Professional Documents
Culture Documents
Cryogenic Valve Test Procedure PDF
Cryogenic Valve Test Procedure PDF
Test Procedure
For Testing of Cryogenic Valves
Based on BS 6364, API 598
and EPA Method 21
1. INTRODUCTION
The following procedure covers the testing for sealing and operability of
This procedure is based on BS 6364, API 598 and EPA Method 21, to the extent
applicable and specified herein.
2. TEST PLAN
The test comprised of the following tests at ambient temperature, then cryogenic or
low temperature and again at ambient temperature as given below:
2.1 Ambient Temperature: - Shell and seat test according to API 598.
- Visual examination.
3. TEST TEMPERATURES
4.2 System proving seat test: Test at high seat test pressure (one increment), as per
table below.
4.3 High pressure seat test: Raise the pressure in increments up to seat test pressure
as given in the table below. Measure and record the
leakage rate at each pressure interval.
4.5.1 Globe and gate valves: Seat test is conducted in the normal flow direction for the
valve.
4.5.2 Check valves: Seat test is conducted in the reverse flow direction of the
valve.
4.5.3 Ball and Butterfly valves: Seat test is conducted in preferred direction.
4.5.4 Bidirectional sealing: For valves, which are capable of sealing in both directions
the normal flow (or preferred) direction only shall be
tested. Upon specific customer request and agreement,
each sealing direction can be tested separately.
5. TEST SEQUENCE
5.2.1 Cool down the valve. When the valve has reached the required test temperature,
allow for one hour of stabilization time, then testing may proceed.
5.2.3 Cycle the valve open-close 5 times. Record the operating torques for the first and last
cycle.
5.2.4 Close the valve and pressurize the valve in the seat test direction to 80 psig. Wait at
least 10 minutes to stabilize helium temperature.
5.2.6 Increase the pressure in stages to the maximum seat pressure as described in
Section 3.3. At each test pressure, wait at least 10 minutes to stabilize the gas
temperature.
5.2.7 Record the seat leakage rate at each of the pressure intervals.
5.2.1 Remove the valve from liquid nitrogen and let it warm back to ambient temperature.
5.4.1 With the valve half-open, pressurize the valve to 90 psig helium gas.
5.4.2 Use the helium mass spectrometer to sniff the leakage around the valve for the entire
test duration. Mass spectrometer is to be calibrated (before use) as per the equipment
manufacturers procedure. A one-time packing-bolt, bonnet-bolt, and end-flange-bolt
re-torque is allowed during testing.
5.4.3 Retainerless design dual plate check valves are exempt from external leakage test. A
retainerless design has no openings through the body, i.e. no packing and no
body/bonnet gasket.
After successful testing, the test flanges on the valve shall be removed and the disc
and seat sealing area shall be visually inspected to the extent possible (practical only
on large valves).
6.2 External leakage: at no time throughout the duration of the test, shall a reading of
>1x10-4 cc/sec or 50 PPM be sustained for more than ten (10)
seconds.
7.1 The valve to be tested is supported in a stainless steel tank. The sides and the
underside of the tank are insulated on the outer faces. The valve is held between
blind flanges with small bore piping leading to the fixtures to supply the pressurising
media. Stem position is vertical for gate, globe, ball and butterfly valves. Check valves
(swing, dual plate, etc.) may be tested in either vertical or horizontal disc position.
7.2 The valve temperature is lowered by filling the tank with the cooling medium (see
below table), to a level above the body-bonnet bolting. The valve temperature is
monitored by a thermocouple, installed on the valve body or end plate. The test is
commenced when the valve temperature is stabilized within the test temperature
tolerances, as indicated by the thermocouples.
7.3 Test gas (see below table) from a fully charged bottle is used to provide test pressure
on the inlet side of the valve. The opening and closing torques for the valve are
measured using a torque wrench.
7.4 All instruments (flowmeter, pressure gauges, torque wrench, etc.) are to be calibrated.
8.1 Flow meter type: Any type may be used, provided that it can be calibrated.
8.2 Flow meter calibration: Are to be calibrated for the test gas (Helium or Nitrogen). Or,
the manufactures correction factor is to be used to relate the calibration gas (typically
Nitrogen) to the test gas.
8.3 Seat leak: Allowed seat leak is measured at the flowmeter, and are at standard
atmospheric conditions.
8.5 Pressure Correction: Pressure of the exiting test gas is near atmospheric, and
assumed to have negligible effect. Therefore, measurement of exiting pressure, as
well as correction is not required.
8.6 Exemption: Some flow meters (eg. Electronic mass flow meters) are not affected by
pressure or temperature changes, and can be calibrated to provide a direct reading at
standard conditions. When such a flow meter is used, test gas pressure and
temperature (as well as correction) at the flow meter is not required.
Test Pressure (psig) Duration of Seat Test (min) Seat Leak Rate (scc/min)
Initial Proving Test Pressure (psig) Duration (min) Seat Leak (cc/min)
Low Pressure Seat Test (psig) Duration (min) Seat Leak (cc/min)
Increment Pressure (psig) Duration (min) Measured Seat Leak (cc/min) Corrected Seat Leak (scc/min)
1st
2nd
3rd
4th
5th
Initial Proving Test Pressure (psig) Duration (min) Seat Leak (scc/min)