Professional Documents
Culture Documents
REF86100 FAQs HumaCycler PDF
REF86100 FAQs HumaCycler PDF
SYSTEM VERSION
COPYRIGHT
Copyright 2015, Human Gesellschaft für Biochemica und Diagnostica mbH, Wiesbaden,
Germany. All rights reserved.
No part of this documentation may be reproduced in any form, nor processed, copied or distrib-
uted by means of electronic systems, without prior permission of HUMAN in writing. Since all
precautionary measures were taken into account in producing these operating instructions, the
manufacturer accepts no responsibility for any errors or omissions. This includes any liability for
damage that could arise from possible incorrect operation based on this information. Subject to
changes without notice as result of technical development.
TABLE OF CONTENTS
1 SAFETY INSTRUCTIONS 5
1.1 INTRODUCTION 5
1.2 USER WARRANTY 5
1.3 INTENDED USE OF THE INSTRUMENT 5
1.4 GENERAL SAFETY WARNINGS 6
1.5 DISPOSAL MANAGEMENT CONCEPT 6
1.6 BIOHAZARD WARNING 7
1.7 INSTRUMENT DISINFECTION 7
1 SAFETY INSTRUCTIONS
1.1 Introduction
This manual is considered part of the instrument and must be available to the
operator and the maintenance personnel. For accurate installation, use and
maintenance, please read the following instructions carefully.
In order to avoid damage to the instrument or personal injury, carefully read
the ”GENERAL SAFETY WARNINGS”, describing the appropriate operating pro-
cedures. Please contact your HUMAN authorised local Technical Service in the
event of instrument failure or other difficulties with the instrument.
Figure 1
Biological Hazard Symbol
Table 2 Channel F1 F2 F3 F4
(nm range) (nm range) (nm range) (nm range)
FAM VIC Cy5 ROX
(492/515) (538/554) (552/570) (568/595)
Texas-Red
SYBR Green- I HEX (535/556)
(595/620)
Fluorescence Dye
TET
(521/536)
JOE
(525/555)
Center Excitation
470 523 550 571
Wavelength (nm)
Half Peak Range (nm) 30 20 20 20
Center Emission
525 564 584 612
Wavelength (nm)
Half Peak Range (nm) 20 20 20 20
1 Control Unit
2 Transmission Unit
3 Computer system
4 Hot-lid Unit
2
5 Thermal Circulation
6 Unit
6 Photoelectric Unit
7 Power Supply Unit
3
7
to protect fluorescence detection. Keep it away from heater, stove and any other
heat source! If the instrument will not be used for a long time, the power supply
should be turned off and power plug shall be withdrawn and the instrument
should be covered with soft cloth or plastic film to avoid dust or dirt come into
the instrument.
Figure 3
Figure 4
Please keep the Fixing Pin tightened during the normal operation of the instru-
ment. Before moving the instrument, turn the instrument on and wait until the
self-test action is completed. Shut down the instrument normally and remove
the Fixing Pin from the back of the instrument and insert it into the hole on the
right side of the instrument and tighten it.
Warning!
1. Before cleaning the instrument, the power supply must be switched off.
2. When cleaning the conical wells of the module, prevent any cleaning
agents dropping into the wells.
3. The surface of the instrument MUST NOT be cleaned with corrosive
cleaning agents.
4. In order to avoid scratches or damages of the optics in the wells, NEVER
use sharp or hard objects to clean the wells.
The maintenance of the HumaCycler should be performed by local Human
authorized Technical Support or well trained users depending of the fre-
quency of use.
wards, presses and the program runs. If the plate spring pressurization structure
is used, only certain deformation can generates pressure to act optic coupling,
so it is a “constant pressure type”. Without any PCR tubes, the hot-lid still keeps
the certain gap with the module when it goes down to lower limit.
2.18 How long does the instrument need for plate scanning?
The relation between theoretical running time and recommended re-scanning
interval settings is shown in Table 6:
Total Channel Parameter 1 Time 2 Times 3 Times 4 Times
Table 5 Scanning Time
Whole Plate Time difference 5.04
Scanning between wells
(sec.)
Theoretical 6.54 13.08 19.62 26.16
cumulative time
(sec.)
Rescanning Recommended 7 14 20 27
Interval minimum setting
(sec.)
Single row scan- Time difference 0.48
ning fluorescence between wells
collection (sec.)
Theoretical cu- 1.98 3.96 5.94 7.92
mulative time
(sec.)
Rescanning Recommended 2 4 6 8
Interval minimum setting
(sec.)
It takes a certain amount of time for the module temperature to conduct to the
reagents in the PCR tube and to stabilize the temperature. This time for heat
conduction depends on the PCR tube type, wall thickness and the material heat
conductivity. Therefore, to develop a PCR amplification program, time for the
heat conduction should be added on the recommended re-scanning time in-
terval to consider the time for a constant temperature at the detection point.
When the program is running, the instrument will forecast before the constant
temperature period on detection point is finished and will scan in advance. If
the time for constant temperature is set shorter than the time that scanning
needs, the instrument will extend the constant temperature period automati-
cally to meet the scanning requirements. The time difference at the detection
cavity must be very small, if the reagent fluorescence detection is required. To
scan the whole plate, the time difference for single color fluorescence detection
between the wells is 5.04 sec. To scan the designated single color in a single row,
the re-scanning interval is recommended to set at 2 sec. The time difference for
the fluorescence detection between wells is 0.48 sec. To scan designated several
rows, the time difference is between 0.48 sec. and 5.04 sec.
1. The Photomultiplier (PMT) gain needs 1.5 sec. for stabilization upon the
channel bank replacement.
2. For the theoretical calculation, the entire X-axis takes 0.48 sec., Y-axis
takes 0.15 sec/line, X+Y=0.563 sec. (the total time for X and Y to complete
one row and move one column). The phase difference of total scanning
time over the whole 96-well plate is 0.63*8=5.04 sec. Since it prepares to
replace the gain during the scanning process, return plus gain stabiliza-
tion is calculated as 1.5 sec.
3. The theoretical cumulative time includes the time for zero return and the
time for data transmission.
4. The time difference between the wells refers to the time difference bet-
ween first and the last cavity. The re-scanning interval time includes the
time for the constant temperature and the time for the temperature
change at detection point.
Figure 5
2.24 Why does the display of the system parameter menu require
the input of a password?
The system parameters are for the instrument manufacture’s internal calibra-
tion and require a special accession password. The function is not required for
the end user and for calibration.
20
3.4 Why does the actual temperature displays 0°C or 100°C after
detecting the sample position?
The temperature module is damaged. It accompanied the panel red lamp
alarm and a message appears in the software. The instrument stops running
automatically. Please, contact the local HUMAN authorized Technical Support.
3.5 Why is the OPEN key not working during the instrument is
running?
While the instrument is running, the hot-lid presses onto the test tubes and the
module is not allowed to move. After the run is completed, the hot-lid lifts to the
right position and the button works again.
22
3.6 Why can’t the module not locked if it’s pulled out and pushed
in quickly?
According to the instrument design, the module locks 0.5 sec. to lift the slide
block to get it out. Please, wait when it is pushed in.
3.7 What are the default temperatures of the hot-lid and the block
in a non-operation state?
The default control temperature of the hot-lid is 105°C and 30°C of the block.
Start the running program, the control temperature of hot-lid will change ac-
cording to the customer’s program. After finishing the PCR run, the instrument
restores to the non-operational status.
5.1 The ports are okay, but the instrument does not communicate
If the software’s running communication fails, use the serial port communica-
tion assistant. A message is shown if COM1 or COM2 are occupied. Another rea-
son could be that software can occupy ports, so that the instrument does not
communicate.
It might be also helpful to install the driver software. Base upon your computer
configurations, use either the CP210xVCPInstaller_64.exe (64 bit) or CP210x-
VCPInstaller_86.exe (32 bit) for driver installation. These files can be found in
the HCy4 software. Restart the HCy4 software program and connect the instru-
ment again.
5.2 Why does the instrument not respond although the flip switch
at the rear was turned on?
The Run Switch at the front of the instrument is not turned on. The internal
switching power has not voltage output. For working with the instrument, the
Run Switch must be pressed and the green indicator lamp should be lit green to
indicate that the control system is energized.
5.3 Why the module temperature runs normally, but the fluore-
scence detection does not scan?
The Fixing Pin is not removed or not inserted and tightened into the rear un-
locker port. Remove the Fixing Pin and tighten it into the unlocker port.
5.4 Why does the step motor not work and the communication
fails at the detecting sample position?
1. The reason could be a poor contact or damage of the interface wire. Please
verify, connect or replace the interface wire.
2. The flip switch is not turned on or it’s turned on only after the program starts
running. Please, turn the flip switch on and restart the program.
3. The step motor or the drive is damaged. Please, contact the local HUMAN
authorized Technical Support.
4. The Fixing Pin is not fully inserted. Tighten it into the unlocker port and
switch the power button on after shutdown.
26
5.5 Why the power lamp fails to light up after turning the flip
switch on?
The Run Switch is not pressed. This switch is for the temporary turning the
ON/OFF of the output power and its equivalent to the standby mode. For a
prolonged shut down, the rear flip switch should be switched off. Anoth-
er reason could be that the power button is not switched on. Please, turn the
flip switch on. In addition, check if the fuses are blown. Therefore, replace the
fuses included in the shipment. In case of one of the switches are damaged,
contact the manufacturer or the HUMAN authorized Technical Support or the
local distributor.
5.7 The indication lamps are always in the ON mode after the
instrument started
If the switch button on communication connection box is at “Update” position,
3 lamps on small panel are always on, which means the CPU program is not run-
ning. This status is for CPU program upgrading. Please keep the switch button at
the “Normal” position for use.
5.14 How to get the reagent tubes back if the power fails during
the PCR run?
If the power supply fails during a PCR run, the instrument cannot be opened and
the reagent tubes cannot taken out. Please act as below:
1. Turn the plastic gear at the instrument is top continuously as instructed; lift
the internal hot-lid and stop when you think it reaches the position.
2. Insert a flat screwdriver (diameter <5mm, length approx. 150 mm) from the
hole at the top behind the instrument until an unlocking sound is heard, the
28
module is bounced out slightly, pull the module out manually, take the rea-
gent tubes out.
3. Push the module in a normal position.
R: Reporter
Q: Quencher
30
Figure 7
the linear line of fluorescence S-shaped curve. In general, to start with Real-Time
PCR, customers can calculate 10 times of the standard deviation of the fluores-
cence background signal before amplification. For further information, please
see the Instruction for Use of the HumaCycler DYE CALIBRATION KIT (REF 86150).
Figure 8
To set up these parameters manually, try to select the early stage that comes to
exponential phase and make sure the regression coefficient is bigger than 0.95.
6.7 What is the relation between the Ct value and the initial
template amount?
The more template amount is in the sample, the less cycles are needed for the
specific fluorescence signal to reach the threshold.
32
Figure 9
Figure 10
6.11 What could be the reason for poor linear regression of the
standard curve?
1. Quantification Standards show no linear regression due to errors of the sam-
ple application.
2. Quantification Standards are degraded. Avoid multiple freezing and thaw-
ing cycles, aliquot the Quantification Standards in convenient portions.
3. Primers or/and probes are not working. Please, check the storage conditions,
expiry dates, aliquot them in convenient sizes and avoid multiple freezing
and thawing cycles.
4. Template contains inhibitors or suffers too high concentration. Dilute the
templates with TE, pH 8.
6.19 Why there are more than one peak in a melting curve?
1. The primer design is not optimized. Avoid primer dimers and pin structures.
2. The primer concentration is too high. Decrease the primer concentration pro-
perly and pay attention to the concentration ratio of the forward and reverse
primers.
3. The concentration of magnesium ions is too high. Decrease the magnesium
ion concentration properly.
4. The template is contaminated. Avoid DNA contamination during the RNA ex-
traction process or avoid non-specific amplification by an optimum primer
design.
5. Some multi-probe reagents may have many Tm peaks.
If using commercial available CE marked assays, just the point 4) and 5) might
be relevant.
is not very ideal. If users are doing an absolute or relative quantification exper-
iment with the HCy4 software, the algorithms to analyze the data are prein-
stalled.
6.24 What are the reasons for false positive results and how to
avoid it?
If a positive result appears in negative controls, the detection result of other
samples in this test might be false positive. The reasons of false positive include
the contamination of samples, amplification reagents or amplification products.
The common contamination source includes lab environment, sample injector,
aerosols generated during operation, reagent or anything in contact with am-
plification product. It is recommended to isolate the working area and to decon-
taminate all items that are used (pipettes, centrifuges, vortex mixers, benches
with fresh prepared 10% bleaching solution and or UV light. Dispose all reagents
of the assay and templates and use a new assay for re-testing. Users should
improve the lab operation, e.g. to avoid the reagent splash during the sample
application, take high-pressure treatment on sucker and centrifugal tube. Don’t
open the lid of the PCR tube after amplification and use filter tips for transfer
the reagent and templates to the PCR tubes, etc. Use 3 or 4 working areas and
wear separate gloves, coats and shoes or shoe covers.
6.25 What are the reasons for false negative results and how to
avoid it?
If a positive control set is not amplified into positive result, the test might have
a false negative result. Besides that, if there are problems with the extraction
steps of specimen nucleic acid purification, the specimen detection might find
false negative even though positive control is positive. So, the reason might be
that there is no DNA or RNA in the sample or after extraction. Standardize the
operation of DNA/RNA extraction, use commercial available extraction assays
and pay attention to clean working. Store the nucleic acids at -20°C or lower in
aliquots. Avoid multiple freezing and thawing cycles.
38
6.27 Why are the first 3 cycles in the amplification curve always
accompanied with a lift or fall?
When the amplification begins, the temperature transfer in the test tube takes
time. If the constant temperature cycle time set in PCR reaction program is short,
the temperature of the block is not transferred to reagent completely, the re-
agent fluorescence is affected by temperature and the PCR tube material and
structure cause thermal conductive coefficient difference. As a result, an in-
creasingly steady process is shown. PCR tubes are recommended, which have a
higher thermal conductive coefficient.
6.29 What are the reasons for getting non S shaped curves and
how to avoid it?
If a positive control curve is not S-shaped, the reason could be that the reagents
are not fully thawed or not mixed uniformly before the sample was added.
Besides, many unknown components in the sample might interrupt the PCR
6.33 Can multi-channel dyes used in the same tube with the
HumaCycler?
Yes, but pay attention to certain situations, the amplification of a kind of probe
component might inhibit the amplification of other component or generation
of fluorescence. Separate dyes to different test tubes, if there are doubts on the
test result. If using CE marked multiplex assays, the reaction systems are opti-
mized to the different targets and dyes.
40
6.36 Why does the dye channel match but names and default not?
User can name the new dye by most suitable ray filter channel wavelength. Se-
lect Customize Dyes from Tools on the HCy4 software, input channel number
and dye name.