You are on page 1of 34

Disassembly Guide for DP6600

Last Revised:

August 14, 2007

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Last Modified: August 14, 2007

Page 2 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

REVISION CONTROL
Document Title:

Disassembly Guide for DP6600

Author:

Mindray Service Department

Date

By

Action

Pages

July 24th,2007

Walker Wu

New document

All

July 25th, 2007

Larry Hao

Add FRU list

Action

Pages

Review/Approval History
Date

By

Last Modified: August 14, 2007

Page 3 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Table of Contents
Page
Note ________________________________________________________________________5
Tools _______________________________________________________________________6
FRU list_____________________________________________________________________7
Disassembly Map ____________________________________________________________12
Removing the top cover _______________________________________________________13
Step1: remove the cable hook ______________________________________________________ 13
Step2: remove the top cover: _______________________________________________________ 13

Removing the power connecting board ___________________________________________15


Removing the rear cover_______________________________________________________16
Removing the power supply board _______________________________________________19
Removing the CPU board and probe board ________________________________________21
Step1: take out the CPU board and probe board_______________________________________ 21
Step2: separate the CPU board and probe board ______________________________________ 22
Step3: disassemble the probe board _________________________________________________ 24

Removing the CRT monitor ____________________________________________________27


Step1: disassemble the front cover __________________________________________________ 27
Step2: remove the CRT monitor ____________________________________________________ 28

Removing the keyboard PCB and trackball________________________________________30


Step1: disassemble the keyboard assembly ___________________________________________ 30
Step2: disassemble the trackball ____________________________________________________ 31
Step3: disassemble the keyboard PCB _______________________________________________ 31

Last Modified: August 14, 2007

Page 4 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Note:

These are the directions for disassembling the base unit. Before
disassembly, make sure the machine is powered off.
---Screw to be removed

---Cable to be disconnected

---moving direction
For any parts you disassembled, please read the guide reversely to
assemble them.
Please keep the screws and dont mix them with each other.

Last Modified: August 14, 2007

Page 5 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Tools

You will need a medium size + screwdriver.

Last Modified: August 14, 2007

Page 6 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

FRU list
DP6600 contains a lot of parts. However, well not replace each single part. Instead, we replace some
combinations of them. We call them Field Replaceable Units or FRUs.
Please note that the name of the FRU might be not exactly the same with service manual. However, the
part number is exactly the same and part number is the only criteria when we talk about parts.
Here is a list of all the FRUs in DP6600:

No. Part number

Part name and description

2300-30-29051

Main board(CPU board)

2300-30-29059

Power supply board

2300-20-29057

Keyboard(Keyboard PCB)

Last Modified: August 14, 2007

Photo

Page in
this
guide

Page 7 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

2300-21-29149

CRT assembly(CRT ASSY)

2300-30-29055

Probe board

2300-20-10893

Tracking ball(Track ball)

2300-30-29053

Connecting board

Last Modified: August 14, 2007

Page 8 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

2300-20-29061

Top cover of keyboard

2300-20-29062

Bottom cover of keyboard

10

2300-30-29156

Top cover keyboard


ASSY(including keyboard &
trackball)

11

2300-30-29155

Bottom cover keyboard


ASSY

12

2102-20-16994

Protection screen

Last Modified: August 14, 2007

Page 9 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

13

2300-20-29081

Handle

14

2102-30-16949

Probe cable Hook

15

2300-20-29082

Transducer Holder

16

2300-20-29078

Front cover

Last Modified: August 14, 2007

Page 10 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

17

2300-20-29079

Top cover

18

2300-20-29080

Bottom cover

Last Modified: August 14, 2007

Page 11 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Disassembly Map

According to the map, if you want to get keyboard out, you dont need to disassemble the covers.
However, if you want to take CPU board out, youll need to disassembly top cover, power
connecting board and rear cover, and then you can disassemble CPU board.

Last Modified: August 14, 2007

Page 12 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Removing the top cover


Step1: remove the cable hook

Step2: remove the top cover:


Disassemble the four rubber caps and four screws, and then you can remove the cover.

Last Modified: August 14, 2007

Page 13 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Remove the top cover

The whole look

Last Modified: August 14, 2007

Page 14 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Removing the power connecting board


Disassemble ten screws and disconnect the power cable for fan as shown in following picture:

Disconnect the power cable, and then take out the cabinet tail board

Last Modified: August 14, 2007

Page 15 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Removing the rear cover


Disconnect the USB cables,

Disassemble the cable hook holder,

Last Modified: August 14, 2007

Page 16 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Disassemble the nine screws:

And disconnect the power cable for fan:

Last Modified: August 14, 2007

Page 17 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Then remove the rear cover.

Last Modified: August 14, 2007

Page 18 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Removing the power supply board


Now you can take out the power box

Disassemble the 6 screws on two sides of the box

Last Modified: August 14, 2007

Page 19 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Disconnect the power cable

Then you can get the power supply board

Last Modified: August 14, 2007

Page 20 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Removing the CPU board and probe board


Step1: take out the CPU board and probe board
Disassemble six screws on the left side

One screw on the right

Last Modified: August 14, 2007

Page 21 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Pull out

Step2: separate the CPU board and probe board


Disassemble the fixing screws

Last Modified: August 14, 2007

Page 22 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Three screws on another side

Unplug

Last Modified: August 14, 2007

Page 23 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

It is like this

Step3: disassemble the probe board


Remove the probe socket fixing plate

Last Modified: August 14, 2007

Page 24 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Remove the connection metal plate

Remove the metal fastener

Last Modified: August 14, 2007

Page 25 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

All the parts relative to probe board

Last Modified: August 14, 2007

Page 26 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Removing the CRT monitor


Step1: disassemble the front cover

Disconnect the cables

Last Modified: August 14, 2007

Page 27 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step2: remove the CRT monitor


Remove all the screws on two sides and on the plate

Last Modified: August 14, 2007

Page 28 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Then you can take away the monitor assembly

Last Modified: August 14, 2007

Page 29 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Removing the keyboard PCB and trackball


Step1: disassemble the keyboard assembly
Remove the rubber caps and two screws

Lift up

Last Modified: August 14, 2007

Page 30 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step2: disassemble the trackball


Remove the 4 screws and cable

The trackball

Last Modified: August 14, 2007

Page 31 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Step3: disassemble the keyboard PCB


Disconnect the buzzer power cable

Remove the 14 screws on the keyboard PCB

Last Modified: August 14, 2007

Page 32 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Pull out the keyboard and silicon rubber cover

Remove the silicon rubber cover

Last Modified: August 14, 2007

Page 33 of 34

SHENZHEN MINDRAY BIO-MEDICAL ELECTRONICS CO., LTD.

Last Modified: August 14, 2007

Page 34 of 34

You might also like