Professional Documents
Culture Documents
Abstract
An automatic monosegmented ow system (MSFA) based on the binary search concept to perform potentiometric titration is
proposed. A tubular hydrogen ion-selective electrode without inner reference solution, consisting of a conducting epoxy
cylinder machined with an axial hole coated with tridodecylamine (TDDA) was employed as a sensor. The titration procedure
was implemented by exploiting the binary search approach, after each analytical cycle data were evaluated to decide the
variation of the titrant volumetric fraction to be inserted for the next cycle. The ow network comprised three-way solenoid
valves controlled by a microcomputer running software in QUICKBASIC 4.5. Combination of binary search and MSFA resulted in
an automatic system providing possibilities to perform ow titrations without any operator assistance. The main features of the
proposed system were veried by titrating hydrochloric and acetic acid solutions and the feasibility of the approach was
ascertained by analyzing vinegar, coke, lemon soda, isotonic, industrial and natural orange juice samples. Results were in
agreement with those obtained with a conventional potentiometric titration, and no signicant difference at 95% condence
level was observed. A 1% standard deviation (n9) in results was also observed. # 1999 Elsevier Science B.V. All rights
reserved.
Keywords: Potentiometric titration; Tubular pH electrode; Monosegmented ow; Binary search; Multicommutation
1. Introduction
Titration is a widely used technique to monitor
industrial process, presenting as disadvantage the long
time required when a manual titrimetric procedure is
employed. To overcome this drawback, automatic
batch titrators and continuous ow techniques have
*Corresponding author. Fax: +55-19-4294610; e-mail:
reis@cena.usp.br
1
Universidade do Estado da Bahia, Salvador BA, Brazil.
2
CEQUP/Faculdade de Farmacia Universidade do Porto,
Portugal.
0003-2670/99/$ see front matter # 1999 Elsevier Science B.V. All rights reserved.
PII: S0003-2670(99)00092-6
166
167
Fig. 1. Schematic representation of the binary search titration: (a) sampling pattern, (b) hypothetical results. Eb average readings of the
baseline. Eb1Ebks and Eb2Ebks.
168
(1)
(2)
V2 V1 V0 ;
(3)
V20
(4)
V1 V0 ;
stoichiometry condition, the variation of titrant volumetric fraction is calculated as indicated in the following equations:
Vn Vn1 =2;
(5)
Vn Vn1 Vn ;
(6)
Vn0
(7)
Vn1 Vn :
This strategy is followed until the condition established in Eq. (1) is attained. The volumetric fraction of
the sample solution is constant to avoid variations of
the sample matrix, therefore, when the titrant volumetric fraction was decreased (Eq. (4)), a diluent
solution aliquot with a equal volume to that of the
variation of titrant was inserted in order to maintain
the concentration and ionic strength of the sample
zone.
All paths followed by the hypothetical titration
based on the binary search process up to the fth trial
is shown in Fig. 1. The end point is between the third
and fourth trials and the end point was attained in the
fth trial.
3. Experimental
3.1. Apparatus
The ow manifold was built up by assembling a set
of three-way solenoid valves (161T031 NResearch),
mixing coils of PTFE tubing (0.8 and 1.6 mm i.d.) and
transmission lines of polyethylene tubing (0.8 mm
i.d.). Signal detection was carried out with a 2002
Crison potentiometer equipped with a double junction
electrode (Orion 900029 Ag/AgCl) as a reference
electrode. A home-made ow through selective electrode for hydrogen ion [20] was employed as working
electrode. The sensor unit was adapted to the ow
system by using a cell constructed as described elsewhere [21]. A 486 microcomputer equipped with an
electronic interface (Advantech PCL-711S), running
software written in QUICKBASIC 4.5, was used to control
all steps of the titration procedures. An Ismatec IPC-4
peristaltic pump was employed and variation of pumping rates was performed by using its serial port (RS232) coupled to the microcomputer. An electronic
interface [18] was used to match the current intensity
required to switch the solenoid valves.
169
Fig. 2. Flow diagram of the system for automatic titration. (a) Sampling pattern. V1V5 three-way solenoid valves, B coiled reactor
(30 cm, i.d. 1.6 mm), C carrier stream, S sample, T titrant, D diluent, X confluent point, ISE ion-selective electrode, REF
reference electrode, W waste, mV potentiometer, P peristaltic pump, flow rate2.0 ml min1. Dashed lines represent the flow paths
when the valves are switched on.
170
171
172
quence, the generated signals exhibited the uctuations pointed out in Fig. 4 caused by the gas bubbles
when they passed through the electrode.
The above experiments were carried out by inserting a sample string with a volume of 512 ml. Further
experiments showed that a string with a volume of
256 ml yielded a signal of similar magnitude (95%),
hence, this volume was employed further.
The software was designed to read the signal as a
function of time and to display it on the video screen to
allow a real time view of the titration. At the same
time, a data string of 120 measurements (average of 10
analog/digital converter readings) were generated.
Afterwards, the measurements related to the gas bubble were eliminated, and the data were processed to
decide about the course of the titration. Also, these
data were saved in a le to allow further analysis.
A baseline value was used as a reference to nd the
end point of titration. Therefore, the carrier stream
solution was passed through the analytical path and
200 measurements of the potentiometer analog output
were collected before the beginning of each trial.
Afterwards, the mean and standard deviations were
calculated. If the r.s.d. was lower than 2%, baseline
and r.s.d. values were stored, otherwise the measurements were repeated.
As indicated in the model of the binary search
titration (Fig. 1), on increasing the number of trials,
the measurements tend to the baseline value. The end
point titration was found when Eq. (1) was satised.
The coefcient k dened as an arbitrary constant was
set as 5. The pattern in Fig. 4 indicates a behavior
similar to that proposed in the model, so that for
additional trials, the measurements tend to the baseline value. In this gure, baseline value and standard
deviation were 86.7 and 0.9 mV, respectively. In this
sense, the titration end point was reached when the
measurement (Es) was comprised in the 82.291.2 mV
range. In Fig. 4, only three trials of the titration are
shown corresponding to titrant solution slugs volume
32, 16 and 24 ml. The potential difference of curve 3
was larger than 91.2 mV, thus for the next trial the
titrant slug volume should be 28 ml. In this situation,
the volume variation was 4 ml and for the next trial it
should be 2 ml. This value must be added if the
measurement was higher than 91.2 mV or subtracted
if the measurement was lower than 82.2 mV, as indicated by Eqs. (6) and (7). This strategy of slug variation was continued to locate the end point or till it
attained the smallest variation that was pre-set as
0.3 ml (loading time variation of 0.01 s).
After the operational conditions were established,
the feasibility of the approach was assessed by analyzing samples of vinegar, coke, lemon soda, isotonic,
industrial and natural orange juice, yielding the results
in Table 1. Accuracy was checked by applying the
paired t-test with those data obtained by a conventional titrimetric procedure [22], and no signicant
differences at the 95% condence level were
observed. A relative standard deviation of 1%
(n9) was achieved.
The time elapsed for performing one trial was 30 s,
therefore, the overall time to carry out the entire
titration depends on the required number of trials.
For the analyzed sample (Table 1), the minimum
number of trials was 5 and the maximum 8.
Table 1
Mean values and uncertainties (n3) for acidity determination in vinegar, coke, lemon soda, isotonic, industrial and natural orange juice
samples
Sample
White vinegar
White vinegar
Color vinegar
Color vinegar
Coke
Lemon soda
Isotonic
Orange juice (natural)
Orange juice (industrial)
7.250.03
7.210.02
7.580.01
7.290.02
0.4630.03
2.640.04
3.550.01
9.080.01
8.880.01
7.430.03
7.220.01
7.580.02
7.280.05
0.4690.07
2.580.07
4.090.01
9.080.03
8.760.04
5. Conclusions
The proposed system was able to perform all steps
required for the titration without any operator assistance. The combination of the binary search concept
and the monosegmented ow approach provided possibilities to locate the titration end point with good
precision. The acidity of the samples was determined
without using analytical curves. The system can work
without removing the gas bubbles. Their passage
through the electrode cause signal uctuations which
do no affect the analytical signal.
Acknowledgements
The authors are grateful to E.A.G. Zagatto for
critical comments. FAPESP, FINEP/PRONEX, CAPES
and CNPq are thanked for nancial support.
References
[1] H. Zeigel, Z. Anal. Chem. 53 (1914) 755.
[2] W.J. Blaedel, R.H. Laessig, Anal. Chem. 36 (1964) 1617.
[3] W.J. Blaedel, R.H. Laessig, Anal. Chem. 37 (1965) 332.
173