NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)

INDUSTRIAL METROLOGY DIRECTORATE

CERTIFICATE
OF CALIBRATION
Date of Issue: 2016-22

This calibration certificate documents the


traceability to national standards, which realize
the units of measurement according to the
international system of units
PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail:info@nmie.net
website: http://www.nmie.net

Certificate number: PMT-TH-1098

Page 1 of 3

Object

Analog thermometer

Manufacturer

--

Type/Model

--

Serial Number

-LASALLE AGRO PROCESSING P L C


08-0489
3
2016-03-21

Customer
Registration No.:
Number of pages of the certificate
Date of Calibration:

This calibration certificate may not be reproduced other than in full except with the permission
of the issuing authority. Calibration certificates without signature and seal are not
valid.
Seal Date
Calibrated By

Approved

By

2016-03-22
Henok Kebede
Ephrem
W/gebreal
______________________________________________________________________________________________

NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)


INDUSTRIAL METROLOGY DIRECTORATE

CERTIFICATE
OF CALIBRATION
Date of Issue: 2016-22

Calibrated
Object

This calibration certificate documents the


traceability to national standards, which realize
the units of measurement according to the
international system of units
PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail:info@nmie.net
website: http://www.nmie.net

Certificate number: PMT-TH-1098

Page 2 of 3

Type:
Identification No.:

Analog thermometer

Measuring range:
Calibration range:
Scale interval:

--65-105 0c
1 oC

TH-01

Standards
Type - Standard Platinum Resistance Thermometer. C.No, TH-872
Calibration procedure The instrument is calibrated with Standard Platinum Resistance
Thermometer. For selected measurement points the standard and
Place of Calibration
Type Of Calibration
Validity of calibration

instrument readings are reordered & reported on the certificate.


NMIE Temperature Laboratory Addis Ababa
In house Calibration
The values in this certificate are correct at the time of calibration.
Subsequently the accuracy will depend on such factors as the care
exercised in handling and use of the instrument and the frequency of use.
Recalibration should be performed after a period, which has been chosen
to ensure that the instruments accuracy remains within the desired limits.

Measuring
Result

Standard
Reading [ oC ]
67.18
77.30
82.43
87.44

Instrument
Reading [ oC ]
66.0
76.0
81.0
86.0

Correction
[ oC ]
1.18
1.30
1.43
1.44

NATIONAL METROLOGY INSTITUTE OF ETHIOPIA (NMIE)


INDUSTRIAL METROLOGY DIRECTORATE

CERTIFICATE
OF CALIBRATION
Date of Issue: 2016-22

This calibration certificate documents the


traceability to national standards, which realize
the units of measurement according to the
international system of units

Certificate number: PMT-TH-1098

PO BOX 5722
Addis Ababa, Ethiopia
Tel: 251-11- 6517985
Fax: 251-11-6459312
e-mail:info@nmie.net
website: http://www.nmie.net

Page 3 of 3

Note: Standard (Actual) value = Instrument reading + Correction


Estimated uncertainty of measurement: 0.06 o C
The uncertainties are based on root sum square of the contributions with a confidence interval
Not less than 95% and coverage factor k=2.

_____________________________End of Certificate______________________________