Professional Documents
Culture Documents
Cibirka 2001
Cibirka 2001
Roman M. Cibirka, DDS, MS,a Steven K. Nelson, DMD,b Brien R. Lang, DDS, MS,c and
Frederick A. Rueggeberg, DDS, MSd
School of Dentistry, Medical College of Georgia, Augusta, Ga., and School of Dentistry, University of
Michigan, Ann Arbor, Mich.
Purpose. This study examined potential differences in detorque values of abutment screws after fatigue
testing when the dimensions between external implant hexagon and internal abutment hexagon were
altered or the implant external hexagonal shape was eliminated.
Material and methods. Three subsets (N = 10) of NobelBiocare implants were assessed: (1) standard
external hexagon (R), (2) modified hexagon (M), and (3) circular (C) platform geometry. Thirty Procera
machined abutments with 25-degree angulated loading platforms were manufactured. Abutments were
retained with gold Unigrip abutment screws tightened to 32 N/cm with an electronic torque controller.
Vertical scribes across the implantabutment interface allowed longitudinal displacement evaluation. A
carousel-type fatigue testing device delivered dynamic loading forces between 20 and 200 N for
5,000,000 cycles, or the approximate equivalent of 5 years in vivo mastication, through a piston to the
abutment platform. Macroscopic and radiographic examination of the implant/abutment specimens was
performed. The abutment screws were removed and the detorque values recorded. Bearing surfaces were
examined microscopically.
Results. No abutment looseness or longitudinal displacements at the implantabutment interface were
noted. Radiographic examination demonstrated no indication of screw bending or displacement. The mean
detorque values for R, M, and C were 14.40 1.84 N/cm, 14.70 1.89 N/cm, and 16.40 2.17 N/cm,
respectively. The analysis of variance demonstrated significant differences between only designs R and C
(P=.031).
Conclusion. Increasing the vertical height, or degree of fit tolerance, between the implant external
hexagon and the abutment internal hexagon or completely eliminating the implant external hexagon did
not produce a significant effect on the detorque values of the abutment screws after 5,000,000 cycles in
fatigue testing, or the equivalent of 5 years of mastication for the implant/abutment specimens evaluated.
(J Prosthet Dent 2001;85:268-75.)
CLINICAL IMPLICATIONS
Clinicians using implant and abutment systems with an external hexagonal design
and gold alloy screw requiring 32 N/cm tightening torque may infer that implant
hexagons only aid surgical placement into bone and perhaps orientation of the abutment to the implant. The implant hexagon may not influence force distribution or
rotation resistance and may therefore affect the attainment of optimum preload during initial abutment screw tightening.
uccessful implant therapy requires a dynamic equilibrium between biologic and mechanical factors. The
biologic factors are generally considered multifactorial,
CIBIRKA ET AL
applied to the screw joint cause thread slippage, contributing to release of the stretch, or preload, of the
screw. The second stage of loosening involves continual preload reduction below a critical level, allowing
threads to turn and loss of intended screw joint function. Screw loosening has been reported as a function
of implant and prosthetic component design.2,3
Loosening with hexagonal screw joint design ranged
between 6% and 48%.4-17 The fit tolerance between
the external implant hex and the internal abutment
hex has been suggested as a primary cause of screw
loosening by clinicians and researchers.
Numerous investigators have examined the
implantabutment screw joint.15-47 Many studies have
dealt with the assembly mechanisms and reported that
an inaccurate interface places excessive stresses on the
abutment screw joint, creating instability.29,45
Binon29,46,47 stated that rotational movement or stability of the abutment screw was directly correlated
with the fit tolerances of the flat-to-flat of the implant
hexagon to the abutment internal hexagon walls.
Binon29,46,47 further suggested that a 0.005 mm mean
flat-to-flat measure on the same hex and a total flat-toflat range of less than 0.015 mm for the entire
hexagonal design would result in a more stable screw
joint. It has also been suggested that the hexagon
height extending from the implant-bearing surface
may contribute to screw joint stability. Ohrnell et al26
recommended that the external hexagon connection
be a minimum of 1.2 mm in height to provide both
lateral and rotational stability, particularly in
single-tooth applications.
Conceptually, a more accurate hex-to-hex fit tolerance will improve force distribution to the supporting
bone, allow less relative movement of the abutment to
the implant fixture, and reduce the potential for screw
loosening and bending. Likewise, a higher hexagon
has been suggested to reduce screw loosening and
bending. Short, narrow external geometry of the
hexagon appears to be particularly vulnerable to loosening because of the limited engagement of the
external member and the presence of a short fulcrum
point.24,25 Overall, most investigators claim the better
the maintenance of preload, the better the long-term
stability of the joint.
Fatigue testing with different fit tolerances may be
used to evaluate the hexagonal screw joint stability. In
this type of assessment, the abutment screw is tightened using the recommended or optimum torque.
After fatigue testing, the screw is detorqued, and the
measured value of the specimen is recorded. In this
protocol, 3 assumptions are necessary: (1) the abutment screws are all loaded with the same preload after
initial tightening; (2) the detorque value is a measure
of the remaining preload in the abutment screw; and
(3) any differences in the detorque value from speciMARCH 2001
CIBIRKA ET AL
Table I. Width and height comparisons for specimens (n = 10 per experimental group)
Specimen
R
M
C
A
*Parentheses
2.700
2.664
2.668
2.718
(0.004)
(0.016)
(0.003)
(0.041)
0.000
0.035
0.031
(0.018)
0.633
0.608
0.668
0.905
(0.005)
(0.003)
(0.003)
(0.005)
0.000
0.025
(0.034)
(0.272)
CIBIRKA ET AL
B
Fig. 4. A, Custom Procera abutment, Unigrip screw, and regular platform Branemark implant
fixture. B, Loading platform of custom Procera abutment.
RESULTS
Examination of specimen interfaces within each
design group demonstrated no abutment loosening or
rotational displacement at the implantabutment
interface when the vertically scribed lines were evaluated under high resolution (25) (Fig. 8). Radiographic examination demonstrated no indication of
screw bending or displacement after fatigue testing.
Space was visually confirmed between the implant
geometry and the walls of the abutments when assessMARCH 2001
ing radiographs for all specimens. Neither the hexagons for designs R and M nor the circular form of
design C was in contact with the floor of the internal
abutment hexagons (Fig. 7).
Mean detorque values and standard deviations
are presented in Table II. For design R, the mean
detorque value was 14.40 N/cm (SD 1.84 N/cm);
for design M, the mean value was 14.70 N/cm
(SD 1.89 N/cm); and for design C, the mean
value was 16.40 N/cm (SD 2.17 N/cm). The
analysis of variance demonstrated significant differences between only designs R and N (P=.031).
DISCUSSION
The effect of interface design on stability of the
screw joint connecting implant and abutment remains
uncertain, as is evidenced by the many configurations
that exist in todays market. There is no compelling
evidence to support 1 interface design over another.
The external hexagon geometry remains the most
common design in use. However, the crucial issue with
any design, be it hexagonal or otherwise, is the effect
of the design and its machining tolerances in maintaining stability of the screw joint.
Developers reported that 1 purpose of the implant
external hexagon was for driver fixation to assist with
surgical placement. Others indicate that this design
merely assists in orientation of the abutment onto the
implant. Still others suggest that the hexagon provides
an important relationship for the abutment to the
implant and the eventual force transfer to the implant
and supporting bone. Many of those who support this
latter role of hexagons further state that lack of precision
fit between the 2 hexagons will generate screw joint
instability over time. This premise suggests that instability between components ultimately may result in screw
loosening, restoration failure, and possibly the loss of
the bony attachment at the implantbone interface.
271
Fig. 5. Holding device for electronic torque control tightening of abutment screws.
CIBIRKA ET AL
Fig. 7. A, Radiographic image perpendicular to offset bearing table of abutment after loading.
B, Radiographic image in line with offset bearing table of abutment after loading.
CIBIRKA ET AL
R-experiment
M-experiment
C-experiment
Mean (N/cm)
SD
14.40
14.70
16.40
1.84
1.89
2.17
CONCLUSIONS
Within the limits of this study, the following conclusions were drawn:
1. Increasing the distance between the implant
external hexagon width and the internal hexagon of an
abutment did not produce a statistically significant
effect on the detorque values of the abutment screw
after 5,000,000 cycles in fatigue testing.
2. Eliminating the implant external hexagon and
274
CIBIRKA ET AL
increasing the circular height form resulted in a significant effect on the gold alloy abutment screw detorque
values after 5,000,000 cycles (P=.034) compared with
implants having regular hexagons.
3. After 5,000,000 fatigue testing cycles of
implant/abutment specimens with and without an
external hexagon design, no subjective clinical signs of
screw instability or loosening were observed.
We would like to thank Mr Rui Wang, Ms Beth Lang, and Mr
Don Mettenburg for their invaluable support in the completion of
this project.
REFERENCES
1. Bickford JH. An introduction to the design and behavior of bolted joints.
New York, NY: Marcel Dekker, Inc; 1981.
2. Binon PP, Weir D, Watanabe L, Walker L. Implant component compatibility. In: Laney WR, Tolman DE, editors. Tissue integration in oral and
maxillofacial reconstruction. Chicago, IL: Quintessence Publishing Co;
1990.
3. Nichols JI, Basten CH. A comparison of three mechanical properties of
four implant designs. Postgrad Dent 1995;2:4-14.
4. Sones AD. Complications with osseointegrated implants. J Prosthet Dent
1989;62:581-5.
5. Zarb GA, Schmitt A. The longitudinal clinical effectiveness of osseointegrated implants: the Toronto study. Part III: problems and complications
encountered. J Prosthet Dent 1990;64:185-94.
6. Jemt T, Linden B, Lekholm U. Failure and complications in 127 consecutively placed fixed partial prostheses supported by Branemark implants:
from prosthetic treatment to first annual checkup. Int J Oral Maxillofac
Implants 1992;7:40-4.
7. Jemt T, Lekholm U. Oral implant treatment in posterior partially edentulous jaws: a 5-year follow-up report. Int J Oral Maxillofac Implants
1993;8:635-40.
8. Kallus T, Bessing C. Loose gold screws frequently occur in full-arch fixed
prostheses supported by osseointegrated implants after 5 years. Int J Oral
Maxillofac Implants 1994;9:169-78.
9. Jemt T. Fixed implant-supported prostheses in the edentulous maxilla. A
five-year follow-up report. Clin Oral Implants Res 1994;5:142-7.
10. Lekholm U, van Steenberghe D, Herrmann I, Bolender C, Folmer T,
Gunne J, et al. Osseointegrated implants in the treatment of partially
edentulous jaws: a prospective 5-year multicenter study. Int J Oral
Maxillofac Implants 1994;9:627-35.
11. Wie H. Registration of localization, occlusion and occluding materials
for failing screw joints in the Branemark implant system. Clin Oral
Implants Res 1995;6:47-53.
12. Becker W, Becker BE. Replacement of maxillary and mandibular molars
with single endosseous implant restorations: a prospective study. J
Prosthet Dent 1995;74:51-5.
13. Balshi TJ. First molar replacement with an osseointegrated implant.
Quintessence Int 1990;21:61-5.
14. Balshi TJ, Hernandez RE, Pryszlak MC, Rangert B. A comparative study
of one implant versus two replacing a single molar. Int J Oral Maxillofac
Implants 1996;11:372-8.
15. Jemt T, Laney WR, Harris D, Henry PJ, Krogh PHJ Jr, Polizzi G, Zarb GA,
Herrmann I. Osseointegrated implants for single tooth replacement: a 1year report from a multicenter prospective study. Int J Oral Maxillofac
Implants 1991;6:29-36.
16. Jemt T, Pettersson P. A 3-year follow-up study on single implant treatment. J Dent 1993;21:203-8.
17. Walton JN, MacEntee MI. A prospective study on the maintenance of
implant prostheses in private practice. Int J Prosthodont 1997;10:453-8.
18. Rangert B, Jemt T, Jorneus L. Forces and moments on Branemark
implants. Int J Oral Maxillofac Implants 1989;4:241-7.
19. Rangert B, Sullivan R. Learning from historythe transition from full arch
to posterior partial restorations. Nobelpharma News 1995;9:6-7.
20. Jorneus L, Jemt T, Carlsson L. Loads and designs of screw joints for single crowns supported by osseointegrated implants. Int J Oral Maxillofac
Implants 1992;7:353-9.
21. Rangert BR, Sullivan RM, Jemt TM. Load factor control for implants in the
VOLUME 85 NUMBER 3
CIBIRKA ET AL
22.
23.
24.
25.
26.
27.
28.
29.
30.
31.
32.
33.
34.
35.
36.
37.
38.
39.
MARCH 2001
40.
41.
42.
43.
44.
45.
46.
47.
48.
49.
50.
51.
52.
doi:10.1067/mpr.2001.114266
275