Professional Documents
Culture Documents
tories and has found broad application to biochemical, biomedical, microbiological, agricultural, and ecological research.
Presently, a large number of lipid analysts use the FID to
measure FAME separated by GC. Although FID has proven to
be a robust tool for FAME determination, the lack of any selectivity can limit the usefulness of this detector when applied
to complicated samples, since only instrument response and retention time information may be gathered. Owing in large part
to this limitation, misidentification of FAME in the presence of
contaminants, artifacts, or coeluting compounds is still of concern when using FID (35). Thus, much work has been done to
maximize the usefulness of retention time for FAME identification through methods such as retention time locking, retention time prediction, and the dependence of retention time on
FAME equivalent chain lengths (68); however, FAME identities assigned by such methods are generally considered tentative (9). Hence, FID analysis of complex biological extracts
may prove inadequate in some situations, particularly for
FAME of relatively low abundance (10).
With the coupling of MS methods to GC, much has been accomplished in the area of qualitative characterization of FAME
mixtures. Since GCMS provides spectrometric information
on separated compounds, it provides a means of analyte selectivity; thus, detection with MS also represents a potentially
powerful tool for quantitative analysis of FAME, especially in
the presence of a convoluted biochemical background. Despite
the prospective benefits of GCMS methodologies for quantitative FAME analysis, the more familiar FID is still favored in
many laboratories, particularly among lipid specialists.
GCMS offers a host of tools for qualitative characterization of FA. For example, standard 70 eV EI ionization of picolinyl, dimethyloxazoline, pyrrolidide, pentafluorodimethylsilyl,
and trimethylsilyl derivatives of FA has been extensively studied
for the purpose of structural determination (1117). EI ionization of these less common FA derivatives yields fragments of diagnostic value in locating the positions of branching and in some
cases the positions of unsaturation in FA (although the geometric configuration of the double bond is not forthcoming).
The simplest method of FAME quantification by EIMS
may be carried out by monitoring a range of mass-to-charge
(m/z) values that will encompass the fragments expected of the
analytes, then determining amount based on integration of
419
420
TABLE 1
Analyte FAME Present in the Calibration Standards and Their Order of Elutiona
Elution
order
FAME
carbon numer
1
2
3
4
5
6
7
8
10
9
11
12
13
15
17
14:0
16:0
16:1n-7
17:0
18:0
18:1n-9
18:1n-7
18:2n-6
18:3n-3
19:0
20:0
20:1n-9
20:2n-6
20:4n-6
20:5n-3
14
16
18
19
21:0
22:0
22:1n-9
22:4n-6
20
22:5n-3
21
22:6n-3
Systematic name
Common name
Methyl myristate
Methyl palmitate
Methyl palmitoleate
Methyl margarate
Methyl stearate
Methyl oleate
Methyl cis-vaccenate
Methyl linoleate
Methyl linolenate
None
Methyl arachidate
Methyl gondoate
None
Methyl arachadonate
ranging from 0.02 to 100.0 g/mL across the series. As an internal standard, 21:0 FAME was added to each mixture at a
concentration of 50.0 g/mL. All standards were analyzed in
quadruplicate by GCFID and each of the GCMS methods.
RF standard. A commercially available standard mixture of
37 FAME (Supelco) was used in the determination of RF. The
original solution was diluted 10-fold to give a final concentration of 1.0 mg/mL total FAME. The RF standard was also
spiked with the FAME of 21:0 to obtain a 50.0 g/mL internal
standard concentration. This preparation was then analyzed in
quadruplicate by GCFID and all GCMS methods under
study.
NIST SRM. The NIST SRM 1946 (Gaithersburg, MD), Lake
Superior fish tissue homogenate, was used to evaluate the usefulness of each method in analysis of a realistic sample (19).
The tissue homogenate was extracted using a Dionex ASE 200
accelerated solvent extractor (Sunnyvale, CA) (20). Conditions
similar to those previously described for lipid extraction were
used (2127).
To prepare for extraction, the sample was removed from
80C storage and allowed to thaw. A 100-mg portion of the
tissue homogenate was weighed and combined with approximately 1 g hydromatrix drying agent (Varian, Walnut Creek,
CA). The tissue and hydromatrix mixture were transferred to
an 11-mL stainless steel extraction cell fitted with three cellulose filters, and additional hydromatrix was added to fill the
cell. The sample was then extracted with 60% chloroform/40%
methanol (obtained, respectively, from EM Science and Fisher
Scientific, Fair Lawn, NJ) at 100C and 13.8 MPa. Two static
extraction cycles were carried out, each 5 min in duration. Both
extraction solvents were residue-analysis grade and were
treated with 100 mg/L BHT as an antioxidant (Sigma). The
crude extract was dried by pouring it through a sintered glass
funnel containing several grams of anhydrous sodium sulfate
(J.T.Baker, Phillipsburg, NJ). This was followed by thorough
rinsing of the funnel with chloroform. The pooled dried extract
and rinsing were then concentrated to approximately 1 mL
under a stream of nitrogen using a TurboVap apparatus set at a
bath temperature of 50C (Zymark, Hopkinton, MA). The extract concentrate was transferred to a round-bottomed reaction
vessel, and the remaining solvent was evaporated by impinging with nitrogen. The recovered lipids were reconstituted in 1
mL 0.5 M methanolic KOH (VWR, West Chester, PA) and hydrolyzed at 80C for 30 min. Once the reaction mixture had
cooled to room temperature, 1 mL of freshly opened 10% BF3
in methanol was added (Supelco). Transesterification was performed at 100C for 10 min. After cooling, 1 mL of distilled
water and 2 mL of hexane were added to the sample with vortexing. Following phase separation, the organic phase was collected and transferred to a new vessel. The solvent exchange
was repeated with a fresh 2-mL portion of hexane. The recovered organic phases were pooled, spiked with the methyl ester
of 21:0 to result in a final concentration of 50.0 g/mL, and diluted to a total final volume of 10 mL with hexane. This sample was analyzed in quadruplicate by each detection method.
421
422
TABLE 2
Quantification Ions Used in SIM and SIEa
FAME
12:0
13:0
14:0
14:1n-9
15:0
15:1n-5
16:0
16:1n-7
17:0
17:1n-7
18:0
18:1n-9 trans
18:1n-9 cis
18:1n-7
18:2n-6 trans
18:2n-6 cis
18:3n-6
18:3n-3
19:0
20:0
20:1n-9
20:2n-6
20:3n-6
20:3n-3
20:4n-6
20:5n-3
21:0
22:0
22:1n-9
22:2n-6
22:4n-6
22:5n-3
22:6n-3
23:0
24:0
24:1n-9
43 + 74 + 87
43 + 74 + 87
43 + 74 + 87
41 + 55 + 67
43 + 74 + 87
41 + 55 + 67
43 + 74 + 87
41 + 55 + 67
43 + 74 + 87
41 + 55 + 67
43 + 74 + 87
41 + 55 + 67
41 + 55 + 67
41 + 55 + 67
67 + 81 + 95
67 + 81 + 95
67 + 79 + 93
67 + 79 + 93
43 + 74 + 87
43 + 74 + 87
41 + 55 + 69
67 + 81 + 95
67 + 79 + 83
67 + 79 + 93
67 + 79 + 91
67 + 79 + 91
43 + 74 + 87
43 + 74 + 87
41 + 55 + 69
67 + 81 + 95
67 + 79 + 91
67 + 79 + 91
67 + 79 + 91
43 + 74 + 87
43 + 74 + 87
41 + 55 + 69
FAME listed here but absent from Table 1 were unique to the response factor standard. SIM, selective ion monitoring; SIE, selected ion extraction; QP,
quadrupole; IT, ion trap.
[1]
3sb
m
[2]
where m is the slope of the linear fit. Since second-order regression was used to fit a number of the MS calibrations, the LOD
for all FAME were based on first-order regression of the low
concentration portions of all standard curves (i.e., the first four
standard levels detectable by a given method), a region in
which linear response was observed regardless of detection
method.
The resultant LOD values for a number of analyte FAME
are given in terms of both solution concentration and the corresponding number of picomoles (Table 4). Each of the quantification methods is fully capable of detecting low-picomole
amounts of a variety of FAME. Detection of subpicomole
quantities of almost all the FAME in Table 4 is readily accomplished by FID, with remarkably similar LOD observed for all
FAME considered; in general, however, MS quantification
methods proved to have LOD that were more sensitive to the
identity of the FAME being detected. Of the MS-based quantification methods, QP-TIC generally had the highest LOD;
however, when this instrument was operated in SIM mode,
LOD were much improved (by as much as a factor of five), rivaling those of any other method. Figure 3 serves to underscore
the advantage of quantification by SIM as opposed to TIC
when using the QP instrument. For the IT instrument, however,
similar improvements in LOD were not achieved in SIE vs.
TIC modes. Thus, whereas quantification based on SIE in this
case yields marginal improvement in LOD as compared with
the TIC and still allows full-scan mass spectra to be acquired,
SIM offers much more significant improvements in the LOD
while precluding the acquisition of full-scan mass spectra.
RF. To investigate the response dependence of each method
on the number of unsaturations as well as the FAME carbon
number, the mean response of each analyte with respect to the
internal standard was used to calculate the analyte RF according to Equation 3:
RFa =
RaCs
Rs Ca
[3]
where the RF of an analyte (RFa) is given in terms of the analyte response (Ra), the internal standard response (Rs), and the
423
FIG. 1. Gas chromatograms of a FAME calibration standard with detection by FID (a), quadrupole (QP) MS (b), and
ion trap (IT) MS (c). For (b) and (c), the total ion counts (TIC) are shown for m/z 40400. All FAME were present at
50.0 g/mL. The peaks numbered in (a) correspond to the compounds listed in Table 1. The same order of elution
applies to (b) and (c). The internal standard is labeled with an asterisk.
[4]
424
FID
QP-TIC
QP-SIM
IT-TIC
IT-SIE
Mean AR RSD (%) Mean AR RSD (%) Mean AR RSD (%) Mean AR RSD (%) Mean AR RSD (%)
0.154
0.165
0.165
0.163
0.171
0.159
0.171
0.180
0.179
0.136
0.158
1.26
0.79
0.44
0.62
0.62
0.55
0.22
0.55
0.46
2.18
2.10
0.146
0.151
0.130
0.151
0.135
0.112
0.108
0.146
0.107
0.072
0.083
2.44
4.29
2.14
3.88
2.95
2.98
3.33
2.27
3.54
5.34
2.23
0.181
0.180
0.089
0.167
0.085
0.074
0.077
0.092
0.070
0.052
0.061
0.67
0.66
0.34
0.87
0.96
1.10
0.99
0.39
1.14
1.34
1.69
0.155
0.179
0.180
0.182
0.194
0.171
0.160
0.163
0.131
0.084
0.102
3.94
4.22
4.03
3.14
3.82
4.19
3.71
0.98
2.85
6.66
3.76
0.272
0.276
0.103
0.251
0.159
0.215
0.176
0.126
0.130
0.101
0.118
3.01
3.74
3.84
2.97
3.52
3.22
3.45
2.34
0.13
2.44
1.58
In all cases, the mean area ratio (AR) vs. the internal standard and relative standard deviation (RSD) of the AR are based on
quadruplicate analyses of a 10 g/mL standard FAME mixture. TIC, total ion counts; for other abbreviations see Table 2.
TABLE 4
Limits of Detection (LOD) for a Suite of the Standard FAME as Determined by the Various Methods of Analysisa
FID
FAME
14:0
16:0
16:1n-7
18:0
18:1n-9
18:1n-7
18:2n-6
18:3n-3
20:0
20:1n-9
20:2n-6
22:1n-9
22:4n-6
22:5n-3
22:6n-3
QP-TIC
QP-SIM
IT-TIC
IT-SIE
LOD
(g/mL)
LOD
(pmol)
LOD
(g/mL)
LOD
(pmol)
LOD
(g/mL)
LOD
(pmol)
LOD
(g/mL)
LOD
(pmol)
LOD
(g/mL)
LOD
(pmol)
0.12
0.05
0.11
0.13
0.16
0.15
0.13
0.14
0.15
0.12
0.11
0.14
0.32
0.24
0.54
0.50
0.19
0.41
0.44
0.54
0.51
0.44
0.48
0.46
0.37
0.34
0.40
0.92
0.70
1.58
0.61
0.47
0.86
0.64
0.56
0.58
0.92
1.15
0.40
0.52
0.47
0.81
0.65
0.85
0.21
2.52
1.74
3.21
2.15
1.89
1.96
3.13
3.94
1.23
1.60
1.46
2.30
1.88
2.47
0.61
0.11
0.16
0.18
0.21
0.23
0.23
0.25
0.27
0.23
0.24
0.24
0.20
0.20
0.22
0.25
0.45
0.59
0.67
0.70
0.78
0.78
0.85
0.92
0.71
0.74
0.75
0.57
0.58
0.64
0.73
0.20
0.20
0.23
0.19
0.32
0.34
0.28
0.24
0.26
0.38
0.40
0.53
0.44
0.50
0.78
0.83
0.74
0.86
0.64
1.08
1.15
0.95
0.82
0.80
1.17
1.24
1.51
1.27
1.45
2.28
0.09
0.08
0.15
0.09
0.17
0.81
0.20
0.19
0.15
0.18
0.19
0.23
0.42
0.55
0.52
0.37
0.30
0.56
0.30
0.57
2.74
0.68
0.65
0.46
0.56
0.59
0.65
1.21
1.60
1.52
425
isfactory (2127), these different approaches may be partly responsible for some of the minor discrepancies between the present results and the NIST SRM certified values; however, since
the same extract was analyzed by all methods, the results are
indeed suitable for direct comparison among the various quantification methods.
Overall, these results demonstrate that appropriately chosen
and properly calibrated MS detection methods are capable of
producing accurate quantitative results for FAME in a biological sample that are of comparable quality to those produced by
the more traditional FID.
DISCUSSION
FIG. 2. Calibration curves for the 18:0 FAME as determined by FID (a),
QP-MS (b), and IT-MS (c). Error bars, where large enough to be visible,
represent the SD. SIM, selective ion monitoring; SIE, selected ion extraction; for other abbreviations see Figure 1.
sample. The present method made use of high-pressure, hightemperature accelerated solvent extraction with chloroform/
methanol, employing less than 20 mL of solvent to extract 100
mg of tissue homogenate in roughly 20 min. By contrast, the
SRM values were determined by 1824 h Soxhlet extractions
of 2.5 g tissue homogenate samples with hexane/acetone. Although the performance of accelerated solvent extraction for
lipid extraction has been addressed elsewhere and deemed sat-
The present findings demonstrate that, although FID and various MS techniques exhibit distinct behavior where response to
FAME for quantification is concerned, the use of calibration
with internal standardization allows the application of MS to
quantitative FAME analysis with adequate precision, low LOD,
and accurate quantitative results. The hallmarks of FID performance in FAME quantification include a broad range of linear
response, excellent precision, and subpicomole LOD. Of
course, the main limitation of this detector is that the response
is nonspecific and provides no information on the identity of
the analyte.
Overall, the best MS performance was obtained using the
QP instrument operated in SIM mode. Although some selectivity was afforded by virtue of the monitoring of selected ions,
the ability to acquire complete mass spectra was precluded.
Whereas this may be satisfactory for quantitative analysis of
well-characterized samples, samples of unknown FAME composition should be screened with full-scan MS prior to SIM
analysis for quantification. QP-SIM allowed better linearity of
response and better precision when compared with acquisition
of TIC on the same instrument and when compared with the IT
instrument. Subpicomole LOD were also obtainable in SIM
mode, a marginal improvement in the TIC LOD (typically, a
few picomoles). Even so, it should be noted that the majority
of these results occur within a rather small range regardless of
detection method, with each method being capable of a level
of performance that is acceptable for many applications.
The response of the IT instrument was less linear and, as has
been reported, gave slightly different spectra from those produced using QP instruments (33). The main advantage of the
IT instrument was that acquisition of full-scan mass spectra
(IT-TIC) gave better quantitative performance than the QPTIC; thus, although QP-SIM offered the best MS performance
overall, IT-TIC was superior to QP-TIC for quantification with
full-scan acquisition. The IT-MS analysis also afforded the
ability to collect full spectra, yet still offered the capability to
extract ions for a given analyte later if enhanced signal-to-noise
was needed. Indeed, SIE of FAME chromatograms did allow
some improvement in LOD and method precision.
These results also support the need for rigorous calibration,
irrespective of the detection method chosen. Whereas appropriate calibration is particularly critical for MS work, the assumption of uniform FID RF is also inadequate for precise
Lipids, Vol. 40, no. 4 (2005)
426
FIG. 3. Expanded chromatograms at the retention time of 18:0 FAME with detection by QP-MS showing the TIC (a)
and SIM of m/z 43 + 74 + 87 (b). Approximately 20 pg of the 18:0 FAME was injected in both cases. For abbreviations see Figures 1 and 2.
REFERENCES
1. James, A.T., and Martin, A.J.P. (1952) GasLiquid Partition
Chromatography: The Separation and Micro-estimation of
Volatile Fatty Acids from Formic Acid to Dodecanoic Acid,
Biochem. J. 50, 679690.
2. Cropper, F.R., and Heywood, A. (1953) Analytical Separation
of the Methyl Esters of the C12C22 Fatty Acids by VapourPhase Chromatography, Nature 172, 11011102.
3. Ackman, R.G. (1990) Misidentification of Fatty Acid Methyl
Ester Peaks in Liquid Canola Shortening, J. Am. Oil Chem. Soc.
67, 1028.
4. Hawrysh, Z.J., Shand, P.J., Lin, C., Tokarska, B., and Hardin,
R.T. (1990) Efficacy of Tertiary Butylhydroquinone on the Storage and Heat Stability of Liquid Canola Shortening, J. Am. Oil
Chem. Soc. 67, 585590.
5. Johnson, A.R., Fogerty, A.C., Hood, R.L., Kozuharov, S., and
Ford, G.L. (1976) GasLiquid Chromatography of Ethyl Ester
Artifacts Formed During the Preparation of Fatty Acid Methyl
Esters, J. Lipid Res. 17, 431432.
6. David, F., Sandra, P., and Wylie, P.L. (2002) Improving Analysis of Fatty Acid Methyl Esters Using Retention Time Locked
Methods and Retention Time Databases, Agilent Technologies
Application Note 5988-5871EN, Palo Alto, CA.
7. Torres, A.G., Trugo, N.M.F., and Trugo, L.C. (2002) Mathematical Method for the Prediction of Retention Times of Fatty Acid
Methyl Esters in Temperature-Programmed Capillary Gas Chromatography, J. Agric. Food Chem. 50, 41564163.
8. Mjos, S.A. (2003) Identification of Fatty Acids in Gas Chromatography by Application of Different Temperature Programs on
a Single Capillary Column, J. Chromatogr. A 1015, 151161.
9. Eder, K. (1995) Gas Chromatographic Analysis of Fatty Acid
Methyl Esters, J. Chromatogr. B 671, 113131.
10. Roach, J.A.G., Yurawecz, M.P., Kramer, J.K.G., Mossoba,
427
FIG. 4. Response factors for a series of saturated FAME (a) and a series of unsaturated FAME (b)
as determined by each detection method. Error bars represent the SD. For abbreviations see
Figure 2.
11.
12.
13.
14.
15.
16.
17.
18.
19.
428
FIG. 5. Mean mass percentage (as TG) of representative certified FA in the NIST Standard Reference Material (SRM) 1946 as determined by each method. Error bars represent the SD, except for those associated with the NIST SRM values; for these values, the error bars represent
the uncertainties reported by NIST. For other abbreviations see Figure 2.
20. Richter, B.E., Jones, B.A., Ezzell, J.L., Porter, N.L., Avdalovic,
N., and Pohl, C. (1996) Accelerated Solvent Extraction: A Technique for Sample Preparation, Anal. Chem. 68, 10331039.
21. Schafer, K. (1998) Accelerated Solvent Extraction of Lipids for
Determining the Fatty Acid Composition of Biological Material, Anal. Chim. Acta 358, 6977.
22. Boselli, E., Velazco, V., Caboni, M.F., and Lercker, G. (2001)
Pressurized Liquid Extraction of Lipids for the Determination
of Oxysterols in Egg-Containing Food, J. Chromatogr. A 917,
239244.
23. Richardson, R.K. (2001) Determination of Fat in Dairy Products
Using Pressurized Solvent Extraction, J. Assoc. Off. Anal.
Chem. Int. 84, 15221533.
24. Moreau, R.A., Powell, M.J., and Singh, V. (2003) Pressurized
Liquid Extraction of Polar and Nonpolar Lipids in Corn and
Oats with Hexane, Methylene Chloride, Isopropanol, and Ethanol, J. Am. Oil Chem. Soc. 80, 10631067.
25. Toschi, T.G., Bendini, A., Ricci, A., and Lercker, G. (2003)
Pressurized Solvent Extraction of Total Lipids in Poultry Meat,
Food Chem. 83, 551555.
26. Zhuang, W., McKague, B., Reeve, D., and Carrey, J. (2004) A
Comparative Evaluation of Accelerated Solvent Extraction and
Polytron Extraction for Quantification of Lipids and Extractable
Organochlorine in Fish, Chemosphere 54, 467480.
27. Dodds, E.D., McCoy, M.R., Geldenhuys, A., Rea, L.D., and
Kennish, J.M. (2004) Microscale Recovery of Total Lipids from
Fish Tissue by Accelerated Solvent Extraction, J. Am. Oil Chem.
Soc. 81, 835840.
28. Miller, J.N., and Miller, J.C. (2000) Statistics and Chemometrics for Analytical Chemistry, 4th edn., pp. 120123, Pearson
Education, Essex.
29. Ackman, R.G. (2002) The Gas Chromatograph in Practical
Analyses of Common and Uncommon Fatty Acids for the 21st
Century, Anal. Chim. Acta 465, 175192.
30. Ulberth, F., Gabernig, R.G., and Schrammel, F. (1999) FlameIonization Detector Response to Methyl, Ethyl, Propyl, and
Butyl Esters of Fatty Acids, J. Am. Oil Chem. Soc. 76, 263266.
31. Ackman, R.G., and Sipos, J.C. (1964) Application of Specific
Response Factors in the Gas Chromatographic Analysis of
Methyl Esters of Fatty Acids with Flame Ionization Detectors,
J. Am. Oil Chem. Soc. 41, 377378.
32. Craske, J.D., and Bannon, C.D. (1987) Gas Liquid Chromatography Analysis of the Fatty Acid Composition of Fats and Oils:
A Total System for High Accuracy, J. Am. Oil Chem. Soc. 64,
14131417.
33. Horman, I., and Traitler, H. (1989) Routine Gas Chromatographic/Mass Spectrometric Analysis of Fatty Acid Methyl Esters Using the Ion Trap Detector, Biomed. Environ. Mass Spectrom. 18, 10161022.
34. Seppanen-Laakso, T., Laakso, I., and Hiltunen, R. (2002)
Analysis of Fatty Acids by Gas Chromatography, and Its Relevance to Research on Health and Nutrition, Anal. Chim. Acta
465, 3962.
[Received November 18, 2004; accepted March 28, 2005]