You are on page 1of 2

MS100 SpO2 Simulator

Introduction
MS100 SpO2 Simulator is a kind of Separated SpO2 simulator, small and light. It can perform a series of tests for the oximeter b
y simulation means, and gives cognizance of veracity about the oximeter. Because the oximeter manufacturer may use different
R-curves, some R-curves used abroad currently have been embedded.

Main Features
■ Separated connection between simulator probe and host, operation and test are more convenient.

■ 262K color and 320*240 TFT, adjustable brightness level.

■ Film key-press, makes the operation more comfortable.

■ Power with chargeable lithium battery, can display the charge information of battery.

■ Key volume, can open or close


Features:
1.Power with chargeable lithium battery, can display the charge information of battery.
2.Separated connection between simulator probe and host, operation and test are more convenient.
3.262K color and 320*240 TFT, adjustable brightness level.
4.Film key-press, makes the operation more comfortable.
5.Key volume, can open or close.

Main Functions
1.Patient state simulation preset
2.Oxygen saturation simulation;
3.Pulse rate simulation.
4.This Spo2 simulator can measure the following items:
-Reaction time of device;
-SpO2 and pulse rate under different amplitude;
-Performance of device under different interference source;
-Select different R-curve.

Main Parameters

1.Measuring range: 35%~100%


2.Resolution:1%
SpO2 3.Error: When SpO2 is 75%~100%, the error is the larger between ±2% and ±P(P denotes the
precision of the tested oximeter); when SpO2 is 74%~50%, the error is the larger between
±2% and ±P; no define when SpO2 is less than 50%.
1.Measuring range: 25bpm~250bpm
Pulse rate 2.Resolution: 5bpm
3.Error: 1%± 1bpm
1.Measuring range:0.000%~20.000%
Amplitude 2.Resolution:1% for 1.000%~20.000%; 0.1% for 0.100%~0.900%; 0.025% for
0.000%~0.075%.
Patient 24 conditions presetting and 8 conditions as default after start. The number of condition can be
state adjusted by Motion level setup.
Voltage DC 3.6V~4.2V
Power
Voltage 3.7 rechargeable lithium battery × 1
Supply
Battery
working Charge and discharge no less than 500 times
life

Accessaries
A MS100 user manual
A 5.0V adaptor