Professional Documents
Culture Documents
Ventilation
Mohammed M. Tamim, MD
Consultant, NICU, Tawam Hospital
Alain, UAE
1
Objectives
Indications of mechanical ventilation
Basics of respiratory mechanics
Modes of conventional ventilation
Modes of HFV
Indications of HFV
2
Introduction
The primary objective of Mechanical Ventilation is to
support breathing until patient respiratory efforts are
sufficient.
First mechanical ventilation for a neonate in 1959.
3
Respiratory Failure
Hypercapnic Respiratory Failure:
Inability to remove CO2 by spontaneous breathing
Caused by hypoventilation or severe V/Q mismatch
in arterial PCO2
in pH
MV is most commonly needed for treatment
Hypoxemia
Usually the result of V/Q mismatch or RL shunt.
Diffusion abnormalities & hypoventilation (apnea)
4
Respiratory Failure
Respiratory failure can occur because of diseases
in the lung, thorax, airway or respiratory muscle.
Indication for assisted ventilation:
Respiratory Acidosis pH < 7.2
Hypoxemia while on 100% O2
Or CPAP of 60 100%
Severe apnea
5
Respiratory Failure
Clinical Manifestation:
Increase or decrease in respiratory rate.
Increase or decrease in respiratory effort.
Apnea.
6
Neonatal Respiratory Physiology
Compliance:-
Distensible nature of lungs and chest wall.
Volume (L)
= ----------------------
Pressure (cm H2O)
Neonates have greater chest wall
compliance.( premature more than FT)
Premature infants with RDS have stiffer lungs
(poorly compliant lungs).
7
Neonatal Respiratory Physiology
Resistance:-
Property of airways and lungs to resist gas.
8
Resistance
9
Neonatal Respiratory Physiology
Time Constant:
An index of how rapidly the lungs can empty.
Time constant = Compliance X Resistance
In BPD time constant is long because of resistance.
10
Time Constant
Inspiratory time must be 3-5 X time constant
One time conststant = time for alveoli to discharge
63% of its volume through the airway.
Two time constant = 84% of the volume leaves
11
Time Constant
12
Conventional Mechanical Ventilation
Mechanical ventilators achieve a pressure
gradient between the airway opening and lungs.
Ventilator for neonates are usually one of the
following types:
Pressure control Ventilators
Volume control Ventilators
13
Conventional Mechanical Ventilation
Pressure Controlled Ventilators:
A constant flow of gas pass through the ventilator.
Pressure is limited to the desired magnitude.
14
Conventional Mechanical Ventilation
Volume Controlled Ventilators:
A preset volume of gas is delivered to the system
after which inspiration is terminated.
When this gas has been delivered by the piston
inspiration is terminated.
Infants TV (4-8 ml/kg)
15
Continuous Positive Airway Pressure
Important tool in management of neonates
Increase alveolar volume
Redistribution of lung water
Cons Pros
Risk of air leak Alveolar Volume & FRC
Over Distension Alveolar Stability
CO2 Retention Redistribution of lung fluids
Cardiovascular impairment Improved V/Q matching
Compliance
16
Pressure Controlled Ventilation
Ventilator Components:
1. Gas mixer.
2. Inspiratory expiratory time adjustment
3. Expiratory relief valve to limit the peak inspiratory
pressure
4. Pressure gauge to measure applied pressure
5. Humidification or nebulization
6. Positive end expiratory pressure to maintain functional
residual capacity.
7. Exhalation assist to reduce the end expiratory pressure
8. Alarms
17
Pressure Controlled Ventilation
Peak Inspiratory Pressure (PIP):
Changes in PIP affect both PaO2 & PaCO2 by altering
the MAP.
Increase in PIP:
Increase in PaO2
Decrease in PaCO2
18
Pressure Controlled Ventilation
Positive End Expiratory Pros Cons
Pressure
Alveolar volume Increased Risk for
Adequate PEEP prevents & FRC air leaks
alveolar collapse and maintain
Alveolar stability Overdistention
lung volume at end of
expiration.
Redistribution of CO2 retention
Improve V/Q matching lung water
Very high PEEP reduce Improved V/Q Cardiovascular
venous return cardiac matching impairment
output decrease oxygen Decreased
transport increase Compliance
pulmonary vascular resistance
19
Pressure Controlled Ventilation
Rate: PROS CONS
Change in rate alter alveolar
minute ventilation Air leak Gas trapping
High rate low TV is
strongly preferred Volutrauma Generalized
atelectasis
Rate change alone with
constant I:E ratio do not CVS side effects Maldistribution of
alter MAP gas
21
Pressure Controlled Ventilation
Inspiratory Expiratory Ratio:
Short Inspiratory time:
Pros:
Faster weaning
Decreased risk for pneumothorax
Allows use of higher ventilator
Cons:
Insufficient tidal volume
May need high flow rate
22
Inspiratory & Expiratory Time
23
Mean Airway Pressure
24
Pressure Controlled Ventilation
FIO2:
Changes alter Alveolar Oxygen Pressure
Flow:
Not well studied in infants
Minimal effects on ABG
26
Carbon Dioxide Elimination
Depends largely on the amount of gas that passes in
and out of the alveoli Minute Ventilation.
Minute Ventilation = TV X Rate
Any increase in TV or Rate will eliminate CO2.
TV may be increased by in PIP or in PEEP
27
Tidal Volume
28
Alternative Modalities of MV
Technology advances including improvement in
flow delivery systems, breath termination criteria,
stability of PEEP, air leak compensation,
prevention of pressure overshoot and triggering
system led to development of new modalities of
mechanical ventilation:
Patient Triggered Ventilation (PTV)
Proportional Assist Ventilation
Tracheal Gas Insufflation
29
Patient Triggered Ventilation
Modified CMV
Neonate is able to initiate ventilatory breath by:
Abdominal motion
Chest wall impedance
Airway flow
30
Patient Triggered Ventilation
Modes:
Synchronized Intermttent Mandatory Ventilation
(SIMV):
Preset rate that is triggered, other patient breath is not
assisted.
Assist Control Mode (A/C):
All breath initiated by patient is triggered.
Weaning accomplished by reducing PIP.
31
Patient Triggered Ventilation
Advantages:
Reduction in cerebral blood flow variability
(Renie et al 1987)
Shorter time on ventilator
(Visveshwara et al 1999)
Improved oxygenation with SIMV
(Cleary et al 1995)
No difference between SIMV & A/C in length of
weaning.
(chon et al 1994)
32
Proportional Assist Ventilation
Synchronize onset & duration of both
inspiratory and expiratory support.
Ventilatory support is in proportion to the
volume and flow of the spontaneous breath.
It will reduce ventilatory pressure while
maintaining or improving gas exchange.
33
Tracheal Gas Insufflation
Reduce anatomical dead space in alveolar
minute ventilation
Gas is delivered to the distal part of the
endotracheal tube
Result in decrease in PCO2 and PIP
Still under study
34
Preparation for MV
Establishment of artificial airway
Tracheal intubation oral vs. nasal
Examine and continue assessing your patient
Use a manometer when bagging
Follow HgB O2 saturation continually & ABGs
Insure that nebulization is adequate
Understand the effect of every ventilator knob
Select ventilator sitting that is appropriate for your patient
35
Suggested CMV Sittings
Normal Lungs HMD
36
High Frequency Ventilation
Definition:
Ventilation at a high rate at least 2 4 times the
natural breathing rate, using a small TV that is less
than anatomic dead space:
Types:
High Frequency Jet Ventilator (HFJV)
Up to 600 breath / min
High Frequency Flow Interrupter (HFFI)
Up to 1200 breath / min
High Frequency Oscillatory Ventilator (HFOV)
Up to 3000 / min
37
High Frequency Ventilation
Introduction
The respiratory insufficiency remains one of the major causes of
neonatal mortality.
Intensification of conventional ventilation with higher rates and
airway pressures leads to an increased incidence of barotrauma.
Either ECMO or high-frequency oscillatory ventilation might
resolve such desperate situations.
Since HFOV was first described by Lunkenheimer in the early
seventies this method of ventilation has been further developed
and is now applied the world over.
38
Definition
39
High Frequency Ventilation
40
Commercial ventilators
41
The "flow-interrupters" chop up the gas flow into the patient circuit
at a high rate, thus causing pressure oscillations. Their power,
however, depends also on the respiratory mechanics of the patient
42
Commercial ventilators
Other devices (e.g. Sensormedics 3100A)
generate oscillations with a large loudspeaker
membrane and are suitable also beyond the
neonatal period.
As with the piston oscillators, a bias flow system
supplies fresh gas.
However, this device cannot combine
conventional and HFO ventilation.
43
The InfantStar interrupts the inspiratory gas flow with
a valve bank. Some authors regard this device as a jet
ventilator because of its principle of operation.
44
Characteristic parameters and
control variables of HFV
47
Mean Airway Pressure (MAP)
The crucial physiologic effect of such continuously
applied (inflation) pressure is the opening of atelectatic
lung areas, resulting in marked recruitment of lung
volume.
48
Mean Airway Pressure (MAP)
Therefore MAP is the crucial parameter to
control oxygenation.
49
50
Amplitude oscillatory volume
51
Amplitude oscillatory volume
In any HF ventilator, the oscillatory volume depends
characteristically on the oscillatory frequency.
52
Amplitude oscillatory volume
The amplitudes and oscillatory volumes vary also with MAP.
Especially at MAP below 8 mbar oscillatory volumes
are markedly reduced.
53
Amplitude oscillatory volume
54
55
Oscillation amplitude and flow as functions of MAP
and frequency with the Babylog 8000:
56
Oscillatory frequency
57
Oscillatory frequency
The choice of an optimal oscillatory frequency is
currently subject of controversial discussion.
58
Oscillatory frequency
59
The gas transport coefficient DCO2
60
the clinical relevance of the gas transport
coefficient
61
HFV: Indications 1
62
Indications 2
Newborns
relative: PIP > 25 mbar
63
Combining HFV and IMV, and
sustained inflation
64
Combining HFV and IMV, and
sustained inflation
66
HFV: Start
67
68
69
70
Strategies for various lung diseases
71
72
73
74
75