You are on page 1of 5

Physical Therapy in Sport 11 (2010) 81e85

Contents lists available at ScienceDirect

Physical Therapy in Sport


journal homepage: www.elsevier.com/ptsp

Original research

Correlation of three different knee joint position sense measures


Dayanand Kiran a, *, Mary Carlson a, Daniel Medrano b, Darla R. Smith b
a
Physical Therapy Program, University of Texas at El Paso, 1101, N Campbell Street, El Paso 79902, USA
b
Department of Kinesiology, University of Texas at El Paso, El Paso 79902, USA

a r t i c l e i n f o a b s t r a c t

Article history: Objective: The purpose of this study was to investigate correlation during concurrent measurement
Received 17 March 2010 among three knee joint position sense (JPS) measures in sitting position and between two measures in
Received in revised form standing position.
27 May 2010
Methods: Isokinetic dynamometer, electrogoniometer, and two dimensional (2D) video analysis were
Accepted 7 June 2010
used for measuring knee JPS. The JPS was measured both in sitting and standing positions. All three
measures were employed concurrently to measure knee JPS in sitting position; however, only the
Keywords:
electrogoniometer and 2D video analysis were concurrently used in the standing position. The knee JPS
Joint position sense
Electrogoniometer
was recorded in sitting position at 15 , 30 , and 45 and in standing at high, mid and low knee exion
Isokinetic dynamometer positions.
Two dimensional video analysis Results: The results of the study suggest excellent correlation (0.94e0.98) between the electrogoniometer
Correlation and 2D video analysis measures in standing position. In sitting position, good to excellent correlation
(0.63e0.92) was found between the isokinetic dynamometer and electrogoniometer; however, fair to
good correlation was found between 2D video analysis and either of the two measures (electro-
goniometer [0.52e0.57] and isokinetic dynamometer [0.41e0.63].
Conclusion: Either 2D video or an electrogoniometer may be used to measure JPS in standing position;
however, in sitting position 2D video should not be used if the camera is required to be placed at 10
from the plane of motion.
2010 Elsevier Ltd. All rights reserved.

1. Introduction the function of older adults but also have impact on the younger
population, especially with anterior cruciate ligament (ACL)
Proprioception is the umbrella term for kinesthesia and joint injuries. Proprioception is important in the prevention of injuries as
position sense (JPS). JPS refers to the awareness of joint position in reduced proprioception is one of the factors contributing to injury
space and is mediated through various receptors called mechano- in the knee, particularly the ACL. Although the causes of ACL injury
receptors (Grob, Kuster, Higgins, Lloyd, & Yata, 2002). These are multi-factorial, poor proprioception is one of the key causative
receptors are located in the joint capsule, ligaments, menisci, factors (Beynnon & Johnson, 1996; Grifn et al., 2000; Taimela,
musculotendinous unit, and skin (Kavounoudias, Roll, & Roll, 2001; Kujala, & Osterman, 1990). Additionally, the restoration of a fully
Lephart, Pincivero, & Rozzi, 1998). functional knee joint after ACL injury depends on regaining
Poor proprioception is suggested as a risk factor for the devel- proprioception as an important component (Aune, Holm, Risberg,
opment of functional inability in patients with knee osteoarthritis Jensen, & Steen, 2001; Friden, Roberts, Ageberg, Walden, &
(Sharma, Cahue, Song, Hayes, Pai, & Dunlop, 2003). The deteriora- Zatterstrom, 2001). Therefore, proprioception appears not only
tion of proprioception results in increased postural sway, decreased important for the prevention of ACL injuries, but also for regaining
balance, increased risk of falls and changes in gait patterns (Bergin, full function after ACL reconstruction.
Bronstein, Murray, Sancovic, & Zeppenfeld, 1995; Manchester, ACL injuries can be treated with reconstruction surgery where
Woollacott, Zederbauer-Hylton, & Marin, 1989; Tinetti & the patellar tendon or the hamstring tendon is used as a graft to
Speechley, 1989). The changes in proprioception not only affect replace the torn ACL ligament. Although reconstruction is
successful in regaining joint stability, the recovery of propriocep-
tive function remains debatable (Henriksson, Ledin, & Good,
2001). MacDonald, Hedden, Pacin, and Sutherland (1996) repor-
* Corresponding author. Tel.: 1 915 747 7218; fax: 1 915 747 8211.
E-mail addresses: dkiran@miners.utep.edu (D. Kiran), mcarlson@utep.edu (M. ted no signicant improvement in proprioceptive decits in
Carlson), danielm@miners.utep.edu (D. Medrano), darsmith@utep.edu (D.R. Smith). patients 31 months after ACL reconstruction by measuring

1466-853X/$ e see front matter 2010 Elsevier Ltd. All rights reserved.
doi:10.1016/j.ptsp.2010.06.002
82 D. Kiran et al. / Physical Therapy in Sport 11 (2010) 81e85

kinesthesia. However, Reider et al. (2003) reported a signicantly 2.3. Sitting measurement
improved level of proprioception by measuring JPS in an ACL
reconstructed knee after six months of rehabilitation when 2.3.1. Calibration of electrogoniometer
compared with the contralateral limb. Furthermore, Hopper, The Penny and Giles M series twin axis electrogoniometer
Creagh, Formby, Goh, Boyle, and Strauss (2003) reported no (Penny Giles, UK) was used to measure JPS. It measures the
signicant difference in knee proprioception after 12e16 months potential difference between the two endblocks, and converts the
of ACL reconstruction by measuring JPS. Reider et al. (2003) used difference in potential to the respective joint angle. Before data
an electrogoniometer to measure the JPS, however, Hopper et al. collection, calibration of the electrogoniometer was done using
(2003) used Peak Motus motion measurement to measure the VICON Motus Motion System (VMMS) version 8.5 (Peak Perfor-
knee JPS. The results regarding proprioceptive function may be mance Technologies, Inc. CO, USA) and a universal goniometer. One
contradictory not only due to differences in measurement endblock of the electrogoniometer was placed on the xed and the
methods but also due to use of different equipment to measure other on the movable arm of the universal goniometer for the
the proprioception. calibration (see Fig. 1). The movable arm was rotated in 10 incre-
Early measurement of JPS used a modied Thomas splint with ments, and the corresponding voltages were recorded from the
a Pearson knee piece or a copper frame with a motor and a cali- VMMS. Further, the following equation was obtained by plotting
brated scale (Barrett, 1991; Corrigan, Cashman, & Brady, 1992). a graph (R2 0.99) between the recorded voltage (x) and the angle
These measurements were subjected to a high degree of intertester (y) to which the universal goniometer arm was moved:
variability as the joint angle was visually estimated. Many studies
have used the isokinetic dynamometer, electrogoniometer, or 2D Angley 88:118x 261:09 (1)
video analysis to measure JPS (Birmingham et al., 1998; Hopper This equation was used to calculate the knee joint angle during JPS
et al., 2003; Reider et al., 2003). testing.
Although all three measures of JPS have been used separately,
no studies have established if concurrent measurement will result 2.3.2. Calibration of the video camera
in similar values of JPS. Additionally, most of the studies tested JPS A JVC camera (Victor Company of Japan, Japan) was used for
in non-weight bearing conditions while functional activities are recording the video for the measurement of JPS by VMMS. Before
performed in weight bearing conditions. It logically follows that testing, the calibration of the cameras spatial eld was done by
JPS should be tested in weight bearing conditions in order to VMMS using a 0.50 m  0.325 m frame with reective markers at
simulate functional activities. Therefore, this study aimed to each corner.
determine if three JPS measures (isokinetic dynamometer, elec-
trogoniometer, and 2D video analysis) produced equivalent 2.3.3. Testing procedure
results. The participants were asked to sit on the isokinetic dynamom-
eter system 3 (Biodex Medical Systems, Shirley, New York) chair
2. Methods with their trunks secured to the chair after calibrating the system as
per manufacturers guidelines. The endblocks of the goniometer
2.1. Participants were placed on the lateral aspect of the dominant limb knee joint.
To measure the knee angle by 2D video analysis, the reective
A group (convenience sampling) of 30 participants (male and markers were placed on the thigh, lateral knee joint line, and lateral
female), ages 18e25 years with no history of injury or surgery to malleolus. The thigh marker was placed slightly forward and
the knee, were recruited from the university student population. downward to the greater trochanter after placing the other two
The number of participants was decided based on previous markers. All the markers were ensured to be in one line and in the
studies (Grob et al., 2002; Reider et al., 2003) on proprioception same plane in knee extended position. Additionally, the reective
in ACL injuries. The inclusion criteria for the study were (a) marker at the knee joint was stemmed to bring all three markers in
current enrollment at the university and (b) participation in one plane. The placement of the endblocks and the reective
sports-related activities at least three times a week. The partici- markers were held constant throughout the experiment in sitting
pation in sports activities was required so that everybody could and standing position.
be at a similar level of proprioception. The exclusion criteria were
(a) any history of injury, pain, or swelling of the knee in the
previous year and (b) any history of medical problems which
could limit proprioception. All the inclusion and exclusion
criteria were determined by the participants answers on a health
questionnaire form. Approval of the proposal from the uni-
versitys ethical review board was obtained prior to data collec-
tion. All the participants were asked to sign an informed consent
form.

2.2. Procedures

Three measures were taken concurrently in sitting position:


isokinetic dynamometer, 2D video analysis, and electrogoniometer.
However, only two measures were taken concurrently in standing
position: 2D video analysis and electrogoniometer. A practice test
was given to familiarize the participant with the procedure before
testing. The order of the testing in sitting or standing positions was
determined randomly. Fig. 1. Placement of elctrogonometer endblocks.
D. Kiran et al. / Physical Therapy in Sport 11 (2010) 81e85 83

Participants were seated with the knee exed to 90 on the


isokinetic dynamometer chair with eyes covered (Fig. 2). Testing
positions of 15 , 30 , and 45 were selected on the isokinetic
dynamometer protocol (Birmingham et al., 1998). The three test
positions were tested in the order of 15 , 30 , and 45 as the
dynamometer protocol does not permit random selection. The
researcher asked the participant to move the knee from the starting
position to the angle being demonstrated by the dynamometer and
then to return the knee to the starting position. The participants
were asked to remember each demonstrated position and then
return the knee joint to the demonstrated position. The participants
indicated the reproduction of the knee angle to the demonstrated
position by pressing a switch at the appropriate angle.
The measurements from camera and electrogoniometer were
also recorded by the VMMS at each corresponding demonstrated
and reproduced position. An error angle between the demonstrated
and the reproduced position was considered for analysis of JPS.

2.4. Standing measurement Fig. 3. Placement of reective markers and endblocks for measurement of JPS in
standing.
The calibration and placement of the endblocks of the electro-
goniometer was the same as it was in the sitting position. However,
tests. The independent variables were the testing procedures with
the cameras spatial eld was recalibrated by using VMMS and
three levels (electrogoniometer, isokinetic dynamometer, and 2D
1.83 m  0.92 m frame with reective markers at each corner.
video analysis), and testing positions with two levels (sitting and
standing). The dependent variable, joint position sense, was
2.4.1. Testing procedure
measured in degrees.
The participants were asked to stand with maximum weight on
the test limb and with eyes covered (Fig. 3). The foot of the non-
tested leg was touching the oor for stability only. They were asked 2.6. Data analysis
to ex the knee to a randomly determined position of around
0 e45 of knee exion. We considered the test position high, mid We conducted all planned comparisons using SPSS 15.0 statis-
and low when the knee was exed around 15 , 30 , and 45 tical software. First of all, we examined all the variables for distri-
respectively. The participants were asked to remember the butional assumptions and potential outliers using frequency
demonstrated position and return to a standing position. They were analysis and histograms. Following these analyses, descriptive
then asked to place the knee joint in the demonstrated position. statistics were reported including means and standard deviations
Both the demonstrated and reproduced positions were recorded by for continuous variables (JPS score).
the electrogoniometer and the camera for the analysis (Hopper
et al., 2003; Reider et al., 2003). 3. Results

2.5. Research design Thirty healthy participants (13 females and 17 males) with no
history of injury in the lower limb were tested on their dominant
The research design was a prospective within subjects design limb. Three trials of demonstrated and reproduced testing were
that used concurrent measurement and random ordering of the JPS done at each position and the average absolute error values are
presented in Table 1. For the knee position at 15 only two absolute

Table 1
presents the descriptive statistics of absolute error of the knee joint position sense
with the mean, standard deviation, minimum and maximum values (in degrees) in
sitting and standing position.

Test Equipment Measured Minimum Maximum Mean


position position error error
Sitting Isokinetic 15 1 10 4.68  2.21
dynamometer 30 1 8 3.85  1.90
45 0.67 12 3.90  2.55
Electrogoniometer 15 0.27 7.30 3.24  1.63
30 0.53 7.10 2.90  1.61
45 0.26 8.53 3.42  2.04
2D video analysis 15 0.14 6.54 2.91  1.65
30 0.27 7.07 2.72  1.49
45 0.14 5.59 2.31  1.47

Standing Electrogoniometer High 0.31 9.01 3.34  2.33


Mid 1.2 9.15 3.85  2.07
Low 0.35 8.77 3.57  2.1
2D video analysis High 0.34 8.67 3.49  2.24
Mid 0.98 8.21 3.58  2.00
Fig. 2. Placement of reective markers and endblocks for measurement of JPS in sitting Low 0.31 7.26 3.24  1.99
position.
84 D. Kiran et al. / Physical Therapy in Sport 11 (2010) 81e85

errors were averaged due to loss of data. In the sitting position, the Table 3
mean absolute errors measured by the isokinetic dynamometer presents the inter-correlations between all the test positions in standing.

were 4.68 (SD 2.21), 3.85 (SD 1.90), and 3.90 (SD 2.55) degrees in Equipment Measured 2D video analysis
the 15 , 30 and 45 test positions respectively. The mean absolute position
High Mid Low
errors measured by the electrogoniometer in the same positions
Electrogoniometer High 0.98
were 3.24 (SD 1.63), 2.90 (SD 1.61), and 3.42 (SD 2.04) degrees. The (r2 0.96)
2D video analysis mean absolute measures of the knee JPS were Mid 0.96
2.91 (SD 1.65), 2.72 (SD 1.49), and 2.31 (SD 1.47) degrees in the (r2 0.92)
Low 0.94
similar test positions. In the standing position, the mean absolute
(r2 0.88)
errors measured by electrogoniometer were 3.34 (SD 2.33), 3.85
(SD 2.07), and 3.57 (SD 2.10) degrees in the high-, mid-, and low-
test positions respectively. In the same test positions, the mean values of JPS. Therefore, researchers may use either measure during
absolute errors measured by 2D video analysis were 3.49 (SD 2.24), knee JPS measurement in standing position with condence.
3.58 (SD 2.00), and 3.24 (SD 1.99) degrees. To assess the concurrent In the standing measurements, we were able to place the
reliability the correlations between average absolute error angles of camera close to a 90 position; however, in sitting position
two measurement techniques were determined at each position in measurement the placement of the camera at 90 was not possible
both sitting (Table 2) and standing (Table 3). Excellent correlation due to the isokinetic dynamometer arm blocking the cameras
(0.94e0.98) in the standing knee measurements is noted at each visual eld. Therefore, we had to place the camera at an angle of 10
joint position between the electrogoniometer and the 2D video where the dynamometer could not obstruct the visibility of the
analysis. However, good to excellent level of correlation reective markers.
(0.63e0.92) is present between the electrogoniometer and iso- In this study, we found that in the sitting position the correlation
kinetic dynamometer measurements at all three of the positions in between the measures of the isokinetic dynamometer and
sitting. Furthermore, fair to good correlation (0.41e0.63) is present the electrogoniometer was in the range of 0.63e0.92 within the
between the two dimensional video and either of the other two different tested positions. However, the correlation between the
measures (electrogoniometer and isokinetic dynamometer measures of isokinetic dynamometer and the 2D video analysis or
measurements). between the 2D video analysis and the electrogoniometer was in
the range of 0.41e0.63 or 0.52e0.57 respectively. The fair to good
correlation between 2D video analysis and either of the other two
4. Discussion measures appeared to be due to the inability of the VICON Motus
motion system (VMMS) to measure joint angles accurately when
Although knee joint position sense (JPS) has been measured by the camera position was close to 10 from the plane of motion. In
different equipment in the research literature, the correlation a previous study, (Kiran, Carlson, Medrano, & Smith, 2010), we
among the measures has not been studied. The present study reported increased error when the camera was placed at 5 e10
aimed to evaluate if the concurrent measures of the knee JPS by the from the plane of motion. Therefore, the increased error in JPS
three measures (isokinetic dynamometer, electrogoniometer, and measurement by 2D video analysis could be the reason for the fair
2D video analysis) were highly correlated. These measures have to good correlation between 2D video analysis measure and iso-
been used by numerous researchers for several years (Birmingham kinetic dynamometer or electrogoniometer measure. However,
et al., 1998; Hopper et al., 2003; Reider et al., 2003). Further, we when the camera was positioned perpendicular to the plane of
tested the reliability of the measures in both sitting and standing motion, the 2D video analysis measures were highly correlated
positions since both positions were used in the research literature. with the electrogoniometer measures as evidenced by the standing
In the standing position, we found excellent correlation JPS correlation values. Further studies may be done to discern the
(0.94e0.98) between the electrogoniometer and 2D video analysis correlation between concurrent JPS measurements by the electro-
measurement (Fleiss, 1981). The excellent correlation in standing goniometer and the 2D video analysis in the sitting position.
position appeared to be due to the minimal level of error in The correlation between the electrogoniometer and the iso-
measurement of joint angles by 2D video analysis when the camera kinetic dynamometer in the sitting position ranged from 0.63 to
was placed at 90 from the plane of motion. Additionally, the 0.92 with an increase in correlation from 15 knee exion to 45
analysis of the measures of JPS in standing suggests that neither knee exion position. The differential in correlation in these posi-
measure is better than the other, as both measures had equivalent tions may be due to more skin movement in the thigh in a fully

Table 2
presents the inter-correlations between all the positions in sitting.

Equipment Measured 2D video analysis Electrogoniometer


position
15 30 45 15 30 45
Isokinetic 15 0.41 0.63
dynamometer (r2 0.17) (r2 0.40)
30 0.53 0.80
(r2 0.28) (r2 0.64)
45 0.63 0.92
(r2 0.40) (r2 0.85)

Electrogoniometer 15 0.55
(r2 0.30)
30 0.52
(r2 0.27)
45 0.57
(r2 0.32)
D. Kiran et al. / Physical Therapy in Sport 11 (2010) 81e85 85

extended position than in a exed position. Kuo, Tully, and Galea Barrett, D. S. (1991). Proprioception and function after anterior cruciate recon-
struction. The Journal of Bone and Joint Surgery. British Volume, 73(5), 833e837.
(2008) reported displacement of skin markers in the lateral thigh
Bergin, P. S., Bronstein, A. M., Murray, N. M., Sancovic, S., & Zeppenfeld, D. K. (1995).
during hipe knee exion in sitting position. The measurement of Body sway and vibration perception thresholds in normal aging and in patients
the joint angle by electrogoniometer depends on the change in with polyneuropathy. Journal of Neurology, Neurosurgery, and Psychiatry, 58(3),
voltage between the two endblocks. The differential skin move- 335e340.
Beynnon, B. D., & Johnson, R. J. (1996). Anterior cruciate ligament injury rehabilita-
ment in the thigh from extension to exion of the knee may tion in athletes. Biomechanical considerations. Sports Medicine (Auckland, N.Z.),
interfere with the change in voltage between the two endblocks, 22(1), 54e64.
yielding altered JPS measurement values. Further research may be Birmingham, T. B., Kramer, J. F., Inglis, J. T., Mooney, C. A., Murray, L. J., Fowler, P. J.,
et al. (1998). Effect of a neoprene sleeve on knee joint position sense during
done to nd out the accuracy of electrogoniometer measures at sitting open kinetic chain and supine closed kinetic chain tests. The American
different degrees of knee exion angle. Additionally, research may Journal of Sports Medicine, 26(4), 562e566.
be done to discover the amount of skin movement throughout the Corrigan, J. P., Cashman, W. F., & Brady, M. P. (1992). Proprioception in the cruciate
decient knee. The Journal of Bone and Joint Surgery. British Volume, 74(2), 247e250.
knee range of motion. The future study may also include correlation Fleiss, J. L. (1981). Statistical methods for rates and proportions (2nd ed.). New York:
in sitting position of knee JPS measures by an electrogoniometer John Wiley & Sons.
and 2D video analysis with a camera perpendicular to the plane of Friden, T., Roberts, D., Ageberg, E., Walden, M., & Zatterstrom, R. (2001). Review of
knee proprioception and the relation to extremity function after an anterior
motion. cruciate ligament rupture. The Journal of Orthopaedic and Sports Physical
In conclusion, since excellent correlation between 2D video Therapy, 31(10), 567e576.
analysis and electrogoniometer was found, either electro- Grifn, L. Y., Agel, J., Albohm, M. J., Arendt, E. A., Dick, R. W., Garrett, W. E., et al.
(2000). Noncontact anterior cruciate ligament injuries: risk factors and
goniometer or 2D video analysis may be used in the standing
prevention strategies. The Journal of the American Academy of Orthopaedic
position to measure JPS; if the camera is placed perpendicular to Surgeons, 8(3), 141e150.
the plane of motion. Accordingly, either isokinetic dynamometer or Grob, K. R., Kuster, M. S., Higgins, S. A., Lloyd, D. G., & Yata, H. (2002). Lack of
electrogoniometer may be used to measure knee JPS in sitting correlation between different measurements of proprioception in the knee. The
Journal of Bone and Joint Surgery. British Volume, 84(4), 614e618.
position as the correlation between those two measures was good Henriksson, M., Ledin, T., & Good, L. (2001). Postural control after anterior cruciate
to excellent. However, in sitting position, the correlation of the 2D ligament reconstruction and functional rehabilitation. The American Journal of
video analysis and either of the other two measures was only fair to Sports Medicine, 29(3), 359e366.
Hopper, D. M., Creagh, M. J., Formby, P. A., Goh, S. C., Boyle, J. J., & Strauss, G. R.
good. Therefore, the selection of 2D video analysis to measure knee (2003). Functional measurement of knee joint position sense after anterior
JPS in sitting position with the camera placed at 10 from the plane cruciate ligament reconstruction. Archives of Physical Medicine and Rehabilita-
of movement should be avoided. tion, 84(6), 868e872.
Kavounoudias, A., Roll, R., & Roll, J. P. (2001). Foot sole and ankle muscle inputs
contribute jointly to human erect posture regulation. The Journal of Physiology,
Acknowledgements 532(Pt 3), 869e878.
Kiran, D., Carlson, M. E., Medrano, & Smith, D. R. (2010). Validity of two dimensional
VICON Motus motion software at varying camera angles. International Journal of
The research was funded by College of Health Sciences, the Exercise Science: Conference Abstract Submissions, 4(1).
University of Texas at El Paso. Kuo, Y. L., Tully, E. A., & Galea, M. P. (2008). Skin movement errors in measurement
of sagittal lumbar and hip angles in young and elderly subjects. Gait & Posture,
27(2), 264e270. doi:10.1016/j.gaitpost.2007.03.016.
Conict of interest Lephart, S. M., Pincivero, D. M., & Rozzi, S. L. (1998). Proprioception of the ankle and
knee. Sports Medicine (Auckland, N.Z.), 25(3), 149e155.
MacDonald, P. B., Hedden, D., Pacin, O., & Sutherland, K. (1996). Proprioception in
None. anterior cruciate ligament-decient and reconstructed knees. The American
Journal of Sports Medicine, 24(6), 774e778.
Manchester, D., Woollacott, M., Zederbauer-Hylton, N., & Marin, O. (1989). Visual,
Ethical statement vestibular and somatosensory contributions to balance control in the older
adult. Journal of Gerontology, 44(4), M118eM127.
Reider, B., Arcand, M., Diehl, L., Mroczek, K., Abulencia, A., Stroud, C. C., et al. (2003).
The study was approved by the University of Texas at El Paso
Proprioception of the knee before and after anterior cruciate ligament recon-
review board (88060-4). struction. Arthroscopy: The Journal of Arthroscopic & Related Surgery, 19(1), 2e12.
Sharma, L., Cahue, S., Song, J., Hayes, K., Pai, Y. C., & Dunlop, D. (2003). Physical
functioning over three years in knee osteoarthritis: role of psychosocial, local
References mechanical, and neuromuscular factors. Arthritis and Rheumatism, 48(12),
3359e3370.
Aune, A. K., Holm, I., Risberg, M. A., Jensen, H. K., & Steen, H. (2001). Four-strand Taimela, S., Kujala, U. M., & Osterman, K. (1990). Intrinsic risk factors and athletic
hamstring tendon autograft compared with patellar tendon-bone autograft for injuries. Sports Medicine (Auckland, N.Z.), 9(4), 205e215.
anterior cruciate ligament reconstruction. A randomized study with two-year Tinetti, M. E., & Speechley, M. (1989). Prevention of falls among the elderly. The New
follow-up. The American Journal of Sports Medicine, 29(6), 722e728. England Journal of Medicine, 320(16), 1055e1059.

You might also like