Decision Criteria
Fisher E600: 50.6 cpm
Fluke 451: 0.109 Sv/h
Efficiencies
Efficiencies are calculated for the E600 unit measurements. Measurements were taken
using button sources at approximately 1 cm from the pancake probes center. Calculating
efficiencies for the 451 unit does not make sense as this unit measures dose rate with an
ionization chamber.
Source
204
Tl
137
Cs
60
Co
54
Mn
22
Na
133
Ba
Efficiency
2.12 0.02 %
9.5191 0.0003%
1.0398 0.0002%
4.37 .005%
6.89 0.003%
0.6527 .00006%
Counts diminish drastically when the pancake probe is turned 180 degrees away from the
source. The sources containing only beta and low energy photons have nearly no signal in this
orientation. It appears that either the beta particles cannot penetrate the plastic or the company
inserts some material to eliminate beta signals for different applications.
m/s. Slower rates may allow for the use of these detectors. But a device to measure precise
sweep speeds should be used for further analysis.