographer
is virtually assured of near 100% recovery of sample from a chromatography. The number of two-phase systems, which can be
employed, is limited only by the imagination of the chromatographer,
Satyajit D. Sarker and Lutfun Nahar (eds.), Natural Products Isolation, Methods in Molecular Biology, vol. 864,
DOI 10.1007/978-1-61779-624-1_9, Springer Science+Business Media, LLC 2012
221
222
and the systems can be explored by any of several simple tests, prior
to a preparative separation, to assure success. Two similar compounds of almost identical polarity can have surprisingly different
partition coefficients in a specific two-phase system resulting in
base-line separation by CCC.
CCC had its beginning in the 1950s with the Craig machine
(1), a mechanical system of sequential separating cells in which one
phase of a two-phase SS could be equilibrated with the other phase
in successive cells thereby carrying a solute along according to its
partition coefficient between the two phases. Solutes would be
washed from the train in the order that their partition coefficients
favored the mobile phase. These instruments were cumbersome,
delicate, and required a major ventilation system as they invariably
leaked organic solvents to some extent. A typical system would
involve 200400 cells and a separation using such an instrument
would take in the order of a week to accomplish. They were used
because they could achieve separations, which could not be otherwise effected. They were both displaced and replaced. Displaced by
the high pressure liquid chromatography (HPLC) and the large
number of possible solid supports available for this methodology,
and replaced by droplet countercurrent (DCCC) and rotary locular
(RLCCC) instruments, which could be used to effect that same
separation method in a fraction of the laboratory space. DCCC
and RLCCC instruments are still available (Eyela). The former
composed of vertical tubes of a diameter that will allow droplets of
one phase of a two-phase SS to rise (or fall) through the other
phase. These mixing tubes are then connected top to bottom, with
fine tubing such that the droplets would completely fill the tubing
and exclude the stationary phase from these interconnecting tubes.
Thus, the mobile phase is added slowly and allowed to percolate
through the mixing chambers under the force of gravity, to achieve
a distribution of solute according to its partition coefficient. DCCC
systems, while sharing the high separatory power of the Craig
machine and the high overall recovery of load sample, still suffer
from the problem of being slow, and this is further complicated by
the need to maintain constant temperature during the course of a
chromatography.
However, a very recent report by Dini demonstrates that this
can still be a highly effective method of separation (2). The instrument used was a Bchi DCCC equipped with 300 2.2 mm i.d.
tubes each with 2.0 mL volume (according to anecdotal evidence,
only around 50 machines were built, and very few are probably still
operable; another one is located in the chapters authors laboratory). Using a stationary phase of n-butanol/acetone/water (60:12:28)
and a mobile phase of n-butanol/acetone/water (14:12:74), three
new dihydrokaempferol glycosides (Fig. 1) were isolated and characterized from a single run (2).
OR
HO
O
223
R=H
R = L-xylopyranoside
OCH3
OH
O
OCH3
O
HO
OH
OH
O
OH
HO
OH
OH
HO
O
OH
OH
O
O
OAc
R"
OH
HO
HO
O
HO
OH
R'
Cinobufatolin
OH
HO
H
Arenobufagin
Fig. 1. Structures of flavonoid glycosides from Pouteria obovata and bufadienolides isolated from Chan Su.
224
2. Materials
2.1. Current
Instruments
Although several minor variants are available, instruments are basically of two types.
(a) The Centrifugal Partition Chromatography instrument as sold
for many years by Sanki Instruments (Kyoto, Japan) and more
recently by Armen (Saint Ave, France) and Rousselet Rotatel
Kromaton (Annonay, France).
(b) The Coil Planet Centrifuge as designed by Yoichiro Ito.
Variants of the latter are currently sold by AECS (Bridgend,
UK), Conway Centri Chrom Inc (Williamsville, NY, USA), Cherry
Instruments (Chicago, USA), Dynamic Extractions Ltd. (Uxbridge,
UK), and Tauto Biotech (Shanghai, China). Originally, the acronym
CPC was used by Ito for his Coil Planet Centrifuge, which acts in
hydrodynamic mode. Only later CPC was used to designate the
hydrostatic Centrifugal Partition Chromatography.
The Centrifugal Partition instruments effectively replace gravity
as the driving force of a DCCC procedure with centrifugal force
and, thus, achieve a remarkable increase in the speed of the process.
This does necessitate the use of two rotating seals. The Teflon
tubing is replaced by small solid blocks or sheets of Tefzel, honeycombed with a channel system analogous to the tubing of a droplet
system. A distinct advantage of this system is the ability to use relatively viscous solvents such as n-butanol at r. t. Although n-butanolcontaining SSs can be used in the Coil Planet Centrifuge, they
usually result in poor retention of stationary phase at ambient
temperatures, and hence sub-optimal performance, unless other
components reduce the viscosity of the n-butanol-containing phase.
An obvious weak point in the design of the Centrifugal Partition
225
226
CCC has not caught the attention of the large instrument makers, and
their manufacture tends to be a cottage industry. Currently, instruments are available from the following suppliers/manufacturers.
(a) AECS, PO Box 80 Bridgend, South Wales, United Kingdom
CF31 4XZ. Tel: 44-783-876-1524, Fax: 44-1656-789-282,
E-mail: aecs@gmx.com.
(b) Armen Instrument, Z.I de Kermelin, 16 Rue Ampre, F-56890
Saint Ave, France, Tel: 33 (0) 2 97 61 84 00, Fax: 33 (0) 2 97
61 85 00, E-mail: contact@armen-instrument.com.
(c) Cherry Instruments 3440 S. Dearborn St. Suite 242 Chicago,
IL 60616. Email: contact@cherryinstruments.com, Tel: 1-800852-0955 or 1-773-260-0204.
(d) Conway Centri Chrom Inc., 52 MacArthur Drive, Williamsville,
NY 14221. Tel: 1-716-635-088 Fax: 1-716-635-0886. E-mail:
wdconway@centrichrom.com.
(e) Dynamic Extractions Ltd Berkshire, UK, Tel: 44-1753 696979.
WWW: http://www.dynamicextractions.com.
(f) Eyela/Tokyo Rikakikai, c/o Dychrom, 190 Mountain ViewAlviso Road, Suite T, Sunnyvale, CA 94089 USA, Tel: 408745-7911, Fax: 408-745-1554, T. 1-800-439-2476, E-mail:
frank.tao@dychrom.com (DCCC, RLCCC).
(g) Rousselet Robatel Kromaton ZI, Marenton 45, Avenue Rhin
& Danube 07100, Annonay, France, Tel: 33(4)-75-69-22-11
Fax: 33(4) 75-67-69-80 E-mail: kromaton@kromaton.com.
227
(h) Shanghai Tauto Biotech Co., Ltd. Room A301 No. 326 Aidisheng
Road, Zhangjiang Hi-Tech Park, Shanghai, Peoples Republic
of China, 201203. Tel: 86-21-51320502. E-mail: export1@
tautobiotech.com.
3. Methods
3.1. Solvent Systems
and GUESS
Methodology
3.1.1. The Nature
of Solvent Systems
The most critical step in a CCC is the choice of the two phase SS
and these commonly consist of an aqueous and an organic phase,
although this is not necessarily the case. Non-aqueous two phase
systems have been used as have systems in which both phases are
aqueous (4, 1620). The chromatography can be run in a normal
phase mode, i.e., the polar (usually aqueous) phase is stationary or
in a reversed-phase mode with the aqueous phase mobile. This
interchange can be achieved by changing the direction of spin or
the direction of flow through the instrument. Dichloromethane
(DCM) and diethyl ether can both be used but the researcher
should be aware that a vapor lock will force the stationary phase
from the system, and abort the chromatography. Hence these
solvents need to be used only if the instrument comes with temperature control or if ambient temperature permits.
The utility of TLC in selection of an SS in the HEMWat and
ChMWat systems has been greatly increased by development of the
GUESS (Generally Useful Estimate of Solvent Systems) methodology (21, 22), which further exemplifies the analogy between CCC
and TLC (see Note 1 and Tables 13). The GUESS methodology
was developed by taking 21 commercially available natural products of widely varying polarity (The GUESS Mix) and demonstrating that their Rf values on a normal phase silica TLC developed
with the lipophilic phase of a potential SS correlated with their
partition coefficient in that SS. Hence the scientist runs a normal
phase silica gel TLCs of the analyte using only hexane and EtOAc
in the ratios they occur in some of the standard HEMWat systems,
say 4, 0 and +4 (i.e., 7:3, 1:1, and 3:7 respectively). The aim is to
find a ratio which results in the analyte having an Rf of ~0.5. The
corresponding HEMWat system will very probably partition the
analyte with a partition coefficient of between 0.25 and 4.0, and
hence will place the analyte in the sweet spot of a CCC run. The
usefulness of the GUESS mix has been expanded to include not
only solvent system prediction in HEMWat solvent systems but
also the systematic analysis and comparison of biphasic solvent systems commonly employed in CCC. The wide polarity range of the
GUESS mix components enables mapping of optimal polarity
ranges for HEMWat, EBuWat, terAcWat, and HterAcWat solvent
system families.
228
Table 1
Abbreviations of common constituents used for the
nomenclature of 2-phase CCC solvent systems (21)
Solvent
Abbreviation
Acetic acid
Acetone
At
Acetonitrile
Ac
n-Butanol
Bu
t-Butylmethyl ether
Ter
Chloroform
Ch
Cyclohexane
Cy (or Cyc)
Dichloromethane
D (or Di)
Diethylether
De
Ethanol
Et (or Eta)
Ethyl acetate
Hexane
H (or He)
Heptane
Hep
Methanol
M (or Me)
Toluene
Tol
Water
Wat
When two abbreviations are given, the second one can be used when it is
necessary to add consonance to the SS name
229
Table 2
The HEMWat family of two-phase solvent systems (SSs)
consisting of n-hexane (H), ethyl acetate (E), methanol (M), and
water (Wat) represents the most widely used SS in the literature
HEMWat#
Wat
+1
+2
+3
+4
+5
+6
+7
+8
10
10
Table 3
The ChMWat family of two-phase SSs All ChMWat systems
include 10 volumes of chloroform
ChMWat#
Wat
10
+1
+2
+3
+4
230
231
Table 4
Application of CCC in natural products isolation
Solvent system
Compounds
References
Light-PetEtOAcCCl4CHCl3MeOHH2O
1:1:4:4:6:2
IsooctaneEtOAcMeOHH2O 7:3:6:4
n-HexaneEtOAcMeOHH2O 70:30:15:6
(43, 44)
Taxol (Fig. 4)
Tirandamycins A and B (Fig. 6)
(57)
(63)
n-HexaneEtOAcMeOHH2O 1:1:1:1
(64, 65)
Squalestatins (Fig. 6)
(66)
n-HexaneEtOAcMeOHH2O 2:3:3:2
Arizonins (Fig. 7)
(64)
n-HexaneEtOAcMeOHH2O 3:7:5:5
Auxins (Fig. 8)
(67)
n-HexaneEtOAcMeOHH2O (25 mM
phosphate buffer pH 6.9) 7:3:5:5
Australifungins (Fig. 9)
(68)
n-HexaneEtOAcMeOHH2O 2:2:2:1
n-HexaneEtOAcMeOHH2O 2:3:1:2
to 10:15:8:10 step-wise
n-HexaneEtOAcEtOHH2O 8:2:7:3
Phomopsolides (Fig. 8)
Bufadienolides (Fig. 1)
(69)
(26)
Tanshinones (Fig. 5)
(60)
n-HexaneEtOAcMeOHEtOHH2O
10:14:10:2:13
n-HexaneCHCl3MeOHH2O 1:1:1:1
(70)
n-HexaneDCMMeOHH2O 5:1:1:1
(63, 71)
n-HexaneDCMMeOHH2O 10:40:17:8
n-Hexanen-BuOHH2O (0.05%TFA) 43:7:50
n-HexaneCH3CNEtOAc 5:5:2
cyclohexaneMeOHH2O 10:9:1
Steroids
Enramycins A and B (Fig. 6)
Diterpenes (Fig. 5)
Massoia lactones (Fig. 5)
(72)
(62)
(59)
(61)
HeptaneEtOAcMeOHH2O 1:1:1:1
CCl4MeOHH2O (0.01 M KHPO4 pH7)
CCl4CH2Cl2MeOHH2O 5:5:6:4
(73)
(58)
(57)
CHCl3EtOAcMeOHH2O 12:8:15:10
EtOAcCCl4MeOHH2O, 1:6:4:1
Pristinamycins (Fig. 3)
Liensinine alkaloids
(55)
(44)
CHCl3MeOHH2O 1:1:1
Coloradocin, 2-norerythromycins,
pentalenolactone (Fig. 12) and
ascelacins (Fig. 13)
(58, 63,
7478)
CHCl3MeOHH2O 7:13:8
(63)
CHCl3MeOHH2O 4:3:3
(79)
(80)
n-HexaneEtOAcCH3CNMeOH 5:4:5:2
(81)
(continued)
232
Table 4
(continued)
Solvent system
Compounds
References
n-HeptaneC6H6IPAacetoneH2O 5:10:3:2:5
n-HexaneEtOAcH2O 3:7:5
(77)
EtOAcEtOHH2O 3:1:2
(83)
EtOAcMeOHH2O
(84)
iso-AmOHn-BuOHn-PrOHH2OHOAc
t-Bu2S 32:48:40:120:1:4
n-BuOHacetoneH2O ~3:1:4
t-BuOMeTHFH2O 2:2:3 (Et3N, HCl)
(85)
Kaempferol glycosides
Harmine, Harmaline (Fig. 2)
(2)
(46)
(67)
(43)
(86)
diverse systems where the activity is all in the aqueous phase of one
and in the organic phase of the other. Now two very simple partitions
will achieve a high level of purification prior to any chromatography. The approach to the ubiquitous HEMWat [hexaneethyl
acetate (EtOAc)methanol (MeOH)water] system, for example,
is to assume that for organics of medium polarity, hexane and water
will be poor solvents, and EtOAc and MeOH good solvents, and
that the lower phase will consist mainly of MeOH and water,
whereas the upper phase will consist mainly of hexane and EtOAc.
Hence if, in the HEMWat-0 (1:1:1:1) system, the desired compound
favors the upper phase, it can be displaced toward the lower, by
increasing the proportion of hexane or MeOH. Increasing the
proportion of MeOH has its limits, particularly in the ChMWat system,
as at some level the system will become monophasic.
Although it is most common, it is not necessary to maintain
the same SS as the mobile phase throughout the chromatography
run. However, preferably any change should not drastically upset
the equilibrium between the two phases. Jia et al. (26) used a stepwise gradient elution though a series of closely related HEMWat
systems to effectively separate six bufadienolides (Fig. 1) from the
Chinese traditional medicine, Chan_Su. They loaded and initially
ran the CCC with HEMWat 2:3:1:2 and the lower phase mobile
for 200 min, then changed the mobile phase to the lower phase of
HEMWat 8:12:5:8 for a further 200 min, and finally to the lower
phase of HEMWat 10:15:8:10 for a further 300 min. A more daring
233
234
OH
235
OH
O
N
R
Crinine R=H
OMe
Crinamidine
Powelline R=OMe
O
N
O
OH
Liensinine R1 = H, R2 = CH3
Isoliensinine R1 = CH3, R2 = H
Neferine
R1 = R2 = CH3
OR1
OR2
O
H3CO
O
N
H3CO
N
O
Nuciferine R= CH3
N-Nornuciferine R = H
Roemerine
H3CO
H3CO
N
H
Harmine
N
H
Harmaline
236
237
When using the Ito coil, the researcher is presented with the making a choice of 3 twofold variables.
(a) The question of which phase to select as the stationary phase,
i.e., which phase is used to fill the column.
(b) The choice of the inlet tube, either the head or the tail of
the column.
(c) The question of which direction to spin the column, i.e., to
have the Archimedean force directed to the inside or the outside of the spiral.
Two of the eight (23) possibilities will usually work well, two
poorly and the other four will result in no retention of stationary
phase. A more detailed description of the binary choices in CCC
instrument design and operation has been recently published (47).
The tubings to the columns are labeled head and tail, and
spin directions as forward and reverse. It is advisable to fill the
column with the stationary phase while it is spinning with an
Archimedean force against the fill. This ensures that the column is
filled without any vapor blocks. The column does not have to be
spun at normal running speed as this may create balancing problems.
When the column is filled with stationary phase, it should be spun
at 8001,000 rpm while pumping the mobile phase. It is possible
to introduce the sample load with the solvent front of the mobile
phase, but in a new system this is usually unwise. The system can
be tested to ascertain the displacement of stationary phase before
introducing the sample. In the better systems only about 10% of
the stationary phase is displaced before breakthrough of the mobile
phase. After that only the mobile phase is eluted from the column.
It is useful to keep in mind the theoretical shape of the elution
curves. Although this varies with the nature of the solute (48), and
to a lesser extent the SS, for most natural products the following
generalization applies. Components with partition coefficients
strongly favoring the mobile phase will be eluted very early in chromatography, and will be in sharp peaks. As the chromatography
continues, the peaks eluted with the mobile phase broaden. In
early use of HSCCC, it was common to pump two to three column
volumes of mobile phase and then to reverse the direction of spin,
and displace the stationary phase either by continuing to pump
mobile phase or with a stream of nitrogen. The stationary phase is
238
With this mode of CCC the choices are more straightforward for
the novice. The only choice that needs to be made in running the
instrument is which phase is to be the mobile phase. The CPC
instruments come with a conveniently labeled switch, ascending
descending, and this can be mentally placed in the old droplet
format, in which the upper phase mobile will be ascending and
the converse. One does not have to carefully estimate counterbalances. CPCs have neoprene or ceramic seals, which require replacement after extensive operation, the ceramic ones lasting much longer
than the neoprene ones, but are more expensive. These instruments
are loaded with a spin speed of approximately 200 rpm for the
200 mL rotor and 900 rpm for the 1 L instruments. A flow rate of
between 1 and 5 mL/min is appropriate for the 200 mL instrument,
dependent of the relative densities of the two phases. However, 1 L
instruments are run with flow rates between 20 and 30 mL/min,
provided that the pressure on the column does not exceed 800 psi.
These instruments are relatively quiet and stable.
For all instruments certain parameters are absolutely critical,
such as direction of spin, direction of flow and choice of stationary/
mobile phases and the separation time (differential density) of the
phases. Others are less critical but nonetheless have a significant
effect on performance. The optimal adjustment of flow rate, temperature, rotational speed, and sample loading in a 240 mL hydrodynamic (J-type) CCC instrument was investigated by Friesen and
Pauli (50) utilizing the complete GUESS mix in HEMWat 4:6:4:6
(=HEMWat +3).
3.8. Detection
239
3.9.1. Separation
of Pristinamycins
240
H
O
O
N
N
O
Pristinamycin IIA
O
OH
N
O
H
O
O
N
Pristinamycin IIB
OH
O
N
NH
N
O
N
H
H
N
O
O
O
N
N
N
O
O
O
O
O
O
O
HO
OH
HO
HO
O
O
O
O
HN
R
Taxol R =
O
OH O
OH O
O
O
Cephalomannine R =
9-Dihydro-13-acetylbaccatin III
241
3.9.4. Isolation of
Kaurenoic and Polyalthic
Acids from Oleoresin
A nonaqueous system, HAcE, 5:5:2, was used to isolate two diterpenes, kaurenoic acid, and polyalthic acid (Fig. 5) from the oleoresin of Copaifera glycycarpa, which is used as a folk medicine in the
Brazilian amazon (59). From 600 mg of resin, 35 mg of kaurenoic
acid and 150 mg of polyalthic acid were obtained using a two-bobbin AECS Quattro instrument with a capacity of 234 mL per bobbin. With the upper phase mobile and pumping at 3 mL/min the
retention of stationary phase was 92%. The two diterpenes showed
moderate activity inhibiting multiple-drug resistant transporters.
3.9.5. Purification of
Tanshinones from Tanshen
242
N
HO
O
H3CO
O
O
OH
OH
OR
O
Niddamycin A1
R = COCH2CH(CH3)2
Niddamycin B
R = COCH2CH2CH3
Niddamycin F = 10,11-Dihydroniddamycin A1
O
H
H
H
COOH
COOH
Kaurenoic acid
Polyalthic acid
O
O
O
Tanshinone I
Tanshinone IIA
Fig. 5. Structures of niddamycins, Copaiba oil diterpenes, tanshinones, and massoia lactones.
HO
243
Tirandamycin A R = H
Tirandamycin B R = OH
NH
O
O
CH2
CO2H
H
CH3
OR
CH3
OR'
HO
CH3
X=
O
CO2H OH
CO2H
CH3
O
HO
O
O
O
HN
HN
N
H
NH
HN
O
NH
HO2C
N
H
NH2 O
OH
HN
NH
O
NH
O
HN
Cl
NH2
O
OH
O
NH
O
NH
NH
O
HN
NH
O
HO
OH
HN
NH
Cl
HO
NH
H
N
N
H
H
N
OH
OH
OH
O
NH
HN
4. Notes
1. Targeted toward high sample diversity, GUESS is a practical
approach for the prediction of CCC distribution constants, K
values, by standard thin layer chromatography (TLC). It allows
a major reduction in workload by direct use of routine TLC
information. As SSs usually involve several solvents and repeated
description of them becomes tiresome, a simplified abbreviation
set has been developed (21) both for the individual components
244
R'O
O
Arizonin A1
R = Me
Arizonin B1
Arizonin C1
R=H
R = Me
R = R = Me
OR
R = H
R'O
O
O
OCH3
OH
Arizonin A2
R = Me
R = H
Arizonin B2
R=H
R = Me
Arizonin C2
R = R = Me
OMe
OH
O
OH
O
O
O
HO
HO
OH
MeO
O
H2N
H
N
OH
COOH
R
Indole-3-acetonitrile R = -CH2CN
Abscisic acid
Indole-3-acetamide R = -CH2CONH2
Indole-3-acetic acid R = -CH2COOH
Indole-3-butyric acid R = -(CH2)3COOH
Indole-3-carboxylic acid R = -COOH
245
OH
O
H
HO
H
O
O
H
OH
Australifungin
O
O
O
O
OH
O
O
O
C
OH
OH
O
O
OH
O
OH
OH
O
E
OH
D
Phomopsolides
O
O
O
O
OH
HO
OH
246
O
O
O
O
NH
Bu 2313 B
O
O
N
N
H
N
5-N-Acetylardeemin
O
O
H
OH
COOH
O
HO
O
O
H
O
O
O
O
Pentalenolactone
Coloradocin
O
OH
R'
OH
HO
O
O
O
R
OH
O
2-Norerythromycin A
R = OMe
R = OH
2-Norerythromycin B
R = OMe
R = H
2-Norerythromycin C
R = OH
R = OH
2-Norerythromycin D
R = OH
R = H
247
H
N
OH
O
O
NH
N
H
HN
O
HN
O
H3C
HN
N
H
H2N
O
OH
Aselacin A
R=
CH3
O
OH
Aselacin B
R=
CH3
O
O
Aselacin C
R=
CH3
OH
OH
NH
HN
OH
OH
OMe
OMe
OH
OMe
OH
HO
OMe
OH
HO
NH
OH
Michellamine A
NH
OH
Michellamine B
248
CH3
HO
OCH3
H3C
H3C
CH3
OH
R2
CH3
HO
H2
C
H3C
CH3
OCH3
CH3
CH3
OH
CH3
Clarithromycin (6-O-Methylerythromycin) R1 = R2 = H
(14R)-14-Hydroxyclarithromycin R1 = OH, R2 = H
15-Hydroxyclarithromycin R1 = H, R2 = OH
O
HN
OH
OH
COOH
O
OCH3
Dorrigocin A
O
HN
OH
OH
COOH
O
OCH3
Dorrigocin B
OHC
249
O
N
OCH3 O
OCH3 O
O
O
NH2
N
O
N
O
NH2
OCH3
HN
O
O
OCH3
O
N
NH2
CO2CH3
CONH2
CONH2
NH2
CO2CH3
N
O
OCH3
OCH3
OH
OH
OH
OH
HO
HN
H
N
O
OH
NH2
OH
OH
HO
OH
OH
OH
Gibberellic acid
250
OH
N
OH
NH2
OH
OH
OH
O
OH
O
OH
OH
N
Oxysporidinone
NH
HO
N
H2N
N
H
OH
1,3-Dimethylguanine
Siderochelin A
CH2
CH2
H3C
H3C
H
CH3
CH3
CH3
CH3
Taraxasterol acetate
O
O
H3 C
CH3
CH3
H3C
CH3
CH3
H
CH3
CH3
H
CH3
-Amyrin acetate
H3C
CH3
H3C
Lup-20(29)en-3b-ol acetate
CH3
H
CH3
CH3
CH3
Taraxasterol acetate
-Amyrin acetate
CH3
References
1. Craig LC, Craig D (1956) In: Weissenberger A
(ed) Techniques in organic chemistry: separation and purification. Interscience Publishers,
New York, pp 247254
2. Dini I (2011) Flavonoid glycosides from Pouteria
obovata (R. Br.) fruit flour. Food Chem
124:884888
3. Hostettmann K, Marston A (1990) Liquidliquid partition chromatography in natural
product isolation. Anal Chim Acta 236:6376
4. Marston A, Hostettmann K (2006) Developments in the application of counter-current
chromatography to plant analysis. J Chromatogr
A 1112:181194
5. Ito Y (2005) Golden rules and pitfalls in selecting
optimum conditions for high-speed countercurrent chromatography. J Chromatogr A
1065:145168
6. Sticher O (2008) Natural product isolation.
Nat Prod Rep 25:517554
7. Sutherland IA, Fisher D (2009) Role of countercurrent chromatography in the modernisation
of Chinese herbal medicines. J Chromatogr A
1216:740753
8. Pan Y, Lu Y (2007) Recent progress in countercurrent chromatography. J Liq Chromatogr
Relat Technol 30:649679
9. Yoon KD, Chin Y-W, Kim J (2010) Centrifugal
partition chromatography: application to natural
products in 1994-2009. J Liq Chromatogr
Relat Technol 33:12081254
10. Pauli GF, Pro SM, Friesen JB (2008) Countercurrent separation of natural products. J Nat
Prod 71:14891508
11. Ito Y (1986) High-speed countercurrent
chromatography. CRC Crit Rev Anal Chem
17:65143
12. Ito Y, Conway WD (1984) Analytical chemistry
applied spectroscopy section, abstract 472.
In: Pittsburgh Conference and Exposition,
Atlantic City
13. Ignatova S, Hawes D, van den Heuvel R,
Hewitson P, Sutherland IA (2010) A new nonsynchronous preparative counter-current
centrifuge the next generation of dynamic
extraction/chromatography devices with independent mixing and settling control, which
offer a step change in efficiency. J Chromatogr
A 1217:3439
14. Sutherland IA, Hewitson P, Ignatova S (2009)
Scale-up of counter-current chromatography:
demonstration of predictable isocratic and
quasi-continuous operating modes from the
test tube to pilot/process scale. J Chromatogr
A 1216:87878792
251
252
253
254