Spesifikasi Ventilator Draeger Savina 300 (CO2)

Items Description
Patient Type Small Pediatric (Infant) to Adult (Bodyweight 2 to 200 kg)

Ventilation Settings
Invasive Ventilation Modes All Ventilation Modes
Non-Invasive Ventilation Modes All Ventilation Modes
Basic Ventilation Modes :
Volume (A)Control VC-CMV / VC-AC
Pressure (A)Control PC-AC
Volume SIMV VC-SIMV
Pressure SIMV PC-SIMV+
Spontaneous & CPAP SPN-CPAP/PS
Apnea Backup Ventilation Yes, with return capability to previous mode
Advance Ventilation Modes :
BIPAP/Assist PC-AC
BIPAP/PS PC-BIPAP
APRV PC-BIPAP with Inverse Ratio
Pressure Control with VG VC-AC/AutoFlow
(BIPAP/Assist - Volume Guarantee)
Pressure SIMV with VG VC-SIMV/AutoFlow
(BIPAP/PS - Volume Guarantee)
Minute Volume Guarantee VC-MMV/AutoFlow
Enhancement AutoFlow (AutoAdaptation of Inspiratory Flow in Volume Modes)
Capnograph with Mainstream etCO2
Graphical Loops Overlaid with Reference for Comparison
Parameters Setting :
Ventilation Frequency (RR) 2/minute to 80/minute
Inspiration Time (Ti) 0.2 second to 10 second
Tidal Volume 20 ml to 2000 ml
Flow Acceleration 5 to 200 cmH2O/second
Inspiratory Pressure 1 cmH2O to 99 cmH2O, absolute (include PEEP, not above PEEP)
PEEP (Intermittent PEEP) 0 to 50 cmH2O
Pressure Support 0 to 50 cmH2O (above PEEP)
O2 Concentration 21% to 100%
Inspiratory Trigger Sensitivity 1 L/minute to 15 L/minute
Termination Criteria (Expiratory Trigger) 5% to 75% of Peak Inspiratory Flow

Page 1 of 3
Items Description
Ventilation Monitoring
Displayed Parameters :
Pressure PEEP,
Mean Pressure (Pmean),
Plateau Pressure (Pplat),
Peak Inspiratory Pressure (PIP)
Minute Volume & Flow Minute Volume (MV),
Spontaneous MV (MVspon),
Leakage Minute Volume (MVleak)
Tidal Volume Volume Tidal (VT),
Spontaneous VT (VTspon),
Expiratory VT (VTe)
Respiratory Rate & Time Respiratory Rate (RR),
Spontaneous Rate (RRspon),
Inspiratory Time (Ti),
Plateau Time (Tplat),
Ratio Insp. To Exp. Time (I:E)
Others Inspiratory O2 Concentration (FiO2),
Rapid Shallow Breathing (RSB),
Compliance (C),
Resistance (R )
end Tidal CO2 Data End Tidal CO2 (etCO2)
Numeric/Graphical Displayed :
Simultaneous, up to 5 Numerical Data Pressure, Minute Volume & Flow, Tidal Volume, RR & Time, etc
Simultaneous, up to 3 Waveforms Selectable, Pressure, Flow, Volume and etCO2
Simultaneous, up to 6 Graphical Trends 2h, 6h, 12h or 24h of Selectable Trend data
Simultaneous, up to 6 Graphical Loops Pressure, Flow & Volume Combination (Overlaid with Reference)

Alarms / Monitoring
Expired minute volume High / Low
Airway pressure High / Low
Inspired O2 concentration High / Low
end Expiratory CO2 Concentration High / Low
Spontaneous Breathing Frequency High
Tidal Volume High / Low
Apnea alarm time 15 to 60 second

Page 2 of 3
Items Description
Performance data
Maximum (continuous) inspiratory flow 250 L/min
Control principle Time cycled, volume controlled, pressure limited
Safety Valve Opening on OverPressure 120 cmH2O
Safety Valve Open if Medical Oxygen and Air supply is not sufficient,
enables spontaneous breathing with ambient air.
Medicament Nebulisation Synchronized with Inspiration
Bronchial Suction Manuever Yes
O2 Sensor/Measurement Up to 2 years lifetime
etCO2 Measurement Mainstream, displayed as Numerical Data and Graphic

Operating data
Main Power Connection 100 to 240 Volt, 50/60 Hz
Current Consumption 1.3 Ampere max. at 240 Volt
Battery Backup Up to 5 hours (also for Internal Compressor), integrated on trolley.
Medical Air Compressor Internal (Power from Ventilator),
Turbin Technology, Lifetime up to 8 years
Gas supply :
Oxygen Pressure 3 to 6 Bar

Physical Specifications
Dimensions ventilator (W x H x D) 460mm x 383mm x 364mm (with trolley)
Diagonal screen size 12" TFT Color Touch Screen
Weight basic unit Approx. 26 kg

Machine outputs
Digital output Digital output and input via an RS232 C interface
Draeger MEDIBUS and MEDIBUS.X protocol

Accessories
Humidification System Humidifier F&P MR850
Breathing Circuit Reusable Breathing Circuit Complete with Chamber for Adult/Pediatric
Reusable Breathing Circuit Complete with Chamber for Small Pediatric
Test Lung for Testing Reusable Test Lung for Adult/Pediatric
Reusable Test Lung for Small Pediatric (Infant)
Ventilation Mask Reusable Size S, M and L
Nebulisation Reusable Nebulizer
Gas Hose Oxygen Supply Hose (White)
Gas Connector CIG / BritishStandard / DIN

Optional (Not Include)


Optional BedCoupling and Cylinder Holder BedCoupling to Ventilator Trolley and Holder for O2 Cylinder Holder.
Need to order for Transport (Room-to-Room) locked to Patient Bed.

Page 3 of 3