Professional Documents
Culture Documents
The Moire
Effect * A grid error that happens with the digital IR systems when the grids lines are taken and scanned
parallel to the scan lines in the IR plate readers
Air Gap
Technique *An alternative to the use of a grid. It involves placing the patient at a greater OID therefore
creating an air gap between the patient and the IR. By moving the patient away from the IR, the
amount of scatter reaching the IR will be reduced.
Notebook #2
*The primary disadvantage of the air gap technique is the loss of sharpness that results from the
increased OID.
DEFINITION
Grid Ratios *The ratio of the height of the lead strips to the distance between the
strips
*The total lead quantity can be found by using information from grid ratio
and grid frequency. Lead content is greater in a grid that has a high grid
ratio and low grid frequency.
Technical *When using a grid, the radiographic technologist must increase the mAs
Adjustments (which increases patient dose). The amount of mass needed can be
found with the Grid Conversion Factor equation (GCF).