UV/Vis
Spectrophotometer
Calibration
Procedures
Published by:
International Accreditation New Zealand
626 Great South Road, Greenlane, Auckland 1005
Private Bag 28 908, Remuera, Auckland 1136, New Zealand
Telephone 64 9 525 6655
Facsimile 64 9 525 2266
Email: info@ianz.govt.nz
Internet: http://www.ianz.govt.nz
ISBN: 0908611 59 5
Contents Page
1 Introduction ......................................................................................................................... 3
1.1 Traceability................................................................................................................ 3
2 Principal calibration requirements ...................................................................................... 3
2.1 Wavelength calibration.............................................................................................. 3
2.2 Absorbance Calibration............................................................................................. 4
2.3 Liquid sample holder absorption............................................................................... 5
2.4 Stray light control ...................................................................................................... 6
3 Uncertainty of measurement .............................................................................................. 7
4 Acknowledgements............................................................................................................. 7
5 References ......................................................................................................................... 7
Tables
Table 1 Wavelengths of the absorption maxima of holmium (III) in perchloric acid for two
values of spectral slit width (reference (g)). 3
Table 2 Nominal wavelengths of selected absorption maxima for holmium and didymium
glass filters (reference (o)). 4
Table 3 Emission lines of zinc, cadmium and mercury (reference(h)). 4
Table 4 Absorbance values for CuSO4 5H2O (20.0g.L-1) in 1% H2SO4 (adapted from
reference (o)). 5
Table 5 Nominal absorbance values for K2Cr2O7 in 5 x 10-3 mol.L-1 H2SO4 (reference (o)) 5
2
UV/Vis Spectrophotometer Calibration Procedures
3
UV/Vis Spectrophotometer Calibration Procedures
(ii) Glass filters containing holmium or Traditionally the minimum recommended number of
didymium (a mixture of praseodymium points to be checked on the wavelength scale has
and neodymium). been one in the UV range and three in the visible
The wavelengths of absorption maxima in holmium range. However modern instruments are very linear
and didymium glasses can vary between batches; and for such instruments two points are likely to be
nominal values are listed in Table 2. sufficient. Some modern instruments have built-in
wavelength checks using deuterium or glass filters
Table 2 Nominal wavelengths of selected
and these should be used between the annual full
absorption maxima for holmium and
service/calibration by the relevant accredited
didymium glass filters (reference (o)).
agency.
Wavelength of maximum (nm) 2.2 Absorbance Calibration
Holmium glass Didymium glass As the concentration of an absorbing species is
241.5 0.2* 573.0 3.0 usually directly proportional to the measured
279.4 0.3 586.0 3.0 absorbance, the linearity and the stability of the
287.5 0.4 685.0 4.5 photometric scale must be confirmed. The stability
333.7 0.6 of the photometric reading must be adequate so
360.9 0.8 variations during the course of the determination do
418.4 1.1 not limit precision. It is recommended that
453.2 1.4 absorbance calibration is carried out at
536.2 2.3 absorbances and wavelengths routinely used by
637.5 3.8 the laboratory. Typically this would include six to
*not readily seen in glass eight absorbance values over the range 0 to 3 at
546 nm for an instrument used in the visible and
(iii) Low pressure discharge lamps four to six absorbance values at one wavelength in
containing mercury, cadmium or zinc. the range 230320 nm for instruments used at
Some spectrophotometers contain lamp housings wavelengths shorter than 330 nm. For newer
for these lamps (Hg, Cd, Zn). Where this facility is instruments it may be possible to verify the
provided, wavelength checks can be made very absorbance scale at only one wavelength.
conveniently by operating the discharge lamp and However, evidence validating that conclusion for a
rotating the monochromator to isolate the emitted number of measurements would need to be
spectral lines. For this and all wavelength provided to IANZ.
calibrations the minimum slit width should be used.
The mercury lamp has the advantage that its The following reference materials are preferred:
emission spectrum contains several lines over a 2.2.1 Visible region
reasonable range of wavelength. Wavelengths in (a) Neutral density glass filters, specifically
air of prominent spectral lines of these elements are at 546 nm.
listed in Table 3. Certified neutral density filters are available
from some NMIs and chemical supply
Table 3 Emission lines of zinc, cadmium and companies. These filters will need a first
mercury (reference(h)). recertification after 5 years and thereafter
every 10 years.
Element Wavelength Element Wavelength (b) Analytical reagent grade copper
(nm) (nm) sulphate in 1% sulphuric acid.
Mercury* 185.0 Mercury 435.8
The required solution is prepared by
Zinc 213.9 Cadmium 467.8
Cadmium 228.8 Cadmium 480.0 dissolving 20.0 g of copper sulphate
Mercury 253.7 Mercury 546.1 pentahydrate (CuSO4.5H2O) in water to
Mercury 365.0 Mercury 579.1 which is carefully added 10 ml sulphuric
Mercury 404.7 Cadmium 643.8 acid (d=1.83) and then diluting to 1 litre.
* wavelength in vacuum; all others in air The reference is a solution of sulphuric acid
of the same concentration. Absorbances of
For spectrophotometers with a deuterium lamp, the this solution in 10 mm cells at different
two deuterium emission lines at 486.0 nm and wavelengths are given in Table 4.
656.1 nm (reference (o)) can also be used for
wavelength calibration. Copper sulphate solutions of different
concentrations can be used to check the
Note: When calibrating a scanning instrument it is linearity of absorbance scales. The use of
most important to use the slowest scanning the absorbances at 600 nm and 650 nm is
speed available in order to minimise the particularly useful for examining the low end
effect of the response time of the of the absorbance scale.
instrument.
4
UV/Vis Spectrophotometer Calibration Procedures
Table 4 Absorbance values for CuSO4.5H2O Table 5 Nominal absorbance values for
(20.0g.L-1) in 1% H2SO4 (adapted from K2Cr2O7 in 5 x 10-3 mol.L-1 H2SO4
reference (o)). (reference (o))
5
UV/Vis Spectrophotometer Calibration Procedures
2.4 Stray light control UV-vis beams. Filters that may be used for this test
Stray radiant energy (SRE), often referred to as and the corresponding wavelengths 0 include;
stray light, falls into two categories: Various sharp-cut filter glasses
Light from the monochromator at Liquid filters
wavelengths different from, and mostly very
KCl 12 g/L aqueous at 175200 nm
different from, the selected passband, often
referred to as heterochromatic stray light. NaI 10 g/L aqueous at 210259 nm
Light in the selected wavelength passband Acetone at 250320 nm
which is scattered into the beam exiting the NaNO2 50 g/L aqueous at 300385 nm.
sample chamber without passing through
the sample, often referred to as FIG. 2 Test for the heterochromatic stray light:
isochromatic stray light. measure the transmittance in the pass band of a
These categories are illustrated in Fig. 1. Tests cut-off filter that stops the monochromator set
should be made for the presence of both of these wavelength (indicated by the arrow)
as part of the operational qualification of a
spectrophotometer.
6
UV/Vis Spectrophotometer Calibration Procedures