You are on page 1of 8

Notebook #8

SIMILARITIES AND DIFFERENCES OF RECORDED DETAIL AND DISTORTION

RECORDED DETAIL DISTORTION (SIZE & SHAPE)


ASSESSME 1. 1 out of 2 geometric properties of image quality A. 2 out of 2 geometric properties of image quality
NT 2. Controls detail B. Controls detail itself
3. Referred to as: C. Distortion = misrepresentation of the size or
A. Definition shape of the structures being imaged
B. Sharpness D. Only careful attention to the distances, directions,
C. Spatial Resolution and angulation between the anatomical part,
D. Detail central ray, and image receptor can minimize
4. High resolution = shorter-wavelength w/ high distortion
frequency
5. SPATIAL RESOLUTION = differentiating between two
points next to one another as being separate
densities
A. Point Spread Function (PSF) = measures
penumbra and is used to quantify digital system
spatial resolution
B. Spatial Frequency
C. Modulation Transfer Function (MTF) = measures
the accuracy of an image compared to the
original object on a scale of 0 to 1
D. Noise (Imaging noise & Quantum noise)
a. Imaging Noise = background information that
the image receptor receives
b. Quantum Noise = insufficient number of
incoming photons reaching the imaging
receptor
6. Digital Sampling (Nyquist Criterion)
A. Processing algorithm that averages the incoming
analog data by using the distance between the
imaging detector elements
7. Aliasing (Moire Pattern)
A. When Nyquist criterion is violated
B. Happens when the spatial frequency is greater
than the Nyquist frequency and the incoming
data are sampled less than twice per cycle
Notebook #8

EFFECTS 1. Resolution demonstrates the fine detail structures A. Size


ON IMAGE A. When fine detail is lacking, the image will be a. A matter of magnification
APPERANC blurred b. All magnification involves a degree of loss
E of resolution
B. Shape
a. Involves both elongation and
foreshortening
b. Serious alteration in the projected image

AFFECTIN 1. Solve resolution problems by: 1. SIZE DISTORTION


G A. Eliminate motion a. Magnification is THE ONLY possible size
FACTORS B. Reduce OID distortion
C. Reduce focal spot size i. Reduced magnification size distortion
D. Reduce intensifying screen phosphor size and increases the resolution
concentration ii. Magnification size distortion is
E. Increase SID controlled by positioning the body part
2. GEOMETRY and tube to maximize SID while
A. Distance minimizing OID
a. Resolution is better when b. SOURCE-IMAGE-DISTANCE (SID)
1. OID decreases (This is why affected side i. The greater the SID, the smaller the
or part is positioned closest to receptor) magnification
2. SID increases ii. SID must be maximized to decrease
b. Making adjustments to improve detail, you magnification
need to change OID 1st c. OBJECT-IMAGE-RECEPTOR-DISTANCE (OID)
c. EX: PA chest over AP chest i. Critical distance in both magnification
d. Once OID is minimized, resolution is and resolution
improved when SID increases ii. When objects in the body are at
a. EX: Lateral Cspine is at 72 instead of different levels, they will be projected
40 bc OID cant be reduced due to onto the image as different sizes
distance between the neck and iii. OID must be minimized to decrease
shoulder magnification
B. Focal Spot Size d. CALCULATING SIZE DISTORTION
1. Major controller of image resolution bc it e. Magnification factor formula
controls penumbra i. M=SID/SOD
2. Penumbra = unsharp edge shadow 2. SHAPE DISTORTION = misrepresentation by unequal
3. As focal spot decreases in size = penumbra magnification of the actual shape pf the structure
also decreases increasing resolution being imaged
4. Attenuation of object when added to a. Occurs because structures lie normally at
Notebook #8

penumbra, causes structures in the human different levels within the body
body to have gradual edges, instead of b. Also occurs because of the divergence of the
sharp abrupt edges x-ray beam
3. IMAGE RECEPTOR 3. ALIGNMENT
1. Film/Screen Systems a. Proper positioning is when the central ray is at
i. Slow film/screen combination will have right angles to the anatomical part and to the
better resolution than a fast image receptor
film/screen b. Incorrect centering may occur from off-
ii. Intensifying screens depend on 3 centering the tube, incorrectly positioning the
things: part, or off-centering the image receptor
1. Phosphor size i. Central ray
2. Phosphor layer thickness 1. Intended to be perpendicular
3. Phosphor concentration both the anatomical part and the
iii. As intensifying screen speed is image receptor
decreased, there is a gain in 2. Any structure that is not
resolution, but patient dose must be positioned at the central ray will
increased be distorted because of the
2. Digital Systems divergence of the beam
i. Indirect digital systems a. The farther from the
1. Intensifying screen with CCD central ray = the greater
detector the distortion
2. Silicon detector with a CsI ii. Anatomical part
scintillator 1. The long axis of the anatomical
3. A high fill factor produces higher part is intended to be positioned
resolution perpendicular to the central ray
4. Major factor that limits indirect and parallel to the image
digital systems is the size of the receptor
detector element 2. Elongation occurs when there is
ii. Direct digital systems poor alignment of the tube and/or
1. Photons traveling through this image receptor
flat panel have a more vertical 3. Foreshortening ONLY occurs when
path than silicon detectors there is poor alignment of the
2. Virtually no blurring part
3. Major factor that limits direct iii. Image receptor
digital systems is the size of the 1. Intended to be positioned
detector elements perpendicular to the central ray
4. MOTION and parallel to the anatomical
1. Voluntary motion: is under the control of the part
Notebook #8

patient 2. As long as part is parallel to IR,


2. Involuntary motion: is not under the the only result of off-centering is
conscious control of the patient clipping anatomy
3. Equipment motion: vibration from a. Will need to be repeated,
reciprocating grid BUT there is no distortion
4. Communication: BEST method of reducing or size or shape
motion is patient communication 4. ANGULATION
5. Exposure Time Reduction: when patient is a. Refers to direction and degree the tube is
not cooperating, reduce exposure time and moved from its normal position perpendicular
increase mAs to the IR
6. Immobilization: foam pads, sponges, b. Angulation of tube is designed to cause a
sandbags controlled or expected amount of shape
distortion to avoid superimposition
i. Angulation changes SID, unless
compensated for by a new SID, will
produce a decrease in IR exposure
5. DIRECTION
a. Most common direction is LONGITUDINAL
b. Termed cephalad or caudad
i. Degree = a method of describing the
exact amount of angulation and is
usually states as the angle between the
central ray and the IR plane
Notebook #8

Divergence of the beam. More tissue is imaged and


exposed at the level of the spine (Point B) than at the level OID
of the collimator (Point A)
Notebook #8

Critical distances in radiography SID

Central Ray Alignment


Notebook #8

7cm OID vs. 14cm OID


Off-centering the image receptor
Notebook #8

Foreshortening and magnification due to anatomical


part and image receptor alignment.

(A) Normal relationship between part and IR.

(B and C) Foreshortening and magnification due to changes


in anatomical part alignment.

(D and E) Elongation and magnification due to changes in


part/IR and central ray/IR alignment.

You might also like